Presentation is loading. Please wait.

Presentation is loading. Please wait.

Adjusting comfort settings on Newport HT70 Series Ventilators

Similar presentations


Presentation on theme: "Adjusting comfort settings on Newport HT70 Series Ventilators"— Presentation transcript:

1 Adjusting comfort settings on Newport HT70 Series Ventilators
Narrated by Cyndy Miller, RRT

2 Comfort and Synchrony HT70 “Comfort” settings impact patient-ventilator synchrony. Synchrony impacts time to liberation for those patients who are weaning from the ventilator. Comfort impacts well being for those who are ventilator dependent. HT70-Adjusting Comfort Settings 0412

3 Dys-synchrony-Discomfort
Patient looks uncomfortable or tells you they are uncomfortable. Patient doesn’t reliably trigger the ventilator when Trigger is set at a normal trigger setting. Unusually high or low respiratory rate. Double triggering. Tidal volume fluctuates a lot from breath to breath. HT70-Adjusting Comfort Settings 0412 3 3

4 Comfort Settings Condition Adjustment Volume Control breaths
Flow & Flow Waveform Pressure Control breaths Slope Rise Pressure Support breaths Slope Rise, Expiratory Threshold, PS Max i Time HT70-Adjusting Comfort Settings 0412

5 Comfort Settings: Volume Control Flow and Flow Waveform
Higher Flow Flow to Patient Lower Flow HT70-Adjusting Comfort Settings 0412

6 Volume Control Breaths Delivery
End How does Volume Control work? Flow delivery is determined by user settings Flow Airway Pressure Pressure is generated by the ventilator to push the set Tidal Volume into the patient PEEP Breath ends when the Inspiratory Time elapses/Tidal Volume has been delivered. HT70-Adjusting Comfort Settings 0412

7 Volume Control Breaths – Comfort Settings
You select the Peak Flow. You select the Flow Pattern. The Peak Flow and Flow Pattern don’t change even if the patient inhales at a higher or lower flow. HT70-Adjusting Comfort Settings 0412

8 Volume Control Breaths – (Peak) Flow
(Peak) Flow is set on the Main Screen HT70-Adjusting Comfort Settings 0412

9 Volume Control Breaths – (Peak) Flow
Patients who are spontaneously breathing will already be inhaling at a particular flowrate. If ventilator Flow is set too low, the patient will work harder to breathe. The (Peak) Flow you set must be at or above the patient’s spontaneous flowrate in order for the ventilator to take the workload away from the patient. Increase the (Peak) Flow setting until you see pressure rise smoothly when the patient inhales and you see the breath end in time to allow full exhalation (avoid auto-PEEP). HT70-Adjusting Comfort Settings 0412

10 Volume Control Breaths – (Peak) Flow
Typical flowrates in Volume Control: Adults L/min Pediatrics: L/min HT70-Adjusting Comfort Settings 0412

11 Volume Control Breaths – (Peak) Flow
Setting (Peak) Flow for patients who are not spontaneously breathing should be focused on achieving appropriate breath timing. Set Flow so that the patient has plenty of time to exhale in order to avoid auto-PEEP. HT70-Adjusting Comfort Settings 0412

12 Volume Control Breaths – (Peak) Flow
Flow and Inspiratory Time are mathematically linked. Setting the Flow at a higher flowrate means making the breath shorter. And vise verse. If you increase the Flow to meet patient’s needs, it could make the Inspiratory Time too short. Flow Time HT70-Adjusting Comfort Settings 0412

13 Volume Control Breaths – (Peak) Flow
Increasing the Flow setting (decreasing Inspiratory Time) to meet the flow demands of an actively breathing patient can cause double triggering and dys-synchrony if patient effort continues past when the breath ends. . V S L/min 1 2 3 4 5 6 The need for higher flow may be remedied by changing from a Square to Descending Ramp Flow Waveform first, then adjusting Flow. When you change to a Descending Ramp, the flowrate is automatically increased without shortening inspiratory time. HT70-Adjusting Comfort Settings 0412

14 Volume Control Breaths - Flow Waveform
Flow Waveform is set on the More Screen (along with other “comfort settings”) HT70-Adjusting Comfort Settings 0412

15 Volume Control Breaths - Flow Waveform
There are two selections: Square Descending Ramp HT70-Adjusting Comfort Settings 0412

16 Volume Control Breaths - Flow Waveform
Challenge Situation: VT 350 mL Square Wave - Flow 30 L/min. Insp Time 0.7 s Patient is stressed and needs more flow. HT70-Adjusting Comfort Settings 0412

