1 一种基于晶体间光分享原理的深度测量PET探测器
Development of a depth-encoding PET detector module based on light sharing between adjacent crystals 邝忠华,杨茜,王晓辉,付鑫,任宁,张春晖,赵斌清,桑子儒,杨永峰* 中国科学院深圳先进技术研究院 保罗C. 劳特伯生物医学成像研究中心

2 Detector Design and Measurements
Outline Introduction Detector Design and Measurements Results Conclusions

3 I. Introduction DOI effect in whole body and total body PET scanner
Whole PET scanner Total Body PET scanner 20 cm None-DOI detector 200 cm DOI detector

4 Timing resolution: 500 ps  7.5 cm
I. Introduction Time of flight (TOF) PET DOI TOF-PET ∆𝒙=𝒄× ∆𝒕 𝟐 Timing resolution: 500 ps  7.5 cm We aim to develop a depth-encoding PET detector with good timing resolution.

5 II. Detector Design and Measurements
𝐷𝑂𝐼 𝑅𝑎𝑡𝑖𝑜= 𝐸 𝑆𝑖𝑃𝑀1 𝐸 𝑆𝑖𝑃𝑀1 + 𝐸 𝑆𝑖𝑃𝑀2 Detector unit Detector module (32 detector units) Couple material optimized in our previous work (Zhonghua Kuang et al 2018 Phys. Med. Biol. 63 045009).

6 II. Detector Design and Measurements
Readout electronics Row and column readout 64 chs-> 16 chs 𝑥= 𝑖=1 𝑛 𝑖× 𝐸 𝑐𝑜𝑙 (𝑖) 𝑖=1 𝑛 𝑖× 𝐸 𝑐𝑜𝑙 (𝑖) + 𝑖=1 𝑛 (𝑛−𝑖+1)× 𝐸 𝑐𝑜𝑙 (𝑖) 𝑦= 𝑖=1 𝑚 𝑖× 𝐸 𝑟𝑜𝑤 (𝑖) 𝑖=1 𝑚 𝑖× 𝐸 𝑟𝑜𝑤 (𝑖) + 𝑖=1 𝑚 (𝑚−𝑖+1)× 𝐸 𝑟𝑜𝑤 (𝑖) 𝐷𝑂𝐼 𝑅𝑎𝑡𝑖𝑜= 𝐸 𝑐𝑜𝑙 (2𝑗) 𝐸 𝑐𝑜𝑙 2𝑗−1 + 𝐸 𝑐𝑜𝑙 (2𝑗)

7 II. Detector Design and Measurements
Experimental setup A LYSO slab detector (1.0 mm thick) and a 22Na point source (0.25 mm diameter) were mounted on a translation table. Experiments were performed in both singles and coincidence modes. In singles mode, the whole array can be irradiated uniformly. In coincidence mode, a specific depth (2mm, 6mm 10mm 14mm and 18mm) is irradiated by using electronic collimation.

8 III. Results Flood & DOI profiles in singles mode Flood histogram
(clearly resolved) DOI histograms (all pairs)

9 III. Results DOI profiles of detector unit 2 (left) and unit 15 (right) measured at five depths of 2, 6, 10, 14 and 18 mm. Unit 2 Unit 15

10 III. Results DOI resolution (E>400 keV)
Average DOI resolution: 4.62±0.59 mm

11 III. Results Energy spectrum and resolution Crystal No. 2
Average energy resolution: 13.5±1.7 %

12 III. Results Timing spectrum and resolution (E> 400 keV)
Whole array Average timing resolution: 518±28 ps

13 IV. Conclusions Performance of the proposed detector module
The SiPM readout circuit will be further simplified and improved to obtain a better timing resolution in the future. The proposed DOI detector module can be used to build high performance whole body and total body PET scanners. DOI resolution All events 5.27 ± 0.62 mm E > 400 keV 4.62 ± 0.59 mm Energy resolution 13.5 ± 1.7 % Timing resolution (E > 400 keV) 518 ± 28 ps Let me make a conclusion for the performance of the three measured detectors by using the following table.

14 Thank you for your attention!