Multi-parameter Water-Quality Probes (YSI/Hydrolab) Lecture 4.

Slides:



Advertisements
Similar presentations
Analyze-Immediately Parameters pH Analyze Immediately Within 15 minutes of collection: pH Residual Chlorine Temperature Chlorine Dioxide Dissolved Oxygen.
Advertisements

Water Quality Testing Made Easy GPS Aquameter and Multiparameter Aquaprobe The perfect solution for your water quality testing needs.
SECTION 11 LEVEL B SUIT OUT. SECTION 12 AIR MONITORING.
Sonde Maintenance and Care March 2014 Sonde Maintenance and Care 10 th Annual Red River Basin Water Quality Monitoring Training March 5, 2014 Bruce Paakh,
Data Quality Issues Cabled observatories will produce long time series of real time physical ocean data -- revealing: –Trends –Long-period fluctuations.
Comparing the Humidity and Dew Point of two different locations
Acidity Constant by pH Titration Curves
Last week: Statistics I (average, standard deviation, standard deviation of the mean) This week: Statistics II (95% Confidence Interval) & Calibration.
Effects of Riparian Forestation on Water Quality in Urban Streams Aana Taylor-Smith, UNCW Honors Student Supervisor: Brian S. Arbogast, PhD. UNCW Department.
©2012 Fluke Corporation SPRT Health 1 Health Monitoring of SPRTs.
The Flow of Food: an Introduction
IC Controls 865 ppb Dissolved Oxygen System pH / ORP Conductivity Dissolved Oxygen Chlorine Standards ppb Galvanic Dissolved Oxygen.
Blood Gases. Specimen Collection and Handling Arterial Specimen - Avoid use of tourniquet Usually drawn by an MD or Respiratory Therapist Radial, Brachial.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering NRP 3: Let’s get started.
Calibration (Qualification) of Centrifuge
Sanitation and HACCP Monitoring with the LIGHTNING MVP BioControl Systems, Inc.
The Flow of Food Objectives: How to prevent cross-contamination
The Flow of Food To keep food safe throughout the flow of food:
Water Quality Supplies Waste bottle 1413 solution pH 10 solution pH 4 solution Solution cups Water sample cup Hanna Meter.
Soil Water Potential Peter Cull ICT International.
Des Murphy Kovara Energy Consultants Establish a Measuring and Monitoring System Energy MAP Step 18 K O V A R A Energy Consultants.
Kuskokwim Watershed Council Water Quality Training Workshop
Verifying the Use of Specific Conductance as a Surrogate for Chloride in Seawater Matrices Rob Mooney Technical Marketing Manager In-Situ ® Inc.
Multi-parameter Water-Quality Probes (YSI/Hydrolab) Lecture 5.
pH and Conductivity Best Practices
10 Energy Saving Advices. 1- Lowering of the temperature at night Limit the lowering of the temperature at night. If the temperature is reduced too much,
Food Thermometers Calibrating Instant-Read Bi-metallic Thermometers
Overview of Continuous Water-Quality Monitoring. Purpose of Monitoring Define the objectives of the water quality monitoring project 1. Environmental.

