Newport HT50 Ventilator Orientation & Training

Slides:



Advertisements
Similar presentations
QUARTERLY SCBA Practice Part 1
Advertisements

Trouble Shooting (Mechanical Ventilation)
SECTION 11 LEVEL B SUIT OUT. SECTION 12 AIR MONITORING.
HFOV – Equipment Required (Filtered circuit)
Bump-test and Calibration Procedures
(PLEASE SWITCH TO THE SLIDE SHOW MODE)
Flight 60 vs. HT Flight Medical Confidential 1.
LogTag Training January 2015
TLA. Replacing The Battery On the Back of the pager press in this button while moving the door away from the pager, then lift up. On the Back of the pager.
Model 754 Portable Ventilator
 Description of PressurePro System  Identify system components  Display and Operational Modes  Sensor alerts  Checking of tire pressures  Manual.
BE-2110i VF/WIDEBAND TRANSMISSION TEST SET OPERATING OVERVIEW Slide 1 of 9.
1 © 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part, except for use as permitted in a license.
 Before you can begin mechanical ventilation your ventilator must be correctly assembled ◦ Step 1: Assemble expiratory and inspiratory filters on the.
1 LIFECARE PLV-100 by Bryce Younger. 2 Classification ► Electrically powered  microprocessor controlled ► Volume-control, time cycled ► Rotary drive.
Initiating Positive Pressure CMV A generic approach!
3100B Theory of Operation and Controls. VIASYS Healthcare, Inc. 3100B Theory of Operation and Controls Approved for sale outside the US in 1998 for patients.
pNeuton Transport Ventilator
Clean Air Box Usage & Care CABs RAMs CO Monitors.
MSA - ALTAIR 4X Gas Meter.
Troubleshooting and Problem Solving
Initiation of Mechanical Ventilation
Bmd Crossvent 3.
OXYTRONIC 600 CONTINOUS FLOW NITROX SYSTEM
RSPT 2414 Mechanical ventilation Review Unit 3 classifications By Elizabeth Kelley Buzbee AAS, RRT- NPS.
Merlin Telemetry Board Dräger PSS 7000 Q & A
DataLogger Operation - 1 DataLogger ® Operation Procedures.
AHP300 VENTILATOR Prepared by Caesar Rondina, EMTP, SCT, EMTP, CES
AccuMax Multi-Point Injection Mechanics
1 Life Products LP-6, LP 6 Plus and LP 10 Home Ventilator By Bryce Younger.
Setting up the NICO monitor to measure pulmonary dead space fraction (V d /V t )
1 Elsevier items and derived items © 2010 by Saunders, an imprint of Elsevier Inc. Chapter 19 Mechanical Ventilation of the Neonate and Pediatric Patient.
MySOS Device User Training
Trouble Shooting (Mechanical Ventilation) Arthur Sadhanandham Medical ICU, CMC.
Ventilators All you need to know is….
Dr Chaitanya Vemuri Int.Med M.D Trainee.  The choice of ventilator settings – guided by clearly defined therapeutic end points.  In most of cases :
SureVent 2 - SV2® Gas Powered Automatic Ventilation Resuscitator
ASSISTED VENTILATION By: Dr.Saif Assistant Professor Of Paediatrics Allied Hospital Faisalabad.
The Unit Between The Device & Patient (NeoPAP)
Without reference, identify principles about volume/pressure and high frequency ventilators with at least 70 percent accuracy.
© 2008 Masimo Corporation RAD-57 EMS Carry Case Lightweight: 13 oz. Small: 6.2” x 3.0” x 1.4” Portable: operates on 4 “AA” batteries Supplied with high-visibility,
Final Considerations in Ventilator Setup Chapter 8.
MSA Five Star 4-Gas Monitor Air Monitoring Instrumentation.
3100B Theory of Operation and Controls. SensorMedics 3100B u Electrically powered, electronically controlled piston-diaphragm oscillator u Paw of 5 -
BMC Medical Co., Ltd. Nov Training Course. 1 Table of Contents 1. CATALOGUE 2. INSTALL& DISASSEMBLE 3. SET UP.
1 Respironics BiPAP-Vision. 2Classification Electrically powered – internal battery powers Vent Inoperative & audible alarms if AC power lost. Error code.
Basic Concepts in Adult Mechanical Ventilation
CLINICAL ALGORITHM FOR THE MANAGEMENT OF INTUBATED PATIENTS PRESENTING WITH CHANGES VISSIBLE ON CxR Next step in the algorithm.
3 nd LECTURE VENTILATORS Part One. Ventilators One of the major life support systems. Ventilators take over the vital role of the respiratory muscles.
BMC Medical Co., Ltd. Nov Training Course. 1 Table of Contents 1. CATALOGUE 2. INSTALL& DISASSEMBLE 3. SET UP.
Ventilatory Modes. Volume Controlled Mandatory Breath Gas is delivered at a constant flow until the set tidal volume is reached. Pressure rises to a.
Mechanical Ventilator
NEW 3.0 SOFTWARE FOR SERVO-i AND SERVO-s Servo-i Ventilator
PRESSURE CONTROL VENTILATION
The Crossvent 2i+ option The Crossvent 2i+ is available in 2 different configurations: Model 2200JC Built-in model for retrofitting incubators that currently.
MSA Passport  Personal Alarm Combustible gases and vapors Oxygen enriched Oxygen deficient Specific Toxic gases Alarm levels are set at the factory.
Mechanical Ventilation
PEEP Residual Volume Forced Vital Capacity EPAP Tidal Volume A-a gradient IPAP PaCO2 RR ARDS BIPAP BiPAP NIV PaO2 IBW Plateau Pressure FiO2 A/C SIMV.
Pressure Regulated Volume Control Mode
Simplified Instruction for use* Evita V300
F60 iO2 and Advanced Modes Assaf Weiss, Product Manager
MACS CPAP System Self Guided Tour.
Mechanical Ventilation
Andrea Ferro Tel: Mob: Skype: andrea.ferro84.
Basic Concepts in Adult Mechanical Ventilation
Simplified Instruction for use* Evita V300
Digital Chart Recorder Operation
Ventilation strategies
Transport Ventilator and Beyond!
Puritan Bennett™ 560 Ventilator
Presentation transcript:

