Qualitative, quantitative analysis and “standardless” analysis NON DESTRUCTIVE CHEMICAL ANALYSIS Notes by: Dr Ivan Gržetić, professor University of Belgrade.

Slides:



Advertisements
Similar presentations
NUCP 2371 Radiation Measurements II
Advertisements

Furnace aa. Today is Guy Fawkes Day L’vov platform furnace Sample is placed on platform Temp of platform rises more slowly than that of walls.
Instrumental Chemistry
Formation Characteristics
Raman Spectroscopy A) Introduction IR Raman
Bragg’s Law nl=2dsinΘ Just needs some satisfaction!! d Θ l
X-ray Diffraction. X-ray Generation X-ray tube (sealed) Pure metal target (Cu) Electrons remover inner-shell electrons from target. Other electrons “fall”
EDX-Spectra Simulation
Internal – External Order We described symmetry of crystal habit (32 point groups) We also looked at internal ordering of atoms in 3-D structure (230 space.
PROVIDED BY EMAD BEHDAD SPECIALIZED PROTECTION AND CORROSION OF MATERIALS LECTURERS H.MONAJATI,PHD JAFARI,PHD ISLAMIC AZAD UNIVERSITY OF NAJAFABAD BARANCH.
Molecular Fluorescence Spectroscopy
Spectrum Identification & Artifacts Peak Identification.
Wolff, John A. and Conrey, Richard M. GeoAnalytical Laboratory School of Earth and Environmental Sciences Washington State University Pullman, WA
E nergy D ispersive X -ray Fluorescence Spectrometer E nergy D ispersive X -ray Fluorescence Spectrometer EDX-720 Training.
Catalysis and Catalysts - XPS X-Ray Electron Spectroscopy (XPS)  Applications: –catalyst composition –chemical nature of active phase –dispersion of active.
X Ray Flouresence Analysis (XRF). XRF X-Ray Fluorescence is used to identify and measure the concentration of elements in a sample X-Ray Fluorescence.
Spectroscopy Chapter 7.
Ge 116 Module 1: Scanning Electron Microscopy
Advanced Higher Chemistry Unit 1 Spectroscopy. Spectroscopy  Spectroscopy is used to give information regarding the structure of atoms or molecules.
NON DESTRUCTIVE CHEMICAL ANALYSIS
Gamma Spectroscopy HPT TVAN Technical Training
Corpuscular induced and X-ray or γ -ray induced fluorescence NON DESTRUCTIVE CHEMICAL ANALYSIS Notes by: Dr Ivan Gržetić, professor University of Belgrade.
Atomic Absorption Spectroscopy
Field Methods XRF Copyright © 2013 by DBS. Atomic Spectroscopy for Metal Analysis Introduction to the Principles of Atomic Spectroscopy Atomic spectroscopy:
Spectroscopy (continued) Last time we discussed what spectroscopy was, and how we could use the interaction of light with atoms and molecules to measure.
Gamma ray spectrum, its acquiring and analysis
Seminar Author: Bojan Hiti Mentor: doc. dr. Matjaž Kavčič Determination of trace impurities on Si wafers with x-ray fluorescence.
Stopping Power The linear stopping power S for charged particles in a given absorber is simply defined as the differential energy loss for that particle.
Common types of spectroscopy
Atomic Emission Spectrometry
References Hans Kuzmany : Solid State Spectroscopy (Springer) Chap 5 S.M. Sze: Physics of semiconductor devices (Wiley) Chap 13 PHOTODETECTORS Detection.
Chemical analysis with EDX On Supra40 How to get good quality quantitative chemical analyses - Understanding the involved physical phenomena - Choice of.
Chapter 12 Atomic X-Ray Spectroscopy
Scanning Electron Microscope (SEM) – Shoots a high energy beam of electrons (waves of electrons) at a target. Electron gun Focusing coil Objective lens.
Unknown at first, these photons from innershell transitions have played a vital role in materials analysis.
Quantitative Analysis Quantitative analysis using the electron microprobe involves measuring the intensities of X-ray lines generated from your unknown.
Advanced Analytical Chemistry – CHM 6157® Y. CAIFlorida International University Updated on 9/28/2006Chapter 6Electron Spectroscopy Chapter 6 Electron.
AAS Atomic Absorption Spectrophotometry. AAS – Widely in clinical laboratories to measure elements such as aluminum, calcium, copper, lead, lithium, magnesium,
Fluorometric determination of riboflavin
Fluorometric determination of riboflavin. Introduction.
Chapter 32 Radiochemical Methods. Introduction… Radiochemical methods tend to be labor intensive and generate liquid waste due to the chemical separations.
PHYS 430/603 material Laszlo Takacs UMBC Department of Physics
Energy-Dispersive X-ray Microanalysis in the TEM Anthony J. Garratt-Reed Neil Rowlands.
Chapter 15 Molecular Luminescence Spectrometry Three types of Luminescence methods are: (i) molecular fluorescence (ii) phosphorescence (iii) chemiluminescence.
Plan : intro Characterization of thin films and bulk materials using x-ray and electron scattering V. Pierron-Bohnes IPCMS-GEMME, BP 43, 23 rue du Loess,
SEM- Schematic Overview. Electron Detection Tungsten Filament Electron Source.
Atomic Fluorescence Spectroscopy. Background l First significant research by Wineforder and Vickers in 1964 as an analytical technique l Used for element.
X-ray microanalysis in the electron microscope
Elemental Analysis using Atomic Absorption and Emission Spectroscopy Bodhisatwa Das.
Determination of Concentration Using Spectrophotometry
Low Angle X-ray Scattering (LAXS) for Tissue Characterization Dr M A Oghabian.
Infrared Spectroscopy
The probe Some material used from:. EPMA - electron probe microanalysis Probe signals.
Fluorometric determination of riboflavin
IB NOTES: Modern Analytical Chemistry. Definitions: Qualitative Analysis: The detection of the __________________ but not the __________ of a substance.
Tube Exposure Factors Math Technique Contrast and Density
EBB 245. Materials Characterisation Lecture 3 X-ray Fluorescence Spectrometry (XRF) Dr Zainovia Lockman PPKBSM, USM (lecture presentations/notes.
Interactions of Ionizing Radiation
Introduction to Spectrophotometry
X-ray spectroscopy Repetition, exercises and more
Flame Emission Spectrometry
Spectroscopy Chapter 7.
Infrared Spectroscopy
Instrument Parameters in WDXRF
Atomic Absorption Spectroscopy
Raman Spectroscopy A) Introduction IR Raman
Atomic Absorption Spectroscopy. Atomic absorption spectroscopy is based on the same principle as the flame test used in qualitative analysis.
X-RAY FLUORESCENCE TECHNIQUE EMMANUEL A. BAMIDELE.
Spectroscopy Uses emission and absorption of light by electrons moving between ground and excited state configuration, hence electronic configuration.
Radiochemical Methods
Presentation transcript:

