Biophan Technologies, Inc. Jeffrey L. Helfer, Vice President of Engineering ISMRM Workshop on MRI Safety November 6, 2005.

Slides:



Advertisements
Similar presentations
C. M. Johnson, P. H. Riley and C. R. Saha Thermo-acoustic engine converts thermal energy into sound energy by transferring heat between the working media.
Advertisements

Mitigation of AC Induced Voltage On Buried Metallic Pipelines
Introduction to RF for Accelerators
Richard J. Mohr President, R. J. Mohr Associates, Inc.
Introductory Circuit Analysis Robert L. Boylestad
MRI Phillip W Patton, Ph.D..
RF Circuit Design Chris Fuller /7/2012.
MRI in patients with pacemakers DR PRADEEP SREEKUMAR.
Electromagnetic Oscillations and Alternating Current
DAREPage 1 Non-Invasive Induction Link Model for Implantable Biomedical Microsystems: Pacemaker to Monitor Arrhythmic Patients in Body Area Networks Prepared.
Radio Frequency Induced Heating Near Medical Devices During MRI Justin Peterson University of Western Ontario Medical Biophysics March. 23, 2011.
UNIVERSITI MALAYSIA PERLIS
Designing a EMC Compatible Electronic Meter using AD7755 a.
RF Safety for Interventional MRI Procedures Ergin Atalar, Ph.D. Bilkent University, Ankara, Turkey Johns Hopkins University, Baltimore MD USA.
Capacitors for RF Applications Michael P. Busse Vice President Dielectric Laboratories, Inc 2777 Rte. 20 East Cazenovia, NY
Week 3 Electrical Circuits and Components. It’s a Radio!
Introduction  Image Guided Adaptive Radiotherapy (IGAR) involves the use of an imaging modality to minimize planning treatment volumes required to deliver.
Electrical eddy currents in the human body: MRI scans and medical implants Brent Hoffmeister Rhodes College Department of Physics.
FMRI: Biological Basis and Experiment Design Lecture 18: Physical practicalities Digression: analysis ICE9: Example for WA8 Safety limits –dB/dt –SAR –Acoustic.
Magnetism The density of a magnetic field (number of magnetic lines passing through a given surface) is the magnetic flux: Units of flux are Webers. Tesla/m.
Electromagnetics (ENGR 367)
Magnetic Resonance Imaging Mary Holleboom ENGR 302 May 7, 2002.
Consider a long solenoid of cross-sectional area A, with number of turns N, and of length l. The flux is The magnitude of B is given by: Therefore, The.
Chapter Two: Radio-Frequency Circuits. Introduction There is a need to modulate a signal using an information signal This signal is referred to as a baseband.
Physiological Effects of Electricity on Human Body by Mohd Yusof Baharuddin.
Conducted Emissions and Susceptibility Conducted emissions are simpler to investigate than radiated emissions because only the product’s power cord needs.
Wireless Power By: Peter Woods Supervising Professor: Dr. Chiao July 16 th 2013.
Copyright ©2011 by Pearson Education, Inc. publishing as Pearson [imprint] Introductory Circuit Analysis, 12/e Boylestad Chapter 20 Resonance.
Copyright ©2011 by Pearson Education, Inc. publishing as Pearson [imprint] Introductory Circuit Analysis, 12/e Boylestad Chapter 20 Resonance.
AFCI Reaction to Utility Events. CenterPoint Energy 35 KV Distribution Capacitor Bank – Switched through an automated system.
Coupling and Cohesion Pfleeger, S., Software Engineering Theory and Practice. Prentice Hall, 2001.
- MRI Safety Update - RF Induced Heating
