Presentation is loading. Please wait.

Presentation is loading. Please wait.

X Series ICP-MS Training Course

Similar presentations


Presentation on theme: "X Series ICP-MS Training Course"— Presentation transcript:

1 X Series ICP-MS Training Course
Introduction to ICP-MS theory and X Series ICP-MS PlasmaLab Analytical Method Development Dealing with interferences

2 Application areas

3 Metals / Industrial Analysis of major, minor and trace elements
Purity of fine metals Solution Analysis via nebulisation Solids Analysis via Laser Ablation Reagents required: 40% v/v/ Primar Nitric acid and deionised water No sample preparation required. Bulk or feature analysis can be performed on metals.

4 Nuclear Fuel quality certification
Environmental and bioassay monitoring of low level actinides Precise and accurate isotope ratio measurement Signal to noise ratio needs to be maximized for ppq & sub ppq LOD’s. Use of high efficiency nebulisers. For waste minimization use of low flow nebulisers, e.g. PFA-50

5 Semiconductor Determination of low levels of impurities in ultra-pure process reagents Impurity analysis of wafers Transition metals suffer from Argon oxide and Argon nitride interferences, which are greatly reduced with the PlasmaScreen Plus and CCT options. Use ultra-clean glassware, cones etc. Recommended Accessories: USN and PFA-50, if low sample volume

6 Biological & Clinical Speciation of metals
Elemental uptake and excretion studies Stable isotope tracer studies Analysis of salvia, serum, blood and urine CCT technique recommended to overcome interferences on key elements such as Cr, Fe, As, Se in heavy biological matrices

7 Geological Rocks, sediments and soils
Direct solid analysis using Laser Ablation on: Individual minerals, thin section, isotope ratios, zonation studies, depth profiling. Oxide interferences within REE e.g. BaO on 151Eu and 153Eu, NdO on 150Tb, 160Gd and 162Dy. Residual HClO4 or saline matrices will interfere with Cr,V and As(ClO & ArCl interferences).

8 Environmental Analysis of drinking and waste water
Seawater and marine life (shellfish, corals etc.) Analysis of industrial effluents Soil/sediment studies Transition metals suffer from Argon oxide and Argon nitride interferences, which are greatly reduced with the CCT option. Reagents required 40% v/v Primar Nitric acid and deionised water Sample dried into a digest ‘bomb’, if required.

9 Minimising contamination
Laboratory environment Water quality Ultrapure reagents Handling of ultrapure reagents Selection of material for containers & cleaning procedures Sample and standard preparation

10 Tracing sources of contamination
The Step-by-Step Approach Instrument blank Water blank Dilute acid blank Analysis blank

11 Causes of interferences

12 Interferences – mass spectroscopic

13 Interferences – non mass spectroscopic

14 Isobaric Interference
ELEMENTI ISOTOPI in ICP-MS Isotopi : Atomi dello stesso elemento con differenti masse (stesso numero di protoni & elettroni ma diverso numero di neutroni) Isotopes for Iron Fe Mass : 55,9116 Isobaric Interference Mass 54 Isotope for Fe Isotope for Cr Isotopes for Chromium Cr Mass : 52,0554

15 INTERFERENZE Le interferenze Molecolari e Isobariche vengono classificate come interferenze spettroscopiche - Interferenze Molecolari (poliatomiche): sono Ossidi ed Idruri, Specie poliatomiche Gli ossidi possono essere minimizzati attraverso il tuning, gli idruri mediante sistemi per l’abbattimento delle interferenze (CCT) -Interferenze Isobariche: questo tipo di interferenze sono dovute alla sovrapposizione di diversi isotopi

16 Interferences Spectrometer Lenses (CCT) Ar , N , O , Ar , etc. Ar Ar
+ , N , O 2 , Ar , etc. ArO , ClO ArNa etc. Ar Ar (0.1%) diss .“metal”, H 0, HNO 3 HCI, etc. dry “metal” or metal-oxide, decomposing anions Metal Atoms Ions (M ) Conventional “Front-End” approaches to reduce interferences “Analyzer approach” to reduce or resolve interferences M o > M + e - “Ionization potential”

