Measuring Volume Biotech I pg. 301 to 323 in Basic Laboratory Methods.
Advertisements

Use and Maintenance of Micropipets
Use and Maintenance of Micro-pipettes
Measuring Small Amounts Part 1: Using pipets and micropipets.
INTRODUCTION TO PIPETTES
Cautions! 1. Set pipet volume only within the range specified for that micropipet 2. Do not attempt to set a volume beyond the pipet’s minimum or maximum.
Oct. 22, 2012 Generosity Ludicrous: causing or deserving laughter because of absurdity Do Now: What is a microliter? How many microliter are in one liter?
What are Micropipettors?! We will be using 3 different sizes of micropipettors: ◦ P20 ◦ P200 ◦ P1000 Very expensive and very fragile – USE Responsibly!!!
Measurements Chemistry Department Minneapolis Community & Technical College Intro to Chemistry Chem1020 Lab 1.
Volumetric and Mass Measurements
Basic Laboratory Skills
Bellwork What is an SOP? What is its purpose?
Use and Functional Verification of Automatic and Micro-Pipettes Module 7: Quality Assurance of Analytical Phase.
Micropipette Tutorial
Pipette Calibration Procedure Pipette, water and balance Test the maximum volume or minimum volume Pre-wet tips Test 5 volumes per setting.
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
The Basic Skills of the Biotechnology Workplace
Measuring volumes which are what we scientists call “very, very small”
Laboratory Glassware and equipment lab 2
Laboratory equipment.
Introduction to Equipment
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
MEASUREMENT OF VOLUME. OVERVIEW First some general principles Then, more detail about micropipettes.
Some Common Laboratory Glassware Please be careful when using glassware in the laboratory. Wet glassware and wet gloves are slippery. Always read a volume.
1. Which tool would you use for picking up 15 microliter? Which tool for 25 ml? Why? 2. Why do the little test tubes look this way? 3. Answer “Micropipette.
4/24/ :54 PM Accuracy = of a measurement is how close that measurement is to the true or “exact” value EX: Standard weight = 5.00g 4.98g more accurate.
Micropipettes Why they are useful How they work How to care for them.
Experiment 1 Basic Laboratory Techniques Parts A, D, E, and an Unknown
Automatic pipettes use and calibration
Biotech 8/20/13 What pump would you use for a 1 or 2 ml pipet? What pump would you use for a 5 or 10 ml pipet?
Topic 7.5. You must control an experiment. You can test only one variable at a time. (How does the amount of water effect plant growth? Keep the light,
Proper Micropipette Technique and Slide Preparation
General Laboratory Techniques Dry Lab Chemistry 1105.
Use and Maintenance of Micro-pipettes
1 REGIONAL BIOTECH AT THE UNIVERSITY OF TEXAS-PAN AMERICAN BIOSCIENCE EDUCATION FOR THE RIO GRANDE VALLEY.
Laboratory Glassware and equipment
Automatic Pipettes Use And Calibration
Making Measurements and Using Numbers
Micropipette Tutorial
How to do a Manual Titration
Techniques in Molecular Biology 2017 Fall Lecture -2-
READING MEASUREMENT SCALE INSTRUMENTS
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
Lab Techniques How to do a Titration.
مهارتك الأساسية في معامل التقنية الحيوية
Techniques in Molecular Biology 2017 Fall Lecture -2-
Measurement of Volume Chapter 20.
Use of the Labnet Micropipettor
Metric System and Lab Skills
Use and Maintenance of Micro-pipettes
Introduction To Medical Technology
Lab Equipment Overview
Laboratory Techniques: Titration
Erlenmeyer Flask An Erlenmeyer Flask is a cone-shaped container with neck, so you can hold the flask or attach a clamp or use a stopper. They are used.
How to do a Manual Titration
Micropipette Tutorial
Identification of the common laboratory glassware, pipettes and instruments . BCH 312 [PRACTICAL]
Measuring volumes which are what we scientists call “very, very small”
Micropipettes.
Pipette Technique.
Common Laboratory Glassware
Do Now How many ul (microliters) is in 1 liter?.
Using the micropipette
Experiment 1 Basic Laboratory Techniques Parts A, D, E, and an Unknown
Do Now How many ul (microliters) are in one liter?
Measurement & Micropipetting
Laboratory Glassware.
Pipette Calibration.
Identification of the common laboratory glassware, pipettes and instruments . BCH 312 [PRACTICAL]
Micropipette Tutorial
Presentation transcript:

Measuring Volume

Volume Liquid volumes are measured in Liters (L), milliliters (mL) or microliters (mL) A liter is slightly more than a quart A microliter is 1/1000 of a mL or smaller than a tiny drop Depending on the volume to be measured, 3 options: graduated cylinder, pipet or micropipet

B S Rule Bigger units left Smaller units right Move decimal point to the right if converting big units to small units Move decimal point to the left is converting small units to big units

B S Rule 1.25 L is equal to how many mLs? How many liters is 60 mLs? 250 uL is how many L? 0.00025 L

Small to Medium Volumes Typically 0.1 mL to 25 mL Plastic disposable or glass pipets Pipette roller or pipette bulb is needed Commonly used pipettes are: 1mL, 5mL, 10mL, and 25 mL Pipets are named by the maximum they deliver

Pipetting Terminology Graduations: lines marked on volume measuring devices that indicate volume Meniscus: a curve formed by the surface of liquids confined in a narrow space, such as in a measuring device

Pipetting Terminology TC: To Contain; will contain the specific amount when filled to the capacity mark. It will not deliver the amount if the liquid is poured out because some of the liquid will adhere to the sides of the container.

