POINT-OF-CARE TESTING and the CLIA SURVEY

Slides:



Advertisements
Similar presentations
Numbers Treasure Hunt Following each question, click on the answer. If correct, the next page will load with a graphic first – these can be used to check.
Advertisements

Basic Principles of GMP
Scenario: EOT/EOT-R/COT Resident admitted March 10th Admitted for PT and OT following knee replacement for patient with CHF, COPD, shortness of breath.
Variations of the Turing Machine
AP STUDY SESSION 2.
1
& dding ubtracting ractions.
Slide 1Fig 39-CO, p Slide 2Fig 39-1, p.1246.
Select from the most commonly used minutes below.
STATISTICS HYPOTHESES TEST (I)
STATISTICS INTERVAL ESTIMATION Professor Ke-Sheng Cheng Department of Bioenvironmental Systems Engineering National Taiwan University.
CAP/POCT 2003 Carol Riley-Hunte, RRT Senior Education Specialist Bayer HealthCare.
RECORD KEEPING Cooperative Development of Operational
David Burdett May 11, 2004 Package Binding for WS CDL.
LAW 11 Offside.
Prepared by: Workforce Enterprise Services For: The Illinois Department of Commerce and Economic Opportunity Bureau of Workforce Development ENTRY OF EMPLOYER.
Local Customization Chapter 2. Local Customization 2-2 Objectives Customization Considerations Types of Data Elements Location for Locally Defined Data.
Create an Application Title 1Y - Youth Chapter 5.
Process a Customer Chapter 2. Process a Customer 2-2 Objectives Understand what defines a Customer Learn how to check for an existing Customer Learn how.
Welcome PC-ACE Pro32 Training
CALENDAR.
1 Click here to End Presentation Software: Installation and Updates Internet Download CD release NACIS Updates.
Site Safety Plans PFN ME 35B.
Engagement in Human Research & Multi-Site Studies K. Lynn Cates, M.D. Assistant Chief Research & Development Officer Director, PRIDE May 30, 2012.
Regulatory Reform (Fire Safety) Order 2005 Fire Safety Order What is the Fire Safety Order? The order was made, under the Regulatory Reform Act 2001.
Welcome. © 2008 ADP, Inc. 2 Overview A Look at the Web Site Question and Answer Session Agenda.
Student & Work Study Employment Facts & Time Card Training
Break Time Remaining 10:00.
Turing Machines.
PP Test Review Sections 6-1 to 6-6
Customer Service.
1 IMDS Tutorial Integrated Microarray Database System.
Preparation Process and Exercise Manual
Hazardous Waste Section
Bellwork Do the following problem on a ½ sheet of paper and turn in.
1 The Royal Doulton Company The Royal Doulton Company is an English company producing tableware and collectables, dating to Operating originally.
Exarte Bezoek aan de Mediacampus Bachelor in de grafische en digitale media April 2014.
Checking & Corrective Action
Copyright © 2012, Elsevier Inc. All rights Reserved. 1 Chapter 7 Modeling Structure with Blocks.
1 RA III - Regional Training Seminar on CLIMAT&CLIMAT TEMP Reporting Buenos Aires, Argentina, 25 – 27 October 2006 Status of observing programmes in RA.
Adding Up In Chunks.
FAFSA on the Web Preview Presentation December 2013.
SLP – Endless Possibilities What can SLP do for your school? Everything you need to know about SLP – past, present and future.
MaK_Full ahead loaded 1 Alarm Page Directory (F11)
1 10 pt 15 pt 20 pt 25 pt 5 pt 10 pt 15 pt 20 pt 25 pt 5 pt 10 pt 15 pt 20 pt 25 pt 5 pt 10 pt 15 pt 20 pt 25 pt 5 pt 10 pt 15 pt 20 pt 25 pt 5 pt Synthetic.
GEtServices Services Training For Suppliers Requests/Proposals.
DSS Decision Support System Tutorial: An Instructional Tool for Using the DSS.
: 3 00.
1 hi at no doifpi me be go we of at be do go hi if me no of pi we Inorder Traversal Inorder traversal. n Visit the left subtree. n Visit the node. n Visit.
Section 404 Audits of Internal Control and Control Risk
Speak Up for Safety Dr. Susan Strauss Harassment & Bullying Consultant November 9, 2012.
1 Phase III: Planning Action Developing Improvement Plans.
Converting a Fraction to %
Clock will move after 1 minute
PSSA Preparation.
Physics for Scientists & Engineers, 3rd Edition
Select a time to count down from the clock above
Copyright Tim Morris/St Stephen's School
1.step PMIT start + initial project data input Concept Concept.
Page 1 Orchard Harvest ™ LIS Find a Patient Training.
Aviation Management System 1 2  Silver Wings Aircraft Aviation Management System represents a functional “high – end” suite of integrated applications.
Dr Samah Kotb Lecturer of Biochemistry 1 CLS 432 Dr. Samah Kotb Nasr El-deen Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Nasr.
CLIA COMPLIANCE. What is CLIA? In 1988 Congress turned its attention to deficiencies in the quality of services provided by the nation’s laboratories.
Unit #7 - Basic Quality Control for the Clinical Laboratory
Inspection Issues in the Analytical Laboratory: An FDA Perspective Yvonne McKnight Chemist US Food and Drug AdministrationPhone: x
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
Introduction to Quality Assurance. Quality assurance vs. Quality control.
CLIA Hot Topics CLIA Gary Yamamoto
Presentation transcript:

POINT-OF-CARE TESTING and the CLIA SURVEY Gary Yamamoto Centers for Medicare & Medicaid Services San Francisco Regional Office CLIA

CLIA Regulatory Requirements Waived Tests: 42 CFR § 493.15(e)(1) Follow manufacturers’ instructions for performing the test CLIA

CLIA Regulatory Requirements Non-Waived Tests (Moderate, PPMP, and High): Proficiency Testing (42 CFR § 493.801) Facility Administration (42 CFR § 493.1100) General Systems (42 CFR § 493.1230) Preanalytic Systems (42 CFR § 493.1240) Analytic Systems (42 CFR § 493.1250) Postanalytic Systems (42 CFR § 493.1290) Personnel (42 CFR § 493.1351) CLIA

CLIA CMS’ Current Survey Approach Survey approach remains educational, but deficiencies identified are cited Laboratory has opportunity to correct Regulatory explanation &/or resources provided In 2006, GAO stated survey consistency concerns Training & guidance provided to surveyors “Mandatory” citations determined Accelerated enforcement for repeat offenders Slight increases in citations occurred Due to mandatory citations & areas of training focus CLIA

CLIA CMS’ Current Survey Approach The primary objective of the CLIA survey process is to determine whether or not the laboratory meets the CLIA requirements. CLIA

CLIA CMS’ Current Survey Approach The surveyor meets this objective by employing an outcome-oriented survey process or approach. The intent of which is to focus the surveyor on the overall performance of the laboratory and the way it monitors itself, rather than on a methodical evaluation of each standard level regulatory requirement. CLIA

ONSITE CLIA SURVEYS CLIA NOT subject to biennial inspections: Certificate of Waiver Certificate of PPMP Subject to biennial inspections: Certificate of Compliance Certificate of Accreditation CLIA

ONSITE CLIA SURVEYS Certificate of Waiver/PPMP Pursuant to 42 CFR § 493.1775(b): CMS/agent may conduct an inspection to: Determine if the laboratory is operated and testing is performed in a manner that does not constitute an imminent and serious risk to public health Evaluate a complaint CLIA

ONSITE CLIA SURVEYS Certificate of Waiver/PPMP Determine whether the laboratory is performing tests beyond the scope of the certificate held by the laboratory Collect information regarding the appropriateness of tests specified as waived tests or PPMP CLIA