17 Idruri e doppie cariche Interferenze poliatomiche
ESEMPI DI INTERFERENZE SPETTROSCOPICHE Interferenze da Ossidi, Idrossidi, Idruri e doppie cariche Interferenze poliatomiche

18 Illustration of overlapping peaks
Interferences Polyatomic peak overlaps Causes/sources : solvent Anions in sample Cations in sample Ar gas Entrained air Signal suppression Ionisation suppression or enhancement Causes/sources : Overloading plasma Overloading nebuliser Cone orifice issues Unintended plasma cooling effects Elemental peak (isobaric) overlaps Generally precludes the use of some isotopes if overlap exists Total measured signal = analyte + polyatomic Analyte portion of total measured signal Illustration of overlapping peaks

19 Correction equations Monoisotopic As in a Cl matrix
Therefore we can calculate the contribution of ArCl at 77 ArCl 40,35 As 75 ArAr 36,40 ArCl 40,37 Se 77 ArAr 38,40 ArAr 40,40 Se (82) 75 76 77 78 79 80 81 82

20 Correction equations Monoisotopic As in a Cl matrix
76 77 78 79 80 81 82 75 Se (82) ArAr 40,40 38,40 36,40 ArCl 40,37 40,35 As Monoisotopic As in a Cl matrix Se(77) = Se(82) x AbSe(77) AbSe(82) Calculate 77Se contribution at mass 77 from 82Se Calculate 77ArCl contribution at mass 77 from 77Se ArCl(77) = Int(77) - Se(77) ArCl(75) = ArCl(77) x Ab(ArCl(77)) Ab(ArCl(75)) Calculate ArCl contribution at mass 75 from 77ArCl As(75) = Int(75) - ArCl(75) Calculate 75As contribution from 75ArCl

21 Instrument performance options
But the X Series ICP-MS, does supply a few ‘hardware’ remedies for these problems….. PlasmaScreenPlus - a)‘cool’ screen operation - analysis of Li, Na, K, Ca, Fe b) ‘hot’ screen operation – enhances sensitivity while maintaining low background CCT - removal of polyatomics Xi – improves matrix tolerance and reduces polyatomics

22 Instrument performance options
Isotope Interferant Technique 6,7 Li Background CP 23 Na 24 Mg C 2 CCT 28 Si N 31 P NOH 56 Fe ArO CP,CCT 39,40 K ArH, Ar 40 Ca Ar 75 As ArCl 51 V ClO 52 Cr ArC 60 Ni CaO Xi 80 Se

23 PlasmaScreen Plus PlasmaScreenTM Torch
- Grounded metal screen inserted between torch and load coil - Reliable automatic switching between ‘hot’ and ‘cool’ plasma conditions Cool screen conditions (~600 Watts) - Minimizes argon-based interferences i.e. 38Ar1H+, 40Ar16O+, 40Ar2+ - Minimizes high backgrounds associated with easily ionizable elements i.e. Li, Na - ng/L LODs for Li, Na, Ca, K, Fe, Cr Hot screen operation - Enhanced sensitivity - Low background maintained - Improved detection limits

24 COOL PLASMA Caratteristiche Vantaggi Limitazioni
RF a bassa potenza: w con nebulizzatore a flusso alto Temperatura del Plasma prossima a K Vantaggi Permette di determinare in ICP-MS elementi come Ca, Mg, Fe, Na, K, Cr, V Limitazioni Plasma non robusto: è necessario analizzare campioni nella medesima matrice Impossilità di analizzare elementi con alto potenziale di eccitazione/ionizazione ( As, Hg, Se)

