BLM FOR THE INJECTORS PROJECT CABLING NON-CONFORMITIES LIU-PSB Working Group Meeting (15/08/2013) Christos Zamantzas (christos.zamantzas@cern.ch)
2
11.5 +/-0.3 8.71 7.94 8.03 7.3 8.07 7.98 7.25 2.6 PE CU AL-PETP-AL PETP High Mu FE PETP Brown polyolefinSI-CU 23x9x0.16 SI-CU 23x9x0.16 Diameter in [mm] Cable: Signal -(1000V)- Screen 1 –(500V)- Screen 2 (Triaxial) Connector male: (receptacles) -Signal -Isolation (1000V) -Screen 1 -Isolation (500V) -Screen 2 (Triaxial) Connector female: (cable) -Signal -Isolation (1000V) -Screen 1 -Isolation (500V) -Screen 2 (Triaxial) 1. Adapter: BNC or SMA Signal - Screen 1 (coaxial) 2. Adapter: BNC / SMA / N-type Signal - Screen 1 & Screen 2(coaxial) CKB50: Coaxial cable christos.zamantzas@cern.ch215/08/2013
3
Cabling Diagram christos.zamantzas@cern.ch3 Screen of HV BNC is open on the IC side to assure there is no ground loop. Internal screen to shield low frequency noise (GND only on electronics side, IC is floating). External screen to shield high frequency noise. BNC SMA TRIAX Connector Types: Ionization Chamber enclosed in cable traytube 50 ÷ 200 m~ 5 m~ 1 m CKB 50 BNCSMA Acquisition Card (BLEDP) External Screen 2 Internal Screen 1 Signal HV Power Supply (~1500 V) HV BNC 15/08/2013
4
11.5 +/-0.3 8.71 7.94 8.03 7.3 8.07 7.98 7.25 2.6 PE CU AL-PETP-AL PETP High Mu FE PETP Brown polyolefinSI-CU 23x9x0.16 SI-CU 23x9x0.16 Diameter in [mm] Cable: Signal -(1000V)- Screen 1 –(500V)- Screen 2 (Triaxial) At random positions discovered shorts between the two shields. Not possible to be used with Triaxial functionality. CKB50: non-conformity christos.zamantzas@cern.ch4 Isolation between the two shields not optimal. Added additional steps for the connector assembly to avoid leaks. 15/08/2013
5
Cabling Non-Conformities christos.zamantzas@cern.ch5 Internal and External screen are shorted in multiple and random places. Non-conformity (most probable) due to thin insulator. BNC SMA TRIAX Connector Types: Ionization Chamber enclosed in cable traytube cable non-conformities 50 ÷ 200 m~ 5 m~ 1 m CKB 50 BNCSMA Acquisition Card (BLEDP) External Screen 2 Internal Screen 1 Signal HV Power Supply (~1500 V) HV BNC 15/08/2013
6
Alternative Cabling Configuration christos.zamantzas@cern.ch6 Connect both shields together on both sides Leave the ground on the detector side floating Configuration will be very similar like in LHC system. BNC SMA TRIAX Connector Types: Ionization Chamber enclosed in cable traytube cable non-conformities 50 ÷ 200 m~ 5 m~ 1 m CKB 50 BNCSMA Acquisition Card (BLEDP) External Screen 2 Internal Screen 1 Signal HV Power Supply (~1500 V) HV BNC 15/08/2013
7
Meeting with manufacturer On the first meeting agreed: Modification of the specifications Two more PET-foils over the magnetic screen. No problems mounting the connector. The electrical properties will not change. Manufacturer to run a trial until end of September We will test the sample and release the cable until mid-October If all ok, manufacture about 6km until end of November Re-installation to begin mid-January. christos.zamantzas@cern.ch715/08/2013
8
Next steps Next week, meet with manufacturer on financial matters. Costs of the new production Handling of the stock already at CERN Decide if we continue with the installation or remove all and wait for the new production. Currently the issue affects 32 detectors During LS2, 80 additional will need to be installed. christos.zamantzas@cern.ch815/08/2013