IMAGE GENERATION IN CT.
Advertisements

Environmental Remote Sensing GEOG 2021
Computers and Computed Tomography
Image Reconstruction.
Computed Tomography
Image Quality Bushong Chapter 7.
BIOE 220/RAD 220 REVIEW SESSION January 30, 2011.
Mark Mirotznik, Ph.D. Associate Professor The University of Delaware
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Computed Tomography II
Image reconstruction and analysis for X-ray computed microtomography Lucia Mancini 1, Francesco Montanari 2, Diego Dreossi 3 1 Elettra - Trieste 2 A.R.P.A.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
CT Physics V.G.Wimalasena Principal School of radiography.
Computed Tomography III
Lecture 4 Linear Filters and Convolution
1 Image Filtering Readings: Ch 5: 5.4, 5.5, 5.6,5.7.3, 5.8 (This lecture does not follow the book.) Images by Pawan SinhaPawan Sinha formal terminology.
Computed Tomography. Introduced in 70’s Principle: Internal structures of an object can be reconstructed from multiple projections of the object.
Optical Density and Brightness
tomos = slice, graphein = to write
Lean Body Mass Assessment: Interpretations of Computed Tomography
LEC ( 2 ) RAD 323. Reconstruction techniques dates back to (1917), when scientist (Radon) developed mathematical solutions to the problem of reconstructing.
Seeram Chapter 11: Image Quality
COMPUTED TOMOGRAPHY (I) RAD 365 CT - Scan
Basic principles Geometry and historical development
Digital Image Characteristic
CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.
Without reference, identify principles relating to Computed Tomography Clinical Applications with at least 70 percent accuracy.
Digital Images Chapter 8 Exploring the Digital Domain.
STEP 1: Determining the exact image width STEP 1: Determining the exact image width Position of X-ray Filter Position of X-ray Filter STEP 5: Crop Extra.
Computers in Imaging Q & A Robert Metzger, Ph.D..
Seeram Chapter 7: Image Reconstruction
An experiment in computed tomography
Schematic Representation o f the Scanning Geometry of a CT System
Integral University EC-024 Digital Image Processing.
Computed Tomography Q & A
PRINCIPLE AND BASIC PHYSICS OF COMPUTED TOMOGRAPHY
Computed Tomography Physics, Instrumentation, and Imaging
PRINCIPLES OF CT Dr Mohamed El Safwany, MD. Intended learning outcome The student should learn at the end of this lecture principles of CT.
Factors affecting CT image RAD
Digital Image Processing Lecture 5: Neighborhood Processing: Spatial Filtering Prof. Charlene Tsai.
CT and MRI. Computed Tomography 1972 __________________ X-ray tube in a tube Rows of _________ opposite x-ray tube Produces ___________________.
DIGITAL IMAGE PROCESING NASA LUNAR AND PLANETARY EXPLORATION.
LEC (1) Design/layout by E Bashir, CAMS, King Saud University, Ref: Seeram, Karthikeyan, internet.
Digital Image Processing Lecture 16: Segmentation: Detection of Discontinuities Prof. Charlene Tsai.
Sounds of Old Technology IB Assessment Statements Topic 14.2., Data Capture and Digital Imaging Using Charge-Coupled Devices (CCDs) Define capacitance.
If F {f(x,y)} = F(u,v) or F( ,  ) then F( ,  ) = F { g  (R) } The Fourier Transform of a projection at angle  is a line in the Fourier transform.
CT IMAGE RECONSTRUCTION  Hounsfield envisioned dividing a slice into a matrix of 3-dimensional rectangular boxes (voxels) of material (tissue). Conventionally,
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
CS TC 22 CT Basics CT Principle Preprocessing. 2 CT Basics CT principle preprocessing CS TC 22 Blockdiagram image processor.
More digital reading explaining LUT RT 244 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
Part No...., Module No....Lesson No
Digital Image Processing Lecture 5: Neighborhood Processing: Spatial Filtering March 9, 2004 Prof. Charlene Tsai.
CT Chapter 4: Principles of Computed Tomography. Radiography vs. CT Both based on differential attenuation of x-rays passing through body Radiography.
Digital Image Processing Lecture 16: Segmentation: Detection of Discontinuities May 2, 2005 Prof. Charlene Tsai.
More digital 244 wk 12 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA,
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Theory of Reconstruction Schematic Representation o f the Scanning Geometry of a CT System What are inside the gantry?
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Introduction to Medical Imaging Week 2: X-ray and CT
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Chapter 10 Digital Signal and Image Processing
Mathematics for Computed Tomography
Environmental Remote Sensing GEOG 2021
Central Section Theorem
Computed Tomography Image Reconstruction
Digital Image Processing Lecture 16: Segmentation: Detection of Discontinuities Prof. Charlene Tsai.
Chapter I, Digital Imaging Fundamentals: Lesson II Capture
Basic principles Geometry and historical development
Basic Principles of CT Chapter 1.
Presentation transcript:

…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography

Correction for star artifacts - Filtering A filter function is applied to each point along the attenuation profile to eliminate these artifacts A filter function is applied to each point along the attenuation profile to eliminate these artifacts Different filter functions are used to create sharper (higher resolution) or smoother (lower noise) images Different filter functions are used to create sharper (higher resolution) or smoother (lower noise) images Attenuation profile Filter function Filtered profile

Convolution? The process of applying the filter function to the attenuation profile is called CONVOLUTION The process of applying the filter function to the attenuation profile is called CONVOLUTION (Super imposition of the filter function and the attenuation profile)

Result of convolution The convolution minimizes the artifacts by changing the back projected information. The convolution minimizes the artifacts by changing the back projected information.

