WBC and Differential measurement

Slides:



Advertisements
Similar presentations
What happens in the aperture?
Advertisements

Unicel DxH 800 Flags, Codes, & Messages
Slide 1, 4/12/2015 of DNA.ppt  Purdue University Cytometry Laboratories DNA1.ppt.
Erythrocyte (RBC) Stacking allows for passage through narrow blood vessels.
Lasers in medicine: blood and blood flow in the tissues Today’s talk... 1.Introduction to blood and blood flow 2.Laser Flow-Cytometry (LFC) of blood...
Buffy coat (leukocytes and platelets) - <1%
DxH 800 Flags, Codes, Messages
Electrical Impedance and Colorimetric Measurements Joanna Ellis, MLS(ASCP)
Automated Hematology Cell Counters
CHAPTER 37 MLAB 1415 HEMATOLOGY JOANNA ELLIS, MLS(ASCP) Optical Light Scatter and Flow Cytometry.
FLOW CYTOMETRY Dr. MOHAMMED H SAIEMA LDAHR KAAU FACULTY OF APPLIED MEDICAL SCIENCES MEDICAL TECHNOLOGY DEPT. 2 ND YEAR MT INSTROMINTATION EXT
FLOW CYTOMETRY Dr. MOHAMMED H SAIEMA LDAHR KAAU FACULTY OF APPLIED MEDICAL SCIENCES MEDICAL TECHNOLOGY DEPT. 2 ND YEAR MT INSTROMINTATION EXT
Automation in hematology Until Slide 23
-Automation blood count -Red and White blood count and differential count (Manual blood count) Prepared by: Mona alharbi Experiment:
Agile - affordable - accurate Optical System and Measurement Technology Presentation.
Applications of flow cytometry in basic immunology Generation and characterization of DC Assays for T cell activation –Cell proliferation – Cell division.
-Automation blood count -Red and White blood count and differential count (Manual blood count) Experiment:
Introduction to Flow Cytometry
A compact 5-diff solution
Agile - affordable - accurate. Hematology, is the part of internal medicine, that is concerned with the study of blood, the blood-forming organs, and.
Cell viability studies Sepideh Khoshnevis. The Goal To distinguish live cells from dead and apoptotic cells in order to calculate the the percentage of.
Analyticon Biotechnologies AG
Clinical Haematology & Cell Counting
Flow Cytometry at Boston University Medical Campus Introduction to some methods that we offer Yan Deng (X4-5225), Gerald Denis (X4-1371),
Introduction To Flow Cytometry By Noha Kamel. Flow cytometry is a method of measuring multiple physical and chemical characteristics of particles by optical.
Dr Gihan Gawish Hydrodynamic focusing is a technique used to provide more accurate results from flow cytometers or Coulter counters for determining the.
Basic Principles in Flow Cytometry
 Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid.
CLS 541-Advanced Medical Haematology M. sc. Program Lab-2-
Automation in haematology
Pictures/Diagrams Copyright © University of Nebraska
Chapter 5 Hematology. Figure 5.1 (a) A double ended needle partially pierces the rubber stopper in an evacuated tube. (b) After the left needle pierces.
Haematology.
Flow Cytometry Principles & practice of “Fluorescence Spectroscopy in Biological Diagnosis & Research” Dr.Hekmatimoghaddam Assistant professor of pathology.
Laser Flow Cytometry Forward Scatter indicates size Forward Scatter.
Flow Cytometry Basic Training. What Is Flow Cytometry? Flow ~ cells in motion Cyto ~ cell Metry ~ measure Measuring properties of cells while in a fluid.
1 Flow Cytometry in the Clinical Laboratory Patricia Aoun, M.D., M. P. H. Jean Bailey, MT-ASCP Kellie Neth, MT-ASCP The Nebraska Medical Center.
Lab Ex. 38, 39 & 40 Blood & Blood Testing. Blood cells.
Flow Cytometry Becton Dickinson Asia Limited Company.
Course title: Hematology (1) Course code : MLHE-201 Supervisor :Prof.Dr Magda Sultan Date : Outcome : The student will be able to use the automated.
Electron Microscope. How do they work Instead of using light they fire a beam of electrons (which have a wavelength less than 1nm compared to light which.
Sysmex XE-2100 Counting of Red Cells and Platelets by impedance & fluorescent techniques.
Part III Automation in Hematology and Hemostasis
HAEMATOLOGY LINE A.C.KAROBIA
Complete Blood Count (CBC)
Hematology Unit 2 Chapter 8 Automated Analyzers Copyright © 2015 by Mosby, an imprint of Elsevier Inc. All rights reserved.
Reticulocyte Hb equivalent and hypochromic red cells in the study of erythropoiesis in pregnancy Eloísa Urrechaga 1, Elia Crespo 1, Luís Borque 2, Jesús.
Automatic Electronic Blood Cell Counting
Medical Laboratory Instrumentation
Flowcytometry.
Flow Cytometry FACS (Fluorescence-Activated Cell Sorter)
اساس اندازه گیری در سل کانتر
FLOWCYTOMETRY.
PENTRA 60 C +.
Patient A.
Flow cytometry of wild‐type and ΔseqA cells.
Flow Cytometry and Sorting Part 3
Fluorescence Flow Cytometry in Haematology
Sysmex XE-2100 NRBC Detection.
Reticulocyte Detection
Optical measurement.
Volume 8, Issue 4, Pages (April 2017)
Differential expression of functional chemokine receptors on human blood and lung group 2 innate lymphoid cells  Cathryn A. Weston, PhD, Batika M.J. Rana,
Flow Cell Injector Tip Fluorescence signals Focused laser beam Sheath
Volume 8, Issue 4, Pages (April 2017)
Complete Blood Count (CBC)
Blood Tests.
Lab Ex. 38, 39 & 40 Blood & Blood Testing
Flow Cell Injector Tip Fluorescence signals Focused laser beam Sheath
Presentation transcript:

