Radiation Exposures Unit 2

Slides:



Advertisements
Similar presentations
RADIOGRAPHIC GRIDS.
Advertisements

Technical Aspects: the machine the image
Advanced Biomedical Imaging Lecture 3
Control of Scatter Radiation
Chapter 11 Prime Factors.
Formation Characteristics
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
X-Ray Beam Restrictors
X-Ray Interaction with Matter & Human Biology
Primary Exposure Factors IV
Components of Radiographic Image Quality
Radiographic Quality Chapter 5.
Resident Physics Lectures
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
Chapter 6 Control of Secondary and Scattered Radiation.
Control of Scatter Radiation Beam Restricting Devices
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
Resident Physics Lectures
Chapter 6, Fauber Supplement-Chapter 14, Bushong
The Grid Kyle Thornton DMI 50B.
Analyzing the Image Density. Density Overall blackening of the image.
Lecture 4 Beam restricting Devices, their Use and Maintenance
Factors affecting the X-Ray output
X-Ray Production & Emission
X-Ray Production & Emission
Conventional and Computed Tomography
Radiographic Grids I By Professor Stelmark.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Control of Scatter Radiation
Computed Tomography Physics, Instrumentation, and Imaging
UNIT 6 Introduction to Radiation Protection
Quality Control.
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
Radiation Monitoring and Protection
IMAGE QUALITY REVIEW RT
Dr Mohamed El Safwany, MD.
Resident Physics Lectures
RAD 354 Chapt. 11 Control of Scatter Break down into: Those that reduce pt. dose and those that are geometrical in nature and those NOT! 3 factors affecting.
Image Quality Radiographic Resolution.
Mammography Questions and Answers
Exposure Factors or Prime Factors
Photon Tissue Interactions
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
Beam Restricting Devices
Radiographic Equipment
Grids George David Associate Professor Department of Radiology
Radiographic Grids I By Professor Stelmark.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
Grid Performance By Prof. Stelmark.
Lavin: Chapter 7 CTVT: pp
RTT 425 Radiation Therapy Physics Radiation Quality, Chapter 4 From Stanton and Stinson: Applied physics for Radiation Oncology.
Tube Exposure Factors Math Technique Contrast and Density
RADIOGRAPHIC GRIDS.
Optimizing The Image RVT: Chapter 7
Exposure Factors or Prime Factors
Chapters 21 & 22.
Resident Physics Lectures
X-Radiation.
Factors Effecting the Production
Resident Physics Lectures (Year 1)
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures (Year 1)
Chapter 5, Fauber Supplement-Chapter 11, Bushong
Filters, Collimators and Grids
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Presentation transcript:

Radiation Exposures Unit 2 Chapters 10, 15, 16, and 18

Objectives Define filtration, inherent filtration, added filtration , compound filtration, compensating filtration and total filtration. Explain concept of half-value layer equivalency measurements of filtration. Appraise various types of filters for specific clinical situations. Describe effect of filtration on entire x-ray beam. Identify the factors that affect scatter. Discuss types of scatter control used by radiographers Explain the purpose, construction, advantages and disadvantages of beam-restricting devices. Describe the effect of beam restriction on image quality and patient dose. Explain the process of attenuation Describe the basic composition of the human body and how they attenuate the x-ray beam Explain the relationship of the patient to density/image receptor exposure, contrast, recorded detail and distortion of the recorded image. Describe the purpose of a grid and its construction Explain: material, ratio, frequency and lead content Differentiate various grid patterns. Describe: parallel vs focused grids and stationary vs moving grids. Explain grid and technique selection based on : patient , procedure, dose and IR exposure. Discuss proper use and errors with grids and how both effect the image.

Chapter 10 Filtration

Filtration The process of eliminating undesirable low-energy photons by inserting absorbing material into the primary beam Shape the photon emissions into a more useful beam “hardens” the beam Removes low energy (soft ) photons Main Purpose Decreases radiation dose to the patient by removing photons that would not enhance the image.

