ProEdit Procedure Management Writing a Verification Procedure For a Megger CM500 Combination Tester - Loop Function
2
ProEdit :: Writing a Procedure Preparing to write a procedure Before beginning the write the procedure the specifications for the instrument are required – these can be found from the following sources : Instrument manufacturers handbook Manufacturers web site Catalogues such as R.S., Farnell or other instrument resellers
3
ProEdit :: Writing a Procedure Step 1 Start ProEdit Select File Procedure Builder Wizard Loop Tester
4
ProCal :: Writing a Procedure Step 2 Enter Manufacturer, Model Number and Version number
5
ProCal :: Writing a Procedure Step 3 SECTION 1 : LOOP RESISTANCE TESTS Select a range from the first drop down list. A set of values is displayed – select the required resistance values It is recommended to set at least 3 test points for each range For each additional range, select from the drop down lists (up to 3 ranges can be set) Tip : Use the Select All or Select None buttons to easily set resistance values. Range 1Range 2Range 3 Note : The resistance values shown are the NOMINAL values. The actual value will be the NOMINAL + SUPPLY LOOP resistance (automatically read back from the 2100 calibrator)
6
ProCal :: Writing a Procedure Step 3 Cont’d SECTION 1 : LOOP RESISTANCE TESTS Select the accuracy for the ranges (use one or more of these boxes) % of Reading % of Range ± Counts
7
ProCal :: Writing a Procedure Step 5 SECTION 6 : ADDITIONAL CHECKS Select from the following type tests Indicator Check Display Check Controls Check Tests are performed manually and either pass or fail selected on screen.
8
ProCal :: Writing a Procedure Step 7 The Use Uncertainty Statements option will automatically select the correct uncertainty statements for the procedure, based on the functions selected.
9
ProCal :: Writing a Procedure Step 8 Click on the Build Procedure button to proceed When this message appears click No to close the screen or Yes to create another procedure
10
ProCal :: Viewing a Procedure Step 1 Now the procedure has been created, it can be viewed using ProEdit. Select File > Edit An Existing Procedure Select Model Number and enter CM500, then click Start Search
11
ProCal :: Modifying a Procedure Step 2 The CM500 procedure will be listed, select the line and click Accept
12
ProCal :: Modifying a Procedure Step 3 The procedure instrument information screen will be displayed – click Next to continue. Uncertainty statements & traceable instruments screen will be displayed – click Next to continue.
13
ProCal :: Modifying a Procedure Step 4 Click the Show All button – a scrollable list will display all the procedure tests
14
ProCal :: Modifying a Procedure Step 5 Test information can be viewed and edited if required Note : The accuracy automatically recalculates based on the change of test value
Similar presentations
© 2018 SlidePlayer.com Inc.
All rights reserved.
Ppt on french revolution class 9th Ppt on hlookup in excel Ppt on telephone exchange in indian railways Ppt on polynomials of 900 Ppt on saas business model Ppt on object oriented concepts Ppt on generating electricity by walking Ppt on ideal gas law Ppt on water resources of maharashtra Ppt on audio visual aids