ONSITE CLIA SURVEYS Certificate of Compliance The inspection sample for review may include testing in the categories of waived tests and PPMP. CLIA

ONSITE CLIA SURVEYS Regulatory Requirements Non-Waived Tests (Moderate, PPMP, and High): Proficiency Testing (42 CFR § 493.801) Facility Administration (42 CFR § 493.1100) General Systems (42 CFR § 493.1230) Preanalytic Systems (42 CFR § 493.1240) Analytic Systems (42 CFR § 493.1250) Postanalytic Systems (42 CFR § 493.1290) Personnel (42 CFR § 493.1351) CLIA

FACILITY ADMINISTRATION Facilities Requirements for transfusion services Retention requirements CLIA

FACILITY ADMINISTRATION Retention Requirements Test Reports: Retain or be able to retrieve a copy of the original report (including final, preliminary, and corrected reports) at least 2 years after the date of reporting (42 CFR §493.1105(a)(6)) A copy, either paper or electronic, includes all information sent to the individual requesting the test or using the test result(s). CLIA

PROFICIENCY TESTING CLIA CLIA proficiency testing requirements pertain to non-waived tests only. Proficiency is required for only the test system, assay, or examination used as the primary method for patient testing during the proficiency testing event. (42 CFR § 493.801(b)(6)) CLIA

PROFICIENCY TESTING CLIA If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites. (42 CFR § 493.1281(a)) CLIA

PROFICIENCY TESTING CLIA Each laboratory must enroll in a proficiency testing program that meets the criteria in subpart I. At least twice annually, the laboratory must verify the accuracy of any test or procedure it performs that is not included in subpart I. (42 CFR § 493.1236(c)(1)) CLIA

GENERAL SYSTEMS CLIA Confidentiality of patient information Specimen identification and integrity Complaint investigations Communications Personnel competency assessment policies Evaluation of proficiency testing Quality assessment program CLIA

GENERAL SYSTEMS Specimen Identification and Integrity The laboratory must establish and follow written policies and procedures that ensure positive identification and optimum integrity of a patient’s specimen from the time of collection or receipt of the specimen through completion of testing and reporting of results. (42 CFR § 493.1232) CLIA

GENERAL SYSTEMS Personnel Competency Assessment Policies As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to asses employee and, if applicable, consultant competency. (42 CFR § 493.1235) CLIA

PREANALYTIC SYSTEMS CLIA Test requisition Specimen submission, handling, and referral policies and procedures Quality assessment program CLIA

PREANALYTIC SYSTEMS Test Requisitions Test requisitions must include: The name and address or other suitable identifiers of the authorized person requesting the test and, if appropriate, the individual responsible for using the test results, or the name and address of the laboratory submitting the specimen. . . (42 CFR § 493.1241(c)(1)) CLIA

PREANALYTIC SYSTEMS Test Requisitions 42 CFR § 493.1241(c)(1) Interpretation: The test requisition must provide the information necessary to identify and send test results to the individual who ordered the test (the authorized person), or, where applicable, to the authorized person’s agent. CLIA

PRETANALYTIC SYSTEMS Test Requisitions “Authorized person” An individual authorized under State law to order tests or received test results, or both “Agent” An individual or entity legally acting on behalf of the authorized person to receive test results CLIA

ANALYTIC SYSTEMS CLIA Procedure manual Test systems, equipment, instruments, reagents, materials, and supplies Performance specifications Maintenance and function checks Calibration and calibration verification procedures Control procedures Test records Quality assessment program CLIA

ANALYTIC SYSTEMS Performance Specifications 42 CFR § 493.1253: Establishment and Verification of Performance Specifications VERIFICATION ESTABLISHMENT CLIA

