EBL: Focus & height rev 0, 6/6/08. Electron gun ZrO/W emitter Suppressor First anode Second anode Acceleration electrodes Ground anode First alignment.

Slides:



Advertisements
Similar presentations
Basic Electron Microscopy
Advertisements

Edge Sensors for the Schwarzschild-Couder Telescopes David Hanna McGill University UCLA AGIS Meeting June 26-27, 2008.
Lecture 11. Microscopy. Optical or light microscopy involves passing visible light transmitted through or reflected from the sample through a single or.
Electron Optics.
Electron Microscopy for Catalyst Characterization Dr. King Lun Yeung Department of Chemical Engineering Hong Kong University of Science and Technology.
Scanning Electron Microscope (SEM)
Groups: WA 2,4,5,7. History  The electron microscope was first invented by a team of German engineers headed by Max Knoll and physicist Ernst Ruska in.
MICROSCOPES.
SEM Magnification Calibration. Magnification Errors Proper calibration of the SEM scans (magnification) is primary to metrology. SEM Magnification requires.
MICROSCOPES. Microscope Quiz Friday –Jan. 28 n Label parts of microscope n How to use (ex: use coarse knob to find object, adjust diaphragm for light)
Components of Initial Eprobe Startup Electron Gun –turning on beam, setting beam current, acceleration, saturation level Focusing lenses –Permit focussed.
Fixed Beam Moving Stage using a stitching free exposure strategy.
Electron Optics Two essential components: 1)Electron source (gun) 2)Focusing system (lenses) Add scanning apparatus for imaging Electron gun Cathode Anode.
Electron Microscope. Light Resolution  The resolution of a microscope is limited by the diffraction of light. Single diffractionSingle diffraction 
Optical Alignment with Computer Generated Holograms
Lab meetings Week of 6 October
Scanning Electron Microscope Jamie Goings. Theory Conventional microscopes use light and glass lenses SEM uses electrons and magnetic lenses to create.
Surface Emissions Specimen current X-rays Cathodoluminescence Pole Piece, etc SE3  ≈ 1 nm for metals up to 10 nm for insulators.
Imaging Science Fundamentals Chester F. Carlson Center for Imaging Science Display Systems Viewing Images.
Introduction to scanning electron microscopy
SEM microscope By: Doug, Holly & Oleg. Scanning Electron Microscope vs. Optical Microscope Advantages Continuously variable magnification High resolution.
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Factors affecting the depth of field for SEM Afshin Jooshesh.
©2001 Timothy G. Standish Psalm 119:99, I have more understanding than all my teachers: for thy testimonies are my meditation. 100 I understand more.
Scanning Electron Microscopy
B. Spatial coherancy & source size Spetial coherancy is related to the size of the source. Source size governs spatial coherancy and maller source sizes.
Notre Dame extended Research Community NANOWeek: The Power of Microscopes Optics Visible light Optical microscopes and telescopes Scanning electron microscope.
Electron Microscopy 1 Electron Microscopy (EM) Applying Atomic Structure Knowledge to Chemical Analysis.
Electron Microscopes Used to count individual atoms What can electron microscopes tell us? Morphology – Size and shape Topography – Surface features (roughness,
Scanning Electron Microscope (SEM)
JEOL JBX-9300FS Electron Beam Lithography System Training
Page 1 Workshop for Electron Beam Lithography System JBX-6300FS By Nelson LI 13 November 2009 Nanoelectronics Fabrication Facility.
Scanning Electron Microscope (SEM)
NANO 225 Micro/NanoFabrication Electron Microscopes 1.
Tools in Science Miss Colabelli. Tools & Techniques Tools are objects to improve the performance of a task. Microscopes are tools that extend human vision.
Transmission Electron Microscope Basic premise of a TEM is to project a magnified image of the specimen onto a fluorescent screen where it can be viewed.
Reminders for this week Homework #4 Due Wednesday (5/20) Lithography Lab Due Thursday (5/21) Quiz #3 on Thursday (5/21) – In Classroom –Covers Lithography,
Microscope Basics.
NANO 230 Micro/Nano Characterization
Notre Dame extended Research Community NANOWeek: The Power of Microscopes Optics Visible light Optical microscopes and telescopes Scanning electron microscope.
SEM- Schematic Overview. Electron Detection Tungsten Filament Electron Source.
SEM Scanning Electron Microscope
Notre Dame extended Research Community NANOWeek: The Power of Microscopes Optics/Use of light Optical microscopes and telescopes Scanning electron microscope.
Questions/Problems on SEM microcharacterization Explain why Field Emission Gun (FEG) SEM is preferred in SEM? How is a contrast generated in an SEM? What.
Scanning Transmission Electron Microscope
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
METHODOLOGY Nanotechnology Prof. Dr. Abdul Majid Department of Physics
SEM Imaging - Magnification and Deflection System
Scanning Electron Microscope Eee-Jay Rodriguez. The Structure of the Microscope Inside Outside.
BY GAJENDRA KUMAR ID- 2011uit1721. WHAT IS SEM  It is a microscope that uses a focused electron probe to extract structural and chemical information.
Introduction to scanning electron microscopy
Presentation on SEM (Scanning of Electron Microscope) Represented by:-Ravi Kumar Roll:- (BT/ME/1601/006)
Introduction to scanning electron microscopy
Laboratory equipment Lecture (3).
Electron-beam lithography with the Raith EBPG
Electron-beam lithography with the Raith EBPG
Introduction to Scanning Electron Microscope by Sameer S
Electron-beam lithography with the Raith EBPG
METALLURGICAL MICROSCOPE
NANO 230 Micro/NanoFabrication
Electron-beam lithography with the Raith EBPG
Topic 1: Introduction to Histology
Measure the width of date and D on a dime
PHOTOCATHODE STUDIES Electron Energy Analyzer NEA Cathodes (GaAs)
Electron-beam lithography with the Raith EBPG
Do it with electrons !.
A C B D 1. Read the information about different sorts of microscope.
Types of Microscopy Type Probe Technique Best Resolution Penetration
SEM Imaging - Magnification and Deflection System
Scanning Electron Microscopy (SEM)
Presentation transcript:

EBL: Focus & height rev 0, 6/6/08

Electron gun ZrO/W emitter Suppressor First anode Second anode Acceleration electrodes Ground anode First alignment coil Second alignment coil Blanking electrode Blanking aperture Second lens Third lens Zoom lenses Dynamic focus correction electrode Third alignment coil Dynamic astigmatism correction electrode Subsidiary deflector (SUBDEF) Electromagnetism astigmatism correction electrode Main deflector (PDEF) Backscattered electron detector Objective aperture Objective lens Workpiece surface

Lens 4 magnet electron beam the Lens 4 magnet is the objective lens it is used to focus the beam best focus beam blow up due to electrons repelling each other

limited depth of focus

where is your sample? ??? beam spot size will change on your sample smaller spot = better pattern resolution

500um field vector scan

correct field field too small field too big

ways to focus SFOCUS manual HEIMAP virtual chip mark height detection

ways to focus SFOCUS uses AE mark to measure beam width and sets focus to minimum beam diameter can be considered in a sense as the "home" reference height does not focus on your sample

ways to focus manual use SEM mode and EOS to focus on your sample can be used for optically transparent substrates potentially more accurate than height detection system, but if your sample has height variation, it is only good at location where you focused does not adjust field distortion correction values, you will get field stitching errors. can use HEIGHT and SUBHEI commands to correct for field distortion, but must manually calculate

ways to focus HEIMAP uses white light height detection system sample must be optically opaque define an array, takes average value, uses this to set focus for entire sample must include in PATH to take effect for your job spot size of light is ~3mm, cannot measure near edge of sample windows average value may not be good enough if your sample has > 5um height variation

ways to focus virtual chip mark height detection same white light height detection system as HEIMAP sample must be optically opaque measures height at location of chips and sets focus individually for each chip whereas HEIMAP uses single average value for all chips cannot measure height near sample window edge (may be a problem on windows like 3A)

example of how height/focus variation affects field stitching on a pattern

SEM inspection right corner bottom corner test pattern

bottom corner right corner

+x gain -x gain original 500um field boundary -y gain +y gain JEOL HEIGHT command HEIGHT f,gx,gy,rx,ry,sx,sy gx = x gain gy = y gain

-y gain +y gain no gain -x gain +x gain no gain

-y gain +y gain 0 gain -x gain +x gain

x direction stitching (focus values shown in microns below reference) best focus

y direction stitching (focus values shown in microns below reference) best focus probably halfway between -10um and -7um focus

1um change in height = 12nm change in field stitching