Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

Slides:



Advertisements
Similar presentations
A few thoughts on gamma densitometry of powder jet Goran Skoro 27 October 2008.
Advertisements

1 Proton detection with the R3B calorimeter, two layer solution IEM-CSIC sept report MINISTERIO DE EDUCACIÓN Y CIENCIA CONSEJO SUPERIOR DE INVESTIGACIONES.
Detector systems 1) Anti-Compton spectrometers 2) Pair spectrometers 3) Crystal spheres, walls, complex set-ups of semiconductor and scintillation detectors.
Quartz Plate Calorimeter Prototype Ugur Akgun The University of Iowa APS April 2006 Meeting Dallas, Texas.
DREAM Collaboration: Recent Results on Dual Readout Calorimetry. F.Lacava for the DREAM Collaboration Cagliari – Cosenza – Iowa State – Pavia – Pisa –
E. Nappi First results from measurements of optical properties of YAP:Ce crystals  Procurement of YAP:Ce crystals Price enquiry for 3.6x3.6x100.
Geant4 simulations for the calorimeter prototypes D. Di Julio, J. Cederkäll, P. Golubev, B. Jakobsson Lund University, Lund, Sweden.
Geant4 simulations for the calorimeter prototypes D. Di Julio, J. Cederkäll, P. Golubev, B. Jakobsson Lund University, Lund, Sweden.
Observation of Fast Scintillation of Cryogenic PbI 2 with VLPCs William W. Moses, 1* W.- Seng Choong, 1 Stephen E. Derenzo, 1 Alan D. Bross, 2 Robert Dysert,
Tagger Electronics Part 1: tagger focal plane microscope Part 2: tagger fixed array Part 3: trigger and digitization Richard Jones, University of Connecticut.
ipno.in2p3.fr Milano - October 2006IPNO-RDD-Jean Peyré1 Measurements of CsI(Tl) Crystals with PMT and APD Jean Peyré Milano - Oct 2006.
R&D on CsI(Tl) + APD Optimisation of the energy resolution M. Gascón I.Durán.
1 SCINTILLATION LIGHT YIELD OF Ce-ACTIVATED LSO, LYSO, LuAP AND LuYAP.
Nustar/crystal simulations B. Genolini Crystal: CsI(Tl) (Saint Gobain), wrapped with reflector (VM2000, Tyveck?) Geometry:
Crystals and related topics J. Gerl, GSI NUSTAR Calorimeter Working Group Meeting June 17, 2005 Valencia.
Lecture 2-Building a Detector George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
ipno.in2p3.fr Lund - June 2007IPNO-RDD-Jean Peyré1 Measurements of some Crystals with PMT R&D on Calorimeters Jean Peyré Lund - June 2007.
An Alternate Perspective Most development on ton detectors –~500 GW-ton-years –Systematics limited for rate measurement Lots of thought about systematics.
PEPPo Weekly Meeting February 23, 2011 PEPPo Electromagnetic Calorimeter First measurements Jean-Sébastien Réal, Eric Voutier Laboratoire de Physique Subatomique.
Photon detection Visible or near-visible wavelengths
Absolute light output determination of crystal scintillators
PPARC forum on developments in scintillator technology R M Brown - RAL 1 Scintillating Crystals for Particle Physics Outline  Performance considerations.
EXL/crystal simulations B. Genolini Simulation of NUSTAR crystals with Litrani Presentation of Litrani: simulation of.
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
ExternalTargetFacility at CSR FRIB-China East Lansing Sun, Zhiyu Institute of Modern Physics, CAS.
DSSD – Si, 300 mkm, 1.85 mm pitch, 59.5x59.5 mm**2. CsI(Tl)/PD, 20x20(front)23x23(back)x100mm**3. Produced by St.Gobain, Khar’kov and China Companies PD.
November 8, 2005JLAB Awake Animal Project New Instrumentation for JLAB Awake Animal Project Smaller Pixel Size requires higher granularity in PMT (2.4.
Innovation is in our genes. 1 Siemens Medical Solutions Molecular Imaging What are SPECT basics?
Annual Modulation Study of Dark Matter Using CsI(Tl) Crystals In KIMS Experiment J.H. Choi (Seoul National University) SUSY2012, Beijing.
Heavy Scintillating Crystal Fibers for calorimetry
2004/Dec/12 Low Radioactivity in CANDLES T. Kishimoto Osaka Univ.
Light Output Uniformity of Long CsI(Tl) crystals - the “bottle-neck” for gamma and proton spectroscopy Light Output Uniformity for CsI(Tl) as supplied.
Setup for hypernuclear gamma-ray spectroscopy at J-PARC K.Shirotori Tohoku Univ. Japan for the Hyperball-J collaboration J-PARC E13 hypernuclear  -ray.
Tests WLS - Readout axial PET - Bari - Janvier 2007 AXIAL PET - HPD AXIAL COORDINATE RECONSTRUCTION WITH WLS STRIPS
Halide Scintillator Growth & Characterization for Rare Decay Search Presented by Gul Rooh Kyungpook National University Republic of Korea.
T. Sugitate / Hiroshima / PHX031 / Nov.01 The Photon Spectrometer for RHIC and beyond PbWO 4 Crystal Density 8.29 g/cm 3 Radiation length 0.89 cm Moliere.
The Simulation of the quenching factor and the channeling effect for the CsI(Tl) and NaI(Tl) crystals Juhee Lee and Sunkee Kim Seoul National University.
Neutron detection in LHe ( HMI run 2004) R.Golub, E. Korobkina, J. Zou M. Hayden, G. Archibold J. Boissevain, W.S.Wilburn C. Gould.
Increase in Photon Collection from a YAP:Ce Matrix Coupled to Wave Lenght Shifting Fibres N. Belcari a, A. Del Guerra a, A. Vaiano a, C. Damiani b, G.
HBD Gas and QE Monitoring Craig Woody BNL HBD Working Group Meeting October 19, 2005.
Development of a Gamma Camera Based on an 88 Array of LaBr3(Ce) Scintillator Pixels Coupled to a 64-channel Multi-anode PMT Hidetoshi Kubo, K.Hattori,
R. Pani Department of Experimental Medicine and Pathology University of Rome La Sapienza-Italy. Flat Panel PMT: advances in position sensitive photodetection.
WIMP search Result from KIMS experiments Kim Seung Cheon (DMRC,SNU)
Measurement Birk’s constant and Ligth Yield Light Yield is the number of optical photons emitted when 1 MeV energy deposited in the scintillaotr γ γ'γ'
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
Detectors for VEPP-2000 B.Khazin Budker Institute of Nuclear Physics 2 March 2006.
Adam Para, Fermilab, February 16, Who Cares? What is the Problem? 2 Dual Readout Total Absorption calorimeter has very good energy resolution.
PoGO_G4_ ppt1 Study of optimized fast scintillator length for the astronomical hard X- ray/soft gamma-ray polarimeter PoGO November 1, 2004 Tsunefumi.
1 Plannar Active Absorber Calorimeter Adam Para, Niki Saoulidou, Hans Wenzel, Shin-Shan Yu Fermialb Tianchi Zhao University of Washington ACFA Meeting.
The BGO matrix readout for the 2009 Test Beam Guido & Alessandro.
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
Sep. 22, 2011 Seoul National University Jae Keum Lee KIMS Background 1 China-Korea Workshop 2011 September 22-23, 2011.
Chapter-2 The Planar Imaging Gamma Camera
Seoul National University Han-wool Ju CUNPA Kick-off Meeting Aug.22-23, 2013.
WP 22 – SciFi Development and Characterisation of Inorganic Scintillating Fibers Made of LuAG:Ce and LYSO:Ce S. Diehl 1, R. Novotny 1, Aubry Nicolas 2.
R&D on Scintillation Crystals and Special Glass at BGRI Mingrong ZHANG, Yuncheng GE (Beijing Glass Reseach Institute) 2nd HHCAL workshop, Beijing, May.
Gamma Spectrometry beyond Chateau Crystal J. Gerl, GSI SPIRAL 2 workshop October 5, 2005 Ideas and suggestions for a calorimeter with spectroscopy capability.
WIMPs Direct Search with Dual Light-emitting Crystals Xilei Sun IHEP International Symposium on Neutrino Physics and Beyond
Nuclear Medicine Physics
The Lund R3B prototype: In-beam proton tests and simulations
Connect Ratios and Rates to Multiplication and Division
Status of 100Mo based DBD experiment
Lecture 1: Development of the Analog Gamma Camera
Dual-Readout Calorimeter: DREAM
Rare earth activated wide band gap oxide crystals can be very useful as detectors of ionizing radiation in a number of applications. Lu3Al5O12:Pr (Pr:LuAG)
LSO/LYSO Crystals for Future High Energy Physics Experiments
A.Takada, A.Takeda, T.Tanimori
HE instrument and in-orbit performance
Characterization of Large CaMoO4 crystals for 0n bb search
一种基于晶体间光分享原理的深度测量PET探测器
Presentation transcript:

Tests of scintillators CsI, CsI(Na), BGO, NaI(Tl) CsI(Tl)

CsI(Na) 170 mm 15 x 15 mm 9 x 9 mm

Energy of PMT 1 (small window) Energy of PMT 2 (big window) CsI(Na) scintillator with 2 PMT readout Full absorption Measurement without collimation 137 Cs, 661 keV

Energy of PMT 1 (small window) Energy of PMT 2 (big window) CsI(Na) measured with a collimated source Sum (energy) Ratio (position)

Energy spectrum (sum of 2 PMT’s energies) Sum of 2 PMTs energies Energy resolution 9.6% at 661 keV Constant for the position of the interaction

Position resolution (ratio of the PMT’s energies) Light division between 2 PMTs 15 cm 10 cm 5 cm

Position resolution

Position resolution 1.5 – 3 cm

Mean absorption distance

Readout with one PMT 5 cm 10 cm 15 cm

CsI, pure

Energy resolution Resolution 17 – 19 % Sum energy

Position resolution Light division between 2 PMTs 15 cm 10 cm 5 cm

Position resolution

Position resolution 2 – 8 cm

BGO

Energy of PMT 1 (small window) Energy of PMT 2 (big window) BGO scintillator with 2 PMT readout Scintillator was illuminated by Cs137 source (661 keV) without a collimator Strong nonuniformity of light yield along the crystal distance

BGO collimated measurement Full absorption

BGO collimated measurement Full absorption

1)Light yield is strongly non uniform along the crystal distance 2)Light separates always in the same proportion between left and right sides 3)Energy resolution ≈18% (big window) ÷ 24% (small window) BGO results

CsI(Tl) 7% at 661 keV 2.5 cm

to continue with..