The dark and white bands in the illustration add up in such a way as to yield an image which accurately represents scanned object The dark and white bands in the illustration add up in such a way as to yield an image which accurately represents scanned object Impulse signals from linear traverses following angular rotations 1 st back projection 2 nd back projection 3 rd back projection Final back projection

Practical method of Generation of Attenuation profiles Attenuation profiles are a recording of X-ray attenuation verses position in the object. Attenuation profiles are a recording of X-ray attenuation verses position in the object. Attenuation is related to the ratio between incident intensity (I 0 ) and transmitted intensity (I t ) Attenuation is related to the ratio between incident intensity (I 0 ) and transmitted intensity (I t ) I0I0 ItIt

X-ray intensity is measured by a DETECTOR which converts x-ray photons to electrical current X-ray intensity is measured by a DETECTOR which converts x-ray photons to electrical current The reference (Input) detector measures the incident intensity (I 0 ) The reference (Input) detector measures the incident intensity (I 0 ) The output detector measures the transmitted intensity (I t ) The output detector measures the transmitted intensity (I t ) I 0 = incident intensity = tube output ItIt

I 0 should remain constant I 0 should remain constant I t will change depending on the attenuation of the object scanned (e.g. centered, round object of uniform density) I t will change depending on the attenuation of the object scanned (e.g. centered, round object of uniform density) I o =1000 I t =1000 ItIt

Calibration profile If the object scanned is a centered, round water bath each attenuation profile is called a CALIBRATION PROFILE. and the set of profiles are called the Calibration File or Cal file If the object scanned is a centered, round water bath each attenuation profile is called a CALIBRATION PROFILE. and the set of profiles are called the Calibration File or Cal file

DATA Calibration Subtraction of calibration file from the attenuation profiles of the object called data calibration Subtraction of calibration file from the attenuation profiles of the object called data calibration (consider an object containing two areas of different densities as shown) (consider an object containing two areas of different densities as shown)

Scan Data File The difference profile are stored as numerical values as a function of position in the profile. The difference profile are stored as numerical values as a function of position in the profile. This data is used in the image reconstruction process This data is used in the image reconstruction process The set of difference profiles for a complete scan is called SCAN DATA FILE or raw data The set of difference profiles for a complete scan is called SCAN DATA FILE or raw data

Back projection of scan data 1. Back projection of the corrected attenuation profiles is accomplished by feeding the numerical values of each point along the profile into a matrix. (Since the scanning motion is circular, the matrix is usually round) --

2. Close rays are averaged to attain a pixel value from the ray that passes closest to its center 3. Every pixel is assigned one value per view. 4. The final value is the sum of these divided by the number of views (as shown)

6. When all the measurements have been projected back into the matrix and all pixel values determined, the image can be displayed

7. The number of pixels in the image corresponds to the MATRIX SIZE. This is given as 256 x 256, 320x 320, 512 x 512. this refers to the number of pixels across a vertical or horizontal diameter of the image. This is given as 256 x 256, 320x 320, 512 x 512. this refers to the number of pixels across a vertical or horizontal diameter of the image.

8. The actual size of a pixel is equal to the size of the area scanned, called the FIELD OF VIEW (FOV) divided by the matrix size. ( Small pixel sizes helps to display small objects that can be resolved by the system) ( Small pixel sizes helps to display small objects that can be resolved by the system)

CT Numbers CT numbers are calculated from the measured attenuation values using the equation given below. CT numbers are calculated from the measured attenuation values using the equation given below.

Range of CT numbers Tremendous range of attenuations that might be displayed in a CT image requires a large number scale. +/ is commonly used and some systems have extended this range further on the high end to differentiate bone densities Tremendous range of attenuations that might be displayed in a CT image requires a large number scale. +/ is commonly used and some systems have extended this range further on the high end to differentiate bone densities

Limitation in display of CT numbers The human eye can distinguish only about 20 levels of gray in the gray scale and cannot appreciate the large range of information contained in CT scan The human eye can distinguish only about 20 levels of gray in the gray scale and cannot appreciate the large range of information contained in CT scan This is solved by using the ‘window width’ & ‘window level’ This is solved by using the ‘window width’ & ‘window level’

Window width This is the range of CT numbers that is displayed using the useful levels of gray scale This is the range of CT numbers that is displayed using the useful levels of gray scale Small window separates one CT number from another Small window separates one CT number from another Large window condenses more than one CT number in each of gray level Large window condenses more than one CT number in each of gray level

Window level This is the centre of the range of CT numbers displayed by the window This is the centre of the range of CT numbers displayed by the window The level control moves the visible gray scale up and down the CT number scale The level control moves the visible gray scale up and down the CT number scale

Hounsfield unit system & CT numbers As an honour to Hounsfield who was the pioneer of invention of CT the CT numbers are called Hounsfield units. As an honour to Hounsfield who was the pioneer of invention of CT the CT numbers are called Hounsfield units. The range of Hounsfield units Is from to The range of Hounsfield units Is from to Standard reference points are for Air and 0 for water. Standard reference points are for Air and 0 for water.