Qualitative, quantitative analysis and “standardless” analysis NON DESTRUCTIVE CHEMICAL ANALYSIS Notes by: Dr Ivan Gržetić, professor University of Belgrade – Faculty of Chemistry

Qualitative, quantitative analysis and “standardless” analysis Every element of the periodic system of elements (except H & He) have its characteristic X-ray emission lines. To be sure that we will obtain line of interest the excitation potential (incident radiation) should be at least three times greater than the energy of emission line in question. For example, if one wants to get Cu K α line (8,04keV) than the excitation potential must be 24 kV or greater.

Qualitative, quantitative analysis and “standardless” analysis Every producer of SEM-EDS, SEM-WDS, WD-XRF, ED- XRF and Handheld ED-XRF has their own program packages for qualitative, quantitative analysis and “standardless” analysis. Usually all these packages have similar purposes, but different names. They are always protected with trademarks. But, once you get acquainted with one set of programs produced by one firm, immediately you gain experience for all of them. Why? Theoretical and scientific background is always the same. The differences appear only for specialized requirements set by users like semiconductor industry, metallurgy, environmental science, inspections, etc.

ED XRF Qualitative analysis Very simple procedure: if one already has a sample one selects: appropriate excitation potential (4-50kV), corresponding current (<2mA) (higher currents Si(Li)-detector cannot sustain) and counting time (~100sec).

ED XRF Qualitative analysis Identification of X-ray emission lines nowadays is fully automated, but no one should take program results as absolutely correct. Overlaps, sum peaks and escape peaks are always present. Escape peak (an artefact): With Si(Li) detectors Si escape peaks occurs 1.74 keV below each 'true' peak.

