AUDIOMETRY An Audiometer is a machine, which is used to determine the hearing loss in an individual.

Slides:



Advertisements
Similar presentations
Pure Tone Audiometry Training Course ——Marketing Dept.
Advertisements

Occupational Audiometric Testing Part 1: Purposes and Procedures Thomas W. Rimmer, ScD, CIH Fay W. Boozman College of Public Health University of Arkansas.
Pure Tone Audiometry Basic Principles of Sound The Audiometer
Foundations of Physics
Masking: Insuring that the non-test cochlea is not helping out. Making sure that the ear you want to test is the one that is responding.
Microphones and Sound Chapter 5. The Basics of Sound Sound itself is a vibration—a specific motion—of air molecules. A sound source creates changes in.
Chapter 6: Masking. Masking Masking: a process in which the threshold of one sound (signal) is raised by the presentation of another sound (masker). Masking.
M.Sc. in Medical Engineering
Audiological Evaultions
1© Manhattan Press (H.K.) Ltd. Loudness, pitch and quality Intensity level and loudness Intensity level and loudness 11.2 Properties of sound.
1 Multimedia Systems 1 Dr Paul Newbury School of Engineering and Information Technology ENGG II - 3A11 Ext: 2615.
ANSI Overview Hearing Aid Testing and Verification What is ANSI?
TOPIC 4 BEHAVIORAL ASSESSMENT MEASURES. The Audiometer Types Clinical Screening.
IT-101 Section 001 Lecture #15 Introduction to Information Technology.
Chapter 6: The Human Ear and Voice
PURE TONE AUDIOMETRY BALASUBRAMANIAN THIAGARAJAN DRTBALU'S OTOLARYNGOLOGY ONLINE.
1 Live Sound Reinforcement Audio measurements. 2 Live Sound Reinforcement One of the most common terms you will come across when handling any type of.
Lecture 12 10/03/2015  The child is conditioned to wait for a frequency specific stimuli in the sound field and respond using a play activity  used.
© 2010 Pearson Education, Inc. Conceptual Physics 11 th Edition Chapter 21: MUSICAL SOUNDS Noise and Music Musical Sounds Pitch Sound Intensity and Loudness.
Measurement of Hearing and Audiogram Interpretation
HISTORY OF HEARING TESTING. Audiology Audio - Latin…to hear, pertaining to hearing Logy - Greek…logus…science Therefore the science of hearing and hearing.
AUDIOMETRIC TESTING PROTOCOLS and TECHNIQUES
Summer 2007 SPA 4302 Pure Tone Audiometry.
Audiology Training Course ——Marketing Dept. Configuration of the ear ① Pinna ② Ear canal ③ Eardrum ④ Malleus ⑤ Incus ⑥ Eustachian tube ⑦ Stapes ⑧ Semicircular.
Sound Sound is a wave that carries vibrations. It is mechanical, longitudinal, and a pressure wave.
1 Special Testing. 2 Site of Lesion Cochlear—sensory Nerve—neural (retrocochlear) Reliability vs. validity.
C-15 Sound Physics Properties of Sound If you could see atoms, the difference between high and low pressure is not as great. The image below is.
EMBEDDED MICROCONTROLLER BASED SYSTEM FOR MEASUREMENT OF NEURO SENSORY HEARING LOSS.
15.1 Properties of Sound  If you could see atoms, the difference between high and low pressure is not as great.  The image below is exaggerated to show.
Sound Waves Sound waves are divided into three categories that cover different frequency ranges Audible waves lie within the range of sensitivity of the.
Test of Hearing And Pure tone Audiometry
Bastaninejad, Shahin, MD, ORL-HNS
Introduction to Audio. What is "Audio"? Audio means "of sound" or "of the reproduction of sound“. Specifically, it refers to the range of frequencies.
Pure Tone Audiometry SPA 4302 Summer A, The Pure-Tone Audiometer Electronic device that generates tones for determining hearing thresholds Manufactured.
