Presentation is loading. Please wait.

Presentation is loading. Please wait.

Grids George David Associate Professor Department of Radiology

Similar presentations


Presentation on theme: "Grids George David Associate Professor Department of Radiology"— Presentation transcript:

1 Grids George David Associate Professor Department of Radiology
Medical College of Georgia

2 Purpose Directional filter for photons Ideal grid
passes all primary photons photons coming from focal spot blocks all secondary photons photons not coming from focal spot Focal Spot “Good” photon Patient “Bad” photon X Grid Film

3 Grid Construction Lead Interspace ~ .05“ thick upright strips (foil)
material between lead strips maintains lead orientation materials fiber aluminum wood Lead Interspace

4 Grid Ratio Ratio of interspace height to width h Grid ratio = h / w w
Lead Interspace h w Grid ratio = h / w

5 Grid Ratio Expressed as X:1 Typical values
8:1 to 12:1 for general work 3:1 to 5:1 for mammography Grid function generally improves with higher ratios h w Grid ratio = h / w

6 Lines per Inch # lead strips per inch grid width Typical: 103 25.4
W + w w W w = thickness of interspace (mm) W = thickness of lead strips (mm)

7 Grid Structure

8 Grid Patterns Orientation of lead strips as seen from above Types
Linear Cross hatched 2 stacked linear grids ratio is sum of ratios of two linear grids very sensitive to positioning & tilting Rare; only found in specials

9 Grid Styles Parallel Focused

10 Parallel Grid lead strips parallel useful only for small field sizes
large source to image distances

11 Focused Grid Slightly angled lead strips
Strip lines converge to a point in space called convergence line Focal distance distance from convergence line to grid plane Focal range working distance range width depends on grid ratio smaller ratio has greater range Focal range Focal distance

12 Grid Cassette Grid built into cassette front
Sometimes used for portables formerly used in mammography low grid ratios focused

13 Ideal Grid passes all primary radiation
Reality: lead strips block some primary Lead Interspace

14 Ideal Grid block all scattered radiation
Reality: lead strips permit some scatter to get through to film Lead Interspace

15 Grid Performance Measurements
Primary Transmission (Tp) Bucky Factor (B) contrast improvement factor (K)

16 Primary Transmission Fraction of a scatter-free beam passed by grid
Ideally 100% (never achieved) Lead Interspace

17 Measuring Primary Transmission
small area beam scatterer in beam far from grid virtually no scatter reaches grid measure radiation intensity with & without grid ratio X 100 is Primary Transmission (Tp) Focal Spot Lead Diaphragm Grid Detector

18 Primary Transmission Typical values: 55 - 75%
Theoretic calculation: (fraction of grid that is interspace) Tp (%)= 100 X W / (W+w) where W = Interspace thickness w = lead strip thickness actual transmission < theoretical primary attenuated by interspace material focusing imperfections w W W+w

19 Bucky Factor Radiation incident on grid transmitted radiation indicates actual increase in exposure because of grid’s presence due to attenuation of both primary & secondary radiation

20 Bucky Factor Measurement
large x-ray field thick phantom ratio of intensity measurement with & without grid Grid Detector

21 Bucky Factor Measures fraction of radiation absorbed by grid
high ratio grids have higher bucky factors

22 Bucky Factor Higher bucky factor means higher x-ray technique
higher patient dose typically 3-6

23 Contrast Improvement Factor
Ratio of contrast with & without grid Scatter reduces appearance of contrast No Scatter Scatter

24 Contrast Improvement Factor
Depends on kVp field size phantom thickness increase in any of above means more scatter less contrast lower contrast improvement factor

25 Contrast Improvement Factor
Better contrast improvement with higher ratio more lead content in grid

26 Lead Content of Grid Definition
weight per unit area grams (Pb) / cm2 of grid

27 More Lines / inch at Same Ratio Means Less Lead Content & Contrast Improvement
thinner lead & same ratio less lead (less thickness, same height) Same interspace dimensions h d Grid ratio = h / d