25 CCT Operating the ion guide with a suitable collision/reaction gas (in CCT mode) Selective attenuation of polyatomic interference ions 40Ar35Cl, 40Ar12C, 40Ar22+, 40Ar2H, 38Ar1H+, 40Ar16O+, 40Ar2+ Innovative, user selectable CCT Plasma operated at normal power - Transmission of the analyte ions remains largely unaffected Ability to switch to CCT mode within sample SKIMMER CONE EXTRACTION LENS SAMPLE CONE HEXAPOLE RODS

26 CCT - ArAr 59Co+ ArAr+ H2 He ArAr(neutral) + H2+

27 CCT – analysis of iron 10ppb Fe - CCT Blank - CCT

28 Collision cell Based on a multipole ion guide contained in a cell into which the collision cell gases flow; cell is located between the interface and the quadrupole analyser Hexapole or octopole ion guides are used Split-flow high-compression turbo pump Rotary Pump Detector Quadrupole COLLISION CELL Plasma MFC 1 MFC 2 Slide Valve Cell Gas 1 Cell Gas 2

29 Collision cell Based on a multipole ion guide contained in a cell into which the collision cell gases flow; cell is located between the interface and the quadrupole analyser Hexapole or octopole ion guides are used Split-flow high-compression turbo pump Rotary Pump Detector Quadrupole COLLISION CELL Plasma MFC 1 MFC 2 Slide Valve Cell Gas 1 Cell Gas 2

30 Principi della tecnologia a cella di collisione
Il fascio ionico viene iniettato in una cella pressurizzata - Tradizionalmente vengono impiegate miscele He, H2 or a He/H2 Gli ioni poliatomici collidono con il gas della celle e vengono dissociati nei loro componenti atomici, ioni etc. 38Ar1H+/39K, 40Ar+/40Ca, 40Ar12C+/52Cr, 40Ar23Na+/63Cu, 40Ar16O+/56Fe, 40Ar35Cl+/75As, 40Ar2+/80Se La trasmissione degli ioni analiti attraverso la cella non viene praticamante influenzata Le condizioni di plasma ad alta temperartura vengono mantenute Ferro Argon Ossigeno Idrogeno ArO+ Ar H2 H2O+ Fe+ Principles of Collision Cell Technology The principle of operation is that molecular fragmentation occurs via a range of possible mechanisms thus removing the isobaric overlap. When operating in CCT mode a collision gas such as H2 and/or He is bled into the cell at a low pressure. Polyatomic ions collide with the cell gas and are dissociated into their component atoms/ions or converted into non interfering species Transmission of analyte ions is largely unaffected The beauty of CCT is that high RF power plasma operation is maintained enabling removal of interferences in complex matrices (e.g. organic solvents) where cool plasma is ineffective In the example shown: - Fe+ ions at mass 56 enter the cell and collide with the H2 molecules. H2 is a multi-purpose collision gas and does not react with the Fe+ ions which simply lose a little energy and pass on through to the quadrupole analyser. H2 is drawn away through intermediate turbo pump. ArO+ (also mass 56) enters the Cell, collides with the H2 gas and the collision dissociates the ArO+ into Ar and H2O+ molecules which are drawn out of the chamber along with the H2. None of the species travel through to the analyser. .All the commonly occurring molecular species shown can be effectively removed or attenuated with the X7 CCT. PE system generates reactive intermediate species which react with the ion beam to form adducts which may interfere in other regions in the spectrum. To counteract this PE need a quad pre-filter to deselect regions of the spectrum to remove these adducts. This leads to a more complicated experimental set-up

31 CCTED – Meccanismi possibili
Dissociazione collisionale? e.g. ArAr+ + He = Ar + Ar+ + He Reazione chimica? e.g. ArAr+ + H2 = ArH + ArH+ Trasferimento di carica? e.g. ArAr+ + H = ArAr + H+ Ritardo collisioneale e successiva flitrazione in termini di energia delle specie? e.g. ArAr+* + He = ArAr+ + He* CCTED – Possible Mechanisms This slide shows possible reaction mechanisms.

