Image Quality Bushong Chapter 7.

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

CT Scanning: Dosimetry and Artefacts
Computers and Computed Tomography
Image Reconstruction.
Formation Characteristics
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Primary Exposure Factors IV
Radiographic Quality Chapter 5.
Resident Physics Lectures
RAD 354 Chapt. 28 The Digital Image Spatial resolution Contrast resolution Contrast-detail curve Pt. dose considerations.
RAD 350 Chapter 17Digital Rad Tech. Spatial Resolution – ability to distinguish small items in close proximity with near the same atomic mass density Spatial.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Image Artifacts Chapter 8 Bushong.
Computed Tomography II
Advanced Biomedical Imaging
Dr. Mustafa Zuhair Mahmoud Mr.Ali B Alhailiy بسم الله الرحمن الرحيم 1.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods
Computed Tomography III
Computed Tomography RAD309
tomos = slice, graphein = to write
Unsharpness Calculations and Resolution By Professor Stelmark.
Chapter 2 Stewart C. Bushong
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Conventional and Computed Tomography
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
Seeram Chapter 11: Image Quality
COMPUTED TOMOGRAPHY (I) RAD 365 CT - Scan
Basic principles Geometry and historical development
Digital Image Characteristic
Image Characteristics. What is an image? Dictionary meaning An optical appearance An optical appearance A form of semblance A form of semblance A mental.
Computers in Imaging Q & A Robert Metzger, Ph.D..
Quality Control.
SPECIAL IMAGING/ADVANCED IMAGING INTRODUCTION TO DIGITAL RADIOGRAPHY
Resident Physics Lectures
Image Quality Radiographic Resolution.
Computed Tomography Q & A
PRINCIPLE AND BASIC PHYSICS OF COMPUTED TOMOGRAPHY
Seeram Chapter 9: Image Manipulation in CT
Computed Tomography Physics, Instrumentation, and Imaging
Factors affecting CT image RAD
Exposure Factors or Prime Factors
Radiographic Quality Visibility and Sharpness
Image Quality The capacity to define, measure, and assess image quality is a primarily responsibility of a CT Technologist.
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Part No...., Module No....Lesson No
CT Chapter 4: Principles of Computed Tomography. Radiography vs. CT Both based on differential attenuation of x-rays passing through body Radiography.
Quality Assurance.
Modulation Transfer Function (MTF)
Interaction of x-ray photons (and gamma ray photons) with matter.
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Exposure Factors or Prime Factors
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Chapter 10 Digital Signal and Image Processing
Computed Tomography Basics
Image quality and Performance Characteristics
Quality Control Testing of Screen Speed
Resident Physics Lectures (Year 1)
Basic principles Geometry and historical development
Resident Physics Lectures
Computed Tomography (CT)
Resident Physics Lectures (Year 1)
Presentation transcript:

Image Quality Bushong Chapter 7

Image Quality Image quality can be described by five characteristics Contrast resolution Spatial resolution Image noise Linearity/Distortion Uniformity/Artifacts

What influences image quality? X-ray beam characteristics Dose Transmissivity of the subject/Attenuation characteristics Slice thickness Scatter Efficiency of analog-to-digital conversion Pixel size Reconstruction algorithm Display resolution

Image Quality Specification of image quality is usually very subjective and is described by such generic terms as detail, sharpness, or blurring. E.g. “the images of the patient appear blurry because she wouldn’t hold her breath”

Image Quality Image quality cannot be represented by a single number There isn’t one all encompassing method that can accurately measure image quality Image quality can be represented by several numbers Image quality can be described numerically by noise, linearity, and uniformity

Contrast Resolution Contrast resolution describes the property of distinguishing between similar tissues, for example gray-white matter in the brain, or differentiating between the liver and spleen. Plain film radiography tends to lack contrast resolution (it has great spatial resolution however)

Plain Film vs. CT Contrast Resolution Examine the following images The liver is very hard to distinguish by looking at the radiography The CT does a much better job at demonstrating anatomical structures associated with the liver This is because liver tissue is similar in density and CT is superior at detecting objects with similar contrast values

Radiograph vs. CT

Contrast Resolution Determined by tissue atomic number (Z), mass density, and electron density Tissues with large difference in atomic number will have high contrast Lead (atomic number 82) Oxygen (atomic number 8) These items have high contrast (basically appearing white and black on a radiograph)

Contrast Resolution Tissues with large differences in mass density will also have high contrast Density is a measure of mass per unit of volume The SI unit is kg/m3 but most people use g/cm3 because it is more convenient Sodium = 968 kg per cubic meter Lead = 11340 kg per cubic meter There would be a large difference in contrast between these two elements

Contrast Resolution During radiographic imaging, contrast resolution is improved with reduced scatter radiation and lower kVp Scatter only contributes to noise and decreases overall image quality When kVp is increased the resultant image has a long scale of contrast (appears very grey with little differentiation between tissue types)

Contrast Resolution CT employs a high kVp because this helps to minimize patient dose by reducing the amount of radiation that is absorbed the patient CT is able to maintain contrast resolution because of the narrow x-ray beam collimation which provides excellent scatter radiation rejection