17 Volume Control Breaths - Flow Waveform
Challenge Situation: VT 350 mL Square Wave - Flow increased to 50 L/min. Caused Insp Time to decrease to 0.4 s Inspiratory Time too short, double triggering occurs. Pt-ventilator dys-synchrony and fatigue follow. This did not solve the problem. HT70-Adjusting Comfort Settings 0412

18 Volume Control Breaths - Flow Waveform
Challenge Situation: VT 350 mL Change to Descending Ramp Flow Wave – Set Flow to 50 L/min. Insp Time is now 0.6 s! Adequate for the patient. Patient is now comfortable. Flow needs are met without the breath being too short! HT70-Adjusting Comfort Settings 0412

19 Comfort Settings: Pressure Control Slope/Rise
Faster (set closer to 1) Slope Rise Slower (set closer to 10) Slope Rise Target Pressure PEEP HT70-Adjusting Comfort Settings 0412

20 Pressure Control Breaths Delivery
How does Pressure Control work? Flow delivery is a result of: Pressure change needed, Slope Rise setting, patient effort and patient pulmonary mechanics. Flow Airway Pressure Pressure delivery is determined by pressure change needed (difference between PC setting and PEEP). PEEP The actual target pressure is the PC setting above room pressure, (not above PEEP) HT70-Adjusting Comfort Settings 0412

21 Pressure Control Breaths End
How does a Pressure Control breath end? Flow Breath ends when the Inspiratory Time elapses. Airway Pressure PEEP The user set Inspiratory Time determines the duration of the breath HT70-Adjusting Comfort Settings 0412

22 Pressure Control Breaths – Slope Rise
The Slope Rise setting determines the rate of pressure rise to the Pressure Control level. Airway Pressure PC setting PEEP HT70-Adjusting Comfort Settings 0412

23 Pressure Control Breaths – Slope Rise
Slope Rise is set on the More Screen (along with other “comfort settings”) HT70-Adjusting Comfort Settings 0412

24 Pressure Control Breaths – Slope Rise
Range= 1-10 HT70 Plus = 1 fastest HT70 = 10 fastest HT70-Adjusting Comfort Settings 0412

25 Pressure Control Breaths – Slope Rise Work of Breathing
A faster Slope-Rise with a squared pressure waveform delivers a higher peak-flow and higher machine assistance for the patient at the onset of the breath. A slower Slope-Rise with a slanted pressure waveform delivers a lower peak-flow and less machine assistance for the patient at the onset of the breath. Target Pressure Airway Pressure PEEP HT70-Adjusting Comfort Settings 0412

26 Pressure Control Breaths – Slope Rise Volume Delivery
A faster Slope Rise can provide the opportunity for earlier volume delivery into the alveoli. Remember that alveolar pressure always lags behind airway pressure. When airway pressure rises slowly, alveolar pressure rises even more slowly. If it rises too slowly, alveolar pressure may not even reach the target peak pressure before the breath ends. Target Pressure PEEP If the user does not change Inspiratory Time, a faster Slope Rise setting may deliver a higher tidal volume. HT70-Adjusting Comfort Settings 0412

27 Pressure Control Breaths – Slope Rise
How can you tell if Slope Rise is set too fast? “Ringing” in the flow waveform Sounds like “ch ch ch ch ch” Flow Slow it down a little HT70 = Set it at lower number HT70 Plus = Set it at higher number HT70-Adjusting Comfort Settings 0412

28 Comfort Settings: Pressure Support
HT70-Adjusting Comfort Settings 0412

29 Pressure Support Breaths Delivery
How does Pressure Support work? Flow delivery is a result of the pressure change that must happen, the Slope Rise setting, the Expiratory Threshold setting, the PS Max i Time setting, patient effort and patient pulmonary mechanics. Flow Airway Pressure Pressure delivery is determined by pressure change needed (the actual PS setting). PEEP The Pressure Support setting IS the pressure change (above PEEP). HT70-Adjusting Comfort Settings 0412

30 Pressure Support Breaths End
How does a Pressure Support breath end? It ends by one of three mechanisms: Flow delivery drops to a user-set percent of the peak flow. Pressure rises 3 cmH2O/mbar above the target pressure. The breath duration reaches the user-set PS Max i Time setting. 100% of Peak Flow Flow Set from 5-85% of Peak Flow Target Pressure + 3 cmH2O Target Pressure PS setting + PEEP PAW PEEP PS Max i time = Maximum inspiratory time for pressure support breaths set from 0.1 to 3 seconds. HT70-Adjusting Comfort Settings 0412

31 Comfort Settings: Pressure Support Slope/Rise
Faster (set closer to 1) Slope Rise Slower (set closer to 10) Slope Rise Target Pressure PEEP HT70-Adjusting Comfort Settings 0412

