Presentation is loading. Please wait.

Presentation is loading. Please wait.

Part III Automation in Hematology and Hemostasis

Similar presentations


Presentation on theme: "Part III Automation in Hematology and Hemostasis"— Presentation transcript:

1 Part III Automation in Hematology and Hemostasis

2 Automation in Hematology and Hemostasis
AUTOMATED BLOOD CELL-COUNTING INSTRUMENTS O before automation - manual hemacytometer blood cell counts - spun hematocrits - spectrophotometrically determined hemoglobins - microscopic blood smear evaluations O using automated instrument - complete blood cell count (CBC) - platelet count - a five-part leukocyte differential - absolute reticulocyte count reticulocyte hemoglobin concentration (CHr or Ret He): -> assessment of iron incorporation into hemoglobin, functional availability of iron reticulocyte mean cell volume (MCVr) immature reticulocyte fraction (IRF) -> indicator of an erythropoietic response

3 Automation in Hematology and Hemostasis
O two principles of blood cell counting 1. impedance (electrical resistance) - based on the increased resistance - the number of pulses indicate the blood cell count - amplitude (height) of each pulse: proportional to the volume (size) of the cell 2. optical light scattering - based on light scattering measurements - forward scatter: cell size - side scatter: cell complexity or granularity

4 Automation in Hematology and Hemostasis
IMPEDANCE INSTRUMENTS Coulter Gen/S O by Beckman-Coulter O determine: CBC, five-part leukocyte differential, reticulocyte count O basic principle of operation - using EDTA anticoagulated blood sample, - four chamber 1. RBC/platelet dilution chamber - external electrode, three apertures, internal electrode (Fig. 36-1) - when blood cells pass through an aperture -> increase in resistance - threshold limit: for enumerating cells based on cell volume more than 36 fL -> erythrocyte 2-20 fL -> platelet the average of the counts from the three apertures

5 Automation in Hematology and Hemostasis

6 Automation in Hematology and Hemostasis
- RBC histogram (size distribution curve) (Fig. 36-2) visualization of changes in size within the erythrocyte population shift to the left: microcyte shift to the right: macrocyte - platelet histogram (Fig. 36-4) mean platelet volume (MPV), platelet distribution width (PDW) - obtainable data RBC count, MCV: by direct measurement RDW: derived from the RBC histogram (Fig. 36-3) Hct (MCV X RBC count): by calculated MCH (hemoglobin conc. / RBC count): by calculated MCHC (hemoglobin conc. / Hct): by calculated * hemoglobin concentration: obtained from the WBC/hemoglobin dilution chamber - platelet count: mean platelet volume(MPV), platelet distribution width(PDW)

7 Automation in Hematology and Hemostasis

8 Automation in Hematology and Hemostasis

9 Automation in Hematology and Hemostasis

10 Automation in Hematology and Hemostasis
2. WBC/hemoglobin dilution chamber - determination of WBC count, hemoglobin concentration - WBC count: add lytic agent, directly measured lytic agent: lyse the erythrocytes, convert released hemoglobin to cyanmethemoglobin shrink the leukocyte cell membrane and cytoplasm - > 35fL -> leukocyte - three aperture - leukocyte histogram (Fig. 36-5) visualization of subpopulation of cells based on their relative sizes. lymphocytes: fL monocytes: fL granulocyte: fL - system of region flags (Table 36-1) - hemoglobin concentration: absorbance of cyanmethemoglobin at 525 nm

11 Automation in Hematology and Hemostasis

12 Automation in Hematology and Hemostasis

13 Automation in Hematology and Hemostasis
3. orbital mixing chamber - blood is mixed by gentle agitation with heated lysing reagent to remove the erythrocytes while leaving the leukocytes in their near native state - sent to the volume.conductivity.scatter (VCS) flow cell for determination of the five-part leukocyte differential volume: by impedance conductivity, scatter: evaluates internal physical and chemical constituents - contour gating (Fig. 36-6)

14 Automation in Hematology and Hemostasis

15 Automation in Hematology and Hemostasis
4. heated reticulocyte dilution chamber - mixed with new methylene blue reagent - stained sample is mixed with an acidic, hypotonic solution elutes hemoglobin from the erythrocytes but permit precipitated RNA to remain spheres the erythrocytes (ligth scatter data -> more reproducible data) - sent to VCS flow cell classify reticulocytes vs mature erythrocytes - data absolute reticulocyte count, reticulocyte percentage, immature reticulocyte fraction reticulocyte mean cell volume (MCVr) O data compiling - corrects coincidence (two or more cells passing through the aperture at the same time) - Fig software-generated flag or user-defined flag (definitive flag)

