 Before you can begin mechanical ventilation your ventilator must be correctly assembled ◦ Step 1: Assemble expiratory and inspiratory filters on the.

Similar presentations


Presentation on theme: " Before you can begin mechanical ventilation your ventilator must be correctly assembled ◦ Step 1: Assemble expiratory and inspiratory filters on the."— Presentation transcript:

1

2  Before you can begin mechanical ventilation your ventilator must be correctly assembled ◦ Step 1: Assemble expiratory and inspiratory filters on the ventilator. (ventilators vary, so this process will also vary and be taught during lab) ◦ Step 2:Attach circuit to ventilator. Acute ventilator circuits will have an inspiratory limb and an expiratory limb so that the circuit is unbroken and the patients returned volumes are assessed on the vent. (long term vents typically have a inspiratory limb only)

3 A expiratory filter with H2O trap for condensation- this is for the 840 ventilator. Attached to the ventilator, prevents H2O from entering the machine A inspiratory filter, attached to the vent and inspiratory limb, prevents small particulates from the machine to the patient

4 Subacute/long term vent circuit, the patients exhalation does not go back into the ventilator. Uses a separate pressure line Pt exhalation Pressure line To pt

5 Wye Inspiratory limb Expiratory limb Extra tubing for heater Heater probe inlet To Pt

6  Step 3: Once circuit is attached, close all open ports on the circuit. Circuits have ports for heated humidification temperature probes  If circuit is a heated circuit attach humidifier and probes  Most adult ventilators are setup with an HME initially. All neonatal and pediatric circuits are setup with heated humidifiers with heated circuits  If you setup a heater, but do not have a heated circuit, you must use water traps inline to catch the condensation that will develop

7 Non-heated circuit with water traps Water trap

8 Concha Heater. Heats up a metal column water Passover heated HME, placed at Wye

9  Ventilator circuits are age specific. For an adult you must use an adult circuit, pediatric and neonates have specific circuits as well  Neonatal and Pediatric circuits are always heated circuits- meaning they have a coil inside to maintain temperature during humidification  Depending on the ventilator used some circuits will have a pressure line inlet attached to the machine and/or a flow sensor adapter

10  Ventilator circuits should not be changed routinely for infection control purposes, however you can change a circuit if soiled  Circuits are sterile, meaning you should not insert a younker inside the circuit to remove secretions.  Once the circuit is attached you can now plug in the electrical and air/O2 connections to your ventilator…  Ventilators are electrically powered and pneumatically driven

11 Long term vent Transport vent with O2/Air cylinders

12  The use of closed suction catheters should be considered part of a VAP prevention strategy, and they do not need to be changed daily for infection control purposes. The maximum duration of time that closed suction catheters can be used safely is unknown Use only inline suction ballards on vents. They make ETT and Trach sizes

13  Once your circuit, filters and humidification is on and your vent is plugged in to electricity and Air and O2, you can now do the pre-use calibration  This varies greatly with all ventilators, but generally you start by turning on your vent (remember, most vents you will take off the stopper at the WYE when you first turn on the vent)  Once the vent is on you choose SST (short self test) option to run the calibration

14  The calibration will assess the circuits compliance/elastance, check for leaks and proper flow  http://www.youtube.com/watch?v=4p0SppVb GMs http://www.youtube.com/watch?v=4p0SppVb GMs

15  Once it has been determined a patient requires a ventilator you must now attach them and apply the proper settings.  When you first turn on the ventilator you will press either “new patient” or “same patient”, verbiage will vary. Most new vents will save the previous settings in case you transfer a patient and reattach them  During the setup the patient is typically being bagged. The MD may or may not give you the settings.

16  Mode is AC, SIMV or CPAP ◦ Assist control (no breathing, or inadequate breathing, patient sedated…) ◦ SIMV (same reasons as AC, but you expect patient will breathe spontaneously soon, post ops) ◦ Spontaneous/CPAP: Patient is breathing spontaneously ◦ http://www.youtube.com/watch?v=IUZ3Plmz_YQ&f eature=related http://www.youtube.com/watch?v=IUZ3Plmz_YQ&f eature=related

17  PCV: Used for patients where you want to control their pressure limit, set on AC or SIMV ◦ Set inspiratory pressure limit, I-time, FIO2, PEEP, Rate  VC: Used in most patients, control minute volume, set on AC or SIMV ◦ Set tidal volume, flow, FIO2, PEEP, rate  PRVC: Becoming a common mode for most patients, limits pressure and gives a minimum volume, set on AC or SIMV ◦ Set minimum VT, pressure limit, FIO2, rate, peep, I- time

