Performances of the COROT CCDs for high accuracy photometry Pernelle Bernardi and the CCD team From Meudon : Tristan Buey, Vincent Lapeyrere, Régis Schmidt,

Slides:



Advertisements
Similar presentations
HgCdTe Avalanche Photodiode Arrays for Wavefront Sensing and Interferometry Applications Ian Baker* and Gert Finger** *SELEX Sensors and Airborne Systems.
Advertisements

Results of Acceptance Tests of Hyper Suprime-Cam CCDs Yukiko Kamata, Hidehiko Nakaya, Satoshi Miyazaki National Astronomical Observatory of Japan.
Kamera CCD Astronomi (2) AS3100 Lab. Astronomi Dasar I Prodi Astronomi 2007/2008 B. Dermawan.
Parameters to choose the CCD The CCD test bench *Temperature range : -55 to +40°C. *Stabilization : < 0.05°C/hour. *5 temperature probes : CCD and electronics.
1 M. Auvergne. Natal October 2004 Instrument performances. Signal perturbations: Radiations. Scattered light. ACS. Temperature. Readout electronic. Calibrations.
CHARGE COUPLING TRUE CDS PIXEL PROCESSING True CDS CMOS pixel noise data 2.8 e- CMOS photon transfer.
Astronomical Detectors
1 CCD RAIN (PHOTONS) BUCKETS (PIXELS) VERTICAL CONVEYOR BELTS (CCD COLUMNS) HORIZONTAL CONVEYOR BELT ( SERIAL REGISTER ) MEASURING CYLINDER (OUTPUT AMPLIFIER)
Semiconductor Technology Associates Siliconus Maximus SDW2005 Scientific Detector Workshop June 19 – 25, 2005 Taormina, Italy Richard A Bredthauer Semiconductor.
SDW20051 Vincent Lapeyrère LESIA – Observatoire de Paris Calibration of flight model CCDs for CoRoT mission.
Barry E. Burke Title Slide.
From CCD to EMCCD Scientific imaging for today’s microscopy.
SDW2005, juin, Taormina The Corot Space instrument.
Charge-Coupled Device (CCD)
Digital Technology 14.2 Data capture; Digital imaging using charge-coupled devices (CCDs)
CCD testing Enver Alagoz 12 April CCD testing goals CCD testing is to learn how to – do dark noise characterization – do gain measurements – determine.
Copyright : Marconi Applied Technologies Limited. All Rights Reserved. CCDs for Adaptive Optics CfAO Workshop: Nov 1999 By P.
Different sources of noise in EM-CCD cameras
CCD Clocking and Biasing CABAC_0 : Design Test_0 : Design Pierre Antilogus (from a Hervé Lebbolo’s talk) BNL, Raft Electronic Workshop January 25 th 2012.
1 COROT: Scientific comity Sept Calibrations during the commissioning and before the first run. Calibrations cover closed Calibrations cover open.
Eddington Kick-Off. Vienna, September 17th, 2001 T.Muñoz/C.Laviada (INTA) 1 EddiCam: The Eddington Photometric Camera Preliminary Design Layout.
1st Eddington Workshop. Córdoba, June 14th, 2001 J. Miguel Mas-Hesse 1 EddiCam: The Eddington Photometric Camera Preliminary design.
SIAMOIS, kickoff meeting, Paris, may, SIAMOIS: optical test bench BOSS (Banc Optique SiamoiS) Pernelle Bernardi Observatoire de Paris-Meudon,
AST3 detector properties
 Johann Kolb, Norbert Hubin  Mark Downing, Olaf Iwert, Dietrich Baade Simulation results:  Richard Clare Detectors for LGS WF sensing on the E-ELT 1AO.
Bulk Silicon CCDs, Point Spread Functions, and Photon Transfer Curves: CCD Testing Activities at ESO Mark Downing, Dietrich Baade, Sebastian Deiries, (ESO/Instrumentation.
Photon detection Visible or near-visible wavelengths
Advanced Concepts & Science Payloads Office Eddicam/EST MeetingPage 1 CCD Procurement Schedule driven Review off-the shelf availability Specific mode of.
Super-hot pixels, hot pixels and DSNU on Hawaii-2RG detector
1 COROT WEAEK Meudon May The CCD performances by Miss Pernelle Bernardi Mr Tristan Buey and the CCD team on stage Régis Schmidt, Bertrand le.
1 Imaging Techniques for Flow and Motion Measurement Lecture 5 Lichuan Gui University of Mississippi 2011 Imaging & Recording Techniques.
STATUS REPORT OF FPC SPICA Task Force Meeting March 29, 2010 MATSUMOTO, Toshio (SNU)