32 Solving the problem of cell generated interferences
Subsequent collisions of cell-generated ions cause them to lose kinetic energy These ions (and also plasma derived polyatomic ions), with their larger collision cross section, lose more energy than the smaller analyte ions Can prevent these lower energy interfering ions from being detected by using an energy barrier between the cell and the quadrupole analyser – this is called (kinetic) energy discrimination (ED or KED) +

33 How is energy discrimination set up?
Changing collision cell and quadrupole bias voltages provides a simple way to set up the required energy barrier Polyatomic ions from plasma or cell reaction products have insufficient energy to pass through the energy discrimination barrier Potential Position Without ED With ED Potential Barrier Collision Cell Lenses Quadrupole -2V -9V -18V -14V Analyte Ion Isobaric Polyatomic ion Key CCTED – Simple Optimization Tuning for energy discrimination with the X Series is extremely simple requiring the cell pole-bias potential to be adjusted to be slightly negative with respect to the quadrupole pole-bias thus creating an energy barrier at the entrance to the quadrupole. The Energy Discrimination setting is then effective for the full mass range removing the need for optimization for each interfered analyte. Typical pole bias values of –2V (without ED) and –10v (with ED) are shown in the slide. During tuning, PlasmaLab software provides a real time display on the computer VDU of user selected analyte and polyatomic responses, enabling the operator to optimise the signal to background ratio.

34 The power of energy discrimination (1)
In environmental and clinical samples, containing Cl and Ca, 75As is interfered by 40Ar35Cl+ and to a lesser extent, 40Ca35Cl+ Collision cells are effective for removing these interferences, however, if ED is not used, a large signal is still observed on 75As This signal does not match the 3:1 m/z 75 to m/z 77 ratio expected if the interference is Cl - related, so it must be another species If Ca is not present, the interference is not observed, so it must be related to Ca CCTED – Simple Optimization Tuning for energy discrimination with the X Series is extremely simple requiring the cell pole-bias potential to be adjusted to be slightly negative with respect to the quadrupole pole-bias thus creating an energy barrier at the entrance to the quadrupole. The Energy Discrimination setting is then effective for the full mass range removing the need for optimization for each interfered analyte. Typical pole bias values of –2V (without ED) and –10v (with ED) are shown in the slide. During tuning, PlasmaLab software provides a real time display on the computer VDU of user selected analyte and polyatomic responses, enabling the operator to optimise the signal to background ratio. Mass 75 = 17,000 cps (Mass 77 = 2,000cps) Scan of 100ppm Ca, collision cell without ED

35 The power of energy discrimination (2)
The interference is believed to be CaOH(H2O)+ or CaO(H3O)+, formed from collision, then reaction, between CaOH+ and H2O in the cell As the Ca-species forms in the cell it’s energy is immediately reduced relative to the As+ analyte ion; further collisions as it passes through the cell reduce the energy further Applying a few volts of ED results in elimination of the species As+ is still transmitted to the quadrupole and can be determined to low ppt levels CCTED – Simple Optimization Tuning for energy discrimination with the X Series is extremely simple requiring the cell pole-bias potential to be adjusted to be slightly negative with respect to the quadrupole pole-bias thus creating an energy barrier at the entrance to the quadrupole. The Energy Discrimination setting is then effective for the full mass range removing the need for optimization for each interfered analyte. Typical pole bias values of –2V (without ED) and –10v (with ED) are shown in the slide. During tuning, PlasmaLab software provides a real time display on the computer VDU of user selected analyte and polyatomic responses, enabling the operator to optimise the signal to background ratio. Mass 75 = 6 cps Scan of 100ppm Ca, collision cell with 2V ED