18  Once you decide what mode and breath type, next you input the settings.  VT: set in VC mode, based on IBW, normal range 8-12 ml/kg, restrictive lungs 5-7 ml/kg  Flow: Set to give an appropriate I:E ratio, typically 40-60 LPM, set higher for COPD patients, watch for airtrapping  Pressure Limit: set to achieve an acceptable VT, typically set between 15-25 cmH2O

19  I-time: Decrease when rate is set high, otherwise start around 0.8-1 second  Rate: Initial rates are 8-12 per min. If you suspect or know patient has high PaCO2 you may start higher  FIO2: typically set at 100% if it’s a new patient and you do not know their PaO2, otherwise set per patients FIO2 before vent if it was acceptable  PEEP: typically started at 0, and added once FIO2 is at 60% and patient has refractory hypoxemia, then initiate at 5

20  Pressure Support: If you started in CPAP or SIMV mode, you will set a PSV. Typically around 10-12 cmH2O, but give to increase spontaneous Vte to acceptable ranges and give enough to over come RAW  Sensitivity: Set as either flow or pressure. In either one it is set between 0.5-3, the higher this number the harder it is for the patient to trigger the breath, the lower =auto trigger

21  Flow pattern: In VC mode you can choose how the set flow will be delivered to the patient. Either as a constant flow (Square) or as a decelerating pattern. In PC and PRVC the flow is not set, so the pattern in always decelerating. A constant flow will increase MAP, this will increase oxygenation but increase PIP

22

23  Any time it is desired to limit inspiratory pressure.  High PIPs  Low Pa02  Need high, variable flow rates

24 Goals in ARDS  ARDS goal - control/minimize lung damage  Literature suggests – low alveolar pressure / low Vt strategy – recently more concerned with lung damage caused by over distention –preventing repeated collapse and reopening of alveoli

25 Treating ARDS Treating ARDS  What are your goals for these patients? –O 2 delivery –oxygenation is more affected by MAP –PCV, BiLevel, or APRV may be a better way to approach these patients  peak airway pressure and MAP is controlled  oxygenation is favored  ventilation is sacrificed if it conflicts with controlling pressure

26  Pressure Control is time cycled.  Rapid rise to set pressure and pressure is maintained for the duration of the inspiratory time.  Inspiratory plateau promotes alveolar recruitment and oxygenation.

27 Pressure Time Curve Volume Expiration 12 3 4 56 30 P aw cmH 2 0 Sec Volume Pressure Pressure Time Curve V. Inspiratory Time

28  Use the plateau pressure obtained during volume ventilation as a starting point (if you started in VC first)  Adjust pressure to obtain desired Vt in the range of 5 – 8 cc/kg

29 0 20 40 60 20 40 60 P aw cmH 2 0 V T LITERS “D” = Lower Inflection Point A B “A” represents PEEP “B” Represents Inspiratory Pressure.6.2.4 “C” = Upper Inflection point

30 Setting I-Time in PCV Setting I-Time in PCV  Observing the Flow and Tidal Volume Time curve during PCV can help determine adequate inspiratory time.  Observing the Pressure Time curve will assist in determining if the I -Time is too long

31 12 3 4 56 SEC 12 3 4 56 VTVT 600 cc -20 120 SEC V. LPM 0 450cc

32

33

34  Flow x Time = Volume  Increasing the I time can increase the Vt. Try this before increasing the inspiratory pressure.  Watch for I time too long, causing auto peep and dyschrony.  Keep I time constant to maintain more stable tidal volumes

35 The Pressure Control Mode Pressure Control  Pressure Control –Set high pressure = P I  P I is PEEP compensated and changes as PEEP level changes –Set low pressure = PEEP –Set breath timing parameters = T I, T E, I:E ratio

36 Rise TIme  Commonly referred to as rise %, or rise time  Available in all pressure breaths (PC, PS, and spontaneous)  Tailors inspiratory rise to match patient demand. Does not change I time.

37 Rise Time %  Commonly referred to as rise %, or rise time  Tailors inspiratory rise to match patient demand

38 Rise Time % Pressure Relief Transient Overshoot FAP = 1 FAP = 50 FAP = 100

39 Expiratory Sensitivity (E Sens ) V T Peak Inspiratory Flow 5% 20% 40%