Presentation is loading. Please wait.

Presentation is loading. Please wait.

The Electrical Management of Cardiac Rhythm Disorders Bradycardia Device Course The Electrical Management of Cardiac Rhythm Disorders, Bradycardia, Slide.

Similar presentations


Presentation on theme: "The Electrical Management of Cardiac Rhythm Disorders Bradycardia Device Course The Electrical Management of Cardiac Rhythm Disorders, Bradycardia, Slide."— Presentation transcript:

1 The Electrical Management of Cardiac Rhythm Disorders Bradycardia Device Course
The Electrical Management of Cardiac Rhythm Disorders, Bradycardia, Slide Presentation 06 Implant Technique Implant Technique

2 Pacemaker Components

3 Leads Epicardial Endocardial

4 Goals of Cardiac Pacing
The electrical management of bradyarrhythmias requires Ability to deliver enough energy to consistently depolarize the heart (capture) Ability to correctly sense intrinsic cardiac activity These functions are affected by many factors Settings of output parameters (pulse amplitude, pulse width) Sensitivity parameter settings Impedance Electrical concepts

5 Capture The capture threshold is defined as the minimum amount of electrical energy required to consistently depolarize the myocardium When a pacemaker output causes a depolarization, that is also called “capture” The capture threshold is also called the pacing threshold or the stimulation threshold The capture threshold is not constant It can change over time (disease, medications, age) It can even change over the course of the day!

6 What Affects the Capture Threshold?
Activity level Posture Time of day Comorbidity Heart failure Meals Drugs Disease progression

7 Capture Threshold Values
Atrial Ventricular Acute Threshold < 1.5 V < 1.0 V Chronic Threshold < 2.5 V Threshold values at implant should be low; expect chronic thresholds to increase These are suggested values and may not be possible for all patients Capture threshold values at implant should be low, usually not more than 1.5 V in the atrium and 1.0 V in the ventricle. The ideal chronic thresholds are below 2.5 V in both chambers. Keep in mind that such values are ideal and may not be possible for all patients.

8 Sensing Sensing refers to how well the pacemaker is able to “listen to” or perceive intrinsic cardiac events Important things to consider when talking about sensing Surface ECG Intracardiac electrogram (EGM) Sensing threshold Sense amplifier Sensitivity setting and sensitivity safety margin Unipolar/bipolar configurations Electromagnet interference

9 Surface ECG/Intracardiac EGM
Surface ECG: graphic depiction of heart’s electrical signals recorded from electrodes on the body’s surface Intracardiac EGM: graphic depiction of the heart’s electrical signals recorded by electrodes from inside the heart (pacing lead)

10 Sensitivity Safety Margin
Sensing thresholds are not constant and vary with many factors The sensitivity safety margin allows reliable sensing even with fluctuations in the sensing threshold Using this formula, the safety margin should be at least 2 at implant Safety margin = Sensing Threshold Sensitivity Setting Like capture thresholds, sensing thresholds can fluctuate with time and other factors, so a safety margin is required. The safety margin should be calculated using this formula and should be 2 or greater. For instance, if a patient’s sensing threshold is 2 mV then the sensitivity setting should be 1 mV. Likewise, if the patient’s sensing threshold is 4 mV, the sensitivity setting should be 8 mV. This is a 2:1 ratio. (Remember, increasing the mV setting decreases sensitivity.)

11 Electromagnetic Interference (EMI)
EMI is defined as electrical signals of nonphysiologic origin May interfere with pacemaker (temporarily or permanently) Common sources of EMI Cardioversion/defibrillation Electrocautery MRIs Extracorporeal shock wave lithotripsy (ESWL) Therapeutic radiation Radiofrequency ablation

