Resident Physics Lectures (Year 1)

Similar presentations


Presentation on theme: "Resident Physics Lectures (Year 1)"— Presentation transcript:

1 Resident Physics Lectures (Year 1)
Christensen, Chapter 5 Attenuation George David Associate Professor Medical College of Georgia Department of Radiology

2 Beam Characteristics Characteristic Radiation Bremsstrahlung
Discrete energies Characteristic of target material Bremsstrahlung Energy range 0 – kVp selected

3 Beam Characteristics 1, 2, 3, ... ~ ~ ~ ~ ~ Quantity
number of photons in beam 1, 2, 3, ... ~ ~ ~ ~ ~

4 Beam Characteristics 1 @ 27 keV, 2 @ 32 keV, 2 at 39 keV, ... ~ ~ ~ ~
Quality energy distribution of photons in beam 27 keV, 32 keV, 2 at 39 keV, ... ~ ~ ~ ~

5 Beam Characteristics ~ ~ ~ ~ ~ ~ ~ ~ Intensity
324 mR Intensity weighted product of # & energy of photons depends on quantity Quality Beam intensity can be quantified by ionizations Meter reading in milliRoentgens (mR) ~ ~ ~ ~ ~ ~ ~ ~

6 So what’s a Roentgen? Unit of measurement for amount of ionizing radiation that produces 2.58 x 10-4 Coulomb/kg of STP 1 C ~ ×1018 electrons

7 Beam Intensity Can be measured in terms of # of ions created in air by beam Valid for monochromatic or for polychromatic beam 324 mR

8 Monochromatic Radiation (Mono-energetic)
Radioisotope Not x-ray beam all photons in beam have same energy attenuation results in Change in beam quantity no change in beam quality # of photons & total energy of beam changes by same fraction

9 Monochromatic radiation beam
Attenuation Formula N = No e -mx where No = number of incident photons N = number of transmitted photons e = base of natural logarithm (2.718…) m = linear attenuation coefficient (1/cm); property of energy material x = absorber thickness (cm) No N x Monochromatic radiation beam

10 Monochromatic radiation beam
If x=0 (no absorber) N = No e -mx where No = number of incident photons N = number of transmitted photons e = base of natural logarithm (2.718…) m = linear attenuation coefficient (1/cm); property of energy material x = absorber thickness (cm) No N X=0 Monochromatic radiation beam

11 Monochromatic Radiation
Let’s graph the attenuation of a monochromatic x-ray beam vs. attenuator thickness 60% removed 40% remain Monochromatic radiation beam

12 Monochromatic Radiation
Yields straight line on semi-log graph 1 .1 .01 .001 Fraction (also fraction of energy) Remaining or Transmitted N / No = e -mx 1 2 3 4 5 Attenuator Thickness Monochromatic radiation beam

13 Polychromatic Radiation (Poly-energetic)
X-Ray beam contains spectrum of photon energies highest energy = peak kilovoltage applied to tube mean energy 1/3 - 1/2 of peak depends on filtration

14 X-Ray Beam Attenuation
reduction in beam intensity by absorption (photoelectric) deflection (scattering) Attenuation alters beam quantity quality higher fraction of low energy photons removed Beam Hardening Higher Energy Lower

15 Monochromatic radiation beam
Half Value Layer (HVL) absorber thickness that reduces beam intensity by exactly half Units of thickness value of “x” which makes N equal to No / 2 HVL = .693 / m N = No e -mx Monochromatic radiation beam

16 Half Value Layer (HVL) Indication of beam quality
Valid concept for all beam types Mono-energetic Poly-energetic Higher HVL means more penetrating beam lower attenuation coefficient

17 Factors Affecting Attenuation
Energy of radiation / beam quality higher energy more penetration less attenuation Matter density atomic number electrons per gram higher density, atomic number, or electrons per gram increases attenuation

18 Polychromatic Attenuation
Yields curved line on semi-log graph line straightens with increasing attenuation slope approaches that of monochromatic beam at peak energy mean energy increases with attenuation beam hardening 1 .1 Polychromatic Fraction Transmitted .01 Monochromatic .001 Attenuator Thickness

19 Applications As photon energy increases
subject (and image) contrast decreases differential absorption decreases at 20 keV bone’s linear attenuation coefficient 6 X water’s at 100 keV bone’s linear attenuation coefficient 1.4 X water’s

20 Scatter Radiation NO Socially Redeeming Qualities
no useful information on image detracts from film quality exposes personnel, public represents 50-90% of photons exiting patient

21 Scatter Factors An increase in any of above increases scatter.
Factors affecting scatter field size thickness of body part kVp An increase in any of above increases scatter.

22 Scatter & Field Size Reducing field size causes significant reduction in scatter radiation II Tube X-Ray II Tube X-Ray One of the most effective ways of minimizing operator exposure is to reduce field size through collimation. Even a relatively small reduction in field size can often result in a substantial reduction in operator exposure. This occurs for two reasons. The first is that a smaller beam irradiates a less volume of tissue so that there is less tissue to act as a scatter radiation source. Secondly reducing beam size means that scatter radiation must travel further through the patient before exiting. The increased travel distance means a less intense scatter field for the operator. A fluoroscopist should always collimate the x-ray beam to a size no larger than is required clinically.

23 Field Size & Scatter Field Size & thickness determine volume of irradiated tissue Scatter increase with increasing field size initially large increase in scatter with increasing field size saturation reached (at ~ 12 X 12 inch field) further field size increase does not increase scatter reaching film scatter shielded within patient

24 Thickness & Scatter Increasing patient thickness leads to increased scatter but saturation point reached scatter photons produced far from film shielded within body

25 Scatter Management Reduce scatter by minimizing field size thickness
within limits of exam thickness mammography compression

26 Scatter Control Techniques: Grid
directional filter for photons Increases patient dose

27 Grid Structure

28 Purpose Directional filter for photons Ideal grid
passes all primary photons photons coming from focal spot blocks all secondary photons photons not coming from focal spot Focal Spot “Good” photon Patient “Bad” photon X Grid Film

29 Scatter Control Techniques: Air Gap
Gap intentionally left between patient & image receptor Grid not used Grid Air Gap Patient Patient Air Gap Grid Image Receptor


Download ppt "Resident Physics Lectures (Year 1)"