Presentation is loading. Please wait.

Presentation is loading. Please wait.

Advanced Biomedical Imaging Lecture 3 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.

Similar presentations


Presentation on theme: "Advanced Biomedical Imaging Lecture 3 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University."— Presentation transcript:

1 Advanced Biomedical Imaging Lecture 3 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University

2 X ray Machine Slide 5 X ray tube Slide 13

3 Effective focal spot: Face of target is slanted from vertical (6-15 0 ). As angle increased effective focal spot increased but (actual focal spot) remains the same. 1. Focal spot Actual focal spot Effective focal spot

4 A small focal spot produces: Sharper images,Sharper images, but unable to dissipate heat rapidly,but unable to dissipate heat rapidly, can not sustain high current required for short exposure required to decrease patient motion,can not sustain high current required for short exposure required to decrease patient motion, used in mammography.used in mammography. A large focal spot: can tolerate more heat,can tolerate more heat, but un-sharpness increases,but un-sharpness increases, used in fluoroscopy.used in fluoroscopy.

5 A quality filter: thickness of material placed in radiation beam to attenuate lower energies photons (responsible for patient dose) more than higher energies (responsible for image). A quality filter: thickness of material placed in radiation beam to attenuate lower energies photons (responsible for patient dose) more than higher energies (responsible for image). This can be done through photoelectric interaction process. 2. Filters Slide 2

6 So filtration:- photon energy, exit dose / entry dose ratio so film dose / skin dose ratio, but it decreases the contrast. Uα1/E 3 As u (attenuation coefficient) is α to 1/E 3, it attenuates low energy photons & does not affect high energy photons

7 3. Collimators & Grids I=I 0 e -ux Slide 10

8 The problem is scattered radiation? Primary radiation carries information to be imaged & but Scattered radiation obscures it. S/P depend on thickness of part, it is 4:1 in PA chest & 9:1 in lateral pelvis. Large patient creates more scatter. It contrast which produced by primary beam.

9 To reduce amount of scatter produced by patient Field size cones & diaphragm volume of scattering tissues scatter & contrast. Field size cones & diaphragm volume of scattering tissues scatter & contrast. Compression of patient: moving overlying tissues laterally reduces volume of scatter & reduce patient dose. scatter & contrast Compression of patient: moving overlying tissues laterally reduces volume of scatter & reduce patient dose. scatter & contrast Kv : Kv μ, scatter in pt and forward scatter reach film. contrast Kv : Kv μ, scatter in pt and forward scatter reach film. contrast

10 To reduce amount of scatter after it left pt: Grid between patient & film, air gap & flat metal filter. Grid between patient & film, air gap & flat metal filter. 1. Grid: anti-scatter grid: thin strips of lead thin strips of lead sandwiched between plastic, sandwiched between plastic, carbon fiber & AL. carbon fiber & AL. Lead absorb 90% of scattered rays while allow 70% of 1ry beam to pass through gap to reach film. contrast Lead absorb 90% of scattered rays while allow 70% of 1ry beam to pass through gap to reach film. contrast The beam becomes more penetrating as low energy beam can not reach the film so the mean of the energy increase. The beam becomes more penetrating as low energy beam can not reach the film so the mean of the energy increase. Slide 7

11 Depth / Width of interspaced channel Grid ratio (GR)= Depth / Width of interspaced channel – GR, efficiency to absorb scatter radiation & scatter / primary rays, contrast., & dose Contrast improvement factor = Contrast with grid / contrast without grid=3-5. = Contrast with grid / contrast without grid=3-5. Bucky factor or grid factor Exp necessary with grid/Exp necessary without grid=2-3 =Exp necessary with grid/Exp necessary without grid=2-3 Frequency is the number of strips per inch. High ratio & frequency more dose required. No grid is used with thin part of body (extremities) & child

12 Film moved 30 cm a way from patient so much oblique scatter will miss the film & so contrast improves Film moved 30 cm a way from patient so much oblique scatter will miss the film & so contrast improves Need Kv or mAS as image magnified (mammography). Need Kv or mAS as image magnified (mammography). Patient dose by grids, air gap, grids require higher dose than air gap so they are not used in pediatrics. Patient dose by grids, air gap, grids require higher dose than air gap so they are not used in pediatrics. 3. Flat metal filter: placed on cassette, absorb softer & obliquely traveling scatter > harder direct rays. 2. Air gap