Pipetting Terminology TD: To Deliver; marked so that it will deliver the specified amount, assuming the liquid is 20C and is poured out. Pipets with a TD on them indicate they are designed “To Deliver” and that the tiny amount left in the bottom of the pipet after dispensing should NOT be blown out. Note: plastic containers are considered non-wetting which means water does not stick to them so there is no difference between TD and TC

Pipetting Terminology Tolerance: how much error is allowed in the calibration or a measuring device. Volumetric: the most accurately calibrated glassware

The “art” of Pipetting Learning how to pipette takes practice and determination Patience and careful lab technique will ensure your results are accurate and reliable

Steps to Proper Pipetting Determine the proper pipette size Determine the proper pipette roller or bulb Blue: 1 mL pipette Green: 5 or 10 mL pipette We will be using pipette pumps in this shop Draw up past your required volume and dispense to your exact fill volume required Pay attention to the meniscus

Pipetting Lab: Precautions Never mouth pipette Hold the bottom of the roller and the top of the pipette when you are assembling them Always keep the pipette vertical when there is fluid in it Always keep the tip of the pipette sterile and take care to not touch surfaces

Really Small Volumes Most QC and R&D laboratories measure very small volumes. Micropipets are used to measure microliter quantities. What is the symbol for microliter? mL Micropipets are expensive instruments which must be handled carefully

Getting to Know your micropipet There are 4 parts to a micropipet Plunger button Ejector button Volume display (setter) Dispensing Tip/ Tip attachment

Getting to Know your micropipet

Name the parts

Getting to Know your micropipet Plunger button Typically there are 2 stops The first evacuates the air in the micropipet The second stop evacuates the volume plus another 50% or so. Practice to feel the difference

Picking the proper size micropipet Look at your plunger It will give you the min. and max. that it can deliver 3 numbers will be displayed The top # is the digit for the maximum I.e. 2 is the top # on the P200

Interpret the following settings How many microliters is this on a m10? on a m100 on a m1000 3 5 3 5 3 5 3.50 uL or 0.00350 mL 35.0 uL or 0.0350 mL 350.0 uL or 0.350 mL

Setting your required volume To set your volume you need to turn the adjusting knob. Rotate slowly to the desired volume setting At no time should the knob be turned past its upper or lower limits. Do not force the settings, the knob will turn easily

Sources of Error for micropipet Bad pipet tip: leaks, clogged tip Damaged pipetor Wrong size tip for pipet size Bubbles in your tip due to poor technique Liquid not at room temperature or it may be viscous

Multi channel pipets In order to increase efficiency, pipets have been designed to hold 4-16 channels at a single dispense. Most commonly used when running ELISA assays 8 channel pipet

Multi Channel pipet

Electronic or Automated pipets Also referred to as a pipette pump Typically they are battery operated Press a button to pick up, press button to dispense Use to increase or decrease volume settings For large volume labs, automated pipetting instruments are used. Multi channel heads

Burettes Long graduated tube with a stop cock at one end which is used to dispense known volumes accurately.

Micropipetter Precautions These are expensive and precise pieces of equipment. DO NOT play with them Handle them with care Keep them clean

Using a micropipet Micropipet operating instructions Choose the proper pipet for the desired volume, choose proper pipet tip Set the volume Depress the plunger to first stop and draw up required volume Wipe tip gently with Kimwipe, do not draw out liquid Depress plunger to expel liquid. Go to the second stop to blow out all of the liquid.

Calibration Definition: adjustment of dispensing devices so they dispense accurate volumes. In the case of pipets we are checking the accuracy and precision of the pipet.

Calibration Accuracy: to deliver the exact quantity required Mean measured volume divided by Nominal or required volume Precision: the ability to deliver that exact quantity time after time The amount of variation between the each volume measured 1.01, 1.01, 1.02

Calibration Procedure Pre Calibration Activities Be sure the Micropipetter is clean Be sure the balance is calibrated and working properly (Verification) Be sure the water you are using is 22 to 25C Ensure the scale is free of drafts Place a wt. on and verify it provides a consistent reading

Calibration Procedure Set the Micropipetter to deliver 2 uL on a 2 to 20 uL pipetor or 25 uL on the 20 to 200 uL pipetor Place a weigh boat on the calibrated balance Add a few drops of water to the weigh boat Tare the balance by pressing the zero button

Calibration Procedure Pipet the appropriate volume and dispense it into the weigh boat. Record your reading Zero the balance (good practice to change tips each time) Repeat until you have 5 readings

Calibration Procedure Set the Micropipetter to 20 uL for 2-20 uL pipet or 200 for 20 to 200 uL pipet Zero the balance Pipet the appropriate volume and dispense it into the weigh boat. Record your reading Repeat until you have 5 readings

Calibration Procedure Calculations: Calculate the mean weight of water Density of water at 22C = 1.0031 g/mL 1 uL would weigh how many mgs? 1.0031 mgs Convert the mean water weight to the mean volume weight Mean Volume of water = Mean weight / density

Calibration Procedure Mean of 5 weights / 1.0031 mgs i.e. if the weights of 2 uL setting were 2.001, 2.001, 2.003, 2.002, 2.002 mg What is the mean weight? 2.018 What is the mean volume of water? 2.018/1.0031 = 2.006 What is the accuracy of the pipetor? 2.006 / 2.000 x 100 = 100.3% or .3% error

Preventive maintenance of pipets Wipe the pipet with alcohol if it gets contaminated with solution Store properly which is in a stand if possible as you do not want the shaft part of the pipet to get bent from laying on a bench