36 The power of energy discrimination (2)
The interference is believed to be CaOH(H2O)+ or CaO(H3O)+, formed from collision, then reaction, between CaOH+ and H2O in the cell As the Ca-species forms in the cell it’s energy is immediately reduced relative to the As+ analyte ion; further collisions as it passes through the cell reduce the energy further Applying a few volts of ED results in elimination of the species As+ is still transmitted to the quadrupole and can be determined to low ppt levels CCTED – Simple Optimization Tuning for energy discrimination with the X Series is extremely simple requiring the cell pole-bias potential to be adjusted to be slightly negative with respect to the quadrupole pole-bias thus creating an energy barrier at the entrance to the quadrupole. The Energy Discrimination setting is then effective for the full mass range removing the need for optimization for each interfered analyte. Typical pole bias values of –2V (without ED) and –10v (with ED) are shown in the slide. During tuning, PlasmaLab software provides a real time display on the computer VDU of user selected analyte and polyatomic responses, enabling the operator to optimise the signal to background ratio. 1ppb 75As = 300 cps Scan of 100ppm Ca + 1ppb As, collision cell with 2V ED

37 Reduction of 40Ar16O+ and Ar2+ using the collision cell
1% HNO3 Blank 1ppb Fe Quadrupole – Standard Mode Quadrupole – CCT Mode (H2/He) Quadrupole – Standard Mode Quadrupole – CCT Mode (H2/He) 10ppb Se 1% HNO3 Blank

38 Collision cell ICP-MS Fe / Se calibration performance
56Fe (40Ar16O+ attenuated) 80Se (40Ar2+ attenuated) 4.0 14.2 3607 56Fe 1.0 Detection limit (3σ, n = 5) (ppt) 4.3 BEC (ppt) Correlation coefficient 1648 Sensitivity (cps/ppb) 80Se Parameter Sample matrix = 2% HNO3, 8% H2 in He collision gas used

39 Why use oxygen as a collision / reaction gas?
Some interferences cannot be efficiently removed using 'standard' cell gases (e.g. He, H2, NH3) Some interfered analytes react (along with their interferences), with certain cell gases (e.g. NH3) Lose analyte sensitivity; degrade detection limit Kinetic energy discrimination (KED) or bandpass application sometimes not sufficiently effective Interference suppression results in too much sensitivity loss BEC's and detection limits not low enough for the analytical requirement

40 XSeriesII – New Proof Data - 1
Fast Cd and Pb in whole blood analysis Blood reference materials measured; diluted 1:50 for analysis Standard addition calibration on one sample used to quantify other samples Unique ability for Thermo! Ideal approach for biomedical samples! Autosampler probe-to-wash-early function used to maximise throughput Unique for Thermo! Sample throughput = 51 per hour, or more than 400 per 8 hour day! 111Cd 499 ± 1 33.3 ± 0.5 Bio Rad 3 Reference 208Pb (ppb) Reference 111Cd (ppb) 254 ± 2 12.5 ± 0.4 Bio Rad 2 412 ± 13 6.2 ± 0.5 Seronorm blood 2 Measured 208Pb (ppb) Measured 111Cd (ppb) Sample identity 208Pb

41 XSeriesII – New Proof Data - 2
Cd in urine in the presence of Mo (removing the MoO interference) Urine reference material diluted 1:20 with 1% (v/v) HNO3 for analysis; spike recovery (0.50 ppb Cd) performed on sample doped with Mo (5 ppm) Standard addition calibration on the sample used to quantify other samples Samples run in standard and collision cell mode with O2 in the cell O2 promotes MoO+ to higher Mo oxides; Cd does not react Without O2 the interference is large 96Mo16O+ 95Mo16O+ Mo interference 0.02 0.04 Detection limit 3σ (ppb) 5.26 7.98 Conc. (ppb) Standard mode Collision cell mode (O2) Spiked Cd value (ppb) Isotope 0.05 0.03 Detection limit 3σ (ppb) 0.49 0.50 113Cd 111Cd Conc. (ppb) O2 in the cell completely removes the MoO interference!