13 A) x ray cassette is a flat light box with film between a pair of screen. A) x ray cassette is a flat light box with film between a pair of screen. Front of cassette (AL low Z=13) to minimize attenuation of beam so decrease required pt dose. Front of cassette (AL low Z=13) to minimize attenuation of beam so decrease required pt dose. Back of cassette is thin lead sheet to absorb remnant radiation so no back scatter irradiated pt nor screen. Back of cassette is thin lead sheet to absorb remnant radiation so no back scatter irradiated pt nor screen. 4. Screen film cassette Slide 2 Slide 2

14 B) Intensifying screen: phosphor crystals (z) Crystals absorb x-ray and emit light of intensity α intensity of x rays. It converts 1 photon to 1500 light photons. So allows Low pt dose & High contrast But with Noise & low resolution

15 C) Films Silver halide crystals Silver halide crystals exposed to light electrons attach to silver ions silver metal. The more silver metal the more film blackening. Screen is more sensitive than film exposed alone: why? – The screen phosphor layer is more effective than film emulsion at absorbing x-ray. Use of screen & film reduces pt dose by a factor of 50. Film is used alone to obtain sharp image in thin part where fine details is required (dental) Film is used alone to obtain sharp image in thin part where fine details is required (dental)

16 Quantity: It is α Kv 2 mAs & z & exp time. Quality: (penetrating power) It is affected by KV & tube filtration Tube filtration: Affects both quantity & quality, it affect quality by removal of low energy photons. Factors affect x ray emission: (quantity & quality)

17 1.Heat with x ray 2. Patient dose 1. Heat with x ray: 1. Heat with x ray: 99.5% of energy converted to heat It is avoid by using: Rotating anode Copper anode & oil circulating through anode. Limitation of x ray tube

18 2. Patient dose: (limiting factor) X ray is absorbed by patient. X ray is absorbed by patient. Emerged x-ray beam carries pattern of intensity dependent on thickness & composition of organs Emerged x-ray beam carries pattern of intensity dependent on thickness & composition of organs Acceptable patient dose of radiation is a limiting factor; Acceptable patient dose of radiation is a limiting factor; it should be as low as possible it should be as low as possible Minimum exit dose emerges from pt is required to produce satisfactory image. Minimum exit dose emerges from pt is required to produce satisfactory image.

19 Factors affect Patient dose Effect of tube Kv: Kv u skin dose, penetration & proportion of high energy photon reach film screen. Kv u skin dose, penetration & proportion of high energy photon reach film screen. So entrance dose decreases to acquire same exit dose. So entrance dose decreases to acquire same exit dose. Effect of tube mA mA increases no of photons, and so signals but also amount of exposure (dose). patient dose mA increases no of photons, and so signals but also amount of exposure (dose). patient dose Effect of Focus – Film distance on patient dose: FFD, beam diverge and no of photons distributes over larger area patient dose FFD, beam diverge and no of photons distributes over larger area patient dose noise in image, to compensate more x rays is needed to maintain (MAS but it skin dose). noise in image, to compensate more x rays is needed to maintain (MAS but it skin dose). U α Z 3 /E 3

20 So to reduce scatter radiation reaching the film: So to reduce scatter radiation reaching the film: – Use low KV, – Moving grid, Air gap. – Conning, – placing thin sheet of zinc on film cassette, To reduce patient exposure: To reduce patient exposure: – adding 2mm Al filter, – increase KV ( increase penetration, if obese) – decrease mA – x-ray target –object distance, – small object film distance, – compress patient.

21 Questions 1.Enumerate the factors that affect patient dose in x ray image? factors that affect x ray quantity & quality? 2.Enumerate the factors that affect x ray quantity & quality? factors thatreduce the effect of scatter radiation on the film? 3. Enumerate the factors that used to reduce the effect of scatter radiation on the film? 4.What are the advantages of using the grids? 5.What is the GR (grid ratio)? 6.Mention one case in which grid is not recommended?

22


Download ppt "Advanced Biomedical Imaging Lecture 3 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University."

Similar presentations


Ads by Google