32 Pressure Support Breaths – Slope Rise
Pressure Support breaths and Pressure Control breaths use the same Slope Rise setting and it works the same for both breath types. Airway Pressure Target Pressure PEEP HT70 = 10 fastest HT70 Plus = 1 fastest HT70-Adjusting Comfort Settings 0412

33 Pressure Support Breaths – Slope Rise
So we don’t need to cover it again, right? HT70-Adjusting Comfort Settings 0412

34 Comfort Settings: Pressure Support Expiratory Threshold & PS Max i Time
5% = longest setting, 85% = shortest setting Flow Expiratory Threshold Set from 5-85% of Peak Flow PS Max I Time Set from seconds HT70-Adjusting Comfort Settings 0412

35 Pressure Support Breaths End
User sets Expiratory Threshold and PS Max i Time 100% of Peak Flow Flow User does not set the pressure cycling off value. It is preset and not user adjustable. The pressure cycling off mechanism is intended to end the breath automatically if the patient exhales and pressure rises. Expiratory Threshold Set from 5-85% of Peak Flow Pressure rises 3 cmH2O above target pressure (not adjustable) PAW Target Pressure PEEP PS Max i time Set from 0.1 to 3 seconds. HT70-Adjusting Comfort Settings 0412

36 Pressure Support Breaths – Expiratory Threshold
Expiratory Threshold is set on the More Screen (along with other “comfort settings”) HT70-Adjusting Comfort Settings 0412 36 36

37 Pressure Support Breaths – Expiratory Threshold
During a pressure support breath, flow starts out at a higher level and decelerates (declines) to a lower level. How many of you know why the flow rate decelerates? Flow Time HT70-Adjusting Comfort Settings 0412

38 Pressure Support Breaths – Expiratory Threshold
If flow keeps going at a preset rate, pressure keeps climbing. . V Paw cmH2O/ mbar In order to keep pressure at the airway from rising above the target pressure, flow delivery from the ventilator must decline in direct proportion to the rate that the lungs fill. HT70-Adjusting Comfort Settings 0412

39 Pressure Support Breaths – Expiratory Threshold
The maximum flow delivered during the breath is called the “Peak Flow” Peak Flow Expiratory Threshold = a % of Peak Flow Inspiration Exhalation Flow When flow drops from the peak flow to the % of peak flow set as the Expiratory Threshold, the breath cycles from inspiration to exhalation Time HT70-Adjusting Comfort Settings 0412

40 Pressure Support Breaths – Expiratory Threshold
Example If the Peak Flow is 100 L/min Peak Flow = 100 L/min Expiratory Threshold = 33% Flow = 33 L/min And Expiratory Threshold is set at 33% Inspiration Exhalation Flow The breath will change from the inspiratory phase to the expiratory phase when flow delivery to the patient drops from 100 L/min down to 33 L/min Time HT70-Adjusting Comfort Settings 0412

41 Pressure Support Breaths – Expiratory Threshold
Causes Pressure Support to cycle to exhalation when the patient stops inhaling Expiratory Threshold is an indirect setting for inspiratory time. The relationship between Expiratory Threshold and inspiratory time is a bit complex…………….. An ideal expiratory threshold setting = HT70-Adjusting Comfort Settings 0412

42 Pressure Support Breaths – Expiratory Threshold
Some people’s lungs fill/empty slowly Some people’s lungs fill/empty quickly Flow This is called a long time constant. This is called a short time constant Time Time Constant is the product of compliance and resistance. HT70-Adjusting Comfort Settings 0412

43 Pressure Support Breaths – Expiratory Threshold
To end up with a normal inspiratory time, the patient on the left (long time constant) needs a much higher Exp Threshold (% of peak flow) setting than the patient on the right (short time constant). Flow Insp Time = ~1 sec Time HT70-Adjusting Comfort Settings 0412

44 Pressure Support Breaths – Expiratory Threshold
If Expiratory Threshold Too Low Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

45 Pressure Support Breaths – Expiratory Threshold
The breath may cycle-off too late Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

46 Pressure Support Breaths – Expiratory Threshold
This can cause the patient to exhale against continuing support Flow Target Pressure Airway Pressure PEEP HT70-Adjusting Comfort Settings 0412

47 Late cycling off can cause an increase in work, fatigue and delayed liberation (delayed weaning) from mechanical ventilatory support. And also cause the ventilator to miss triggering a breath in response to the patient’s next breathing effort Flow Missed Triggers are one of the most common causes of weaning delay Target Pressure Airway Pressure PEEP HT70-Adjusting Comfort Settings 0412