Newport HT50 Ventilator Orientation & Training Pandemic Working Group RT Education Subcommittee

Newport HT50 Easy to operate Capable of ventilating patients >10 kg Able to operate without a compressed gas source Long lasting internal battery (8-10 hrs) Delivers pressure or volume breaths Public Works and Government Services Canada has purchased HT50 ventilators to be used in a pandemic.

Overview To change any value/ setting, push blue soft key associated with value/ setting until the value flashes. Use up/ down arrows to adjust. To lock in new value, push soft key again to confirm before selecting another value to adjust. If left alone for approx. 3 seconds, the ventilator will accept the new value automatically If changing between modes of ventilation, the blue soft key needs to be pushed twice. The ventilator will prompt you to “Press again” in message display window. (If not pressed twice, it will revert back to previous setting)

Front Panel The green values displayed (Ptrigger, PEEP, Psupport, f, ti, pressure control or volume control) are settings that can be manipulated using the blue soft keys.

“Blue Soft Keys” Manual Inflation / Calibration Rate Trigger sensitivity Ti PEEP/ CPAP Mode selections: CMV/ SIMV/ Spont Pressure Support Up/ down adjustments On/ Standby Volume (Vt) / Pressure (PIP) controls

The displayed red values (High and low airway pressures) are alarm settings that can be manipulated by using the grey soft keys. The displayed orange values are calculated values based on settings (with the exception of high & low minute volume alarms which are set values) and cannot be changed unless other settings are changed. These values are also selected using the grey soft keys.

