Design and Calibration of a Dissolution Test Equipment Training Workshop on Dissolution, Pharmaceutical Product Interchangeability and Biopharmaceutical.

Slides:



Advertisements
Similar presentations
Symantec 2010 Windows 7 Migration EMEA Results. Methodology Applied Research performed survey 1,360 enterprises worldwide SMBs and enterprises Cross-industry.
Advertisements

Symantec 2010 Windows 7 Migration Global Results.
1 A B C
AGVISE Laboratories %Zone or Grid Samples – Northwood laboratory
Trend for Precision Soil Testing % Zone or Grid Samples Tested compared to Total Samples.
Simplifications of Context-Free Grammars
Variations of the Turing Machine
AP STUDY SESSION 2.
1
Select from the most commonly used minutes below.
Copyright © 2013 Elsevier Inc. All rights reserved.
STATISTICS HYPOTHESES TEST (I)
STATISTICS INTERVAL ESTIMATION Professor Ke-Sheng Cheng Department of Bioenvironmental Systems Engineering National Taiwan University.
David Burdett May 11, 2004 Package Binding for WS CDL.
NTDB ® Annual Report 2009 © American College of Surgeons All Rights Reserved Worldwide Percent of Hospitals Submitting Data to NTDB by State and.
1. 2 Begin with the end in mind! 3 Understand Audience Needs Stakeholder Analysis WIIFM Typical Presentations Expert Peer Junior.
Whiteboardmaths.com © 2004 All rights reserved
Create an Application Title 1Y - Youth Chapter 5.
CALENDAR.
The 5S numbers game..
Inspections on an iPad, iPhone, iPod Touch, Android Tablet or Android Phone.
Media-Monitoring Final Report April - May 2010 News.
Welcome. © 2008 ADP, Inc. 2 Overview A Look at the Web Site Question and Answer Session Agenda.
Break Time Remaining 10:00.
The basics for simulations
Factoring Quadratics — ax² + bx + c Topic
PP Test Review Sections 6-1 to 6-6
Regression with Panel Data
Operating Systems Operating Systems - Winter 2010 Chapter 3 – Input/Output Vrije Universiteit Amsterdam.
Exarte Bezoek aan de Mediacampus Bachelor in de grafische en digitale media April 2014.
| Slide 1 of 63 April 2007 Training Workshop on Pharmaceutical Development with focus on Paediatric Formulations Protea Hotel Victoria Junction, Waterfront.
Copyright © 2012, Elsevier Inc. All rights Reserved. 1 Chapter 7 Modeling Structure with Blocks.
Adding Up In Chunks.
SLP – Endless Possibilities What can SLP do for your school? Everything you need to know about SLP – past, present and future.
MaK_Full ahead loaded 1 Alarm Page Directory (F11)
Mid-Term Review For each statement, write true or false
TCCI Barometer September “Establishing a reliable tool for monitoring the financial, business and social activity in the Prefecture of Thessaloniki”
When you see… Find the zeros You think….
Before Between After.
1 Lab 17-1 ONLINE LESSON. 2 If viewing this lesson in Powerpoint Use down or up arrows to navigate.
: 3 00.
5 minutes.
1 hi at no doifpi me be go we of at be do go hi if me no of pi we Inorder Traversal Inorder traversal. n Visit the left subtree. n Visit the node. n Visit.
Types of selection structures
Speak Up for Safety Dr. Susan Strauss Harassment & Bullying Consultant November 9, 2012.
1 Titre de la diapositive SDMO Industries – Training Département MICS KERYS 09- MICS KERYS – WEBSITE.
Essential Cell Biology
Converting a Fraction to %
Numerical Analysis 1 EE, NCKU Tien-Hao Chang (Darby Chang)
CSE20 Lecture 15 Karnaugh Maps Professor CK Cheng CSE Dept. UC San Diego 1.
Clock will move after 1 minute
famous photographer Ara Guler famous photographer ARA GULER.
Flat Flow by Kamila Součková 11. Task Fill a thin gap between two large transparent horizontal parallel plates with a liquid and make a little hole in.
Physics for Scientists & Engineers, 3rd Edition
Select a time to count down from the clock above
Copyright Tim Morris/St Stephen's School
1.step PMIT start + initial project data input Concept Concept.
9. Two Functions of Two Random Variables
1 DIGITAL INTERACTIVE MEDIA Wednesday, October 28, 2009.
1 Dr. Scott Schaefer Least Squares Curves, Rational Representations, Splines and Continuity.
Chapter 4 FUGACITY.
1 Dissolution Measurement System: Current State and Opportunities for Improvement Dr. Lucinda Buhse Director, Division of Pharmaceutical Analysis.
Pharmacopoeial tests (USP)
Dissolution and drug release testing
Bureau of Drug and Narcotic Mechanical Calibration Temperature control  Calibrated thermometer  ˚ C Distance between the inside bottom of the.
Evaluation of tablet dosage form 5) dissolution
Presentation transcript:

Design and Calibration of a Dissolution Test Equipment Training Workshop on Dissolution, Pharmaceutical Product Interchangeability and Biopharmaceutical Classification System. Kyiv, Ukraine, June 25 - 27 2007 Dr. Sandra Klein, Institute of Pharmaceutical Technology, Johann Wolfgang Goethe University Frankfurt

Dosage forms to be tested immediate release dosage forms powders, granules / beads, tablets, capsules controlled release dosage forms transdermal systems implants

Official Dissolution Monographs United States Pharmacopeia USP XXX (30) European Pharmacopoeia Ph. Eur. 5th Edition, Supplement 5.3 British Pharmacopoeia BP 2007 Japanese Pharmacopoeia JP XIV (14) http://jpdb.nihs.go.jp/jp14e/contents.html

Official dissolution apparatus USP 30 classification Rotating Basket (Ph.Eur./BP/JP) Paddle (Ph.Eur./BP/JP) Reciprocating Cylinder (Ph.Eur.) Flow Through Cell (Ph.Eur./BP/JP) Paddle Over Disk (Ph.Eur.) Rotating Cylinder (Ph.Eur.) Reciprocating Holder

Which type of dissolution apparatus ? Depends on intention Quality control examining batch homogeneity examining batch to batch conformity examining stability Research & Development examining drug release behavior of preformulations in vitro simulation of the gastrointestinal passage IVIVC

Apparatus 1 - Basket Useful for Standard volume capsules beads delayed release / enteric coated dosage forms floating dosage forms surfactants in media Standard volume 900/1000 ml 1, 2, 4 liter vessels

Apparatus 1 - Basket Advantages breadth of experience (more than 200 monographs) full pH change during the test can be easily automated which is important for routine investigations

Apparatus 1 - Basket Disadvantages disintegration-dissolution interaction hydrodynamic „dead zone“ under the basket degassing is particularly important limited volume  sink conditions for poorly soluble drugs ?

Apparatus 1 - Basket

Apparatus 2 - Paddle Useful for Standard volume tablets capsules beads delayed release / enteric coated dosage forms Standard volume 900/1000 ml Method of first choice !!!

Apparatus 2 - Paddle Advantages easy to use robust can be easily adapted to apparatus 5 long experience pH change possible can be easily automated which is important for routine investigations

Apparatus 2 - Paddle Disadvantages pH/media change is often difficult limited volume  sink conditions for poorly soluble drugs ? hydrodynamics are complex, they vary with site of the dosage form in the vessel (sticking,floating) and therefore may significantly affect drug dissolution „coning“ sinkers for floating dosage forms

Sinker types JP/ USP / Ph. Eur. 5.3 Sinker „a small loose piece of nonreactive material such as not more than a few turns of wire helix may be attached to dosage units that would otherwise float …“ „…. other validated sinker devices may be used“

Coning

Apparatus 2 - Paddle

Apparatus 3 – Reciprocating cylinder Useful for tablets beads controlled release formulations Standard volume 200-250 ml per station

Apparatus 3 – Reciprocating cylinder Advantages easy to change the pH pH-profiles hydrodynamics can be directly influenced by varying the dip rate Disadvantages small volume (max. 250 ml) little experience limited data

Apparatus 3 – Reciprocating cylinder

Apparatus 4 – Flow-Through Cell Useful for low solubility drugs microparticulates implants suppositories controlled release formulations Variations open system closed system

Cell types Tablets 12 mm Tablets 22,6 mm Powders / Granules Implants Suppositories / Soft gelatine capsules

Apparatus 4 – Flow-Through Cell Advantages easy to change media pH pH-profile possible sink conditions different modes a) open system b) closed system Disadvantages Deaeration necessary high volumes of media labor intensive

Apparatus 4 – Flow-Through Cell

Apparatus 5 – Paddle over disk Useful for transdermal patches Standard volume 900 ml

Apparatus 5 – Paddle over disk Advantages standard equipment (paddle) can be used, only add a stainless steel disk assembly Disadvantages disk assembly restricts patch size

Apparatus 6 – Rotating cylinder USP apparatus 7 – Reciprocating holder most probably will be removed from the USP !!!