16 Automation in Hematology and Hemostasis

17 Automation in Hematology and Hemostasis
Coulter LH Series O provide a CBC, five-part leukocyte differential, reticulocyte count Wright-stained peripheral smear O utilize basic principle of the Gen.S instrument O additional parameter 1. nucleated red blood cell count (NRBC%), corrected WBC count (Fig. 36-8) - WBC scatterplot: detection of cells between the lymphocyte population and RBC ghosts - WBC histogram: the presence of cells to the left of the lymphocyte population - role out giant platelets or small lymphocytes - NRBC%: the number of nucleated red blood cells per 100 leukocytes - NRBC# = NRBC% X WBC count - NRBC: increase -> WBC count: falsely increase -> corrected WBC count

18 Automation in Hematology and Hemostasis

19 Automation in Hematology and Hemostasis
2. RDW-SD - Fig. 36-9 - reflects the degree of anisocytosis - MCV does not affect the RDW-SD -> better indicator - reported in fL, reference interval: fL - used in the differentiating iron deficiency anemia from heterozygous beta thalassemia - Fig

20 Automation in Hematology and Hemostasis

21 Automation in Hematology and Hemostasis
Sysmex XE-2100 O impedence, radio frequency, absorption spectrophotometry, flow cytometry -> CBC, 5-part leukocyte differential, reticulocyte count O hemoglobin: measured using the sodium lauryl sulfate (SLS) method - lyse erythrocytes - convert ferrous iron to ferric iron (methemoglobin) by SLS - methemoglobin: combine with SLS -> SLS-hemichrome molecule - at 555 nm - advantage: decrease measurement time, reagent is cyanide-free O in the RBC/PLT channel: RBC, platelet count - using impedance with hydrodynamic focusing - MCV, MCH, MCHC, RDW-SD, RDW-CV, MPV

22 Automation in Hematology and Hemostasis
O in the DIFF channel: leukocyte differential - hole in the cytoplasmic membrane of nucleated cells allowing a polymethine dye to enter the cell and bind to its DNA and RNA - forward scatter (cell size), side scatter (cell complexity) fluorescent intensity - Fig immature granulocyte (IG) count: metamyelocyte, myelocyte, promyelocyte O in the WBC/BASO channel - lyse erythrocytes - shear all leukocytes except basophils - Fig O in the IMI channel - mature leukocytes with phospholipid-rich membranes are completely lysed - immature myeloid cells (blasts) remain intact - provide the hematopoietic progenitor cell (HPC)

23 Automation in Hematology and Hemostasis

24 Automation in Hematology and Hemostasis

25 Automation in Hematology and Hemostasis
O in the NRBC channel - determine the presence of nucleated erythrocytes - NRBC and lymphocytes fall in the same cluster on the DIFF scattergram - remove the NRBC cytoplasmic membrane, shrink the nucleus - perforates only WBC cytoplasmic membrane, shape is not altered - fluorescent dye binds to nucleus and intracytoplasmic organelle - leukocyte: high fluorescent intensity and high forward scatter - Fig O in the RET channel - oxazine: binds to residual RNA within the reticulocytes - polymethine: binds to RNA and DNA within nucleated cells - differentiating mature erythrocytes, reticulocytes, platelets (Fig ) - immature platelet fraction (IPF): reticulated platelets increased size, higher fluorescence (Fig ) useful in diagnosis of thrombocytopenia associated with increased platelet destruction versus bone marrow failure - Fig

26 Automation in Hematology and Hemostasis

27 Automation in Hematology and Hemostasis

28 Automation in Hematology and Hemostasis

29 Automation in Hematology and Hemostasis
Abbott CELL-DYN Sapphire O for CBC, a five-part leukocyte differential, reticulocyte O flow cytometry, fluorescence staining, impedance O in the hemoglobin dilution cup - lyse erythrocytes - convert hemoglobin to a single chromogen by forming a complex with imidazole cyanide-free reaction - determined spectrophotometrically at 540 nm

30 Automation in Hematology and Hemostasis
O in the WBC dilution cup - total WBC count, nucleated RBC, five-part leukocyte differential - flow cytometry (light scattering, fluorescence (for NRBC) - using multi-angle polarized scatter separation (MAPSS) (Fig ) 0 degree light scatter (forward scatter): cell size 7 degree light scatter: cell complexity 90 degree light scatter (side scatter): nuclear lobularity 90 degree depolarized light scatter: cytoplasmic granularity - 0 degree (size) versus 7 degree (complexity): differentiating of neutrophil, monocyte, lymphocyte (Fig ) 90 degree D (granularity) vs 90 degree (lobularity) differentiating of eosinophil, neutrophil - NRBC is clearly distinguished from leukocytes strip the cytoplasmic membrane from nucleated erythrocytes

31 Automation in Hematology and Hemostasis

32 Automation in Hematology and Hemostasis

33 Automation in Hematology and Hemostasis
O in the RBC/PLT dilution cup - erythrocyte/platelet dilution sphere the erythrocytes, RBC, platelet count by impedance, light scattering - reticulocyte dilution fluorescence (RNA), 7 degree light scatter (cell complexity) (Fig ) O Fig O fluorescence immunophenotyping method O immuno T-cell assay - using anti-CD3 (T cell), CD4 (T helper cell), CD8 (T cytotoxic cell) - scatter plot of 0 degree vs 7 degree O immuno PLT (CD61) assay - using anti-CD61 - when platelet count is very low, leukocyte or erythrocyte fragments interfere