Test of Hearing And Pure-tone Audiometry
Physics of Sound Part 1 Sound waves How they are generated and travel.
Chapter 5: Normal Hearing. Objectives (1) Define threshold and minimum auditory sensitivity The normal hearing range for humans Define minimum audible.
Pure Tone Audiometry most commonly used test for evaluating auditory sensitivity delivered primarily through air conduction and bone conduction displayed.
1 RF (Radio Frequency) technology BASIC TELECOMMUNICATIONS.
Hearing Lab Tests. Rinne Test rinne.htm rinne.htm
Loudness level (phon) An equal-loudness contour is a measure of sound pressure (dB SPL), over the frequency spectrum, for which a listener perceives a.
HEARING Structure and Function. Function of auditory system Analyse Discriminate -sounds -widely varying intensity widely varying frequencies any spatial.
MASKING BASIC PRINCIPLES CLINICAL APPROACHES. Masking = Preventing Crossover Given enough intensity any transducer can stimulate the opposite cochlea.
Sound and Its System. What is sound? basically a waveform of energy that is produced by some form of a mechanical vibration (ex: a tuning fork), and has.
1 Hearing Sound is created by vibrations from a source and is transmitted through a media (such as the atmosphere) to the ear. Sound has two main attributes:
Threshold of detection Established according to some criterion.
Speech Audiometry. Pure Tone Audiometer Schematic of Speech Audiometer.
Main Components of the Hearing Mechanism: Outer Ear Middle Ear Inner Ear Central Auditory Nervous System.
AUDITORY FUNCTION.  Audition results from sound conduction by either air or bones of the skull or both. Sound waves are converted (mechanically in.
Introduction to Audiology How to Read an Audiogram Degrees of Hearing Loss Types, Causes and Prevalence of Hearing Loss Types of Tests.
15.1 Properties of Sound. Chapter 15 Objectives  Explain how the pitch, loudness, and speed of sound are related to properties of waves.  Describe how.
Chapter 18 Oscillation, Wave and Sound. Oscillation equation Pendulum Wave equation.
Physics Mrs. Dimler SOUND.  Every sound wave begins with a vibrating object, such as the vibrating prong of a tuning fork. Tuning fork and air molecules.
Conductive hearing loss: there is obstruction of external auditory canal or disease of middle ear. Sensorineural hearing loss: disease affecting cochlea.
MULTIWAY SPEAKER SYSTEM
HEARING- 3. LEARNING OBJECTIVES LEARNING OBJECTIVES Discuss the principles used in performing tests of hearing Discuss the principles used in performing.
Narnarayan Shastri Institute Of Technology SUBJECT:- AVS FACULTY:- Malhar Chauhan FIELD:- E.C SEM-5 TH TOPIC:- Types Of Microphones Prepared By, PATEL.
FUNCTION GENERATOR.
IT-101 Section 001 Lecture #15 Introduction to Information Technology.
Equipment for Measuring Hearing and Calibration
Hearing tests.
Loudness level (phon) An equal-loudness contour is a measure of sound pressure (dB SPL), over the frequency spectrum, for which a listener perceives a.
Loudness level (phon) An equal-loudness contour is a measure of sound pressure (dB SPL), over the frequency spectrum, for which a listener perceives a.
Two Vacuums Shopvac Bosch Dept. of Mech. Engineering 1
Structure and Function
Overview Communication is the transfer of information from one place to another. This should be done - as efficiently as possible - with as much fidelity/reliability.
"Digital Media Primer" Yue-Ling Wong, Copyright (c)2013 by Pearson Education, Inc. All rights reserved.
C-15 Sound Physics 1.
HEARING ASSESSMENT: PURE TONE AUDIOMETRY
Sound Waves and Ultrasound
Presentation transcript:

AUDIOMETRY An Audiometer is a machine, which is used to determine the hearing loss in an individual.

Classification Audiometers Pure tone audiometer Speech audiometer Air conduction Bone conduction Speech thresholds

Audiometer Generally employed transducers in audiometer are the following: Earphone Microphone Bone-vibrator Loud speakers

Earphones: Earphones are usually of the moving coil type and gives reasonably flat frequency response upto 6 KHz after which their sensitivity decreases rapidly. They are specially designed for audiometric applications rather than for communication purposes.

Microphones: These are used to translate wave motion in space into electrical signal. Two types which are carbon button changes resistance with the pressure. The second one is the electrodynamic type in which the voltage is induced in a coil by its motion relative to a magnet

Simple audiometer An audiometer will essentially have an oscillator along with a pair of head phones. Pure tone audiometers and speech audiometers are two main groups of audiometers and are grouped according to the basis of the stimulus they provide to elicit audio response. The intensity range of most audiometers starts from approximately 15 dB above normal to 95 below normal over a frequency range from approximately 500 to 4000 Hz.

Pure tone audiometer Pure Tone Audiometer works on the principle of presenting specific pure tone signals to the subject and determining the intensity at which they can barely hear these signals. They are calibrated in terms of frequency and output A pure tone is the simplest type of auditory stimulus. generate test tones in octave steps from 125 to 80000Hz, the signal intensity ranging from – 10 dB to dB. frequency range of Hz. Changes in threshold sensitivity associated with various middle ear surgical procedures can be monitored more accurately with pure tone than speech tests.

AIR & BONE CONDUCTION: Air conduction is the transmission of sound through the external & middle ear to the internal ear. Bone conduction refers to the transmission of sound to the internal ear mediated by mechanical vibration of the cranial bones & soft tissues.

Speech audiometer To carry out tests with spoken voices. These tests are particularly important before prescribing hearing-aids. CONSTRUCTION: A double band tape recorder is preferred to interface the two channel audiometer units. Masking noise is supplied by the noise generator. The two channels supply the two head-phones or the two loud speakers of 25 W each.

Bekesy Audiometer System OPERATION: The instrument generates a pure-tone signal which is presented to him through an air-conduction earphone. The subject is asked to press a switch when the tone is heard and to release the switch when it is not heard. A pen connected to the attenuator traces a continuous record of the patient’s intensity adjustments on an audiogram chart, producing a graphic representation.

Bekesy Audiometer System

ELECTRICAL SECTION Oscillator circuit Modulator circuit. Automatic attenuator. Control circuits Master clock generator MECHANICAL SECTION Carriage device Writing system

Oscillator circuits This oscillator generates test signals with frequencies of 125, 250, 500, 1000, 1500, 2000, 3000, 4000, 6000, and 8000Hz. This sequence is first presented to the left ear automatically, each tone for 30s, and then to the right ear, the shift between the frequencies being noiseless. After both ears have been tested, a 1 kHz tone is presented to the right ear to provide a useful indication of the test reliability.

Modulators The models of modulators are available “Pulse” or “Cont”. In the ‘Pulse’ mode the test signal is modulated giving a signal which is easily recognized by the patient. In the ‘Cont’ mode no modulation is applied, giving a signal suitable for use, while calibrating the audiometer.

Attenuator The attenuation range is 100dB, thereby covering the range of hearing levels from – 10 to + 90 dB. When the test is initiated, the attenuator starts at its top position of –dB and then increases the level with a rate of 5 dB/s. The pen drive is controlled by means of the hand switch operated by the patient. Pressing the switch decreases the output from the potentiometer and thereby the level in the ear phones.

Mechanical section: Mechanical carriage with the writing system is driven by a stepping motor via a toothed belt.

The speed and direction of rotation of the motor are automatically controlled via the logic control system. When the test is initiated and the patient indicates that he hears the signal by pressing hand switch, the carriage moves along the X-axis (Frequency axis) of the audiogram in tune with the frequency of the test signal. When the complete test is finished the carriage and writing system returns to the start position.

Writing System Operation: operated by the pen drive, which is driven by a stepping motor. The pen drive moves the pen, and with it the wiper of the automatic attenuator, along the Y-axis (hearing level axis) with a constant speed corresponding to the change in attenuation of 5dB/s. The direction of movement of the pen is determined by the position of the hands witch operated by the patient. Limit switches are also included with the pen drive.

Audiogram Chart The audiogram is printed in standard 5 format (148*210mm).

MASKING IN AUDIOMETRY Need for masking: In case of monaural & asymmetrical binaural hearing losses, there is a serious difficulty in obtaining accurate measures of hearing for the poorer ear. This problem can be overcome by eliminating responses from the better ear by masking in order, to shift the threshold to high level, permitting greater intensities to be presented to the poorer ear without any danger of cross-over.

MASKING IN AUDIOMETRY Efficiency: Masking efficiency depends upon the nature of masking sound as well as intensity. A pure tone can be used to mask other pure tones.

Noise White noise: White noise is a noise containing all frequencies in the audible spectrum at approximately equal intensities. Saw tooth noise: Saw tooth noise is a noise in which the basic repetition rate is usually that of the mains voltage & contains only those frequencies that are multiplies of the fundamental.