Download presentation
Presentation is loading. Please wait.
Published byAnissa Lyons Modified over 4 years ago
1
Anaesthetic considerations in patients with permanent pacemaker
Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab.DCA, Dip. Software statistics PhD (physio) Mahatma gandhi medical college and research institute, puducherry, India Dr.S.Parthasarathy MD DNB PhD
2
Dr.S.Parthasarathy MD DNB PhD
Nomenclature cardiac implantable electronic device (CIED) Dr.S.Parthasarathy MD DNB PhD
3
Dr.S.Parthasarathy MD DNB PhD
Indications Artificial pacing is generally indicated for the treatment of symptomatic bradycardia of any origin. Innumerable !! two major indications 1. failure of impulse formation and 2.failure of cardiac conduction. Clinically , sick sinus syndrome and complete heart block Dr.S.Parthasarathy MD DNB PhD
4
Dr.S.Parthasarathy MD DNB PhD
Pacemaker What is the normal physiological pacemaker ? Sinus node If it fails AV node If it also … Dr.S.Parthasarathy MD DNB PhD
5
Dr.S.Parthasarathy MD DNB PhD
Conduction – normal Dr.S.Parthasarathy MD DNB PhD
6
Three basic techniques
Application of external pacing pads, insertion of a transvenous pacing lead via central venous access, implantation of permanent intracardiac leads along with a pulse generator Dr.S.Parthasarathy MD DNB PhD
7
Dr.S.Parthasarathy MD DNB PhD
Terms Pulse Generator Leads Electrode Pacing threshold Endocardial pacing Epicardial pacing Dr.S.Parthasarathy MD DNB PhD
8
Dr.S.Parthasarathy MD DNB PhD
Pacing code North American Society of Pacing and Electrophysiology (NASPE) British Pacing and Electrophysiology Group (BPEG) 3 letter previously to 5 letter now AAI, VVI , DDD Dr.S.Parthasarathy MD DNB PhD
9
What do the letters tell ??
Dr.S.Parthasarathy MD DNB PhD
10
Dr.S.Parthasarathy MD DNB PhD
Examples The AAI mode can be programmed for a patient with normal AV conduction with symptomatic sinus bradycardia to ensure adequate heart rate. The VVI mode can be programmed for a patient with AV block to ensure adequate ventricular rate. Dr.S.Parthasarathy MD DNB PhD
11
Dr.S.Parthasarathy MD DNB PhD
Dual modes the DDD mode can be programmed for patients with complete AV block and normal sinus node function to ensure that each spontaneous atrial depolarization is followed by a ventricular depolarization. This mode will provide AV sequential or ‘physiological pacing’ because the paced ventricular rate tracks the spontaneous atrial rate. Dr.S.Parthasarathy MD DNB PhD
12
Dr.S.Parthasarathy MD DNB PhD
AAI Block Dr.S.Parthasarathy MD DNB PhD
13
DDD or VDD depends on atrial rate
Dr.S.Parthasarathy MD DNB PhD
14
Image dual chamber pacing
Dr.S.Parthasarathy MD DNB PhD
15
The fourth and fifth positions
Rate modulation – in exercise LV and RV pacing CRT Dr.S.Parthasarathy MD DNB PhD
16
Dr.S.Parthasarathy MD DNB PhD
ICDs Implantable cardioverter defibrillators (ICDs) are devices capable of detecting a ventricular arrhythmia and delivering a defibrillatory shock. Devices are now implanted in a subcutaneous pectoral pocket with transvenous leads. Dr.S.Parthasarathy MD DNB PhD
17
Dr.S.Parthasarathy MD DNB PhD
Life The lithium-powered pacemakers can last 5 to 10 years for dual-chamber pacing and 7 to 12 years for single-chamber pacing. The silver-vanadium oxide batteries used in ICDs typically last for 5 to 8 years. Minimum check 12 months before surgery Dr.S.Parthasarathy MD DNB PhD
18
Dr.S.Parthasarathy MD DNB PhD
19
Dr.S.Parthasarathy MD DNB PhD
How to identify ?? Before p wave – spike - normal QRS – atria QRS wide with spike – ventricular pacing Two spikes in dual pacing Dr.S.Parthasarathy MD DNB PhD
20
sense RV and single, dual and triple chamber available
Dr.S.Parthasarathy MD DNB PhD
21
Dr.S.Parthasarathy MD DNB PhD
Pacemaker syndrome hypotension, syncope, vertigo, light-headedness,fatigue, exercise intolerance, malaise, weakness, lethargy, dyspnoea, and induction of congestive heart failure Possibly normal VA conduction interference with AV conduction May be due to atrial dilation Dr.S.Parthasarathy MD DNB PhD
22
Perioperative considerations
Dr.S.Parthasarathy MD DNB PhD
23
Preoperative evaluation
Systemic illness especially CVS review of all available medical records, ECGs, and chest X-rays.- patient and attender Cardiac consultation obtaining the manufacturer's identification card complete blood count, urinalysis, coagulation screening, serum electrolytes, BUN, blood sugar, Dr.S.Parthasarathy MD DNB PhD
24
Preoperative considerations
A preoperative history of vertigo, syncope in a patient with a pacemaker could reflect pacemaker dysfunction Pulse rate 10 % less than the set rate – battery ? Irregular pulse – sensing ?? Also due to inflammation and irritation causing multiple VPCs Dr.S.Parthasarathy MD DNB PhD
