Topic 8. Gamma Camera (II)

Slides:



Advertisements
Similar presentations
Instruments for Radiation Detection and Measurement Lab # 4.
Advertisements

NUCP 2371 Radiation Measurements II
Chapter 9. PN-junction diodes: Applications
Image Reconstruction.
Formation Characteristics
Study of plastic scintillators for fast neutron measurements
Geiger-Muller detector and Ionization chamber
INSTITUT MAX VON LAUE - PAUL LANGEVIN Fast Real-time SANS Detectors Charge Division in Individual, 1-D Position- sensitive Gas Detectors Patrick Van Esch.
Chapter 8 Planar Scintigaraphy
1. 2  Ray Imaging u Inject a pharmaceutical labelled with a radioactive nuclide  Detect where  -rays are coming from u Local concentration in patient.
Medical Imaging Mohammad Dawood Department of Computer Science University of Münster Germany.
QC And NEMA In The Nuclear Arena
Computed Tomography RAD309
The Camera and Photography
Lecture 2-Building a Detector George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
Instruments for Radiation Detection and Measurement
Radiation Sensors Zachariadou K. | TEI of Piraeus.
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
Radiation Detection and Measurement II IRAD 2731.
lecture 2, linear imaging systems Linear Imaging Systems Example: The Pinhole camera Outline  General goals, definitions  Linear Imaging Systems.
Seeram Chapter 11: Image Quality
8/18/2015G.A. Fornaro Characterization of diffractive optical elements for improving the performance of an endoscopic TOF- PET detector head Student: G.
Quality Control.
EDS Energy Dispersive Spectroscopy
Innovation is in our genes. 1 Siemens Medical Solutions Molecular Imaging What are SPECT basics?
Fundamental Limits of Positron Emission Tomography
Scanning Electron Microscope (SEM)
Factors affecting CT image RAD
Energy-Dispersive X-ray Microanalysis in the TEM Anthony J. Garratt-Reed Neil Rowlands.
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.
Active Pixel Sensors in Nuclear Medicine Imaging RJ Ott, N Evans, P Evans, J Osmond, A Clark, R Turchetta Physics Department Institute of Cancer Research.
Li HAN and Neal H. Clinthorne University of Michigan, Ann Arbor, MI, USA Performance comparison and system modeling of a Compton medical imaging system.
8 th Aug 2009 ANL User Workshop Towards Nuclear Spectroscopy using a Position Sensitive Avalanche Photo Diode S. Lakshmi University of Massachusetts Lowell.
Nuclear Medicine Principles & Technology_I
R. Pani Department of Experimental Medicine and Pathology University of Rome La Sapienza-Italy. Flat Panel PMT: advances in position sensitive photodetection.
Medical applications of particle physics General characteristics of detectors (5 th Chapter) ASLI YILDIRIM.
Lecture 3-Building a Detector (cont’d) George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
Modification of Geant4 simulation Detailed crystal structures included reference materials real geometry.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
Calibration of energies at the photon collider Valery Telnov Budker INP, Novosibirsk TILC09, Tsukuba April 18, 2009.
Lecture 12  Last Week Summary  Sources of Image Degradation  Quality Control  The Sinogram  Introduction to Image Processing.
Acquisition time6 min1 min 12 s Collimator height25 mm (Anger)12 mm (HiSens) Detector1 layer, 1 pixel / hole3 layers, 1 pixel / hole3 layers, 4 pixels.
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
Nuclear Medicine Physics & Instrumentation I
Introduction to Gamma Camera
Chapter-2 The Planar Imaging Important points in chapter 2 (chapter 13 from the book) The gamma camera (the basic principles of the gamma camera) The types.
PET Imaging Positron Emission Tomography
Muhammad Musaddiq.
Chapter-5 Positron emission tomography (PET)
Radiation detectors Ion chamber 2. Geiger Muller counter (GM).
Nuclear Medicine Physics
CEPC ScECAL Optimization for the 3th CEPC Physics Software Meeting
Computed Tomography Data Acquisition
Image quality and Performance Characteristics
Image quality and Performance Characteristics
Unit V: Introduction to Gamma Camera
Lecture 1: Development of the Analog Gamma Camera
X-Ray Physics Assumptions:
CLRS 321 Nuclear Medicine Physics & Instrumentation I
Review of Ultrasonic Imaging
Resident Physics Lectures (Year 1)
QC And NEMA In The Nuclear Arena
Nuclear Medicine Physics & Instrumentation I
Resident Physics Lectures
Resident Physics Lectures (Year 1)
Gamma Camera Ilker Ozsahin Oct
HE instrument and in-orbit performance
Assist. Prof. Dr. Ilker Ozsahin Oct
Presentation transcript:

Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance Characteristics of Converging, Diverging, and Pinhole Collimators Measurements of Gamma Camera Performance.

Intrinsic Spatial Resolution Limits Intrinsic spatial resolution refers to the limit of spatial resolution achievable by the detector and the electronics Intrinsic spatial resolution is limited primarily by two factors: multiple scattering and statistical fluctuation in the distribution of light photons (the later is the main factor). Intrinsic resolution depends on detector crystal thickness and γ ray energy. A thinner crystal is used for Anger camera to achieve better spatial resolution. Larger number of PM tubes and improvement of light collection efficiency result in better resolution.

Intrinsic Resolution vs Photon Energy

Intrinsic Resolution vs Crystal Thickness

Detection Efficiency The crystal thickness of NaI(Tl) in Anger camera is smaller (6-12mm) than that of a well counter (2-5cm). Anger camera is designed for optimal detection of γ ray energies of 100-200keV.

Energy Resolution

Problems at High Counting Rates Pulse pile-up is the problem at high count rate and it results in count loss and image distortion. Count loss depends on the whole energy spectrum but the apparent dead-time depends on the window fraction. Scatter causes narrower of the window fraction therefore longer dead-time.

Problems at High Counting Rates Pulse pile-up causes image distortion. Two scattered event may be added to form a photo-peak which produces a location between the two scattered events.

Problems at High Counting Rates The general effect of the pulse pile-up is to cause a loss of image contrast and details

Problems at High Counting Rates Pulse pile-up rejection circuit can be used to improve the pulse pile-up but it will reduce the maximum achievable count rate Deadtime can be improved by shortening the effective charge integration time for the output signal from the PM tube but it will reduce the amount of light collected by the PM tube, therefore, degrade the intrinsic resolution.

Problems at High Counting Rates Count rate performance should be one of the important factors for Anger cameras

Pulse Pile-up Correction

Image Non-Linearity Non-linearity refers to that a straight line object appear as curved line image Non-linearity occurs when the X and Y positional signals do not change linearly with the displacement distance. “Pincushion” distortion is an inward “bowing” image and “Barrel” distortion is an outward “bowing” image A PM tube may have high light collection efficiency in the centre which may result in a “pincushion” in the centre and “barrel” between PM tubes. This kind of images could results in a “wavy” line pattern.

Nonlinearity Example

Image Non-Uniformity Non-Uniformity refers to the intensity variation from a uniform flood source The causes could be the non-uniform detection efficiency (small differences in pulse height spectrum) and non-linearity of PM tubes (more server) or instrument malfunctions. “Edge packing” refers to the bright ring around the edge of the image. It is caused by the internal light reflection of the detector crystal. “Edge packing” is usually masked and the useful field of view is smaller than the actual detector size

Non-Uniformity Correction Techniques Cosmetic approach: Adjust individual PM tube gains (compensate for the detection efficiency difference) “Correction factors” matrix. Test image is normalised and the correction factor for each matrix element is used to add or subtract from the image. Advanced approach: A set of microprocessors are used to store correction matrices for regional differences in pulse height spectra and for position distortion. Images are corrected on event by event basis (on fly).