34 Automation in Hematology and Hemostasis

35 Automation in Hematology and Hemostasis

36 Automation in Hematology and Hemostasis
LIGHT-SCATTERING INSTRUMENTS Siemens Healthcare ADVIA 120 O CBC, five-part leukocyte differential, reticulocyte count O five measurement channel (1) the erythrocyte/platelet channel - RBC, platelet count - isovolumetric sphering (sphere and fix the blood cells without altering their volume) of the erythrocytes and platelets -> eliminate cell volume errors due to variation in erythrocyte shape - counted and sized by high-angle (5-15): hemoglobin concentration low-angle (2-3): cell volume (size) (2) the hemoglobin channel - lyse erythrocyte - converted to cyanmethemoglobin

37 Automation in Hematology and Hemostasis
(3) the peroxidase channel - identify neutrophil, monocyte, eosinophil by the degree of peroxidase positivity and forward light scatter - lymphocyte and large unstained cell (LUC): forward light scatter - neutrophil: upper right quadrant eosinophil: lower right quadrant monocyte: the cental triangle region lymphocyte: adjacent to the y-axis in the center left quadrant LUC: upper left quadrant (4) the basophil/lobularity channel - lyse erythrocyte and platelet - strips all leukocytes except basophils of their cytoplasm - determined by high and low-angle scatter - each population is identified by its position, area, and density - normal: worm type mononuclear cell: head, neutrophil: tail, basophil: large, above the worm

38 Automation in Hematology and Hemostasis
(5) the reticulocyte channel - stained with oxazine light scattering: high-angle light -> hemoglobin concentration low-angle light -> size - absorbance: RNA content O no reporting NRBC count O Fig Siemens Medical ADVIA 2120 O hemoglobin determination: cyanide-free methodology

39 Automation in Hematology and Hemostasis

40 Automation in Hematology and Hemostasis
AUTOMATED COAGULATION INSTRUMENTS O based on electromechanical or optical density methods of clot detection O plus chromogenic and immunologic methods ELECTROMECHANICAL INSTRUMENTS BBL FibroSystem O detect completion of an electrical circuit between two electrodes when a clot forms O fibrometer probe - clot detector - two sensory electrodes (stationary and moving) - fibrin clot formation -> creation of electric circuit: from stationary electrode through the reaction mixture to fibrin clot to moving electrode

41 Automation in Hematology and Hemostasis
Diagnostica Stago STArt 4 Clot Detection Instrument O based on the increasing viscosity of plasma as clot formation occurs O detect decrease in movement of an iron ball in an electromagnetic field when a clot forms - increasing viscosity is detected by the movement of an iron ball located at the bottom of the reaction cuvette - increase viscosity -> decrease in ball movement OPTICAL DENSITY INSTRUMENTS Trinity Biotech Coag-A-Mate XM O detect sudden increase in optical density(absorbance) occurring with fibrin formation

42 Automation in Hematology and Hemostasis
CHROMOGENIC/CLOT DETECTION INSTRUMENTS O combine clot-based detection principle with chromogenic analysis and immunological techniques O the enzyme of interest cleaves the chromogenic substrate at a specific site, releasing the chromophore tag Trinity Biotech AMAX Destiny Plus O electromechanical, optical density, chromogenic, and immunologic O electromechanical measurement of clot formation - as the fibrin clot forms, fibrin strands draw the ball away from its steady-state position - accurate results are obtained on icteric samples, lipemic samples, and samples from patients on medication that affect optical measurement

43 Automation in Hematology and Hemostasis
O optical density - monitor the change in absorbance over time at 405 nm - the effects of icteria and lipemia on the clotting time is minimized O chromogenic assay - paranitroaniline (chromophore tag) at 405 nm O immunologic assay - detect and quantitate the presence of certain coagulation proteins or products - D-dimer assay detect the increase in absorbance that occurs as immune complex form between the sample's D-dimers and the monoclonal antibodies, thus aggregating the microparticles

44 Automation in Hematology and Hemostasis
Diagnostica Stago STA Compact O electromechanical principle, chromogenic and immunological assay O electromechanical clotting assay - as the fibrin clot forms, the increasing viscosity of the reaction mixture -> decrease the ball's oscillation - clotting time are not affected by lipemic or icteric samples because the determination does not depend on light absorbance O chromogenic assay - chromogen: paranitroaniline O immunologic assay - measure the concentration of von Willebrand factor - microlatex beads coated with rabbit anti human vWF to aggregate or clump -> increasing the absorbance


Download ppt "Part III Automation in Hematology and Hemostasis"

Similar presentations


Ads by Google