Contrast Resolution The larger the dynamic range, for example 1024 versus 512, the less contrast in the image At low spatial frequencies the modulation transfer function (MTF) is a measure of contrast resolution Contrast resolution is the ability to image adjacent tissues having a similar mass density and effective atomic number

Contrast Resolution Contrast resolution is improved by using higher mA Contrast resolution is improved by imaging thicker slices Contrast resolution is improved when imaging smaller patients Contrast resolution is better with low noise imagers

Contrast Resolution Contrast resolution is improved with larger field of view (FOV) and smaller matrix size, hence, larger pixels Contrast resolution can be improved with a smoothing reconstruction filter

Spatial Resolution Spatial resolution refers to the ability to faithfully reproduce small objects having high subject contrast The bone – soft tissue interface represents very high subject contrast The liver – spleen interface represents very low subject contrast Spatial resolution is often described by the degree of blurring in an image

Spatial Resolution A bone – soft tissue interface will be very sharp and is described as a high spatial frequency interface High spatial frequency objects are more difficult to image than low spatial frequency objects Small, high contrast objects are more difficult to image than large, low contrast objects

Spatial Resolution Larger pixel size results in poorer spatial resolution Lower subject contrast results in poorer spatial resolution Larger detector size results in poorer spatial resolution Larger prepatient and postpatient collimation results in more scatter radiation

Scatter Radiation Scatter radiation reduces contrast and results in less contrast resolution A smaller x-ray tube focal spot improves spatial resolution because of the sharper image projection, not the geometry of a shadow Point response function (PRF) is one method of evaluating spatial resolution

Spatial Resolution Edge response function (ERF) is one method of evaluating spatial resolution Line spread function (LSF) is one method of evaluating spatial resolution The Fourier transform (FT) is a mathematical manipulation to convert an intensity versus distance relationship into an intensity versus 1/distance (spatial frequency) relationship

Spatial Resolution The FT of a PRF results in the MTF The FT of an ERF results in the MTF The FT of a LSF results in the MTF Spatial resolution is best described by the limiting spatial frequency (lp/cm) The MTF is obtained from the FT of the PSF, LSR, or the ERF The MTF is useful when evaluating components of a system or comparing similar systems

Spatial Resolution Spatial resolution is improved with a smaller x-ray focal spot size Spatial resolution is improved with thinner section imaging Spatial resolution is improved with narrow predetector collimation Spatial resolution can be improved with the use of an edge enhancement (convolution) reconstruction algorithm

Spatial Resolution Spatial resolution is improved by increasing the number of projection profiles acquired per scan Spatial resolution is improved when small FOV or larger matrix size is employed At high spatial frequencies the MTF is a measure of spatial resolution The MTF is the principal means of expressing CT spatial resolution

Spatial Resolution To understand MTF, one must first understand spatial frequency, which has units of line pairs/mm (lp/mm) One line and a line-sized interspace is a one line pair (lp) High spatial frequencies represent small objects Low spatial frequencies represent large objects

Spatial Resolution An MTF value of 1.0 represents an absolutely perfect image As MTF value is reduced, image blur increases and therefore image quality is reduced Usually, the spatial frequency at the .1 (10%) MTF is identified as the limiting resolution The total MTF of an imager is the product of component MTFs

Spatial Resolution Computed Tomography imagers are capable of approximately 10 lp/cm (1 lp/mm) in normal mode and up to approximately 20 lp/cm (2 lp/mm) in the high resolution mode Z-axis resolution is better with spiral CT compared to conventional CT

Image Noise Contrast resolution is limited by image noise Scatter radiation results in image noise An increase in slice thickness results in less noise An increase in slice thickness usually results in lower patient dose An increase in pixel size results in less noise

Image Noise An increase in patient dose results in less noise In statistics, noise is called standard deviation High noise images appear blotchy, grainy, or spotty Low noise images appear very smooth Noise in a CT image comes from the scanner electronics and the random nature of x-ray interaction with a detector

Image Noise Anything that reduces CT noise will improve contrast resolution Increased image noise at low mAs can usually be accommodated by using a low-pass convolution filter Image noise can be reduced by using 360 degree interpolation

Linearity When water = 0, bone = 1000, and air = -1000, a CT imager exhibits perfect linearity The CT number for a given tissue is determined by the x-ray linear attenuation coefficient (LAC) A plot of CT number versus LAC should be a straight line passing through water = 0

Linearity Good linearity is essential for quantitative computed tomography (QCT) Linearity is the ability of the CT image to assign the correct Hounsfield unit (HU) to a given tissue Linearity is monitored by imaging the 5-pin insert of the American Association of Physicists in Medicine (AAPM) and plotting HU versus linear attenuation coefficient

Uniformity When a test object made of one substance is imaged, the value of each pixel should be the same Cupping is the reduction in CT number toward the middle of a uniform test object Peaking is the increase in CT number toward the middle of a uniform test object Cupping and peaking are signs of poor image uniformity

Uniformity Image uniformity is essential for QCT Uniformity is the ability of the CT imager to assign the same HU to a uniform phantom (water) over the entire FOV

Contrast Detail Contrast detail plots are instructive when evaluating contrast resolution and spatial resolution Contrast detail plots are useful when comparing different scanning protocols The low contrast region of the contrast detail curve is noise limited The high contrast region of the contrast detail curve is determined by the MTR of the imager components