DAQ Systems and Instruments for Hx Performance Testing and Periodic Monitoring.
Mercury Spill Response Instrumentation. Mercury Vapor Analyzers Jerome 431-X MVA (gold film) Lumex 915+ MVA (Atomic Absorption) Jerome J405 MVA.
Continuous monitoring in the benthic boundary layer off the Northwestern Florida shelf. William M. Landing, Stephanie Fahrny, Kevin Speer, Markus Huettel.
Foods II Chapter 5 The Flow of Food. Physical Barriers to Prevent Cross-Contamination Assign specific equipment to each type of food product Assign specific.
THE MEASUREMENT OF pH. I GENERAL PRINCIPLES Convenient way to express hydrogen ion concentration, or acidity pH = - log[ H +
© 2007 Multigas Detector Reliability in a small package.
Chapter 5 The Flow of Food: An Introduction
When Do I Turn On the Fans... And For How LONG??? Carol Jones, PhD. Stored Product Engineering Biosystems and Ag. Engr. Dept.
Laboratory Certification Update Part 2 Common Findings KWWOA Louisville April 15, 2015 Presented by Frank Hall, Laboratory Certification Coordinator.
Observing the States of Matter: Solid, Liquid, and Gas Inquiry Lesson: Reference pages in the “Chemistry Matters” AIMS book.
Relative Humidity. Humidity  the measure of the amount of water in the air  high humidity =lots of moisture in the air.
GLOBE PROTOCOLS By: Angela Gonyer, Noah Hughes, Tyy Williams, Eli Lopez, and Ariana Viray.
Monitoring Temperatures and Conduct Inspections
Estimating and measuring time, temperature, mass, and weight Lesson 4 (4 th 6 Weeks) TEKS 6.8 A/B.
Experiment 1. Temperature and Pressure Calibration Temperature measurement tips are illustrated here for one of the setups at Station 4. The reference.
Data is transmitted from monitoring stations and posted on our internet website. Wallops, VAKlamath, CA.
WHAT TO DO AND WHAT NOT TO DO. How to Write a Lab Report.
Objectives: How to prevent cross-contamination How to prevent time-temperature abuse How to use the correct kinds of thermometers to take temperatures.
Goals of FBIRN ( )  Develop multi-site functional neuroimaging tools. Calibration: Assess and correct for the major sources of variation in multi-site.
General Laboratory Techniques Dry Lab Chemistry 1105.
GREENHOUSE EFFECT LAB Procedures.  Assign roles to your group members:  Temperature monitors (2 students)  Timekeeper  Recorder.
1. Swab throat. 5. Read Results Positive = 2 lines. Strep A Rapid Test Strip Procedure For details, consult Lab Manual located at: P:\Clinical\Lab Manual.
Building an Integrated Ocean Color Sensor Web at the Land-Sea Interface UCSC UARC/ARC Ames GSFC BSI SARP.
Satlantic HyperPro II Freefalling hyperspectral radiometer Can be deployed in “floater” mode 16 bit ADC at 12 Hz (maximum) 0 - ~100 m depth rating, descent.
Testing Requirements in the Boiler House Requirements for regular testing will vary according to national regulations, and the type of equipment installed.
LABORATORY ACTIVITIES FOR REGENTS CHEMISTRY. Many laboratory activities require measurements. Science uses the S.I. (Metric System) of measurements.
Equipment & Utensils Using the Right Tools. Our Goal To provide information about the best selection, use and maintenance of food preparation equipment.
Field Equipment Calibration PH, Temperature, Conductivity, Turbidity, Dissolved Oxygen, Free & Total Chlorine Gregory Langland CCSF.
Chapter 4: The Flow of Food
Using Ratio Reasoning to Convert Measurement Units
EMSA 22 Lab Module 1 Field Equipment Calibration
The Flow of Food: an Introduction
Which material is the best Insulator?
Overview of Continuous Water-Quality Monitoring
Probeware 101.
Electrode Training HORIBA Instruments, Inc. Pamela Millett Scientific
When Do I Turn On the Fans...
Southern Blotting DNA Fingerprinting
Southern Blotting DNA Fingerprinting
Getting the YSI Sonde Ready For Use
Pneumatic System Conditioning
Presentation transcript:

Multi-parameter Water-Quality Probes (YSI/Hydrolab) Lecture 4

Multi-parameter probes PROS –Collect multiple metrics of water quality at the same time –Internal data logger – can collect in unattended (log) mode CONS –Require calibration and maintenance –Expensive ($5-10K and up)

YSI (Yellow Springs Inc)

YSI sensors

HACH - Hydrolab

Some Hydrolab sensors

Calibration For every sensor (except temperature) Follow manual and fill out log (keeps track of sensor condition/status) Required before every deployment and should be tested again after retrieval Should test standards using a separate meter/test strip if possible Most calibrations are one-point, some 2- point

2 point calibration Data in this range (interpolated)

Sequence Change D.O. membrane day before!!! Check temperature against thermometer Conductivity first – 1 pt cal (typically 50mS) pH second – 2 pt cal (typically pH 7 and 10) should be about 180mV difference Turbidity and/or chlorophyll – 2 pt cal, see manual for correct protocol and standards. D.O. last, needs to equilibrate in H 2 O- saturated air. Most likely to drift over time.

Calibration tips ALWAYS RINSE between calibration standards with water and some old standard. AVOID contamination. Save $$$ by knowing your chemistry…BUT check self-made standards with a different instrument. pH sensors go bad – CHECK for a 180mV difference, if less than 165mV then replace.

D.O. issues Frequent membrane changes ensure good data! Become good at this procedure! No bubbles! No wrinkles! Membrane should be firm, but not Allow membrane to stabilize overnight for best results. Calibrate in water-saturated air, membrane must be dry.

D.O. tests Check D.O. sensor function by: –Disconnect power and wait 2 mins. Readings should start high and drop towards 100% saturation (see D.O. sat table for mg/L value at that temperature) –Sensor should read 100% in water-sat air, and drop to 30% or less in STILL water. Sensor/sonde must be in moving water for accurate measurement. At least 1 ft per sec (30cm per sec). Use stirrer attachment, or raise-lower sonde the necessary amount during measurement. During this period the D.O. reading will increase to a plateau, this is the correct reading. Should be within % of saturation value for that temperature (see D.O. sat table!) Keep sonde cool and wet (wrapped in a wet white towel) to easily check if sensor needs to be recalibrated.

Pre-deployment Calibrate at least 24hrs before deployment. Set up to log at your desired interval (usually every hour on the hour) Leave instrument logging in the water-filled cup in the lab. This is a pre-deployment dataset. Do this again after retrieval for a 24 hr post- deployment dataset. The two should be similar, if not, there were sensor problems that occurred during the deployment.

Lab Calibrate 2 YSI’s. Setup for unattended monitoring (hourly) Leave in lab overnight for pre-deployment data Deploy one at pier for 1 weeks.