Filter Any material designed to selectively absorb photons Placed between the tube( source) and the patient Aluminum Most common / standard filter material used All other materials are measured against it ability to filter- Aluminum equivalency (Al/Eq) Alternated Materials glass, oil , copper and tin

Measurement of Filtration Express in half-value layer ( HVL ) – amount of absorbing material to reduce the intensity of the primary beam by ½ the original value Also expressed in Al/Eq example : HVL = 2.0 mm Al/Eq Federal govt. specifies min HVL for diagnostic tubes: 2.5 mm Al/Eq

Types of Filtration 2 types: Inherent – in the design of the tube Added- between tube and image receptor

Inherent filtration Result of composition of tube and housing Glass envelope Oil surrounding tube Glass window – where most of the filtration comes from Total inherent is 0.5-1.0 Increases as tube ages because of vaporization of tungsten HVL testing for QC tests for reduced tube efficiency because of the added filtration

Added Filtration Any filtration that occurs outside the tube but before the IR Absorb as many low-energy photons as possible while allowing max amount of high-energy photons through. Aluminum used as a low energy absorber Collimators are considered added_ 1.0 mm Al/Eq Most comes from the silver on the mirror that reflects the positioning light.

Types of Added Filtration Compound filtration ( AKA: K-Edge) Uses two or more material that work together well to absorb High atomic number material sandwiched with a lower atomic number material Cooper ( 29) with Al ( 13) Built to absorb characteristic photons ( K shell photons) created by the previous layer The higher atomic number will absorb higher energies but Al is needed to absorb the lower energy photons created. Examples: Thoreaus filter- used in radiation therapy Some QC testing uses copper + Al filters

Types of Added Filtration Compensation Filters Designed to solve a problem of unequal subject densities Produces a more uniform IR exposure Components: AL, Lead, leaded plastic or plastic – saline solution bag Common types Wedge- thicker portion is placed over thinner part- foot, t-spine Trough ( double wedge) – chest to even out mediastinum

Total Filtration Equal to the sum of inherent + added filtration but DOES not include compensating filters Thickness depends on use of equipment Above 70 kVp 2.5 mm Al

Effect on Output As filtration increases- technical factors must increase to maintain IR exposure Even though there is an increase in technique necessary for compensation- the ESE is decreased as compared to no filtration in place at lower techniques See table 10-4 page 176

Beam Restriction Chapter 15

Scatter Product of Compton interactions ( scatter that contributes to fog and tech dose) Not part of the useful beam Impair image quality by adding exposure to the IR unrelated to patient anatomy Best way to control scatter reaching the IR is to use a grid. Best way to control scatter to the patient is use beam restriction. Principle factors affecting scatter 1. kilovoltage 2. irradiated material

1. Kilovoltage As kVp increases More photons pass through the patient to interact with the image receptor Image quality is increased with less scatter Photons that do not pass through the matter interact as Compton and photoelectric interactions Higher kVp will increase Compton( scatter) and decrease photoelectric ( pt dose) Since photoelectric contribute to patient dose, dose will be decreased kVp should be based on part/anatomy thickness /size and the amount of contrast desired Increased kVp with NO change in mAs : increased scatter Increased kVp and reduction in mAs: decreased scatter

2. Irradiated Material Scatter is affected by the volume and atomic number of the material being x-rayed. Volume is controlled by field size and patient thickness Increased volume increases scatter ( more interactions) Volume increases as part increases and as field size increase 14 x17 has more volume than 10 x 12 – increased scatter , so smallest field size should be used Large patients should have better collimation to decrease interaction ( decrease scatter) When collimation is used- more technique is needed ( 14x17 to 10x12- aprox increase is 25% , 14x17 to 8 x 10 aprox increase of 40%) Higher atomic numbers will have less scatter- they absorb more through photoelectric absorption- more contrast ( black and white) Bone absorbs more than soft tissue- less scatter Iodine, Ba and Pb absorbs more- less scatter