42 Detection of Pt in 10 ppm Hf solutions
The problems: HfO+ and HfOH+ interferences on all Pt isotopes (m/z 190 to 198) m/z 190 and 198 interferences low abundance but 190Pt very low abundance (0.01%) and 198Pt also low (7.2%) Must suppress HfO+ / HfOH+ interferences on most abundant remaining Pt isotopes (194Pt, 195Pt and 196Pt) The proposed solution: Use O2 as the collision gas, to promote formation of higher Hf oxides / hydroxides Pt less reactive with O2 HfO+, HfOH+ interferences, 10ppm Hf 179Hf16O+ = 3 Mcps Pt isotopes, 12ppb Pt 195Pt+ = 250 Kcps

43 Pt in 10 ppm Hf, standard mode and O2 data
195Pt standard mode 195Pt collision cell mode (O2) 9 5 7 Detection limit 3σ (ppb) 467* 154* 281* BEC (ppb)* Standard mode Standard BEC / collision cell BEC Collision cell mode (O2)# Hf interference Isotope 0.054 0.006 0.003 12289 0.04 180Hf16O+ 196Pt 17111 0.01 179Hf16O+ 195Pt 16529 0.02 178Hf16O+ 194Pt BEC (ppb) * estimated using the 198Pt (linear) calibration sensitivity # O2 flow rate = 2.7 mL/min, non-KED

44 Detection of Hg in 10 ppm W solutions
The problems: WO+ and WOH+ interferences on all Hg isotopes (m/z 196 to 204) m/z 196 and 204 interferences low abundance, but 196Hg very low abundance (0.14%) and 204Hg also low (6.8%) Also have 204Pb interference on 204Hg Must suppress WO+ / WOH+ interferences on most abundant remaining Hg isotopes (198Hg, 199Hg, 200Hg, 201Hg and 202Hg) The proposed solution: Use O2 as the collision gas, to promote formation of higher W oxides / hydroxides Hg less reactive with O2 WO+, WOH+ interferences, 10ppm W 184W16O+ = 2.4 Mcps Hg isotopes, 20ppb Hg 200Hg+ = 140 Kcps

45 Hg in 10 ppm W, standard mode and O2 data
200Hg standard mode 200Hg collision cell mode (O2) 1670 0.02 0.2 4.2 334* 184W16O+ 200Hg 813 0.03 0.3 1.0 244* 183W16O+ 199Hg 0.4 13 Detection limit 3σ (ppb) 247* 12 640* BEC (ppb)* Standard mode Standard BEC / collision cell BEC Collision cell mode (O2)# W interference Isotope 0.05 1235 186W16O+ 202Hg 40 184W16O1H+ 201Hg 3200 182W16O+ 198Hg BEC (ppb) * estimated using the 201Hg (linear) calibration sensitivity # O2 flow rate = 3.4 mL/min, non-KED

46 Detection of Cu and Zn in 10 ppm Ti solutions
The problems: TiO+ (and TiOH+) interferences on all Cu and Zn isotopes (m/z 63 to 70) m/z 68 and 70 interferences low abundance but 70Zn also low (0.6%) 68Zn most useful in absence of collision cell Must suppress TiO+ / TiOH+ interferences on Cu and most abundant Zn isotopes The proposed solutions: Use O2 as the collision gas to try to promote formation of higher Ti oxides / hydroxides Compare with kinetic energy discrimination (KED) approach, using H2/He collision gas TiO+ interferences, 10ppm Ti 47Ti16O+ = 13 Kcps 48Ti16O+ = 130 Kcps 49Ti16O+ = 11 Kcps Cu / Zn isotopes, 10ppb Cu / Zn 63Cu+ = 85 Kcps 64Zn+ = 68 Kcps 65Cu+ = 42 Kcps

47 Cu, Zn in 10 ppm Ti, standard mode and O2 data
64Zn standard mode 64Zn collision cell mode (O2) 1 0.15 0.17 0.03 0.2 50Ti18O+ 68Zn 9 0.05 0.34 0.04 2.9 50Ti16O+ 66Zn 0.38 0.06 0.20 Detection limit 3σ (ppb) 20.3 3.1 2.1 BEC (ppb) Standard mode Standard BEC / collision cell BEC Collision cell mode (O2)# Ti interference Isotope 0.23 0.11 15 1.35 48Ti16O+ 64Zn 17 0.18 49Ti16O+ 65Cu 13 0.16 47Ti16O+ 63Cu # O2 flow rate = 7.0 mL/min, non-KED