12 What About Cardioversion/Defibrillation?
May permanently damage the pulse generator Can temporarily inhibit or reprogram the pacemaker Backup or noise reversion mode Myocardial thermal damage secondary to shock which may result in ventricular fibrillation, myocardial infarction, or both Guidelines Evaluate potential device interactions Place paddles 4 to 6 inches away from implanted pacemaker Orient paddles in anterior/posterior position, if possible This is a complicated subject that can only be addressed briefly. Cardioversion or defibrillation can cause a number of undesirable effects in a pacemaker patient. First, the shock energy can reprogram the pacemaker, might damage the pacemaker permanently, and thermal damage secondary to the shock traveling via the lead into the heart can cause VF, MI, or both. The pacemaker may revert to a backup mode, sometimes called a “noise reversion mode.” These vary by manufacturers and devices but should be described in the device manual or known by the manufacturer’s technical services hot line or representative. It is important to rule out possible noise reversion mode when unexpected pacemaker behavior is observed. In most cases, noise reversion mode remains in force until the device is reprogrammed. The reprogramming step is simple. Guidelines for pacemaker patients under consideration for cardioversion and defibrillation include how the paddles are placed (4 to 6 inches away from the implanted pacemaker) and oriented (AP, if possible). Careful consideration to potential device interactions should be weighed.

13 What About Electrocautery?
May reprogram or permanently damage the pacemaker May inhibit the pacemaker May cause the device to go into backup or noise reversion mode Myocardial thermal damage secondary to the transmission of the electrical energy may result in VF, MI, or both Guidelines Contraindicated

14 What About MRI? The magnet in the MRI device can cause asynchronous pacing (pacing without sensing) Guidelines Generally contraindicated Magna-Safe Study

15 What About Lithotripsy?
The vibrations in extracorporeal shock wave lithotripsy can damage the pacemaker (especially pacemakers with sensors, i.e. rate-adaptive units) Guidelines Program to VVI or VOO mode Keep focal point of lithotripter at least 6 inches away from the implanted pacemaker Monitor the heart throughout the procedure

16 What About Therapeutic Radiation?
Damage depends on dose Damage is cumulative; monitor device throughout course of radiation therapy Transistors may fail Pacemakers may fail but mode of failure cannot be predicted Guidelines Therapeutic ionizing radiation is contraindicated If therapeutic radiation is used, pacemaker should be shielded or moved to a less vulnerable location Note that there can be a difference between diagnostic radiation treatments and therapeutic radiation treatments. Some pacemakers utilize complimentary metal oxide semiconductors or CMOS technology. A CMOS-type pacemaker can be affected by therapeutic radiation but will not be affected by diagnostic radiation. To find out more, contact the technical services group of the manufacturer.

17 What About Radiofrequency Ablation?
RF ablation can temporarily or permanently reprogram the pulse generator Guidelines Interrogate the pacemaker following the procedure to verify proper function If necessary, reprogram RF ablation is unlikely to cause permanent damage to the device.

18 Myopotentials Myopotentials are muscle noises that are sensed by the pacemaker Can inhibit pacing The pacemaker senses the myopotential and inhibits the output, thinking the heart has beat on its own! Can interfere with sensing Can cause inappropriate pacing The pacemaker senses myopotential noise and inappropriately “thinks” it is atrial activity; it then tries to pace the ventricle to keep up or track that atrial activity

19 More EMI Sources Arc welding Automobile alternators Cell phones
Phone antenna should not overlap area of implanted pacemaker Talk on other side from implanted device Do not carry an activated cell phone near the implanted pacemaker May cause inappropriate inhibition, asynchronous pacing, backup mode, inappropriate rate adaptation, and mode switching Cellular Tested only from St. Jude Medical There are other sources of EMI but cell phones continue to be of major interest. After all, how many pacemaker patients are arc welders? But just about everyone uses cell phones. A cell phone may indeed interfere with a pacemaker. The results of that interference can include inapporpriate inhibition of the device, asynchronous pacing (no sensing), backup mode, and inappropriate behavior in terms of rate adaptation and mode switching. General guidelines for all pacemaker patients: do not hold the cell phone on the same side of the body as the implanted device and do not carry a cell phone that’s ON near the pacemaker implant site. St. Jude Medical is the world’s only pacemaker company to receive FDA approval to label its devices “Cellular Tested.” Since the Trilogy® pacemakers, all St. Jude Medical devices incorporate a protective filter which has been shown to help prevent unwanted signals from directly entering the pacemaker circuitry.

20 EMI in the Medical Environment
Electrocoagulation from electrocautery Defibrillation Electroconvulsive therapy Diathermy MRI Stimulators (e.g. transcutaneous nerve) Dental equipment Diagnostic ultrasound Low-frequency acupuncture Lithotripsy

21 EMI in the Industrial Environment
Arc welding Power lines Transformers Radio and TV transmitters Static charge Large metal frames in magnetic fields Induction furnaces and heaters Electrical switches

22 EMI in the Public Environment
CB radio Radiofrequency transmissions Telecommunications antennas Airport metal detectors Anti-theft detectors in stores These may not be marked! Digital cell phones Anti-theft detectors include detectors at store exits. In many stores, it is clear where these are located (if you are looking for them). However, pacemaker patients should know that some stores may conceal these detectors. These pose more of a risk to the pacemaker patient in that the patient may stand or linger near a source of EMI without knowing it.