48 Pressure Support Breaths – Expiratory Threshold
If Expiratory Threshold Too High Flow Early cycling off may occur and this commonly causes double triggering, followed by a “missed trigger”. Target Pressure Airway Pressure PEEP Time (s) Both early and late cycling off can cause missed triggers and discomfort/delayed liberation from mechanical ventilation. HT70-Adjusting Comfort Settings 0412

49 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Setting for Slope Rise and setting for Expiratory Threshold are interrelated Flow Time HT70-Adjusting Comfort Settings 0412

50 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
If you have a patient who can benefit from extra assistance at the beginning of the breath, you need make the Slope Rise faster, right? Flow Time HT70-Adjusting Comfort Settings 0412

51 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Be careful. Making the Slope Rise faster will cause the Expiratory Threshold setting to be reached sooner, so the breath will get shorter if you don’t lower the Expiratory Threshold setting at the same time. Flow Time HT70-Adjusting Comfort Settings 0412

52 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Watch this example! Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

53 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

54 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

55 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

56 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

57 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Flow Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

58 Pressure Support Breaths – Slope Rise & Expiratory Threshold Interactivity
Flow If you make Slope Rise faster and that change results in double triggering, it means that you need to lower the Expiratory Threshold setting. Target Pressure Airway Pressure PEEP Time (s) HT70-Adjusting Comfort Settings 0412

59 Early cycling off Late cycling off Ch ch ch ch ch ch
That’s the sound of double triggering!! Early cycling off Ch ch ch ch ch ch Late cycling off Tidal volume fluctuate a lot. Rate is lower. That’s missed triggering and auto-PEEP! HT70-Adjusting Comfort Settings 0412

60 Pressure Support Breaths – PS Max i Time
What happens if there is a leak in the system? Flow may not decline during Pressure Support because the ventilator is trying to keep pressure in the “Big Lung” at the target pressure. PS Max i Time to the rescue! PS Max i Time set at 1.2 seconds Flow Target Pressure PAW PEEP Hint: Big Lung is the room! HT70-Adjusting Comfort Settings 0412 60 60

61 Pressure Support Breaths – PS Max i Time
PS Max i Time is set on the More Screen (along with other “comfort settings”) HT70-Adjusting Comfort Settings 0412 61 61

62 Pressure Support Breaths – PS Max i Time
This setting determines the maximum time allowable for a Pressure Support Breath. Even if flow never declines and the Expiratory Threshold is never met, the breath will stop when this time setting elapses. HT70-Adjusting Comfort Settings 0412 62 62

63 Pressure Support Breaths – PS Max i Time versus Expiratory Threshold
So which has the ultimate control? PS Max i Time or PS Expiratory Threshold? PS Max I Time Expiratory Threshold HT70-Adjusting Comfort Settings 0412 63 63

64 Pressure Support Breaths – PS Max i Time versus Expiratory Threshold
Whichever is achieved first! HT70-Adjusting Comfort Settings 0412 64 64

65 How to fix early cycling off
Review Decrease the Expiratory Threshold Setting to lengthen the breath Lengthen the PS Max i Time Setting to stop from cutting the breath short For mandatory breaths, lengthen the inspiratory time HT70-Adjusting Comfort Settings 0412

66 How to fix late cycling off
Review Increase the Expiratory Threshold Setting to shorten the breath Shorten the PS Max i Time Setting to cut the breath off in case of a leak For mandatory breaths, shorten the inspiratory time HT70-Adjusting Comfort Settings 0412

67 Quick Comfort Summary Adjust Flow and Flow Waveform to ensure optimal flow support (ventilator does the work) and the most compatible inspiratory time during Volume Control breaths. Adjust Slope Rise to ensure optimal flow support (ventilator does the work) at the onset of every Pressure Support and Pressure Control breath. Adjust Expiratory Threshold to create a compatible inspiratory time for all Pressure Support breaths. Adjust PS Max i Time to create a failsafe back up cycling off mechanism in case a leak in the circuit or around the patient’s mask or artificial airway causes flow to remain high during Pressure Support breaths. HT70-Adjusting Comfort Settings 0412

68 Getting Help is Easy Urgent! NON-Urgent
Clinical questions - Technical questions - Call Cyndy Miller ext. 218 NON-Urgent Call Clinical Assistance 24/7 Hotline: Ext. 123 Technical Service 24/7 Hotline: Ext. 500 Urgent! HT70-Adjusting Comfort Settings 0412

69 HT70-Adjusting Comfort Settings 0412

70 End of the Presentation
HT70-Adjusting Comfort Settings 0412


Download ppt "Adjusting comfort settings on Newport HT70 Series Ventilators"

Similar presentations


Ads by Google