28 More Lines / inch at Same Ratio Means Less Lead Content & Contrast Improvement
thinner interspace & less height to maintain ratio less lead (less height, same thickness) h d Grid ratio = h / d

29 Lead Content of Grid more lines / inch for same ratio means less lead content & thus less contrast improvement puts practical limit on lines per inch same contrast improvement for 133 line 10:1 and 80 line 8:1 grids h d Grid ratio = h / d

30 Grid Disadvantages Increased patient dose Positioning critical
poor positioning results in grid cutoff loss of primary radiation because images of lead strips projecte wider

31 Grid Cutoff focused grids used upside down
lateral decentering (or angulation) focus- grid distance decentering combined lateral & focus-grid distance decentering

32 Upside Down Focused Grid
Dark exposed band in center Severe peripheral cutoff

33 Lateral Decentering uniform loss of radiation over entire film
uniformly light radiograph no recognizable characteristic (dangerous)

34 Lateral Decentering also occurs when grid at correct position but tilted both result in uniform loss of intensity no other clinical clues may be mistaken for technique problems Can be compensated for by over-exposing patient

35 Lateral Decentering cutoff increases with Higher grid ratio
Greater decentering distance smaller focal distances r b L = X 100 fo L = loss of primary radiation (%) r = grid ratio b = lateral decentering distance (inches) fo = focal distance of grid (inches)

36 Lateral Decentering Significant problem in portable radiography
Compensate by over-exposing patient exact centering not possible minimizing lateral decentering low ratio grids long focal distances

37 Distance Decentering Grid too close or too far from focal spot
Darker center All parallel grids have some degree of distance decentering Focused to infinity

38 X Near focus-grid decentering Far focus-grid decentering
target below convergent line cutoff more severe than far decentering Far focus-grid decentering target above convergent line X

39 Near focus-grid decentering Far focus-grid decentering
cutoff at periphery dark center cutoff proportional to grid ratio decentering distance

40 Minimizing Distance Decentering Cutoff
low grid ratio small fields

41 Combined lateral and focus-grid distance decentering
Easy to recognize uneven exposure film light on one side, dark on the other

42 Combined lateral and focus-grid distance decentering
Cutoff proportional to grid ratio decentering distance Cutoff inversely proportional to grid focal distance Less cutoff for longer focus grids cutoff greater for near than for far distance decentering

43 Moving Grids Motion starts with second trigger Grids move ~1- 3 inches
must be fast enough not to see grid lines for short exposures Motion blurs out lead strip shadows for single phase generators grid motion must not synchronize with pulses note error in book, page 111 (omits “not”)

44 Moving Grid Disadvantages
$$$ Vibration Potential May limit minimum exposure time Increases patient dose lateral decentering from motion up to 20% loss of primary evenly distributes radiation on film stationary grid makes interspace gaps darker for same amount of radiation

45 Grid Tradeoff Advantage Disadvantage cleanup / scatter rejection
increased patient dose increased exposure time increase tube loading positioning & centering more critical $$$

46 Grid Selection use low ratios for low kVp, high ratios for high kVp
book recommends 8:1 below 90 kVp 12:1 above 90 kVp

47 Air Gap Techniques Principle Negligible attenuation in air gap
radiation scatters uniformly decrease in scatter (most scatter misses film) air gap decreases angle of capture; increases angle of escape Negligible attenuation in air gap Angles of escape

48 Air Gap air gap very effective in removing scatter originating closest to film much of scatter nearest tube doesn’t reach film Much attenuation of scatter in the body Air gap decreases capture angle

49 Air Gap Applications Magnification Radiography including mammography
geometry causes air gap Air Gap Chest Radiography air gap used as alternative to grid SID increased from 6 feet to 10 feet to maintain geometric unsharpness Grid not used with air gap

50 Air Gap Optimization Air gap more effective for thicker body parts
first inch of air gap most effective in contrast improvement image sharpness deteriorates with increasing gap (magnification) compensate with greater SID smaller focal spot

51 Mammo Cellular Grid


Download ppt "Grids George David Associate Professor Department of Radiology"

Similar presentations


Ads by Google