48 Cu, Zn in 10 ppm Ti, standard mode and H2/He KED data
64Zn standard mode 64Zn collision cell mode H2/He KED 1 0.13 0.25 0.03 0.2 50Ti18O+ 68Zn 12 0.02 0.04 2.9 50Ti16O+ 66Zn 0.38 0.06 0.20 Detection limit 3σ (ppb) 20.3 3.1 2.1 BEC (ppb) Standard mode Standard BEC / collision cell BEC Collision cell (KED) mode Ti interference Isotope 0.05 0.07 62 0.33 48Ti16O+ 64Zn 22 0.14 49Ti16O+ 65Cu 15 47Ti16O+ 63Cu Data obtained using a 8% H2/He gas flow of 5.5 ml/min and a KED barrier of 3.5V

49 CCTED – Evaluation of Performance
4 Standard Reference Waters were analyzed as ‘unknown’ samples over a 10- hour period Wide range of typical environmental analytes were measured - 30 analytes, 55 isotopes Several analytes had associated interference problems…. Experiment- Analytical considerations In order to evaluate the performance of the X7 ICP-MS, 4 Standard Reference Waters were analyzed as ‘unknown’ samples over a 10-hour period A number of analytes typically of interest in environmental analysis were monitored. Over 50 analytical isotopes were used since many the elements of have more than one ‘useful’ isotope for measurement. Most analytes have few interference problems and are measured in the ‘Standard’ ICP- MS Mode, i.e. with high power plasma without any collision or reaction gas admitted into the cell (cell unpressurized).

50 Experimental - analytical profile
Li Be Na Rb Sr Mo Ag Cd Sn Cs Ba Tl Pb U Mg Al K Ca Cr Fe Mn Ni Cu Zn Ga As Se Analytes V Cr Experimental - analytical profile The software groups elements together according to the measurement mode selected such that all elements in Standard mode are measured together, sequentially, in increasing mass, with similar groupings for CCT/ H2 Mode and CCT/NH3 Mode. This slide shows the analytical profile for this experiment: - as can be seen the in- sample switching with rapid stabilisation between measurement modes combined with the intelligent sample uptake and wash ensures that the X7 ICP-MS provides the fastest possible sample throughput. The total sample to sample cycle time for analysis of the 55 isotope suite was 3 minutes and 45 seconds. . Uptake 25s NH3/He 3x1.6s reps Settle Delay 30s H2/He 3x18s reps Settle Delay 30s Standard Mode 3x18s reps Wash 25s Time Profile Total Time Per Sample = 3 minutes, 45 seconds

51 Experimental - procedure
Calibration: Blank 1ppb multi-element solution 10ppb multi-element solution 4 reference water samples analysed as QC checks 12 times over a period of 10 hours against 12 different calibrations NIST 1640 diluted 1+9 and spiked to 2% HCl CNRC SLRS-2 spiked to 2% HCl CNRC SLRS-3 spiked to 2% HCl CNRC SLRS-4 diluted 1+1 and spiked to 2% HCl Detection limits calculated by running 10-replicate blank samples 12 times against 12 different calibrations Experimental – procedure The autosampler was set up with the following standards and samples. Twelve replicate analyses were performed over a period of 12 hours, recalibrating at the beginning of each cycle. Calibration was performed with Blank 1 ppb multi-element standard 10 ppb multi-element standard Samples River water reference samples were prepared as follows: NIST 1640 diluted 1+9 and spiked to 2% HCl CNRC SLRS-2 spiked to 2% HCl CNRC SLRS-3 spiked to 2% HCl CNRC SLRS-4 diluted 1+4 and spiked to 2% HCl The NIST 1640 and SLRS-4 samples were diluted to reduce the analyte concentrations, providing a greater analytical challenge. All samples were spiked with UPA hydrochloric acid to 2% to increase the presence of chloride-based interference Detection limits were calculated by running 10-replicate blank samples 12 times against 12 different calibrations