ECE & TCOM 590 Microwave Transmission for Telecommunications Introduction to Microwaves January 29, 2004.
Influence of Pacing Lead Design on MRI-Induced Lead Heating Presenter: Stuart MacDonald VP Research and Development Heart Rhythm 2006 Boston, MA May 18,
Human Subjects in fMRI Research Credits: Robert Savoy, Ph.D. Franz Schmitt, Ph.D.
MR Scanner Transmit Gain as an Indicator For Lead Heating November, 2006.
Transformers ET 242 Circuit Analysis II
J1879 Robustness Validation Hand Book A Joint SAE, ZVEI, JSAE, AEC Automotive Electronics Robustness Validation Plan The current qualification and verification.
Nelson Research, Inc – N. 88 th St. Seattle, WA USA aol.com Non-Contacting Eddy Current Conductivity Sensor for Measuring.
Copyright © 2009 Pearson Education, Inc. Chapter 32: Inductance, Electromagnetic Oscillations, and AC Circuits.
Course no. 9 Technical University of Cluj-Napoca Author: Prof. Radu Ciupa The Theory of Electric Circuits.
Pacemakers and MRI An outdated contraindication? Emily Myers Research in Allied Health East Tennessee State University.
Laser Treated Metallic Probes for Cancer Treatment in MRI Systems July 08, Advance Materials Processing and Analysis Center (AMPAC) Department of.
Magnetic Resonance Imaging
Assessment of MRI Induced Heating of Cardiac Pacing Leads: Relevance to Target Patient Populations Presenter: Daniel G. Hullihen Jr. Director of Business.
Part I: Dipoles by Marc C. Tarplee Ph.D. N4UFP
Doc.: IEEE Submission Nov 2009 Project: IEEE P Working Group for Wireless Personal Area Networks (WPANs) Submission Title:
Thermal Analysis Assumptions: Body Temperature (Environment) is 37˚C Heat distribution on outside of device will be modeled via FEA Heat transfer method.
 A medium that provides a means for transferring energy from one place to another  It is not expensive or difficult to control the risk of an electrical.
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
Survey of Smart Grid concepts and demonstrations Smart substation Ari Nikander.
FEASIBILITY ANALYS OF AN MHD INDUCTIVE GENERATOR COUPLED WITH A THERMO - ACOUSTIC ENERGY CONVERSION SYSTEM S. Carcangiu 1, R. Forcinetti 1, A. Montisci.
Electrical and RF Safety. Electrical Safety Generators Generators convert mechanical energy to electrical energy. They actually consist of two devices:
Coupling and Cohesion Pfleeger, S., Software Engineering Theory and Practice. Prentice Hall, 2001.
TRANSFORMER A transformer is a static device.
QUANTUM-EFFECT DEVICES (QED)
MR Scanner Transmit Gain as a Potential Indicator For Lead Heating
Coupling and Cohesion 1.
SHORT RANGE WIRELESS POWER TRANSMISSION
Maintenance schedule for different equipment and their records in a hospital By Mr. Rahul kumar.
ELECTRICAL INSTALLATION AND MAINTENANCE (EIM) 5 INSTALL ELECTRICAL PROTECTIVE DEVICES FOR DISTRIBUTION, POWER, LIGHTING, AUXILIARY, LIGHTNING PROTECTION.
POWER SYSTEM ANALYSIS INTRODUCTION.
Date Submitted: [xx Nov, 2009]
J1879 Robustness Validation Hand Book A Joint SAE, ZVEI, JSAE, AEC Automotive Electronics Robustness Validation Plan Robustness Diagram Trends and Challenges.
Human Subjects in fMRI Research
Equations, Performance, Electrical Equivalent Circuits
Exercise Session 11 Power systems.
AC current.
(4)ELECTRONIC SUPPORT SYSTEM
Presentation transcript:

Biophan Technologies, Inc. Jeffrey L. Helfer, Vice President of Engineering ISMRM Workshop on MRI Safety November 6, 2005

- Acknowledgements - Robert Gray Xingwu Wang, Ph.D. W. Timothy Bibens Mark Bocko, Ph.D. Stuart G. MacDonald Jeffrey L. Helfer University Medical Imaging, Rochester, New York

- Introduction - Pacemakers and other devices can create risks to their patients when exposed to MRI 1.Excessive heating of the device (multiple causes) capable of producing uncontrolled tissue heating and thermogenic damage. 2. Induced voltages in the device that can interfere with organ function and device diagnostic and therapeutic capabilities. 3. MR image disruption and distortion that prevents the visualization of tissues “close” to the device.

Introduction - continued While it is relatively easy to demonstrate a heating or induced voltage problem, it is far more difficult to prove a solution to these problems, due to their complex and unpredictable nature, which includes factors such as: RF field strength Patient position in the coil Type of imaging sequence Patient characteristics Duration of imaging procedure Body structure being imaged Lead design Specific type of medical device Lead orientation within patient The degree of perfusion near the device Temp. measurement procedure Respiratory phase Managing MRI-induced Patient Risk is a Very Difficult Task! Many of these parameters are either not recognized or poorly addressed by existing testing methods (i.e. ASTM 2182)

Proper understanding of the MRI safety situation is further exacerbated by the underreporting of adverse events, due to: Physician reluctance to report adverse events Litigation that shrouds the dissemination of circumstances surrounding adverse events MR systems using higher and faster gradient fields, and stronger RF fields will become increasingly common (e.g. move to 3T), maintaining the potential for insufficient safety awareness and risk to patients. Guidelines alone do not guarantee patient safety. We believe that patients deserve devices that are inherently safe! Introduction - continued

To examine the effects of lead design on MRI-induced heating To utilize these insights to develop inherently safe lead designs - Purpose of this Investigation -

- Key Assumptions - MRI energy is coupled into conductive leads in two major ways: - Antennae effect - Electrical potential induced within the body (Implant acts as an electrical “short circuit”) High electrical current densities at the lead-tissue interface induces resistive heating in tissue

Coiled Lead Wires - Hypothesis - Tissue heating can be substantially reduced by increasing the high frequency (i.e. 64MHz) electrical impedance of the lead

Proprietary design bipolar pacing lead prototypes, 52 cm in length. Connected to IPG. Standard active fixation bipolar pacing lead, 52 cm in length. Connected to IPG. (Control) Luxtron® fluoroptic thermometry system Head/torso phantom Gelled-saline solution: 5.8 g PAA, 0.8g NaCl per liter of de-ionized water GE 1.5-T MR system (GE), FSE-XL, Whole body avg. SAR: 1.79 W/kg - Materials and Methods - at

Experimental Setup

Source: R.Ludwig, P. Bretchko, RF Circuit Design Theory and Applications, Prentice Hall, 1999 Theory: Air Core Coils Simplified Impedance Equation Resonance Condition Maximum coil impedance occurs at “self” resonance. R d ≡ Distributed Resistance C d ≡ Distributed Capacitance R s ≡ Series Resistance C s ≡ Parasitic Shunt Capacitance

Theory: Shifting Self Resonance Of Lead Maximum impedance at “self” resonance. MR scanner’s frequency is fixed. So, need to shift lead’s self-resonance frequency by changing coil inductance and capacitance properties.

Simple Model Of Bipolar Lead Circuit Diagram Circuit of pacing lead in MRI scanner is not simple… IPG

MRI Heating of Pacing Leads Leads designed with different inductance and capacitance. - Results - Two leads had less than 0.5°C temp. increase. Control Changing the wire form design changes the capacitance-inductance characteristics of the lead and its impedance

MRI Heating of Pacing Leads - Results - Control #1 (Vendor A) Control #2 (Vendor B) 6 modified leads had < 1° C temp. increase. Leads designed with different inductance and capacitance. Adding a discrete component, high frequency resonator to the lead changes the capacitance - inductance characteristics of the lead and its impedance

- Conclusions - Lead design geometry has a strong influence on worst case MRI-induced heating at 1.5T Worst-case lead heating can be reduced to acceptable levels in several ways: ● Proper choice of wire form design geometry ● Use of discrete component resonator

- Commentary - When implanted, these designs provide the potential to: Provide a greater margin of patient safety Allow a greater number of patients access to MRI Minimally disruptive lead design options are available to reduce worst-case lead heating to acceptable levels We believe that these design options can also be applied to other similar design conductive implants such as ICD and DBS leads, guidewires, catheters, etc. Biophan has also developed easy to implement solutions for reducing or eliminating MRI-induced voltages in leads