1 Sysmex XE-2100 NRBC Detection

2 NRBC Counts - Methodology
From semiconductor laser flow cytometry. STROMATOLYSER-NR (diluent) STROMATOLYSER-NR dye (stain) Passed through the laser flow cytometer for analysis. Counted using side fluorescence and forward light scatter.

3 Flow Cytometry Forward light scatter - cell size
Side fluorescence - RNA/DNA information (nucleus)

4 NRBC Channel Explained

5 Therefore the NRBC Scattergram…

6 Resolution of NRBC and WBC
Clear & reliable separation of NRBC and WBC Wbc take upp more dye than NRBC - so higher fluorescence Smaller volume of the NRBC nucleus gives reduced forward light scatter.

7 CORRECTING WBC COUNT

8 The NRBC Count Algorithm
NRBC% = NRBC# x 100 NRBC# + WBC# (NRBC/ 100 WBC) WBC# NRBC#

9 The Corrected WBC Count
Eo Ba Mo Ne Ly NRBC NRBC

10 NRBC # Revised Algorithm
WBC region in WBC/Baso and Diff scattergram include NRBC if NRBC appear

11 WBC & DIFF channels: Correction for NRBC’s
Correct WBC Minus Correct WBC Differential Minus

12 NRBC Flags Swiching Algorithm
NRBC-CH is measured “NRBC present” flag is monitored (WBC abnormal flag category) Detection from “NRBC%” NRBC-CH is not measured “NRBC?” flag is monitored (WBC suspect flag category) Detection from DIFF scattergram