52 Results: stability in different measurement modes -10ppb standard
Each isotope measured in a different mode, switching from NH3 to H2 to Std mode in- sample 10ppb Solution No internal standard correction Results: stability in different measurement modes -10ppb standard In order to compare the relative stabilities and highlight any drift in each measurement mode the 3 Pb isotopes were each measured in different mode during each analytical cycle. 208Pb in standard mode, 207Pb in CCT / H2 mode and 206Pb in CCT / NH3 mode. Excellent stability of 0.6% RSD was achieved for all isotopes showing that there is no loss of precision when measuring in either of the 2 CCT modes in order to achieve optimum detection limits. i.e The precision and stability are shown to be independent of the measurement mode required to achieve optimum detection limits and no hysteresis effects are observed when switching modes.

53 Results: stability of real sample in different modes
This slide shows the precisions achieved for the analysis of the NIST 1640 standard with V measured in CCT/NH3 mode, Se measured in CCT/ H2 mode and Cd measured in Normal mode again without reference to any internal standard. The data shows that there is no loss of data quality due to the continual switching between modes. Sample diluted and spiked to 2% HCl

54 Results – Method detection limits (g/L)
Results – Method detection limits (ug/L) Method detection limits were determined by taking 10 replicate blank readings prior to each of the calibration blocks. The mean 3 sigma MDLs shown here are in the ppt to sub ppt for almost all elements and are well below the regulatory requirements Based on 3s on 12x10replicates of blank, each from a new calibration

55 Results – accuracy of reference samples
Accuracy is shown here by comparing the measured and certified values for the various CRMs – As can be seen excellent recoveries are obtained all are within +/-10% All concentrations in g/L n=12

56 Results – accuracy of reference samples
In summary: - The analyte concentrations measured in the CRMs in this experiment ranged from ppt to tens of ppm. This slide shows that that excellent accuracy was achieved over this wide dynamic range and was independent of detector mode (analog or pulse counting), and measurement mode, Normal or CCT Mode.

57 Analysis of Seawater Reference Materials
Direct analysis of seawater is challenging: Extreme matrix Severe interferences Ultra trace analyte concentrations Analysed against a matched external calibration H2/He KED CCT mode used for ALL analytes Rapid method (~3 min/sample) No gas switching required No instrument settings switching required Easy set-up

58 XSeriesII – New Proof Data (ppt)

59 XSeriesII – New Proof Data (ppt)

60 XSeriesII – New Proof Data (ppt)

61 CCTED Evaluation of Performance - Summary
Provides Ease of Set-up Auto-tune, performance reports allow fully automated set-up and analysis Ultimate Flexibility Single mode runs ( e.g.. Seawater example) Fast analysis Mode Switching runs Ultimate Detection Limits Handles the most extreme matrices (seawater) to give interference free measurements on difficult analytes Retains excellent stability and accuracy Results - Summary The X Series ICP-MS combines research performance with routine operation. Challenging environmental samples can be analysed with optimised CCT settings with rapid switching between settings. Two different cell gases can be combined with normal mode analysis in a single sample acquisition to provide optimum sample throughput with protocol compliance. Typical sample throughput is < 4 minutes per sample with up to 55 isotopes per sample. Excellent stability is obtained independent of the measurement mode with RSDs <1% over a 12-hour period.  Within sample mode switching allows the optimum conditions for each analyte to be used, resulting in the ultimate analytical performance resulting in detection limits in the ppt to sub-ppt range for most elements. The CCT technique ensures results are obtained with freedom from many interferences. In this study, accuracy to within +/-5% for the vast majority of analytes has been achieved even after spiking to 700ppm chloride.


Download ppt "X Series ICP-MS Training Course"

Similar presentations


Ads by Google