Summary Immediate release dosage forms:  apparatus 1 or 2 (preferably 2) Controlled release dosage forms:  apparatus 1 or 2 using different media for QC  apparatus 3 or 4 for R&D purposes Beside the selection of an adequate dissolution apparatus, adequate test conditions are crucial for all purposes !

Qualification of Dissolution Systems

Calibration Why ? How ? When ? to confirm suitability of the equipment and proper operation of the apparatus How ? mechanical calibration (verification of physical parameters) chemical calibration („Apparatus Suitability Test“ – USP) When ? before using new test equipment after relocation or major maintenance at regular intervals („every 6 months“)

Factors that may affect reliability of the test Proper alignment/geometry of dissolution apparatus dimensions of vessels, paddles, baskets, cylinders height, centering and wobble Proper conditions during dissolution test temperature agitation speed degassing sampling (sampling zone, timing, filtration, dilution) vibration Proper validation of analytical method specified in USP Chapter <1225>

Mechanical calibration Verification of physical parameters specified in the pharmacopoeia: USP apparatus 1 and 2

Mechanical calibration - Parameters Height – Vertical Position of the Paddle or Basket the vertical position of paddle or basket affects the hydrodynamics condition in the vessel each paddle or basket should be individually adjusted to the compendial distance in the pharmacopoeia, a distance of 2.5 + 0.2 cm is specified different kinds of height gauges can be used to align or check* this parameter *

Mechanical calibration - Parameters Rotational Speed – Stirring Rate input variable that affects the hydrodynamics changes in the rotational speed result in a changing liquid-solid interface between the solvent and the dosage form the rotational speed can be checked by using a digital tachometer* the compendia specify a rotational speed tolerance of + 4 % *

Mechanical calibration - Parameters Shaft Wobble – Eccentricity of Stirring Device assumed to alter the pattern of fluid movement in both paddle and basket apparatus and therefore may influence the dissolution rate can be measured with a micrometer* measured is the sum of distance between both sides (180°) of the axis of rotation *

Mechanical calibration - Parameters Centering (Vessel / Shaft) the axis of the rotating shaft must coincide at all points with the axis of the vessel to within + 1 mm “the shaft has to positioned so that is axis is not more than 2^mm at any point from the vertical axis of the vessel and rotates smoothly without significant wobble” *

Mechanical calibration Measurement tools all mechanical tools used for calibration should be certified to assure their reliability the results of mechanical calibration have to be documented

Apparatus suitability test (USP) if all parts ( apparatus, geometry, test conditions, analytical method) are within compliance – why perform an apparatus suitability test? the apparatus suitability is to check for parameters that can not be conveniently measured (vibration, vessel cleanliness, medium degassing ...) and also to provide an overall check of the system

Apparatus suitability test (USP) first established in 1978 routine test in most pharmaceutical laboratories calibration at regular intervals (every 6 months) standard calibrator substances according USP chapter <711> only the method(s) to be used have to be calibrated ! if six units are tested – all have to pass

Apparatus suitability test (USP) Standard calibrators according to USP chapter <711> Apparatus I, II and V: disintegrating type USP Prednisone Tablets nondisintegrating type USP Salicylic acid Tablets Apparatus III: USP Chlorpheniramine Maleate Extended-Release Tablets

Information supplied with calibrators http:/www.usp.org/referenceStandards/useAndstorage/calibrators.html

Apparatus suitability test (USP) USP Prednisone Tablets RS – current lot P0E203 (10 mg nominal prednisone content per tablet) disintegrating type paddle and basket, 50 rpm 500 ml deaerated water, 37°C quantity of prednisone released after 30 minutes is determined specified ranges Lot P0E203: Apparatus 1: 47-82 % Apparatus 2: 37-70 %

Apparatus suitability test (USP) USP Salicylic acid Tablets RS – current lot Q0D200 (300 mg nominal salicylic acid content per tablet) nondisintegrating type paddle and basket, 100 rpm 900 ml deaerated phosphate buffer, 37°C quantity of salicylic acid, released after 30 minutes is determined specified ranges Lot Q0D200: Apparatus 1: 23-30 % Apparatus 2: 17-25 %

Apparatus suitability test (USP) Controversies regarding the current test the variability in the intrinsic performance of the USP calibrator tablets is so great that it exceeds the variability in intrinsic performance of modern test dissolution assemblies this variability becomes obvious in both vessel-to-vessel variability and inter-laboratory variability of results for a given lot of calibrators

Troubleshooting Calibrator Tablets: always check the incoming tablets ! right lot of calibrators ? are the tablets broken, fused or severely chipped ? particularly salicylic acid tablets are often subject to sublimation ( dust on the tablets and the inner surface of the container) use correct storage conditions ! take the tablets out of the original container immediately before test !