Escape Peak in EDS spectrum

Overlaps in EDS spectrum

Sum peaks

WD XRF Qualitative analysis Procedure that requires knowledge. If one already has a sample one selects: appropriate diffraction crystal, collimator, increment with (a fraction of 2 Θ ), counting time per increment (~1-2 sec), excitation potential (~60kV) and corresponding current (~50mA). The whole qualitative analysis depends on number of selected diffraction crystals, increment with and counting time. An average time for qualitative analysis is usually between min (5-10min per crystal).

WD XRF Qualitative analysis Identification of X-ray emission lines nowadays is fully automated, but no one should take program results as absolutely correct. Overlaps and peaks of the X-ray tube material are usually present in WDS spectrum.

WD XRF Qualitative spectrum

WD XRF Qualitative spectrum light elements

XRF Quantitative analysis X-ray fluorescence analysis belongs to a group of very advanced analytical methods. It requires wide and profound knowledge background in chemistry and physics. Quantitative analysis its theoretical background and corresponding calculation steps are the same for both EDS and WDS analysis. But, there are differences between calculation approaches for corpuscular and photon (X- ray) induced X-ray fluorescence.

XRF Quantitative analysis In both cases, corpuscular and photon (X-ray) induced X-ray fluorescence, there are three important matrix efects that should be taken in account in the case of quantitative analysis: ◦ Effect of the atomic number (Z), ◦ Absorption (A), and ◦ Enhanced Fluorescence (F). These effect are different when, for example, excitation is performed with electrons or with X-rays because electrons and X-rays have different penetration depths in the „body of sample“ and as a result we have different effects of ZAF in these two cases.

XRF Quantitative analysis - Effect of the atomic number (Z) Generally speaking, the matrix effects in X-ray fluorescence analysis result from the influence of the variations of chemical compositions of the sample matrix on the fluorescent intensity of the wanted element. These effects can manifest themselves either via a difference in the absorption of both the primary and fluorescence radiations in samples of different matrix composition (absorption effect) or via an increase of the radiation intensity (enhancement effect) due to the fluorescence radiation of some of the inter elements. Penetration depth is also important, higher penetration depth means more extensive effects of absorption and fluorescence.

XRF Quantitative analysis – Absorption (A) Absorption: Any element can absorb or scatter the fluorescence of the element of interest. This effect occurs when the variations in the matrix chemical composition result in changes of the mean absorption coefficients of both the primary radiation of the source and the fluorescence radiation of the wanted element. The absorption effects occurring in a matrix may either decrease or increase the intensity of the fluorescence radiation of the element under determination, depending on whether the matrix composition changes diminish or augment the mass absorption coefficient. A strong decrease of the fluorescence radiation of the wanted element will be observed if the concentrations of disturbing elements of slightly lower atomic numbers become larger.

XRF Quantitative analysis – Enhanced Florescence (F) Enhancement: Characteristic x-rays of one element excite another element in the sample, enhancing its signal. This effect involves an extra excitation of the atoms of the wanted element by the fluorescence radiation of some of the matrix elements, which in this case become interferents. The mechanism of the enhancement effect involves retransmission of the energy of the primary radiation of the source in the form of secondary (fluorescence) radiation of interelements. The energy of this secondary radiation is just slightly higher than the absorption edge of the wanted element (Figure VI.3), the latter will be excited more efficiency than by the primary radiation of the source, whose energy is higher than that of the secondary radiation and, consequently, further from the absorption edge.

Absorption-Enhancement Affects

QUANTITATIVE ANALYSIS Calibration-Standard Methods. The analyte-line intensity from samples is compared with that from standards having the same form as the samples and, nearly as possible, the same matrix. Internal Standardisation. The calibration-standard method is improved by quantitative addition to all samples of an internal standard element having excitation, absorption and enhancement characteristics similar to those of the analyte in the particular matrix. The calibration function involves measuring the intensity ratio of the analyte and internal standard lines.

QUANTITATIVE ANALYSIS “STANDARDLESS ANALYSIS” Mathematical Corrections. Absorption-enhancement effects are corrected mathematically by the use of influence coefficients for each element present (these are derived experimentally from reference samples). The basic approach is that the XRF intensity at a particular wavelength will in some way be affected by each element in the sample. Standardless analysis – the major advantages of X-ray fluorescence analysis (XRF) – allows fast and easy determination of the chemical composition without performing a calibration. Due to powerful matrix correction based on variable alphas (coefficients for the correction of matrix effects) every kind of sample can be analyzed with optimized measurement parameters for the chemical composition, no matter which kind of sample preparation has been used.

Now it is really THE END