23 Effects of EMI Pacemaker protection
Hardware backup circuits (to protect against loss of memory or software errors) Shields Effects EMI inhibition: pulse-to-pulse interval extends to the point that the pacemaker does not pace as often as it should. Noise reversion: change in mode (typically to asynchronous pacing at the programmed rate) which may require reprogramming. EMI tracking: acceleration of pacing as the pacemaker tries to track electromagnetic signals (“thinking” they are atrial signals) First, pacemakers do have protection against EMI. After all, we can’t eliminate EMI—it’s everywhere. Pacemakers have shields and also backup circuitry. There are three main effects of EMI. EMI may inhibit the device, it may cause a mode change (backup or noise reversion pacing), or it may accelerate the device. It is difficult to predict how a particular exposure to EMI will cause the device to function.

24 NASPE / BPEG (NBG) Pacemaker Code
Pacemaker Overview NASPE / BPEG (NBG) Pacemaker Code

25 NAPSE/BPEG Generic (NBG) Code
Position I II III IV V Category Chamber(s) Paced Chamber(s) Sensed Response to Sensing Rate modulation Multisite Pacing Letters Used O-None A-Atrium V-Ventricle D-Dual (A+V) O-None A-Atrium V-Ventricle D-Dual (A+V) O-None T-Triggered I-Inhibited D-Dual (T+I) O-None R-Rate modulation O-None A-Atrium V-Ventricle D-Dual (A+V) Manufacturer’s Designation Only S- Single (A or V) S- Single (A or V)

26 Magnet Use Pacemakers Pace Asynchronously (VOO or DOO) at the given battery rate (Temporarily) Device will revert back to exactly the same parameters it was programmed to once the magnet is removed ICD Will disable ICD Shock Therapy (Temporarily) Does not affect pacing Device will revert back to exactly the same parameters it was programmed to once the magnet is removed. Magnet must be placed over the device in order for temporary changes to occur.

27 A Systematic Approach to Diagnosing Rhythm Strips
Measure Base Rate Measure AV/PV Interval Verify Atrial capture Verify Atrial sensing Verify Ventricular capture Verify Ventricular sensing Verify Underlying rhythm Document

28 Dual Chamber ECG Analysis
What is the Analysis? Base Rate 60 ppm MTR 120 ppm AVD 200 ms PVARP 250 ms ECG # 1

29 Dual Chamber ECG Analysis
Base Rate 60 ppm MTR 120 ppm AVD 200 ms PVARP 250 ms What is the analysis? ECG # 2

30 Dual Chamber ECG Analysis
What is the analysis Base Rate 60 ppm MTR 120 ppm AV 200 ms PV 200 ms PVARP 250 ms ECG # 3

31 Dual Chamber ECG Analysis
What is the analysis? Base Rate 60 ppm MTR 120 ppm AV 200 ms PV 200 ms PVARP 250 ms ECG # 4

32 ECG Tracing Results!!! #1- Normal ECG –Dual chamber pacing and Atrial pacing w/ Ventricular (intrinsic) sensing. #2- Loss of atrial capture. #3- Normal ECG #4-No ventricular sensing and loss of ventricular capture.

33 ECG #5 Appropriate atrial sensing Appropriate ventricular sensing
No paced events, so capture cannot be evaluated

34 Answer Slide #5 Normal Sinus Rhythm
Can not determine any pacemaker function Pacers are usually set to pace above 50 or 60 bpm Single Chamber ICD- Pace above 40bpm Pacemakers only work when? Native heart rate goes below the base rate An intrinsic beat does not occur before the set Paced and Sensed AV Delays. Set at an Asynchronous Mode (VOO or DOO)

35 Thank you for your time!!!


Download ppt "The Electrical Management of Cardiac Rhythm Disorders Bradycardia Device Course The Electrical Management of Cardiac Rhythm Disorders, Bradycardia, Slide."

Similar presentations


Ads by Google