NASSP Masters 5003F - Computational Astronomy Lecture 19 EPIC background Event lists and selection The RGA Calibration quantities Exposure calculations.
CMOS Image Sensor Design. M. Wäny Nov EMVA Standard 1288 Standard for Measurement and Presentation of Specifications for Machine Vision Sensors and.
CCD Detectors CCD=“charge coupled device” Readout method:
LCFI Collaboration Status Report LCUK Meeting Oxford, 29/1/2004 Joel Goldstein for the LCFI Collaboration Bristol, Lancaster, Liverpool, Oxford, QMUL,
Testing and characterizing the CCDs of the OmegaCam wide-angle camera F. Christen 2, C. Cavadore 1, B. Gaillard, O. Iwert 1, K. Kuijken 2, D. Baade 1,
Eusoballoon optics test Baptiste Mot, Gilles Roudil, Camille Catalano, Peter von Ballmoos Test configuration Calibration of the light beam Exploration.
10/26/20151 Observational Astrophysics I Astronomical detectors Kitchin pp
Plato meeting MSSL Wed+Thur 15+16/Oct/08 UK interests/activities Alan Smith and Dave Walton UCL/MSSL.
Fully depleted MAPS: Pegasus and MIMOSA 33 Maciej Kachel, Wojciech Duliński PICSEL group, IPHC Strasbourg 1 For low energy X-ray applications.
1 Leonardo Pinheiro da Silva Corot-Brazil Workshop – October 31, 2004 Corot Instrument Characterization based on in-flight collected data Leonardo Pinheiro.
SNAP CCD Development Progress Hakeem Oluseyi January 9, 2003.
Astronomical Institute University of Bern 31th IADC Meeting, April , 2013, ESOC, Darmstadt, Germany Improved Space Object Observation Techniques.
Observational Astrophysics I
CorotWeek 3, Liège 4-7/12/20021 (working on Saturday!!!) The CCD flight models Miss Pernelle Bernardi Mr Vincent Lapeyrere Mr Tristan Buey and the CCD.
MPPC status M.Taguchi(kyoto) T2K ND /7/7.
TIPS - Oct 13, 2005 M. Sirianni Temperature change for ACS CCDs: initial study on scientific performance M. Sirianni, T. Wheeler, C.Cox, M. Mutchler, A.
Fig. 1: Cross section of a circular DEPMOS- FET pixel cell. Charges collected in the “in- ternal gate’ modulate the transistor current. DEPMOSFET team,
PPLC – ESA FEB 27th 2009 CCD thermal specs G Epstein CCD thermal specifications Divers for Telescope thermal architecture.
Multipixel Geiger mode photo-sensors (MRS APD’s) Yury Kudenko ISS meeting, KEK, 25 January 2006 INR, Moscow.
Basic Detector Measurements: Photon Transfer Curve, Read Noise, Dark Current, Intrapixel Capacitance, Nonlinearity, Reference Pixels MR – May 19, 2014.
ICCD of HERD Le WANG, XIOPM , XI’AN The 3 rd HERD Workshop.
` Background photo courtesy of NOAO. Taken with LBNL p-channel CCD with extended red sensitivity. Pixel Area Variation in CCDs and Implications for Precision.
General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays.
February 21, 2002TIPS meeting1 "Data contained herein is exempt from ITAR regulations under CFR 125.4(13) -- data approved for public disclosure." TIPS.
The Field Camera Unit Results from technical meeting S. Scuderi INAF – Catania.
1 Performance of a CCD tracker at room temperature T. Tsukamoto (Saga Univ.) T. Kuniya, H. Watanabe (Saga Univ.); A. Miyamoto, Y. Sugimoto (KEK); S. Takahashi,
Cameras For Microscopy Ryan McGorty 26 March 2013.
# x pixels Geometry # Detector elements Detector Element Sizes Array Size Detector Element Sizes # Detector elements Pictorial diagram showing detector.
Comparison of a CCD and the Vanilla CMOS APS for Soft X-ray Diffraction Graeme Stewart a, R. Bates a, A. Blue a, A. Clark c, S. Dhesi b, D. Maneuski a,
Miss Pernelle Bernardi
X-ray CCD with low noise charge injection.
Charge Transfer Efficiency of Charge Coupled Device
First Testbeam results
TEMPO Instrument Update
LSST Camera Detector Status
Fabio de Oliveira Fialho Michel Auvergne
Karen Meech Institute for Astronomy TOPS 2003
MAIN CONTRIBUTOR NOISES
Presentation transcript:

Performances of the COROT CCDs for high accuracy photometry Pernelle Bernardi and the CCD team From Meudon : Tristan Buey, Vincent Lapeyrere, Régis Schmidt, Bertrand le Ruyet, Jêrome Parisot, Didier Tiphène From CNES : Olivier Gilard, Guy Rolland (for irradiation tests) 2 nd Eddington Worskshop, Mondello (Italy), 9-11 April 2003

2 nd Eddington Workshop, Mondello, 9-11 April Corot CCD E2V Back-thinned, MPP, anti-reflection coating Operating in frame transfer mode Size of image area : 2048 rows, 2048 columns Pixel pitch: 13.5µm 2 outputs, dump drain register Readout frequency: 100kHz Readout timing: 100µs to transfer a row from the image area to the storage area 150µs to transfer a row in the register Three sides buttable  No specific technological developments  Space qualification

2 nd Eddington Workshop, Mondello, 9-11 April Geometrical requirements and packaging Flatness of image section:  3µm Top of view: 14mm  15µm (spec)  10 µm (meas) e(Invar) = 10 mm e(silicon) ~ 625 µm e(thinned silicon) ~ 15 µm 3 shims silicon invar thinned silicon 50µm min (spec) 500µm max 300µm max With this packaging: distance between image areas of 2 CCDs > 1mm

2 nd Eddington Workshop, Mondello, 9-11 April Working point 3 bias voltages are optimised: V OD, V RD, V OG Working range: measurement of the video signal for different values of the bias voltages  5 CCDs have the same working ranges for the 3 bias voltages  1 has a different behaviour: excluded from the flight Sensitivity of the video signal to the bias voltages :  About same order at worst frequency (100kHz, readout frequency).  Higher sensitivities = 4e-/mV  Specification for the electronics: 1mV peak to peak ---> equivalent noise of few e -.

2 nd Eddington Workshop, Mondello, 9-11 April CCD gain (µV/e-) Specification: 3.8 < G CCD < 6.0 µV/e- CCD gain is measured independently of Qe Good homogeneity of 5 CCDs Global gain (e-/ADU) measured at different temperatures from –45°C to 20°C Temperature coefficient of the gain: ~ -900ppm/K CCD Gain at –40°C Temp coefficient of the CCD gain (ppm/K) Left outputRight output

2 nd Eddington Workshop, Mondello, 9-11 April Quantum Efficiency E2V data: Qe max = 0.90 at 650nm average Qe = 0.62 Temperature coefficient of Qe: –Measure of the flux at: - different wavelengths from 400nm to 1000nm - different temperatures from –45°C to –30°C –Calculation of the temperature coefficient of the CCD response –Correction by the temperature coefficient of the CCD gain (  G ~ -900ppm/K)  Temperature coefficient of Qe ~ 2000ppm/K for the center of the bandwith

2 nd Eddington Workshop, Mondello, 9-11 April Dark current CCD is divided into windows of 32*32 pixels For 5 CCDs: at –40°C –Mean dark current value < 0.5e-/px/s –90% of the windows < 0.5e-/px/s Map of dark current Mean dark current = 0.18e-/px/s

2 nd Eddington Workshop, Mondello, 9-11 April Pixel Response Non Uniformity (1/2) = 420nm,  = 10nm = 700nm,  = 10nm = 900nm,  = 10nm Achieved with flat illumination Shows the uniformity of the state of surface and AR coating  Surface Pattern Shows the uniformity of the physical characteristics  High homogeneity Shows the uniformity of the thickness  Fringing Pattern = 900nm  10nm = 900nm  100nm

2 nd Eddington Workshop, Mondello, 9-11 April Pixel Response Non Uniformity (2/2) CCD is divided into windows of 32*32 pixels. Local PRNU = dispersion of the pixel response in a 32*32 pixel window. For 5 CCDs: local PRNU is 5 times lower than specification. Stable with temperature

2 nd Eddington Workshop, Mondello, 9-11 April Cosmetics Definitions: –White spot: > 100e-/px/s at –40°C –Black spot: when pixel response is less than 50% of the local mean response, at -40°C. –Column defect: more than 100 contiguous blemish elements (white and black spots) Specifications: –Less than 750 black and white spots –Less than 6 column defects Measurements: specifications are fullfilled –5 CCDs have no white spots, 1 CCD has 2 white spots –The worst CCD has 2 columns defects (black columns)

2 nd Eddington Workshop, Mondello, 9-11 April Maps of defects 2 maps calculated per CCD: one for each scientific program. –Images of dark current and flat fields in three parts of the bandwith –2 shapes of PSF (seismo and exoplanets) –Performances of the instrument for each program Seismology channel Exoplanet channel Same CCD

2 nd Eddington Workshop, Mondello, 9-11 April Results of the irradiation tests 3 chips tested (4210) Tested characteristics : –Working point. –Gain and Full Well Capacity. –Dark Current (mean value, defects). –Pixel Response Non Uniformity. Irradiation induces local defects (spatial and temporal) in the dark current Global evolution Beginning of life: < 0.5e-/px/s  End of life: 3 to 10e-/px/s Rapid evolution of the dark current with temperature gives strong constraint on the CCD operational temperature  below -40°C.  Appart from the dark current all the other characteristics of the CCD will not be degraded enough to impact on the system performances.

2 nd Eddington Workshop, Mondello, 9-11 April Conclusion Sixth CCD currently under tests Calibration of 10 CCDs FM will be finished before summer Then: sort the 10 CCDs select 4 CCDs for the flight using software tools If necessary, some specific measurements can be done on selected CCDs