Troubleshooting Standard / Standard solution: USP Standard used ? drying procedure conducted ? standard solution prepared on day of test ? standard solution filtered in the same manner as sample ? amount of alcohol used in the standard < 5% ?

Troubleshooting Vibration vibration produces unwanted variation in dissolution data and mostly results in an increased dissolution rate internal vibration may be caused e.g. from frayed drive belts external vibration may be caused by e.g. magnetic stirrers, centrifuges, vacuum pumps, old fridges, nearby construction, ... inability to properly measure vibration levels at various points within an apparatus is the main reason why calibrator tablets were originally developed Vibration is a variable that can seriously distort the data from any dissolution system. Therefore the system must be free of vibration that significantly interferes with the test. Vibration may be caused internal for example from frayed drive belts or the circulation of the water bath but rather by external sources like magnetic stirrers, centrifuges, nearby construction and so on. The inability to properly measure vibration levels at various points within an apparatus was the main reason why calibrator tablets were originally developed.

Troubleshooting Vibration effects – case example: Effect of vibration levels of the dissolution apparatus on the dissolution rate of enteric coated granules of Cefalexin. The vertical dotted line indicates 0.05 m/s2. Kaniwa N. et al. (1998) Int J Pharm, 175, 119-129 „Low vibration“: < 0.05 m/s2 „High vibration“: > 0.05 m/s2 A possible effect of high vibration levels on the dissolution rate was also shown in a collaborative study of Kaniwa et al. (1998). In order to develop a dissolution standard for evaluating vibration levels for dissolution apparatus, a series of tests, examining enteric coated granules of cefalexin was performed at seven independent laboratories. The vibration of the dissolution apparatus during the test was measured with a vibration meter and the resulting values were plotted against the corresponding release rates of cefalexin. The enteric coated granules worked very well in detection of high vibration levels (especially in the basket apparatus). There seems to be a critical value between so called low and high vibration apparatus at about 0.05 m/s2. Dissolution apparatus that belong to the high vibration category according this classification do not seem to provide accurate and reproducible results! Based on the results of this study, in the event of dissolution results that show unexpectedly high release rates, troubleshooting concerning vibration is crucially important.

Troubleshooting Vibration: dissolution equipment placed planar ? drive chain or belt free of tension and/or dirt ? torn parts replaced ? correctly functioning gear plates ? individual spindles are not surging ? bench/table stable ? no sources of vibration nearby ?

Troubleshooting Dissolution medium: correctly degassed ? correct amount used (900/500 ml) ? correct amount dosed (weight/volume) ? dosing procedure gentle (resaturation/spillage) ? buffer correct (pH + 0.05 units, buffer salts, molarity) ? correct temperature during test (32°C / 37°C + 0.5°C)? evaporation during test negligible ?

Troubleshooting Importance of degassing: insufficient degassing may result in decreased dissolution rates of several drugs e.g. prednisone tablets but also a range of poorly soluble drugs are very sensitive to the amount of dissolved gases in the dissolution medium the degassing procedure should therefore be efficient and reproducible for every test I really want to point out the importance of degassing, because ….

Troubleshooting Deaeration method USP heat the dissolution medium to about 41°C vacuum filter through a 0.45-µm-porosity membrane into a flask, stirring with a magnetic stirrer continue to draw a vacuum and stir for an additional 5 min gently transfer the medium directly into the vessel rotating the apparatus 2 shafts to speed equilibration to 37°C is discouraged!!! use medium promptly after equilibration An adequate dearation method is decribed in the USP: First, a suitable volume of medium is heated to 41°C. With the aid of vacuum it is filtered through a 0.45 µm porosity membrane into a filtering flask equipped with a stirring device. The flask is sealed and while stirring for an additional 5 minutes, the vacuum is been continued. Then the medium is gently transferred directly into the vessel and promptly used after equilibration to a temperature of 37°C.

