MPPC mass check Jan9/2007 Hidetoshi OTONO. Our plan for mass check with relays MPPC relayOn MPPC relayOFF MPPC relayOFF DC board AMPADC relayOFF 測定対象.

Similar presentations


Presentation on theme: "MPPC mass check Jan9/2007 Hidetoshi OTONO. Our plan for mass check with relays MPPC relayOn MPPC relayOFF MPPC relayOFF DC board AMPADC relayOFF 測定対象."— Presentation transcript:

1 MPPC mass check Jan9/2007 Hidetoshi OTONO

2 Our plan for mass check with relays MPPC relayOn MPPC relayOFF MPPC relayOFF DC board AMPADC relayOFF 測定対象 relayOFF Control PC PCI bus HV RS232C

3 PC System NFS Analysis PC Control PC DC board HV RS232C PCI bus RELAY CAMAC CC/NET

4 MPPC Relay MPPC Relay MPPC Relay HV Adapter DC board RS232C Control PC Dark Box CAMAC Menu de OTONO System - Version 4 Ch1 Ch2 ChN - Control HV - Turn ON/OFF relays. - Number of channel can be increased easily. AMP Charge ADC CCNET Ethernet x 63 Scaler x 10 Disc. G.G Disc. 6ns gate Clock Thermal measurement Analysis PC Output register veto AMP NIM Ethernet ADC input

5 1 p.e rate and 2 p.e rate measurement threshold MPPC out Disc. out ADC input Scaler input scaler rate = scaler counts / clock counts 1 p.e rate = scaler rate x {(whole counts – noise counts) / whole counts} 2 p.e rate = scaler rate x {(whole counts – noise counts – 1 p.e counts ) / whole count} noise 1 p.e 2 p.e ADC gate These counts are calculated from ADC histogram.

6 #12 Vop = 78.05 V Noise count 1 p.e count 2 p.e count 3 p.e count Noise 1 p.e 2 p.e 3 p.e 1 p.e rate = 57.4 kHz 2 p.e rate = 1.24 kHz Our measurements are about the same values at another University …..

7 #12 Vop +1.5V = 79.55V What are these ?? Anyway our measurements are about the same values at another University ….. We need more information of sample MPPCs for comparison.

8 -Relay OMRON G3VM-41LR10 -HV for MPPC KIKUSUI PAN110-1.5A (HV) PIA4810 (Adapter) OP01-PIA (Control Board) TU02-PIA (for HVreadback) -Amp HAMAMATSU C5594-44 -DC board CONTEC DO32-B(PCI)H -ADC LeCroy 2249A (Charge ADC) -Thermal measurement Agilent 34970A (with Pt100) Requirements for measurement