1 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Requirements of.

Similar presentations


Presentation on theme: "1 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Requirements of."— Presentation transcript:

1 1 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Requirements of European Standards on Quality Assurance and Quality Control 1. Gaseous Pollutants Wolfgang Garche Saxony-Anhalt Environmental Protection Agency Department Air Quality Monitoring, Information, Assessment

2 2 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards European standards – Ambient air quality – Standard methods for the  EN 14211 measurement of NO 2 and NO by chemiluminescence  EN 14212measurement of SO 2 by UV fluorescence  EN 14625measurement of O 3 by UV photometry  EN 14626measurement of CO by NDIR spectroscopy  EN 14662measurement of benzene

3 3 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards  descriptions of the measuring principles  demands on specific compounds of the analysers and the sampling system  the minimum performance criteria for analysers  descriptions of the type approval test procedures  requirements and recommendations for QA and QC  guidance for determining the accuracy of measurements European standards – Contents

4 4 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Quality control is necessary in order to ensure that the uncertainties of the measured values are kept within the stated limits during continuous monitoring periods in the field. This requires that maintenance, test and calibration procedures shall be followed which are essential for obtaining accurate and traceable air quality data. In the EN-standards, procedures for maintenance, checks and calibration are given. The following procedures are regarded as a minimum necessary for maintaining the required quality level. Statement - Quality control

5 5 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Compound Uncertainty of individual measurements (on a 95% confidence interval) Minimum data capture SO 2 15 %90 % NO 2 / NO x 15 %90 % O3O3 15 % 90 % (Sommer) 75 % (Winter) CO15 %90 % C6H6C6H6 25 %90 % PM 10 + Pb 25 %90 % Data quality objectives

6 6 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Requirements on Ongoing Quality Control MeasuresFrequencyAction criteria Calibration of the analyserAt least every three months and after repair Certification of test gasesAt least every six months Zero: ≥ detection limit Span: ≥ 5,0 % from last certified value Zero and span checkAt least every two weeks (Recommended every 23 h or 25 h.) Zero: ≥ 5 nmol/mol Span: ≥ 5,0 % of initial span value Lack of fit checkWithin 1 year after installation and after repair; further frequency depending on the result of test lack of fit > 6,0 % of the measured value Testing sample manifold -influence of pressure drop induced by the manifold pump - sample collection efficiency At least every three years influence > 1 % of measured value influence > 2 % of the measured value Test of the sampling linesAt least every six months ≥ 2 % sample loss Regular maintenance of the components of the analyser As required by manufacturer As required Changing of (if applicable) drying material and other consumables: At least every six months As required Change of particulate filtersAt least every three months Response to span gas passing the filter is ≤ 97 %

7 7 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards  every three months  at a recommended concentration of 70 % to 80 % of the certification range  used calibration gases shall be traceable to national standards (maximum permitted uncertainty is ± 5 % with a level of confidence of 95 %)  zero gas shall give no instrument reading higher than the detection limit  calibration gases shall be introduced over the filter to check and if necessary to correct for contamination of the filter Calibration of the analysers

8 8 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Zero and span checks  Zero and span gas can be supplied by gas cylinder, or generated by an external calibrator unit or internally in the analyser.  The zero and span gas shall pass through the particulate filter. Frequency of test: At least every two weeks. Recommended every 23 h or 25 h. Action criteria: Zero ≥ 5 ppb (CO ≥ 0,2 ppm, Benzene ≥ 0,5µg/m³) of the initial zero span; Span ≥ 5 % of initial span value. Appropriate action: The site should be visited and the correct function of the analyser determined by calibration.

9 9 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Certification of test gases  Gases for span and zero checks shall be verified at least every six months with use of reference gases traceable to national standards. Action criteria: Zero ≥ detection limit; Span ≥ 5 % of previous verification. Appropriate action: Replace the cylinder.

10 10 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Lack of fit check  The lack of fit of the analyser shall be tested using the following concentrations: 0 %, 20 %, 60 % and 95 % of the maximum of the certification range. At each concentration (including zero) at least two individual readings shall be performed.  Lack of fit may be checked either in the laboratory or on site. Frequency of test: After installation and after repair the analyser shall be tested within 1 year; if the result of this test is ≤ 2,0 % of the measured value the next test shall be carried out within 3 years; if the result is > 2,0 % but ≤ 6,0 % the analyser shall be retested within 1 year; if the result is > 6,0 % appropriate action shall be taken; Action criteria: > 6,0 % of the measured value. Appropriate action: Remove analyser from site for further testing and repair if necessary.

11 11 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Zero gas NO CO / mln / min CO SO 2 Mixing chamber Ozone generator to analysers PC controlled gas mixing system

12 12 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Test of sampling lines Testing comprises two measurements of the test gas, direct sampling of test gas and the sampling of test gas through the sampling line. Frequency: At least every six month Action criteria: Sample loss > 2 %. Appropriate action: Clean or replace the sampling lines and retest.

13 13 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Testing of the sample manifold  Measurement of the pressure drop in the sample manifold.  Calculation of the induced effect on the analyser’s respond. Action criteria: Influence of the pressure drop induced by the manifold sampling pump on the measured concentration < 1 %. Appropriate action: Reduce flow through manifold to reduce the induced pressure drop. Influence of pressure drop: Sample collection efficiency of the sampling system:  Testing comprises two measurements of the test gas, direct sampling of test gas and the sampling of test gas from the complete sampling system. Action criteria: Sample loss > 2 %. Appropriate action: Clean/replace/repair manifold as necessary and retest.

14 14 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards pSpS pApA  R - is the change in the analyser’s response due to the influence of the pressure drop b gp - is the sensitivity coefficient of the analyser to sample gas pressure change (should be known from the type approval test) p A - is the ambient pressure p S - is the pressure in the sampling manifold Influence of pressure drop

15 15 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Test of converter efficiency (only for NO/NO2-analysers)  measurements with calculated amounts of NO 2. This can be achieved by means of gas-phase titration.  two concentration levels: at about 50 % and about 95 % of the maximum of the certification range of NO 2 Frequency: At least every year Action criteria: converter efficiency < 95 %. Appropriate action: replace the converter

16 16 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Many Thanks for your Attention!


Download ppt "1 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop 22.05.2007 European Standards Requirements of."