ANALYTIC SYSTEMS Performance Specifications VERIFICATION of Performance Specifications Each laboratory that introduces an unmodified, FDA-cleared or approved test system must do the following before reporting patient test results: CLIA

ANALYTIC SYSTEMS Performance Specifications Demonstrate that it can obtain performance specifications comparable to those established by the manufacturer for the following performance characteristics: Accuracy Precision Reportable range of test results for the test system. Verify that the manufacturer’s reference intervals (normal values) are appropriate for the laboratory’s patient population. CLIA

ANALYTIC SYSTEMS Performance Specifications ESTABLISHMENT of Performance Specifications Each laboratory that modifies an FDA-cleared or approved test system, or introduces a test system not subject to FDA clearance or approval (including methods developed in-house and standardized methods such as text book procedures)., OR. . . CLIA

ANALYTIC SYSTEMS Performance Specifications . . .uses a test system in which performance specifications are not provided by the manufacturer must, before reporting patient test results, establish for each test system the performance specifications for the following performance characteristics, as applicable: CLIA

ANALYTIC SYSTEMS Performance Specifications Accuracy Precision Analytical sensitivity Analytical specificity to include interfering substances Reportable range of test results for the test system Reference intervals (normal values) Any other performance characteristics required for test performance. CLIA

ANALYTIC SYSTEMS Performance Specifications Establish/Verify Performance Specification: When MULTIPLE INSTRUMENTS (including the same make and model, e.g., point-of-care instruments) are used to perform the same test, the laboratory must establish or verify, as applicable, performance specifications for each instrument. CLIA

ANALYTIC SYSTEMS Performance Specifications When a temporary replacement (loaner) instrument is received which is identical (i.e., same make and model, and method for the same analyte) to the instrument which is being replaced. . . CLIA

ANALYTIC SYSTEMS Performance Specifications . . .the laboratory must verify comparable performance by comparing, at a minimum, results of two or more levels of controls AND either previously tested proficiency testing samples or previously tested patient specimens. CLIA

ANALYTIC SYSTEMS Quality Control Procedures Quality Control Procedures must detect immediate errors that occur due to: Test System Failures Environmental Conditions Operator Performance (42 CFR § 493.1256(c)(1)) CLIA

ANALYTIC SYSTEMS Quality Control Procedures Test System Failures Reagent contamination or deterioration Reagent lot variation Reaction temperature fluctuations Inadequate sampling Improper or loss of calibration Electronic or mechanical failure Power supply variances CLIA

ANALYTIC SYSTEMS Quality Control Procedures Environmental Condition Changes Temperature Airflow Light intensity Humidity Altitude CLIA

ANALYTIC SYSTEMS Quality Control Procedures Operator Performance Improper specimen preparation and handling Incorrect test interpretation Failure to follow the manufacturer’s test system instructions CLIA

ANALYTIC SYSTEMS Quality Control Procedures Operator training prior to testing is critical and competency assessment over time is necessary to ensure continued appropriate test performance. CLIA

ANALYTIC SYSTEMS Quality Control Procedures In general, at least ONCE each day patient specimens are assayed or examined perform the following for – Each quantitative procedure, include 2 control materials of different concentrations; Each qualitative procedure, include a negative and positive control material; Test procedures producing graded/titered results, include a negative control material and a control material with graded/titered reactivity respectively. (42 CFR § 493.1256(d)(3)(i-iii)) CLIA

