1 X Ray Flouresence Analysis (XRF)

2 XRF X-Ray Fluorescence is used to identify and measure the concentration of elements in a sample X-Ray Fluorescence is used to identify and measure the concentration of elements in a sample

3 XRF instrumental parameters x-ray tube kv x-ray tube mA primary beam filters collimator masks x-ray tube kv x-ray tube mA primary beam filters collimator masks collimator crystal detector path collimator crystal detector path

4 Sampel METAL POWDERLIQUID XRF only XRD and XRF

5 XRF Working Concept In X-ray fluorescence spectroscopy, the process begins by exposing the sample to a source of x-rays. As these high energy photons strike the sample, they tend to knock electrons out of their orbits around the nuclei of the atoms that make up the sample. When this occurs, an electron from an outer orbit, or “shell”, of the atom will fall into the shell of the missing electron. Since outer shell electrons are more energetic than inner shell electrons, the relocated electron has an excess of energy that is expended as an x-ray fluorescence photon. This fluorescence is unique to the composition of the sample. The detector collects this spectrum and converts them to electrical impulses that are proportional to the energies of the various x-rays in the sample’s spectrum.

6 Factor of errors in Sample Preparation  Grain size and surface roughness  Uniformity of sample  Contamination through the sample preparation

7 Grain size and surface roughness

8 Uniformity of sample Metallic Sample Casting condition of the sample in the molding. Sand molding Metal molding X-ray intensities differ according to the molding method which comes In the measurement of light elements. Quenching casting which makes the metallic composition fine produces good results Sample polishing NiK  intensityCrK  intensity 50# emery paper0.6860.974 100# emery paper0.6990.983 240# emery paper0.7040.989 Mirror polishing0.7090.993

9 Uniformity of sample Contamination during polishing Contamination effect when carbon steel and Ni-Cr alloy polish after polishing stainless steel. As the contamination form the polishing belt to the sample, the re contamination from The material of the polishing belt and from the remaining trace elements of polished Sample. NiCrFe % Conc0.550.212.10 % Contamination0.050.030.38

10 Powder Sample Grinding Condition Different grinding condition cause variation in particle size distribution which leads to variation in X-Ray intensity.

11 user benefits of wavelength dispersive XRF versatile accurate reproducible fast non destructive versatile accurate reproducible fast non destructive

12 XRF is versatile  element range is Be to U atomic numbers (Z) of 4 to 92  concentration range covers 0.1 ppm to 100 %  samples can be in the form of solids, liquids, powders or fragments  element range is Be to U atomic numbers (Z) of 4 to 92  concentration range covers 0.1 ppm to 100 %  samples can be in the form of solids, liquids, powders or fragments

13 XRF is accurate  generally better than 1 % relative (i.e. 10% ± 0.1%)  accuracy is limited by calibration standards, sample preparation, sample matrix, sampling, instrumental errors & statistics  generally better than 1 % relative (i.e. 10% ± 0.1%)  accuracy is limited by calibration standards, sample preparation, sample matrix, sampling, instrumental errors & statistics

14 XRF is reproducible generally within  0.1% relative good reproducibility requires high quality mechanics, stable electronics and careful construction techniques generally within  0.1% relative good reproducibility requires high quality mechanics, stable electronics and careful construction techniques

15 XRF is fast counting times generally between 1 & 50 seconds for each element semi-quant analysis of all matrix elements in 10 to 20 minutes overnight un-attended operation counting times generally between 1 & 50 seconds for each element semi-quant analysis of all matrix elements in 10 to 20 minutes overnight un-attended operation

16 XRF is non-destructive standards are permanent measured samples can be stored and re-analysed at a later date precious samples are not damaged standards are permanent measured samples can be stored and re-analysed at a later date precious samples are not damaged

17 properties of x-rays the following four slides list some of the more important properties of x-rays that contribute to the nature of XRF analysis

18 XRF analytical envelope elemental range detection limits analysis times accuracy reproducibility

19 elemental range beryllium (4) to uranium (92) in solids fluorine (9) to uranium (92) in liquids

20 range of elements in solid samples are shown in green (Be to U)

21 range of elements in liquid samples are shown in green (Na to U)

22 detection limits (LLD) function of atomic number (Z) & the mix of elements within the sample (sample matrix) < 1 ppm for high Z in a light matrix (e.g. Pb in petrol) or > 10 ppm for low Z in a heavy matrix (Na in slag)

23 XRF applications summary Na to U in all sample types Be to U in solid samples accuracy generally 0.1 to 1 % relative reproducibility typically < 0.5% relative typical LLD is normally 1 - 10 ppm (depends on element being measured and the sample matrix)

24 XRF errors the following section describes major source of errors in XRF analysis, and investigates how these errors can be minimized to achieve maximize accuracy

25 overview of XRF methodology good accuracy requires careful sample preparation fused beads for light elements accurate standards selection of optimum instrument parameters collection of enough counts to avoid statistical errors

26 Methods of Analysis the following presentation describes the requirements for quantitative and semi-quantitative analysis

27 XRF analytical methods the atomic number (Z) of each of the elements to be determined will have an influence on the type of sample preparation to be used, and the quantitative or semi-quantitative method that will be the most suitable

28 XRF analytical methods the quantitative method is the most accurate, but requires calibration standards semi-quantitative method is less accurate, but does not require standards the quantitative method is the most accurate, but requires calibration standards semi-quantitative method is less accurate, but does not require standards

29 overview of XRF methodology elements to be measured low Z will require careful preparation low Z may have lower accuracy low Z may require fusion of powders semi-quant does not measure the very light elements (Be to N)

30 overview of XRF methodology concentration ranges as the concentration range for each element increased, accuracy generally decreases large concentration ranges will require more standards

31 XRF applications summary Na to U in all sample types Be to U in solid samples accuracy typically 0.1 to 1 % relative typical LLD is between 1 - 10 ppm