Types of Beam Restrictors 1. Collimators 2. Aperture diaphragms and cones / cylinders 3 . Ancillary devices

1. Collimator Most common beam restrictor – results in filtration of the beam – aprox 1 mm of Al Multiple size configurations- but should always be minimized to the part being radiographed Lead shutters at right angles and move in opposing pairs 3 purposes: Regulate field size Reduce penumbra Act as a light field

Field Size Various square or rectangle shapes PBL device - positive beam limitation device- automatic collimation of the beam size- works with a sensor in the Bucky tray to determine film size being used. Collimator accuracy must be checked yearly – Within 2 % of the distance

Penumbra Geometric unsharpness around the edge of the image Result of off focus radiation – photons emerge at different angles from the tube and interact with the tissue at various angles and create a “shadow “ image outside the exposed field of radiation Reducing penumbra will increase recorded detail

Light Field Provides an exposure field with crosshair Some units will have an AEC chamber outline Lightbulb is placed at a 450 to a mirror – improper placement of the mirror can cause a mis-projection of the light and misrepresent the field size

2. Aperture diaphragms and cones / cylinders Aperture diaphragm- flat sheet of metal( Pb) with opening cut in the middle Attached to the bottom of the collimator box Simplest of all beam limiting devices Different sizes for different receptor sizes at different distances Mostly seen in mammography because they don’t use collimators Cones and cylinders Circular aperture diaphragm with metal extensions Can flare or diverges- small at top than bottom Most effective use of scatter control

3. Ancillary Devices Designed for a specific need and are tailored for specific use during a given procedure Lead blockers or lead masks Blockers Lead impregnated rubber Used to absorb scatter created from soft tissue- used a lot on larger patients May be placed behind a patient’s back on an lateral L-spine Masks cut to a specific shape and attached to collimator

The patient as a beam emitter Chapter 16 The patient as a beam emitter

Attenuation The reduction in the total number of x-ray photons remaining in the beam after passing through an object Result of : x-rays interacting with matter and being absorbed or scattered The thicker the body part, the more attenuation Photoelectric absorption( provides information) and Compton scattering ( no useful info- only dose to tech ) Determined by the amount or type of irradiated material High atomic number = higher attenuation – bone , barium, lead Low atomic number= low attenuation- hydrogen, carbon, oxygen Lowes to highest attenuators in body by density: Air, fat, water, muscle, bone

The Patient The greatest variable Atomic makeup : hydrogen (atomic # 1) , carbon(6), nitrogen(7) and oxygen(8) calcium (20) 4 Major substances account for most variations in absorption: Air Fat Muscle Bone

1. Air Higher atomic number than fat or muscle but lower tissue density Absorbs fewer photons as it passes through allowing more photons to reach IR Present in : ????? Lungs Sinuses GI tract

2. Fat Similar to muscle- both soft tissue structures- fat has less tissue density Effective atomic number similar to water- water is used to simulate fat in experiments Varies from patient to patient Fat sometimes allow for better visualization of certain organs during x-raying kidneys

3. Muscle Higher atomic number and density than fat- cells are packed closer together Greater attenuator Can often see psoas muscles on abdomen radiograph

4. Bone Calcium level in bones makes them higher attenuators Highest effective atomic number and tissue density of the four basic substances of the body Results in few photons reaching the IR

Patients Relationship to Image Quality Patient has affect on all properties of image quality Density/image receptor exposure ( subject density) Contrast ( subject contrast) Recorded detail ( subject detail ) Distortion ( subject distortion)

1. Subject Density The impact the subject has on the resultant radiographic density or IR exposure Varies depending on the amount or type of tissue being irradiated Thicker the part/ more dense the tissue = less IR exposure

2. Subject Contrast The difference in densities of a recorded image The degree of differential absorption resulting from the differing absorption characteristics of the tissue in the body Dependent on patient’s tissues Less difference in adjacent densities- fat and muscle- less contrast More difference in adjacent densities- bone and air- more contrast

