A Single Photon Emission Computer Tomograph for breast cancer imaging S. Vecchio a, N. Belcari a, P. Bennati b, M. Camarda a, R. Campanini c, M. N. Cinti.

Slides:



Advertisements
Similar presentations
Topic 8. Gamma Camera (II)
Advertisements

What are PET basics?.
Image Reconstruction.
Sezione di Napoli Univ. “Federico II” Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del.
Advanced GAmma Tracking Array
Chapter 8 Planar Scintigaraphy
GRAPE(Gamma-Ray detector Array with Position and Energy sensitivity) Developed at CNS for in-beam  -ray spectroscopy with High Resolution M.
Semiconductor detectors for Compton imaging in nuclear medicine
Kirov A S, MSKCC Overview of Geant4 Use and Issues in Imaging: Emission Tomography (PET and SPECT) Assen S. Kirov Department of Medical Physics Memorial.
Lecture 2-Building a Detector George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
Seeram Chapter 11: Image Quality
8/18/2015G.A. Fornaro Characterization of diffractive optical elements for improving the performance of an endoscopic TOF- PET detector head Student: G.
Coincidence imaging today
Novel high resolution detectors for Positron Emission Tomography (PET)
ON THE EFFICIENCY OF A LIDAR-TYPE SINGLE-SIDED GAMMA-RAY TOMOGRAPHY APPROACH Tanja Dreischuh, Ljuan Gurdev, Dimitar Stoyanov, Christo Protochristov*, Orlin.
Applications of a Pixellated Detection System to Digital Mammography
November 8, 2005JLAB Awake Animal Project New Instrumentation for JLAB Awake Animal Project Smaller Pixel Size requires higher granularity in PMT (2.4.
Innovation is in our genes. 1 Siemens Medical Solutions Molecular Imaging What are SPECT basics?
Fundamental Limits of Positron Emission Tomography
M. Alnafea1*, K. Wells1, N.M. Spyrou1 & M. Guy2
Single Photon Emission Computed Tomography
HEALTH Novel MR-compatible PET detectors for simultaneous PET/MRI imaging FP7-HEALTH-2009-single-stage The focus should be to develop novel.
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.
BMI I FS05 – Class 4 “Nuclear Imaging: Math” Slide 1 Biomedical Imaging I Class 5 – Radionuclide Imaging (PET, SPECT), Part 3: Attenuation and Scatter.
Professor Brian F Hutton Institute of Nuclear Medicine University College London Emission Tomography Principles and Reconstruction.
O AK R IDGE N ATIONAL LABORATORY U.S. DEPARTMENT OF ENERGY Image Reconstruction of Restraint-Free Small Animals with Parallel and Multipinhole Collimation:
F.Murtas1IMAGEM DDG GEM technology for X-ray and Gamma imaging IMAGEM l Detector setup l First results on X-ray imaging l First results on Gamma ray imaging.
Li HAN and Neal H. Clinthorne University of Michigan, Ann Arbor, MI, USA Performance comparison and system modeling of a Compton medical imaging system.
Increase in Photon Collection from a YAP:Ce Matrix Coupled to Wave Lenght Shifting Fibres N. Belcari a, A. Del Guerra a, A. Vaiano a, C. Damiani b, G.
Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist.