WBC and Differential measurement The XE-2100 WBC and Differential measurement

Principles of cell detection Wbc and differential counts are determined using: Flow Cytometry Method Semiconductor Laser. Blood cell information is obtained using: Forward light scatter Lateral light scatter Lateral fluorescent light

Optics

Specific Measurements

The Stain • Staining Reagent: – Polymethine Dye (Group of staining reagents): X = O, S, Se, N-alkyl or C(Ch 3 ) n R 1,2 = alkyl (1-6C) ; R 3,4 = Alkyl, Methoxy or NO n = 0-1

The Wavelength The new stain opens up the possibility of using a different laser wavelength on a compact laser system with little need of increases in power  = 633nm Red laser beam

Methodologies

WBC/BASO Channel Blood sample is aspirated, diluted to a ratio of 1:50 with STOMATOLYSER-FB. Diluted sample is sent to WBC/BASO detection chamber. In this channel, forward and side scatter signals are used to derive the WBC and BASO count.

Measurements Forward Scatter - Measurement of size Side Scatter - Measurement of internal structure i.e size of nucleus

The WBC/BASO Channel Lysing Surfactant BASO Other WBC RBC

WBC/BASO Scattergram

WBC Classification Blood sample is aspirated and diluted with: STOMATOLYSER-4DL and STOMATOLYSER -4DS. Diluted sample is sent to the detector chamber Cells are classified according to side scatter and fluorescence signals in the Diff channel.

Measurements Side scatter - measurement of internal structure. Fluorescence - measurement of nucleic acid content.

The DIFF Channel Abnormal Lymph/ Lymphoblast 1.Lysing 2.Staining High Fluorescence Intensity High Fluorescence Low Fluorescence Blast, I.G. Normal Cell STROM-4DL STROM-4DS Lysing surfactant Polymethine Dye +Organic Acid

Diff Scattergram

Neutrophil Count This is a calculated parameter # Neut = Neut/Baso count(diff channel) - Baso count (WBC/Baso channel)

IMI Channel discriminates between immature and mature white blood cells. Detection is performed using RF/DC detection method.

Methodology Blood is aspirated and diluted to a ratio of 1:250 with STOMATOLYSER-IM. Diluted sample is sent to the IMI detector for analysis. A two-dimensional distribution of cell size and cell structure is drawn according to RF/DC detection method.

Measurements Direct current - measurement of cell volume Radio frequency - measurement of internal cell structure.

IMI Scattergram Platelet Clumps

ACAS & Advanced Discrimination

Adaptive Cluster Analysis System Cells are classified by the process of cluster analysis which is based on learning algorithms. (An astronomical technique!) This allows optimal adaptation to biological differences between samples including abnormal cells

Advanced Discrimination DIFF-CH WBC/BASO-CH 150+-1 Gst/NLME Bas/NLME NLME:WBC except BASO Ly/Ne2 Ly/Mo2 150+-1 Mo/Eo Ly/Mo1 Mo/NE Ly/Ne1 Ne/Eo Ly/Ba Ly/Gst Ba/Gst Ne/Gst

Extended ACAS Principle High adaptation to individual changes Moving clusters Moving discriminators Enhanced separation in abnormal cases Low “Vote Out” rate