3. Subject Detail Primary factor affecting sharpness in distance from structure to IR Dependent on where structure is in the body Even size of body- larger patients the part is further away from IR How the patient is positioned PA vs AP, RT vs LT lateral

4. Subject Distortion The misrepresentation of size or shape of the structure of interest Varies with patient position

Chapter 18 The grid

Invention of the Grid 1913- Gustav Bucky, an American radiologist, designed the first grid Designed in a cross-hatched pattern using lead strips Grid showed on image but decrease scatter and improved contrast 1920- Hollis Potter, a Chicago radiologist, improved Bucky’s design Redesigned lead to run in one direction Made lead thinner and less obvious on image Potter also designed the Potter-Bucky diaphragm- allowed for the grid move during exposure Lead strips became blurred and were no longer visible

Purpose Improves contrast of the image Absorbs scatter before reaching the IR Used with thicker body parts and higher kVp since scatter increases as both of these increase Greater than 10cm kVp above 60

Construction Series of radiopaque strips alternated with interspace strips Radiopaque strips absorb scatter and are made of high atomic number material Interspace is radiolucent – allows radiation to pass through without interaction Construction Involves: materials grid ratio grid frequency

1. Materials Interspace- aluminum and plastic fiber Should not absorb any radiation However, Al has a high atomic number and actually absorbs primary photons- disadvantage at low-kVp techniques- fiber is better for low kVp techniques

2. Grid Ratio Major influence on the ability of the grid to improve contrast Defined as the “ ratio of the height of the lead strips to the distance between the strips” Expressed as a formula: Grid ratio = h h=lead strip height 3 mm height and .25 mm interspace = D D= interspace width 3/ .25 = 12:1 grid Height and distance are inversely proportional: If height is constant but we decrease the interspace distance, we increase ratio 3 mm height and .25 = 12:1 grid If height is constant but we increase the interspace distance, we decrease ratio 3 mm height and .50 = 6:1 grid ratio

Comparing grid ratios Higher grid ratios allow less scatter to pass through- more effective at removing scatter Disadvantage to high grid ratio: must be more precise positioning of grid so primary isn’t absorbed – higher grid error

3. Grid Frequency The number of grid lines per inch or cm Higher grid frequencies have thinner lead strips Range from 60-200 lines/inch Most common 85-103 lines / inch Very high frequencies 178-200 lines/inch are common for digital to minimize grid lines on image( because digital is more sensitive)

Grid Ratio and Grid Frequency The combination of the two variable will determine the total quantity of lead in the grid. Lead content is the most important in determining grid efficiency Lead content is greater in a grid with high radio and lower frequency – because the strips will be thicker in lower frequency

Grid Patterns: 2 types Linear grids- lead runs in one direction Most common- used in x-ray table Allow for tube angle along the direction of the lines ( long axis) – head to feet Can have short axis grids- helpful for Port CXR where gird is place crosswise. Criss-crossed or cross hatched- lead strips run at right angles Do not allow for tube angle Limited applications Grid Cut-off- can occur with ANY grid when the lead strips “cut-off” or absorb the primary beam- not allowing for it to reach IR- no image in that area of the IR

Grid types 1. Parallel – least common 2. Focused

1. Parallel grids Lead and interspace run parallel to each other will never intersect Strips do not coincide with divergence of the beam Absorb primary beam as a result-more along lateral edges Worse at short SIDs Better at long SIDs because beam is more straight

2. Focused Grid Lead and interspace are still parallel but are angle to match the divergence of the beam at different distances Would eventually intersect at a convergence line The distance from the face of the grid to the point of convergence is called grid radius Tube must be located across the convergence line Focused grids are grouped by focal ranges ( distances) 60-72 inch focus, 30-48 inch focus – grid will be marked Lower grid ratios allow for greater latitude in tube alignment Higher grid ratios proper tube alignment is critical