Troubleshooting Alternative deaeration methods the USP states that „other validated deaeration techniques for removal of dissolved gases may be used“ other techniques include: heating sonication vacuum helium sparging (expensive)

Troubleshooting Sampling take each sample at the correct time point  sampling time points (+ 2%) use a single glass syringe for each vessel sample from the right location within the vessel  between media surface and top of the paddle blade  n.l.t. 10 mm from vessel wall

Troubleshooting Sampling always use a suitable filter  check filter adsorption check the clearity of the filtered sample filter the sample immediately after sampling for automated sampling also check the tubings

Troubleshooting Physical conditions of the apparatus vessels scrupulously clean ? vessel surface smooth and curvature appropriate ? Apparatus 1 the conditions of the baskets, particularly of their clips is critical check all baskets for corrosion and blocked meshes before using them align the air holes to prevent air cushions emerging during the test

(Semi) Automation Advantages high throughput of samples minimizes analyst-to-analyst variability in sampling and filtration reduces the average costs per analysis very promising for QC purposes

Media preparation automated mixing of water and concentrate preheating of medium deaeration (vacuum and stirring) media can be dispensed directly into the vessel In this slide you can see a media preparation system that offers you a time- and effort- saving way of mixing, preheating, deaerating and dipensing of your medium. The test medium can be mixed from concentrate (buffer or acid) and water, preheated to a defined temperature, deaerated under vacuum over a particular time and then be directly dipensed into the vessel. Altogether this procedure should result in a reproducible degassing process.

Offline system Here you can see an Offline system for apparatus 1 and 2. This system consists of a dissolution bath, coupled with a sampling unit. Samples can be taken at predetermined time points. If necessary they can be diluted immediately and after that collected in test tubes for UV-analysis or in HPLC vials respectively.

Online system If you have a high throughput of samples that do not require dilution before UV analysis, a so-called Online system will make sense. This system consists of a dissolution bath that is directly coupled with an UV-spectrophotometer. Using this setup, it is possible to record multiple-point (real time) dissolution curves using a flow through-cuvette-system (closed system).

On- / Offline system A combination of On- and Offline System combines two major advantages of single On- and Offline System by offering the possibility of an online-UV-measurement of automatically diluted samples. The system also can be run in single offline or online mode like described before.

HPLC - online system A further possibility for online-detection is represented by an HPLC – online system that combines the dissolution bath with a complete HPLC system. This system would be helpful in case of a high throughput in samples that cannot be measured by simple UV-analysis. For example this will be the case as you have excipients that would interfere with perturb UV-analysis or as you use biorelevant media like e.g. FaSSIF or FeSSIF as test media.

Automation always validate automated methods, including analytical and sampling methods validation should be performed using manual analysis, withdrawing samples at the same times and comparing to the automated results: not highly variable dissolution results: two concurrent runs highly variable dissolution results: simultaneous sampling pay attention to automated dilution and filtering processes What do you have to consider when thinking about (semi-) automation? Automated methods always have to be validated. The validation should include analytical and sampling methods. Validation should be performed using manual analysis, withdrawing samples at the same times and comparing to automated results. If the dissolution results are not highly variable, it is possible to compare two concurrent runs. If they are highly variable, simultaneous sampling has to take place. Special attention should be paid to automated dilution and filtering processes!

Suggested reading FIP Guidelines for dissolution testing of solid oral products. Dissolution Technologies 4:5-14 (1997).    SM Diebold, JB Dressman. Dissolved oxygen as a measure for de- and re-aeration of aqueous media for dissolution testing. Dissolution Technologies 5: 13-16 (1998). S Qureshi. Calibration – the USP dissolution apparatus suitability test. Drug Inf. J. 30, 1055-1061 (1996). N Kaniwa et al. Collaborative study on the development of a standard for evaluation of vibration levels for dissolution apparatus. Int. J. Pharm. 175: 119-129 (1998). VA Gray, CK, Brown, JB Dressman, LJ Leeson. A new general information chapter on dissolution. Pharmacopoeial Forum 27 (6) [Nov.-Dec. 2001] W Brown. General information <1092> The dissolution procedure: development and validation Pharmacopoeial Forum 30 (1) [Jan.-Feb. 2004]   Of general interest: Dissolution Technologies: http://www.dissolutiontech.com With this slide, I want to end my talk. You can see some references for more information regarding calibration of dissolution systems. It is also worth to have a view at the homepage of „Dissolution technologies“. This journal is published quarterly and contains a lot of useful informations for all areas of dissolution testing.

Johann Wolfgang Goethe University Dr. Sandra Klein Johann Wolfgang Goethe University Institute of Pharmaceutical Technology 9 Max von Laue Street Frankfurt, 60438, Germany e-mail: Sandra.Klein@em.uni-frankfurt.de With this slide, I want to end my talk. You can see some references for more information regarding calibration of dissolution systems. It is also worth to have a view at the homepage of „Dissolution technologies“. This journal is published quarterly and contains a lot of useful informations for all areas of dissolution testing.