25
Dr.S.Parthasarathy MD DNB PhD
Pacemaker is working ?? Pacing spikes Get the rate slowed down , carotid massage, valsalva or edrophonium Not slowing down – then place a magnet over it to convert to fixed rate device Dr.S.Parthasarathy MD DNB PhD
26
Dr.S.Parthasarathy MD DNB PhD
Pacemaker – preop (a) battery life, (b) programmed pacing mode such as VVIR, DDDR, C)pacemaker dependency, (d) intrinsic rhythm, (e) magnet rate and rhythm -- newer programmable devices ?? (f) prior recorded arrhythmic events, and (g) pacemaker lead parameters Dr.S.Parthasarathy MD DNB PhD
27
Dr.S.Parthasarathy MD DNB PhD
What to know preop Dr.S.Parthasarathy MD DNB PhD
28
So basically – preop is divided
Patient Pacemaker Dr.S.Parthasarathy MD DNB PhD
29
Dr.S.Parthasarathy MD DNB PhD
Plan electromagnetic interference (EMI) there ?? Rate adaptive functions and anti tachyarrhythmia programming – suspend availability of temporary pacing and defibrillation equipment Dr.S.Parthasarathy MD DNB PhD
30
Dr.S.Parthasarathy MD DNB PhD
Monitors Hand on pulse Others (routine ) Pacemaker mediated tachycardia has been induced by intraoperative somatosensory evoked potential PA catheter and CVP are not directly related to PPM but the disease and surgery – beware of guidewire -- paced PA catheters !! Dr.S.Parthasarathy MD DNB PhD
31
Dr.S.Parthasarathy MD DNB PhD
EMI anaesthetic technique should be used according to the need of the patient Electromagnetic interference Dr.S.Parthasarathy MD DNB PhD
32
Dr.S.Parthasarathy MD DNB PhD
Diathermy – PPM Senses but uncertain – it does go to asynchronous mode PPM senses as R wave and stops pacing ?? Danger Dr.S.Parthasarathy MD DNB PhD
33
Dr.S.Parthasarathy MD DNB PhD
Diathermy Assure that the electrosurgical receiving plate is positioned so the current pathway does not pass through or near the CIED system. NO to cautery’s electrical field to the pulse generator or leads. short, intermittent and irregular bursts at the lowest feasible energy levels. bipolar electrocautery system or ultrasonic (harmonic) scalpel Dr.S.Parthasarathy MD DNB PhD
34
Dr.S.Parthasarathy MD DNB PhD
Other considerations Radiofrequency ablation Avoid direct contact between the ablation catheter and the pulse generator and leads. For electroconvulsive therapy, disable the ICD Radiation therapy is not associated with EMI Avoid focusing the lithotripsy beam near the pulse generator. Pacemaker malfunction can occur in patients undergoing ESWL, MRI is generally contraindicated in patients with CIEDs. Dr.S.Parthasarathy MD DNB PhD
35
Dr.S.Parthasarathy MD DNB PhD
TURP Coagulation current used during TURP procedure has no effect, but the cutting current at high frequencies (up to 2500 kc/sec) can suppress the output of a bipolar demand ventricular pacemaker. Dr.S.Parthasarathy MD DNB PhD
36
Dr.S.Parthasarathy MD DNB PhD
Bipolar PPM – better Dr.S.Parthasarathy MD DNB PhD
37
Dr.S.Parthasarathy MD DNB PhD
Be ready 1. atropine 2. isoprenaline Dr.S.Parthasarathy MD DNB PhD
38
Dr.S.Parthasarathy MD DNB PhD
Potassium An acute increase in extracellular potassium concentration as in patients with myocardial ischemia Rapid potassium replacement in chronic hypokalaemic patients use of depolarizing muscle relaxants in patients with burns, trauma or neuromuscular disease may increase the RMP to less negative value, thus making the capture easier. Dr.S.Parthasarathy MD DNB PhD
39
Dr.S.Parthasarathy MD DNB PhD
Other intraop factors Significant acid–base, electrolyte problems transfusion of blood, myocardial ischaemia, high blood concentrations of local anaesthetics can increase capture thresholds of the leads and alter lead impedance. Dr.S.Parthasarathy MD DNB PhD
40
Defib intraop required ??
1.Place the paddles more than 15cm away from the device 2.The paddles perpendicular to the path of the implanted leads 3.Use the lowest possible energy Dr.S.Parthasarathy MD DNB PhD
41
Postoperative management
Cardiac rate and rhythm should be monitored continuously throughout the immediate postoperative period. Postoperative interrogation and restoration of CIED function are basic elements of postoperative management. Dr.S.Parthasarathy MD DNB PhD
42
Dr.S.Parthasarathy MD DNB PhD
Summary Indications Coding Life Pacemaker syndrome Preop evaluation Monitors Anaesthesia – EMI- asynchronous – Rate mod Post op Dr.S.Parthasarathy MD DNB PhD
43
Dr.S.Parthasarathy MD DNB PhD
Thank you all Dr.S.Parthasarathy MD DNB PhD
Similar presentations
© 2024 SlidePlayer.com Inc.
All rights reserved.