Grid Uses Stationary Mounted moving- AKA Bucky Come in various sizes to match cassette size Grid lines are more noticeable Low frequency grids( thicker strips) worse than high frequency ( thinner strips) Used with portable or in rooms when patient is not on x-ray table- gurney or hosp bed Grid cassettes- for film/screen- grid was build into cassette Mounted moving- AKA Bucky Mounted below table top ( or upright holder ) and hold cassette in place below (behind) the grid Moves during exposure so grid lines will be blurred and not seen Larger than largest cassette size used in machine Grid lines run along the axis of table ( head to foot) Movement moves the grid at right angles to the grid- side to side Two movement types Reciprocating- side to side- motor driven Oscillating- circular movement- electromagnet driven

Grid Selection Made in equipment purchase- not normally interchangeable Stationary grids will be selected by size/ focus/ grid ratio

Grid Conversion Formula: Compensation of technique (mAs) must be made between type of grid selected Same technique can’t be used for no grid vs 16:1 Formula: mAs 1 = GCF 1 mAs 2 GCF 2 Image 1 = 8:1 grid 35 mAs Image 2 = 12:1 grid ?? mAs = 4 X 6 4x=210 x= 52.5 mAs Grid Ratio Conversion Factor No grid 1 5:1 2 6:1 3 8:1 4 10:1 or 12:1 5 16:1 6 This chart is different than in text- eliminates need to figure in kVp ( easier to apply and remember! )

Grid Performance Evaluation The measuring of a grid in cleanup or removing scatter Determined by two factors: Selectivity Contrast improvement ability

1. Selectivity The ratio between the quantity (%) of the primary photons transmitted through the grid to the quantity (%) of scatter photons transmitted - The better the grid is at removing scatter ; the greater the selectivity- higher grid ratios have higher selectivity Formula: selectivity = % of primary radiation transmitted % of scatter radiation transmitted

2. Contrast Improvement Ability Best way to determine grid function Contrast improvement factor ( K) – dependent on the amount of scatter produced As scatter increases , contrast decrease, thus lowering the improvement factor The higher the K factor the greater the contrast improvement Formula: K= Radiographic contrast with the grid Radiographic contrast w/out the grid

Grid Errors Result of improper use More frequent with focused grids- tube must be centered and at correct SID Also has tube side and IR side Types of Errors Off- Level Off-Center Off-Focus Upside-down Moiré Effect

Off-Level Occurs when tube is angled across the long axis of the grid strips Result of improper tube or grid position Most common with stationary grid- portable procedures Result Absorption of primary beam & decrease in overall exposure – light image

Off-Center When tube is not centered to grid Does not line up with divergence of beam The greater degree of lateral off-centering the greater the cut-off Result : Decreased exposure across entire image ( light image)

Off-Focus When a grid is used at incorrect SID Higher grid ratios require greater accuracy to prevent cutoff Result: Grid cutoff along the peripheral edges of the image

Upside-Down Focused grids have an identified tube side Result: If used upside-down- severe peripheral grid cut- off will occur

The Moire Effect Occurs with digital image receptors when grid lines are captured and scanned parallel to the scan lines of the imaging plate reader Grid lines run parallel with the laser Happens with stationary grids- ran across short axis Corrected by high-frequency grids w/ 103 lines per inch or higher

Alternate to Grid Restricting beam size Compress body part- compression band Air-Gap Technique Increased OID – same amount of scatter will leave patient , but won’t reach patient 10” air gap has similar clean-up of 15:1 grid References sheet for all documents

This workforce solution was funded by a grant awarded by the U. S This workforce solution was funded by a grant awarded by the U.S. Department of Labor's Employment and Training Administration. The solution was created by the grantee and does not necessarily reflect the official position of the U.S. Department of Labor. The Department of Labor makes no guarantees, warranties, or assurances of any kind, express or implied, with respect to such information, including any information on linked sites and including, but not limited to, accuracy of the information or its completeness, timeliness, usefulness, adequacy, continued availability, or ownership.