1 Melvin Rivera Suárez Project Manager Alexis Ortiz Pérez Rafael Vega Díaz Wilfredo Bermúdez Soto

2  An estimated 80,700,000 American adults (one in three) have one or more types of cardiovascular disease  Every minute counts. For every minute heart failure goes untreated, unrecoverable damage occurs to the heart muscle

3  Current heart monitoring equipment is commonly stationary, available to patients at hospitals or at their house  Patients’ lifestyle often makes it difficult for them to access heart monitoring equipment.  Health care professionals need live data on the health of their patients to help them during emergencies

4  Device  Measure the ECG wave form and analyze  Portable, wireless  Detect if the user is having a heart attack or any other heart problem.  Mobile phone  Receive the alert of the device and send a message to the persons previously selected  Beyond our scope  Power consumption optimization of the Analog circuit

5  Software optimization. Should be easy installable within a 15 minutes by Nov 2.  Have a working prototype by Nov 21.  Hand in a written system manual to the project manager for review. It should describe all the system functionalities and legal disclaimers by Nov 24.  Tests in different scenarios. This will help to detect any situation not considered in the design process by Nov 25.

6

7

8  Electrocardiogram:  Measure of the heart’s electrical activity  Heart Rate  PR Interval  QRS Duration  QT Interval  QRS Axis

9  Ideally, any type of electric contact could be used  However, wearable, safe and reliable electrodes are desired for the final product  Standard medical ECG electrodes will be used  Two electrodes to be placed on the patient’s chest  Third electrode will be internal to the circuit: Virtual right leg driven

10  Analog Circuit  Instrumentation Amplifier: AD624  Appropriate for amplifying small signals  High Gain: Adjustable up to 10K  High Common Mode Rejection Ratio: 80 dB @ unity gain  Low Noise: 4 nV/√Hz @1 kHz  Photo coupler: PS2506  Electrically isolates the amplifier output signal from the rest of the device  System security measure  Band Pass Filter  Filters signals below 0.04 Hz to 150 Hz  Average heart signal is 50Hz to 70 Hz

11

12  MSP430F149: Ultra low power 16 bit microcontroller  Responsible for all needed calculation of the ECG wave form  Provides the necessary synchronization of the system  Specs:  Supports clock rates up to 8MHz  60 KB of flash ROM  2 KB of RAM  12-bit ADC  2 DAC outputs, 2 UART (serial) ports, and 48 General Purpose IO (GPIO) pins

13  KCWireFree KC21Bluetooth module  Class 2 module  Average power dissipation: 2.5mW  Small and power efficient  10m Range  Built in antenna

14  Interface between the monitor and the patient  Holds emergency contacts, and alerts them in case of a heart attack  Development Platform: J2ME  JAVA as a development tool for Mobile devices  Development Tools  Eclipse, MicroEmulator, NetBeans

15

16

17

18

19  Hardware  Melvin Rivera  Rafael Vega  Software  Alexis Ortiz  Wilfredo Bermúdez

20 Total cost Equipments and parts$454.95 Employees$40,551.71 Subtotal$41,006.66 Total after Overhead 150%$102,516.65 Employee Salary per hour Work hours requires Total cost Software Engineer$19.752805,530 Software Engineer$19.752805,530 Hardware Engineer $19.172805,367.60 Project Manager$37.0928010,386 Subtotal Cost 26,813.60 Bonuses (1.9%) 509.46 Unemployment Insurance (1.40%) 375.3904 Disability (1.1%) 294.9496 S.S 5.3% $1,421.12 Healthcare (7.4%) $1,984.21 401K/403b (4.3%) 1152.9848 Sub total$32,551.71 Professional consultant $100.0080$8,000 Total Employee Cost$40,551.71

21  Hardware components damaged  Unavailability of components  Data lost or damaged  Computer Crash  Loss of a Team Member

22  Environmental Issues  Battery Disposal  Legal Issues  Harmful Electromagnetic Interference  Social Impacts  Mobile Phone Radiation  Might give a false sense of security to a patient which condition really needs a direct hospital care

23  Monitor heart status anytime, anyplace  Portable device can be worn and is easy to carry  Provides patients with reliable heart monitoring  Mobile phone software displays heart status  Provides patients with prompt medical assistance  Send an emergency message if user has a heart attack or any emergency situation.  Improve the time of an emergency response

24