Presentation is loading. Please wait.

Presentation is loading. Please wait.

QC And NEMA In The Nuclear Arena

Similar presentations


Presentation on theme: "QC And NEMA In The Nuclear Arena"— Presentation transcript:

1 QC And NEMA In The Nuclear Arena
Gamma Cameras

2 Floods – Uniformity Testing
Types of floods Co57 Impregnated plastic Old source vs. the new source Refillable flood Adding the Tc99m Mixing and spilling It will bow if its too big Point source at least 5 diameters away from the surface Key Points Setting the peak How many millions of counts? Size of the detector Complete daily Amount of activity

3 Corrected Vs. Uncorrected
Setting your correction matrix Depends on SPECT 30 M vs. 120 M vs. the size of the head Do one for each radiopharmaceutical Correction will either subtract or add counts Take un/corrected flood and bars Correction for each PMT is no more than 10% Acquire at least once or month or when needed

4 Intrinsic Uniformity Determine Integral and Differential Uniformity
Remove collimator place a 3mm Pb “ring” around the UFOV (example next side) Set 20% window 99mTc with source (~200 μCi) 5 UFOV away Set matrix at 64 x 64 to collect no more than 20,000 cps Acquisition - center pixels must collect 4000counts/pixe; (~16 M count flood) Complete a 9 pt smooth and calculate the following

5 Calculations

6 What are we looking at? Integral uniformity looks at maximum deviation for both CFOV and the UFOV Differential uniformity looks at the maximum change of counts over a range of any 5 pixels in every column and row Excessive changes in any of the numbers is an indication non-uniformity problems Time to call service guy!

7 Flood Sources Co57 Flood Source Rectangular or Circular 5 to 20 mCi
Non-uniformity at <1.0% No mixing required – reduces radiation exposure Expensive since it must be repurchased based on Co57 decay Refillable Flood Phantom Rectangular or Circular Different sizes to fit a specific type of detector Inject mCi amount of Tc99m and mix in the water filled container Concerns Mixing can get messy Radiation exposure when mixing Air bubble will distort the image

8 Typical Bars – Four Quadrant
Multiple types of phantoms are available to match the size of the detector Bars contain Pb embedded in plastic. Bars attenuate the gamma radiation being emitted from the flood source Bars vary in size within each quadrant Is usually done with the collimator on Known as an extrinsic bar Tests system resolution Imaging system should resolve between the 3rd and 4th smallest quadrant Smallest bars are 2 mm, however, it depends on the type of 4 quadrant bar purchased

9 Setting Up The Bar Phantom
Comments How would imagine resolution be effected by LEHS vs. LEHR collimation? How does increased counts effect image resolution? How often is this procedure done on the imaging system? Why are the bars rotated every time this procedure is done?

10 Other Types of Bar Phantoms
BHR Test Pattern Lead plate that is 3.2 mm thick is placed between 2 sheets of Plastic Orthogonal-holes are then drilled in an array pattern Emission source can be extrinsic or intrinsic (intrinsic usually done) Clearly defines system resolution Over the entire camera Linearity also identified PLES (Equally Spaced Parallel-Bar) Phantom Checks linearity and resolution Phantom is placed with flood source on top (just like the Bar and Flood) 5 mm and 3 mm bars are available Linearity is the key X and Y axis- linearity is tested Looking for a straight line Lines should not be wavy

11 BHR Phantom Results Example of a BHR phantom with 2.5 cm orthogonal-holes

12 More On Bars Do weekly and rotate each week How many counts?
Size of detector Increased counts improves resolution Collimation Changes in resolution can be seen and identifies a loss of system resolution Intrinsic or extrinsic?

13 Systems Uniformity/Linearity/Resolution
Uncorrected Flood And Bars 4000k/image Corrected Flood And Bars 4000k/image

14 Edge Packing Intrinsic flood uses a mask on top of the crystal to
eliminate this artifact Extrinsic flood may show this artifact

15 Setting The Peak Centering the peak is essential
Off peaking causes imaging artifact What would happen if you imaged a patient with Co57 with Tc99m on board?

16 Example Of Being Off-Peak
Creation of lesions in the liver Could this lead to an inappropriate diagnosis?

17 Distance And Resolution
These images show the effect of distance on image resolution Surface vs. 1 inch 2 inch vs. 3 inches What conclusion can you make when it comes to imaging a patient? Does count density also add to this issue?

18 Problem With A PMT What causes this artifact?
How do both image sets differ? How might this effect the a patient’s diagnosis?

19 Problems With Liquid Filled Floods
Image on the left shows a flood that was note mixed completely Image on the right shows an air bubble in the flood

20 Collimator Damage Image on the right indicates damage to the collimator septa Image on the left shows the effects of septa damage on a diagnostic study

21 More Issues Inappropriate X/Y coordinates set on the a formatter
Cracked crystal

22 Types Of Bars A – Parallel-line equal-space (PLES) phantom: Small lead bars Test Camera’s spatial resolution. Two transmission images are Taken at right angles. Consider the Moire pattern B – Four quadrant bar phantom: Show wholes of equal size. The smallest bars can go down to 2 mm in size. C – Orthogonal-hole (OH) phantom Comes in different sizes (0.64 cm, 0.48 cm, and 0.32 cm Diameter spaced at 1 cm, 0.96 cm, and 0.64 cm respectively. Matching the lower end of system resolution is important.

23 Results Of Excessive Activity
Activity from the source should never exceed 20k/second The image on the left shows the results of excessive activity The image on the right is the same, however, the activity has been reduced to less than 20k per second

24 Intrinsic Uniformity With a point source collect 64 x 64 flood where the center pixel receives a minimum of 4000 cts (16 million count flood) Filter with a 9 point smooth Integral uniformity is the maximum pixel deviation Differential uniformity max change of cts over range of any 5 pixels in all rows and in all columns

25 Key Points Identifies the amount of nonuniformity with all areas of the flood Variations between un/corrected should be no greater than 15% Should be done weekly to determine when the camera needs a PM “Measles” occur when water gets into the crystal Floods should be done with the collimator on to see if there are any defects in it A crystals may crack when there is change of 8 degrees F Room temperature should range between 40 and 110 degrees F

26 Measles Image is off-peak
Loss of the hermetic seal causes water to enter and dissolve the crystal

27 System Sensitivity Is a parameter of camera’s ability to efficiently detect the incident gamma with the collimator on Changes in sensitivity result from Nonuniformities in the crystal Decreased energy resolution Incorrect collimation Improper setting of the PHA Recording the time it takes to collect a Co57 flood corrected for decay will identify any change in system sensitivity

28 System Spatial Resolution
Is defined as camera’s ability to accurately determine the original location of a gamma ray on an X-Y plan with collimation

29 Moiré Pattern When a bar phantom is not lined (oriented) up septa in collimator this patter will occur Bars travel at angle to the line of wholes (septa) in the collimator This can be further seen on the next two slides

30 Moiré Continued Moiré seen on bar Images Bands represent
the bars and the holes represent the septa Images across From each other Identify the pat- tern

31 Moiré Continued This shows a different moiré pattern because of the change In rotation of the bars to the collimator septa

32 Instrumentation and QC
Consider the following procedures that involve QC and instrumentation Uniformity Matrix Floods and Bars Other types of Bars - what is there purpose? Linearity Off peak or on peak Effects of Distance on image resolution Count density PMT artifacts Crystal artifacts Collimator artifacts Flood field artifacts

33 QC On The Dose Calibrator
Constancy Accuracy Linearity Geometry Return to the Table of Content


Download ppt "QC And NEMA In The Nuclear Arena"

Similar presentations


Ads by Google