Measurement accuracy requirements for Reference Meters used in the calibration of dialysis machines This slide shows the IBP‘s products. Rigth: HDM99XP Middle: HDM97Pocket Left: Heat Exchanger for Dialysis Machines 100508
2
Recommendations If no manufacturer recommendation is available, a risk analysis must be carried out by the dialysis provider. B.Braun IBP Meters FMC UMED and IBP Meters Gambro Mesalab meters and IBP Meters Nikkiso IBP Meters
3
What accuracy do we need?
This is the most important question in a risk analysis. The Guidance Section of the international standard ISO "General Quality Assurance Requirements for Measuring Equipment" gives detailed information on this question: "The error attributable to calibration should be as small as possible. In most areas of measurement, it should be no more than one third and preferably one tenth of the permissible error of the confirmed equipment when in use.“ This means that the measuring equipment for the calibration of dialysis machines must have a measuring accuracy which is at least three times higher than the measuring accuracy of the dialysis machines.
4
References to calibrate
The goal of calibration is … … to connect the dialysis machine to the national standard. Chain of accuracy Dialysis Machine Referenc meters References to calibrate National Primary Standard Fresenius B.Braun Gambro Nikkiso IBP Mesalabs Temperature Conductivity Pressure NIST PTB etc. To be connected to the national standard the accuracy of each step in the chain must be at least 3 time higher.
5
References to calibrate
Conductivity chain of accuracy Dialysis Machine Referenc meters References to calibrate National Primary Standard 0.1 mS/cm 0.03 mS/cm 0.01 mS/cm mS/cm 0.01 mS/cm is already the technical border for meteorological reference institutes like NIST and PTB. This shows already the technical border for conductivity measurement.
6
Conductivity Measurement Accuracy Temperature Measurement Accuracy
Measuring Accuracy of Dialysis Machines Machine Conductivity Measurement Accuracy Temperature Measurement Accuracy FMC S ±0.1 mS/cm ±0.5°C Gambro AK200 +0.5/-1.5°C Gambro Phoenix ±0.2 mS/cm B.Braun Dialog ±0.5°C* Machine * Not available, value estimated
7
Temperature to compensate
Influence of Temperature Measurement Accuracy Conductivity mS/cm Temperature °C Reference Temperature 25 °C TPC 2,07 %/°C As the typical dialysate temperature is 37°C the meter needs to compensate 12 °C. Meter Temp. Accuracy Temperature to compensate Possible Error With correct TPC 0.05 °C 12 °C 0.017 mS/cm 0.5 °C 0.17 mS/cm 0.1 °C 0.35 ms/cm 0.2 °C 0.70 ms/cm Meter Accuracy at 37°C HDM97 – HDM99 90XL 90DX - NEO2 This example illustrates how important the accuracy of the temperature measurement is.
8
Dialysis Machine Manufacturer Temperature Coefficient
Conductivity Temperature Compensation To make technicians life interesting, different manufacturers of dialysing machines use different temperature coefficients. The following values are used: Dialysis Machine Manufacturer Temperature Coefficient Baxter (European Machine) 2.20%/°C B.Braun 2.10%/°C Fresenius Gambro (not the US Phoenix Machine) 1.80%/°C Hospal 2.07%/°C Nikkiso Bicarbonate conductivity Total conductivity 2.02%/°C 2.05%/°C
9
Conductivity of Solution Temperature of Solution
Conductivity Temperature Coefficient - Differences As the typical dialysate temperature ist 37°C the temperature compensation coefficient becomes very important. Conductivity of Solution Temperature of Solution Temp. Coefficient Machine Instrument Reading Difference 14.00 mS/cm 37.0 °C 2.07 %/°C 0.00 % 1.97 %/°C 14.17 mS/cm 1.21 % Thus, this example illustrates that the user must set the temperature coefficient correctly in order to ensure accurate conductivity measurements. The temperature coefficient must be selectable!
10
Temperature coefficient
Conductivity Temperature Coefficient Selection Display To use for Temperature coefficient at 25°C at 37°C BAXTER Baxter (Europe Dialysis-Machine) 2.000 %/K 2,200 %/K BELLCO Bellco 2,000 %/K 2,100 %/K B.BRAUN B.Braun FRESENI Fresenius GAMBRO Gambro AK-Type and Phoenix Machines 2,070 %/K HOSPAL Hospal NIK ALL Nikkiso Overall Conductivity 2,050 %/K NIK BIC Nikkiso Bicarbonate Conductivity STANDARD Other Dialysis Machines USER TC Adjustable 0 … 5,000 %/K nLF-ISO Non Linear Function for tap, purified and waste water. According Standard ISO7888 – DIN EN27888 Reference temperature 25°C RO-DWA Reverse Osmosis DWA DWA GmbH & Co. KG Ubstadter Straße 28 76698 Ubstadt-Weiher 0,00 %/K RO-HERCO Reverse Osmosis Herco Herco Wassertechnik GmbH Planckstraße 26 71691 Freiberg Reference temperature 25°C RO-LAUER Reverse Osmosis Lauer LAUER MEMBRAN WASSERTECHNIK GmbH Industriegebiet Speichermatt 79599 Wittlingen nLF-ISO according Standard ISO7888 – DIN EN27888 Reference temperature 25°C RO-PPW Reverse Osmosis PPW Phoenix Pure Water GmbH Siemens Technopark Werner-von-Siemens-Str. 2-6 D Bruchsal - Germany 2,200 %/K Reference temperature 25°C
11
Accuracy Dialysis Machine 1/3 accuracy according to the Standard
Machines vs. the Standard vs. Reference Meters Accuracy Dialysis Machine ±0.2 mS/cm ±0.1 mS/cm ±0.5°C 1/3 accuracy according to the Standard ±0.07 mS/cm ±0.03mS/cm ±0.17°C Accuracy IBP Meters ±0.03 mS/cm ±0.05°C Accuracy MESA 90XL ±0.03 mS/cm ±0.1°C Accuracy MESA NEO-2 ±0.05 mS/cm ±0.1°C Accuracy MESA 90DX ±0.1 mS/cm ±0.2°C The displayed accuracy values apply to 14 mS/cm and 37°C.
12
Which meter can be used with which machine
IBP Meters MESA 90XL NEO-2 MESA 90DX B.Braun FMC Gambro AK200 Gambro Phoenix Water Treatment Suitable - accuracy level of all channels higher than required Suitable for temperature but for conductivity only if temperature reading is be better than technical data Not suitable - accuracy below level required
13
Features comes from innovation
Dialysis Reference Meter This slide shows the IBP‘s products. Rigth: HDM99XP Middle: HDM97Pocket Left: Heat Exchanger for Dialysis Machines Heat Exchanger for Dialysis Machines 100508
14