ANALYTIC SYSTEMS Quality Control Procedures For CLIA purposes, “control materials” are defined as: External quality control materials Have a similar matrix to that of patient specimens Treated in the same manner as patient specimens Go through all analytic phases of testing CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC However, a laboratory can perform EQUIVALENT QUALITY TESTING (EQC) as specified in Appendix C of the State Operations Manual (CMS Pub.7) (42 CFR § 493.1256(d)) CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC EQC may only be used for laboratory testing subject to the following control procedure requirements: 42 CFR § 493.1256(d)(3)(i-iii) 42 CFR § 493.1256(d)(3)(iv) – test procedures that include an extraction phase (limited to Options 1 and 2) 42 CFR §§ 493.1267 to 493.1269 – control requirements for routine chemistry and hematology (limited to Options 1 and 2) CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC OPTION 1 – Test system uses one or more internal/procedural control(s) to monitor all of its analytic components Evaluation Process: For ten (10) consecutive days of testing, perform the test system’s internal control procedures per the manufacturer’s instructions and test two levels of external control material daily. If acceptable. . . The laboratory may reduce the frequency of testing two levels of external control material from daily to once per calendar month. CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC OPTION 2 – Test system uses some internal/procedural controls to monitor only certain components Evaluation Process: For thirty (30) consecutive days of testing, perform the test system’s internal control procedures per the manufacturer’s instructions and test two levels of external control material daily. If acceptable. . . The laboratory may reduce the frequency of testing two levels of external control material from daily to once per calendar week. CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC OPTION 3 – Test systems without internal/procedural controls Evaluation Process: For sixty (60) consecutive days of testing, perform the test system’s control procedures per the manufacturer’s instructions and, at a minimum, test two levels of external control material daily. If acceptable. . . The laboratory may reduce the frequency of testing two levels of external control material from daily to once per calendar week. CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC If any internal/external control results are unacceptable during the evaluation process OR after the laboratory has reduced the frequency for testing external control materials, the laboratory must repeat the testing of the unacceptable control. If repeat control result is acceptable, no further corrective action is necessary. Resume evaluation process or reduced external control testing frequency. CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC If repeat control result is NOT acceptable, the laboratory must: Identify the problem and take appropriate corrective action. Evaluate all patient test results obtained in the unacceptable test run and since the last acceptable test run to determine if patient test results have been adversely affected. (42 CFR 493.1282(b)(2)) Restart and successfully complete the evaluation process before reducing the frequency of testing external control materials. CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC Remember, the following ongoing assessment activities are also required: Proficiency testing results MUST demonstrate acceptable/satisfactory performance Analytic system quality assessment activities must demonstrate problems are not occurring Competency assessment evaluations must demonstrate testing personnel are accurately performing testing CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC If unacceptable results are obtained for any of these assessment activities, the laboratory must: Investigate Identify the problem Document the corrective action(s) taken RESTART the evaluation process CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC If the laboratory chooses to implement the reduced QC frequency (EQC) for MULTIPLE INSTRUMENTS (including the same make and model used to perform the same test) a successful evaluation process must be performed for each instrument for which the QC frequency applies. CLIA

ANALYTIC SYSTEMS Quality Control Procedures - EQC OTHER POSSIBLE OPTIONS??? CLIA

POSTANALYTIC SYSTEMS Test Reports The laboratory must have an adequate manual or electronic system(s) in place to ensure test results and other patient-specific data are accurately and reliably sent from the point of data entry (whether interfaced or entered manually) to final report destination, in a timely manner. (42 CFR § 493.1291(a)) CLIA

POSTANALYTIC SYSTEMS Test Reports Test results must be released only to authorized persons and, if applicable, the individual responsible for using the test results and the laboratory that initially requested the test. (42 CFR § 493.1291(f)) CLIA

POSTANALYTIC SYSTEMS Test Reports 42 CFR § 493.1291(f) Interpretation: Test results must be released to the authorized person, or, if applicable, their agent. Test results must also be released to any additional individuals/entities designated on the test requisition. These entities are understood to be “responsible for using” the test results. CLIA

POSTANALYTIC SYSTEMS Test Reports For CLIA purposes, when the authorized person, designated agent, or individual responsible for using the test result receives the results, whichever is last, the laboratory’s CLIA responsibility ends. CLIA

Information can be found at the CMS Website: CLIA WEBSITE Information can be found at the CMS Website: www.cms.gov/clia CLIA