“Grey Keys” Alarm Silence/ Reset Internal Battery Flowrate- calculated based on f and Vt I : E ratio High airway pressure Low airway pressure High minute volume Low minute volume

Indicators Pressure gauge Alarms Message display window

Fresh Gas Intake & filter- Filter should be changed when dirty or contaminated Oxygen Blender/ FiO2 adjustment- No O2 analyzer built in therefore, an inline oxygen analyzer must be used. A/C connection- should be plugged in at all times to charge internal battery. Fully charged the battery offers 10 hrs of operation. Low battery alarm will sound when battery life is down to approx. 20%.

Humidification/ Oxygenation The HT50 ventilators purchased by Public Works and Government Services Canada will not come with a heated humidifier. We will need to use a HMEF with these ventilators. The HT50 ventilators will have an oxygen blender attached to the fresh air intake that will be used to set FiO2. We will need to use an external oxygen analyzer with these ventilators as there is no internal analyzer.

Ventilator Set-up Gas Output/ circuit connection Proximal airway pressure line connection Exhalation Drive line connection

With Circuit Attached…

Calibrate the exhalation valve prior to use as follows: Press On/Standby once to turn on ventilator Occlude end of circuit or attach circuit to an adult test lung Press Manual inflation TWICE within 3 seconds to begin Calibration procedure - ventilator will perform calibration and display ‘cal completed’ when finished. If calibration fails ( ‘cal failed’), press silence/reset and check circuit & test lung for leaks then press manual inflation twice to perform calibration again. You can now set desired ventilation parameters.

Breath type… volume or pressure? To choose Volume controlled ventilation : Select Mode (A/CMV or SIMV) Press volume control soft key Use arrows to set desired Vt Press volume control soft key again to confirm. To choose Pressure controlled ventilation: Press pressure control soft key Use arrows to set desired Pressure control level Press pressure control soft key again to confirm. *Pressure set is above ambient (atmospheric) in PC*

Modes… A/CMV Can be used with pressure or volume ventilation Set: PEEP f Ti Vt or PIP (above ambient P) Alarms Press On/Standby to start ventilator then attach circuit to patient.

SIMV Can be used with pressure or volume ventilation Set: SIMV mode Vt or PIP f Ti Ptrig PEEP Psupport (above PEEP) Alarms (Vi and Paw Press On/Standby to start ventilator then attach circuit to patient .

Spont Can be used to deliver CPAP or BiPAP (invasive or noninvasively) Set: Spont Ptrig PEEP/CPAP Psupport Alarms Press On/Standby to start ventilator then attach circuit to patient

Message Display Window Displays alarm details as well as measured patient parameters When ventilator is ventilating: press the Up arrow once to display Vt, Vi, f Press the Up arrow twice to display airway pressures (p- peak, m- mean, b- base (PEEP) *Measurements are updated breath by breath (Pressing the up arrow a third time will display total ventilator hours) If there is an alarm condition displayed, it will need to be fixed and reset before pt. measurements can be displayed. .

Alarms When an alarm is active there will be a flashing light in the alarm panel and further info printed in the display window. Pressing silence/reset soft key will silence the audible alarm for 60 seconds. The alarm indicator and displayed message will remain when the alarm condition no longer exists. Press silence/reset to clear the alarm panel and message display.

Minute volume alarms . Set using Grey softkeys Press appropriate grey softkey, use arrows to adjust and then press softkey again. Alarm will be indicated in alarm panel and on message display window If Low Vi alarm is activated and not immediately corrected, back up ventilation will start. Therefore… Ensure low Vi alarm is set appropriately . . .

Back Up Ventilation If minute volume falls below Vi alarm limit, back up ventilation will begin and ventilator will continue to alarm. In A/CMV or SIMV mode: all settings will remain the same with the exception of f. f will increase to 1.5 times the set f, or a minimum of 15 In Spont mode: ventilator will begin in SIMV mode with a Psupport of 25 and f of 15 with other settings unchanged The ventilator will revert back to previous mode/ settings once the back up ventilation has achieved minute ventilation above the alarm limit. .

Testing Internal Battery Unplug ventilator to activate internal battery (Internal battery indicator will light and alarm will sound). Silence alarm. Press and hold Int. Battery softkey Battery strength will be displayed on pressure gauge. Battery strength gauge Press to test internal battery