1 User Guide KI 734x Series Two way Loss & ORL Tester www.kingfisher.com.au

2 Two-way Loss Tester The fastest and easiest two-way loss tester Real time display of bi-directional average multi- link loss on both instruments 3 year warranty & calibration cycle No warm up required 360 hours battery life & i nterchangeable connectors 12 00 storage space for dual test results

3 Fast & Easy For Bi-directional Testing Step 1: Set references Step 2: Press the Auto test button Real time bi-directional average multi- link loss display on both instruments

4 Step 1 - Set Reference At Both Ends Two-way portMeter port Test lead End A End B An optic fiber link

5 Source off Source Return Loss Power Meter Auto test - 52.73 dBm 1310 nm Source on Auto test Menu Abs / Rel Hold - 4.01 dBm 1310 nm - + -4.00 dBm Source on Auto test Menu Set Ref Abs / Rel Hold - 0.00 dB R 1310 nm - + -4.00 dBm Source off Source Return Loss Power Meter Auto test - 52.73 dBm 1310 nm Pushing AUTOTEST Pushing Abs / Rel Alternates between 1310 and 1550 nm if 1310 & 1550 nm lasers are installed Holding Set Ref for 3 seconds to set reference for multi- Pushing Menu to be back to the turn-on display Set Reference Key Press Sequence

6 An optic fiber link Step 2 - Measure Insertion And Return Loss End A End B Source off Source Return Loss Power Meter Auto test - 52.73 dBm 1310 nm Pushing Auto test at one End.

7 Source on Auto test Menu Hold - 34.01 1310 nm - + -4.00 dBm Alternates between 1310 and 1550 if 1310 & 1550 nm lasers are installed - + Source on Auto test Menu Hold - 54.17 - + -4.00 dBm 1310 nm Testing is completed for the link. A B dB A B dB RL STORE Pushing STORE to store average Loss, A-to-B Loss, B-to-A Loss, ORL at end A, ORL at end B of multi- Displays the average bi-directional link loss of multi- Displays Loss/Optical Power from A-to-B (A B) Displays Loss/Optical Power from B-to-A (A B) Displays ORL at A end (A ) Displays ORL at B end ( B) Back to the average bi-directional link loss display Press Select Press Menu Measure Insertion And Return Loss - Continue

8 End AEnd B An optic fiber link Communication During Testing Source on Auto test Menu Hold - 34.01 1310 nm - + -4.00 dBm A B dB After completing one link test, pushing Hold will freeze the display on both units and the unit at End B beeps twice. This tells End B that a test has been completed. Source on Auto test Menu Hold - 34.01 - + -4.00 dBm A B dB Then, End B connects to the next link and pushes Hold. This releases the display on both units and the unit at End A beeps once. This tells End A that he is ready for the next test. Increases the productivity in high fiber-count systems.

9 Set Ref In Autotest Source on Auto test Menu Set Ref Hold - 34.01 1310 nm - + A B dB Displays average link loss for multi-. Hold Set Ref for 3 seconds will set references on both units for multi-.

10 View Ref Value In Autotest Source on Auto test Menu Abs / Rel Hold - 34.01 1310 nm - + A B dB R Displays link loss and reference value for multi- for A-to-B. Source on Auto test Menu Abs / Rel Hold - 32.24 1310 nm - + A B dB R Displays link loss and reference value for multi- for B-to-A. -8.20 dBm -7.80 dBm Displays the average bi-directional link loss of multi- Displays Loss/Optical Power from A-to-B (A B) Displays Loss/Optical Power from B-to-A (A B) Displays ORL at A end (A ) Displays ORL at B end ( B) Back to the average bi-directional link loss display Press Select Press Menu - + Unit A reference value Unit B reference value

11 Pass/Fail Assessment In Two-way testing mode, press Level to enter Pass/Fail display mode. To exit press Level again, or press Select. Source on <<< Auto test Menu Hold PASS 1310 nm - + A Level Select

12 Insertion Loss Pass/Fail Threshold Set Up Multi- of Insertion Loss Pass/Fail criteria can be set. Shift Default Level Cancel Set Shift Source off SEt Menu 15.73 1310 nm - + dB Select another two-way operational wavelength to set up Average Insertion Loss Pass/Fail threshold. Enter Power Meter mode. Select two-way operational wavelength. Use Shift + Level to set Average Insertion Loss Pass/Fail threshold, push Set to store. - +

13 Return Loss Pass/Fail Threshold Set Up Multi- of Return Loss Pass/Fail criteria can be set. Shift Default Level Cancel Set Select another wavelength to set up Return Loss Pass/Fail threshold. Use Shift + Level to set Return Loss Pass/Fail threshold, push Set to store. - + Shift Source off SEt Menu -10.57 1310 nm - + dB RL Enter Return Loss mode. Select wavelength.

14 The Contents of User Memory Average loss for multi- Reference of both units for multi- ORL for multi- Nominal output power for multi- Location info by flashing symbol of either A or B