Non-uniformity Correction Example

Uniformity Profiles

Gamma Camera Tuning

Design and Performance Characteristics of Parallel Hole Collimators Basic limitations in collimator performance Septal thickness Geometry of collimator holes System resolution

Basic Limitation in Collimator Performance Collimator is a limiting factor in camera system performance. Collimator resolution refers to the sharpness or details of the γ ray image projected onto the detector --- worse than the intrinsic resolution. Collimator efficiency refers to the fraction of γ rays that pass through it -- a few percent or less.

Septal Thickness Septal thickness is designed to prevent γ rays from penetrating from one hole to another (allow less than 5% to pass through). Septal thickness should be as small as possible in order to gain maximum efficiency High atomic number and density material should be used (lead is the choice)

Septal Thickness The required thickness t=2dw/(l-w) where (t/w)=(2d+t)/l if θ is small If 5% penetration is allowed, t>[6d/μ]/[l-(3/ μ)] because e-μw<0.05 therefore μw>3

General Comments The septal thickness depends on the γ ray energies to be used because μ depends on the γ ray energy. Energy ranges in nuclear medicine are often classified as: low energy <150keV; medium energy <400keV and high energy <1Mev. Low energy collimators are generally fragile because they are only a few tenth of mm. Low energy collimators are used whenever possible to obtain maximum collimator efficiency (“foggy” background may be superimposed on the image for high energy).

Geometry of Collimator Holes The performance of a collimator is affected by its shape, length and diameter (round or hexagonal are the best) Spatial resolution and detection efficiency are the two important performance parameters Collimator spatial resolution is defined as the FWHM of the radiation profile from a point or line source of radiation projected by the collimator onto the detector.

Collimator Resolution A parallel holes collimator resolution is given by Rc=[d(le+b)]/le where le =l-2μ-1 is the effective collimator length (taking into account of the γ ray penetration)

Collimator Resolution (2) Spatial resolution of a parallel collimator increases (worse) as the distance between the collimator and the source increased

Collimator Efficiency The detection efficiency is given by g=K2(d/le)2[d2/(d+t)2] where K is a constant that depends on the holes shape. The first part is the geometric factor (solid angle subtended by a collimator hole) and the second part is the fraction that is not covered by the septa (the area ratio of the holes and total detector).

Collimator Efficiency Collimator efficiency for a source in air is independent of source-to-collimator distance b.

Collimator Efficiency Invariance of collimator efficiency with source to collimator distance applies to point sources, line sources and uniform sheet source in air with parallel hole collimators.

Resolution and Efficiency The relationship between the resolution and efficiency is gœRc2 For a given septal thickness, collimator resolution is improved only at the expense of decreased collimator efficiency and vice versa.

System Resolution System resolution depends on a number of factors such as, scattering, septal penetration, intrinsic resolution and collimator resolution with the collimator and intrinsic resolutions are the main factors. System resolution (consider intrinsic and collimator only) can be expressed as Rs2=Ri2+Rc2 System resolution is determined primarily by collimator resolution.

Performance Characteristics of Converging, Diverging and Pinhole Collimators

Performance Characteristics of Converging Collimator

Performance Characteristics of Diverging Collimator

Performance Characteristics of Pinhole Collimator

Performance Characteristics of Converging, Diverging and Pinhole Collimators

Some Comments Resolution always is best with the source as close as possible to the collimator For point source in air, the efficiency increases as the source to collimator distance increase with converging collimator (maximum at focus point) and decreases for diverging and pinhole collimators. No change for flood source as long as the source cover the entire detector. Diverging, converging and pinhole collimators may be useful for the change of field of view but the image distortion caused by the magnification with depth may be a problem.

Measurements of Gamma Camera Performance Intrinsic Resolution System Resolution Spatial Linearity Uniformity Counting Rate Performance Energy Resolution System Sensitivity