Development of a Gamma Camera Based on an 88 Array of LaBr3(Ce) Scintillator Pixels Coupled to a 64-channel Multi-anode PMT Hidetoshi Kubo, K.Hattori,
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Filipe Castro Departamento de Física & i3n Universidade de Aveiro C S I(N A ) W AVELENGTH - SHIFTING FIBER GAMMA CAMERA USING S I PM S LIP Lisboa Apr.
Development of a Segmented Planar Germanium Imaging Detector
R. Pani Department of Experimental Medicine and Pathology University of Rome La Sapienza-Italy. Flat Panel PMT: advances in position sensitive photodetection.
Part No...., Module No....Lesson No
Impact of Axial Compression for the mMR Simultaneous PET-MR Scanner Martin A Belzunce, Jim O’Doherty and Andrew J Reader King's College London, Division.
Electron tracking Compton camera NASA/WMAP Science Team  -PIC We report on an improvement on data acquisition for a Time Projection Chamber (TPC) based.
Prostate probe with SPECT technique NSS – MIC November 5 - KnoxvilleF. Garibaldi- INFN – Roma1 – gr. Coll. ISS  the medical problem  the proposal.
Improved Hybrid PET Imaging Using Variable Focal Length Fan-Slat Collimators Thomas C. Rust and Dan J. Kadrmas, Ph.D. Medical Imaging Research Laboratory.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
Peter Dendooven LaBr 3 and LYSO monolithic crystals coupled to photosensor arrays for TOF-PET Physics for Health in Europe Workshop February 2-4, 2010,
Acquisition time6 min1 min 12 s Collimator height25 mm (Anger)12 mm (HiSens) Detector1 layer, 1 pixel / hole3 layers, 1 pixel / hole3 layers, 4 pixels.
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
a Very Interesting Presentation
PET Imaging Positron Emission Tomography
Prostate probe with SPECT technique NSS – MIC November 5 - KnoxvilleF. Garibaldi- INFN – Roma1 – gr. Coll. ISS  the medical problem  the proposal.
Simulations in Medical Physics Y. TOUFIQUE*, R.CHERKAOUI EL MOURSLI*, M.KACI**, G.AMOROS**, *Université Mohammed V –Agdal, Faculté des Sciences de Rabat,
Muhammad Musaddiq.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
6:th IWORID, Glasgow, Scotland, July 2004 Energy Dependence in Dental Imaging with Medipix 2 Börje Norlin & Christer Fröjdh Mid Sweden University.
Nuclear Medicine Physics
Prof. Stefaan Vandenberghe Enrico Clementel
Imaging molecolare ad alta risoluzione spaziale ed alta efficienza
Frontier Detectors for Frontier Physics
Image quality and Performance Characteristics
Development of a Compton Camera for online range monitoring
Low-power Wide-dynamic-range Readout System
Image quality and Performance Characteristics
HIGH RESOLUTION, HIGH MOLECULAR IMAGING CODED APERTURE THE DETECTOR
Chapter 8: Generic performance measures
Application of Nuclear Physics
Single Photon Emission Tomography
A.Takada, A.Takeda, T.Tanimori
HE instrument and in-orbit performance
First demonstration of portable Compton camera to visualize 223-Ra concentration for radionuclide therapy Kazuya Fujieda (Waseda University) J. Kataoka,
一种基于晶体间光分享原理的深度测量PET探测器
Assist. Prof. Dr. Ilker Ozsahin Oct
Presentation transcript:

A Single Photon Emission Computer Tomograph for breast cancer imaging S. Vecchio a, N. Belcari a, P. Bennati b, M. Camarda a, R. Campanini c, M. N. Cinti b, A. Del Guerra a, E. Iampieri c, N. Lanconelli c, R. Pani b, L. Spontoni a. a University of Pisa and INFN-Pisa, Pisa, Italy b University ”La Sapienza” and INFN-Roma 1, Rome, Italy c University of Bologna and INFN-Bologna, Bologna, Italy Functional Imaging and Instrumentation Group

S. Vecchio et al., XI th VCI, Feb MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion Single photon emission imaging (SPECT) of the breast for the detection of small size tumors. Aim: to overcome the present clinical sensitivity limit (1 cm Ø) for the detection of small size tumors. Imaging of the cellular steady state model, which would confirm the presence of higher radioactivity uptake from malignant lesions. Future perspective: analysis of 99m Tc radiopharmaceutical perfusion uptake due to neoangiogenesys generated from the presence of the lesion.

S. Vecchio et al., XI th VCI, Feb MammoSPECT Design MammoSPECT Design Reconstruction Head design Monte Carlo Readout Results Conclusion gamma cameras rotating gantry breast phantom Two gamma cameras, each about 15 cm x 5 cm wide. The SPECT is mounted on a ring that will be rotating around the breast. The patient is prone with a pendulus breast. Radius of rotation: 6.5 cm. Field of View (FOV): 13 cm Ø x 5 cm h

S. Vecchio et al., XI th VCI, Feb Hamamatsu 64-anode PMT H8500 (5 cm x 5 cm). large NaI(Tl) scintillator crystal (75 cm 2 ). “General Purpose” lead collimator. MammoSPECT Head Design MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion PMTs scintillator crystal collimator Each head is about 10 kg total weight

S. Vecchio et al., XI th VCI, Feb Window 2.0 mm MammoSPECT MA-PMT: Hamamatsu “Flat Panel” PMT H mm×52.0 mm×12.7 mm MammoSPECT Design Head design MA-PMT crystal Collimator Reconstruction Monte Carlo Readout Results Conclusion Typical anode luminosity is 55 A/lm Typical anode gain spread is about 3:1

S. Vecchio et al., XI th VCI, Feb MammoSPECT MA-PMT: Hamamatsu “Flat Panel” PMT H cm Total Active Area : 138 x 46 mm 2 (79%) Recovered detector area: 150 x 46 mm 2 (85%) MammoSPECT Design Head design MA-PMT crystal Collimator Reconstruction Monte Carlo Readout Results Conclusion > 3 mm

S. Vecchio et al., XI th VCI, Feb mm 63.5 mm Active Area : 151 x 50 mm 2 69 x 23 pixel MammoSPECT NaI(Tl) Crystal (Saint Gobain) Pixel : 2 mm, Pitch : 2.2 mm Crystal length: 6 mm Window Thickness: 3 mm MammoSPECT Design Head design MA-PMT crystal Collimator Reconstruction Monte Carlo Readout Results Conclusion

S. Vecchio et al., XI th VCI, Feb MammoSPECT Collimator Holes diameter 1.5 mm Septa thickness 0.2 mm Length 22 mm Efficiency [cpm/uCi] 508 Resolution 10 cm) 8.1 mm MammoSPECT Design Head design MA-PMT crystal Collimator Reconstruction Monte Carlo Readout Results Conclusion (back view) Chosen on the basis of dedicated simulations to compare the best alternatives Lead collimator Compromise between sensitivity & 140 keV

S. Vecchio et al., XI th VCI, Feb MammoSPECT Reconstruction algorithm Iterative reconstruction –Simulated annealing –Minimization of a cost function: MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion  N p : number of image pixels;  A ave,n : average activity in a region n x n around the pixel of interest;  S: smoothing parameter;  P k,m P rec k,m : measured projections and pseudo-projections for the k th detection element at angle m.

S. Vecchio et al., XI th VCI, Feb MammoSPECT Correction factors for reconstruction Collimator response Variable spatial resolution Attenuation due to breast tissue Scattering due to breast tissue Corrections are incorporated directly in the projector and used for calculating the pseudo-projections at each iteration of the reconstruction procedure MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion

S. Vecchio et al., XI th VCI, Feb h =  TUMOR  5-8 mm BREAST  BREAST  8-13cm DETECTOR 6.5 cm DETECTOR EGSnrc code Radionuclide: 99m Tc ( 140 keV ) Detector: NaI(Tl) crystal matrix 50x150x6 mm 3, 22x68 pixel. Phantom: breast tissue, cylindrical, Ø 8 cm, 10 cm, 13 cm. specific activity: 100 nCi/cc Tumour: spherical Ø 5 mm, 8 mm specific activity: 1  Ci/cc, ( 0.5  Ci/cc, 1.5  Ci/cc ) T/B = 10:1, (5:1, 15:1) MammoSPECT Monte Carlo simulation MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion

S. Vecchio et al., XI th VCI, Feb Rotation radius: 6.5 cm 36 projection on 360° Total scan duration: 20 minutes Selected event: E > (E photopeak – 2  E ) Iterative reconstruction –Reconstructed FOV: 15 cm x 15 cm x 5 cm –Reconstructed image: 69 x 69 pixel –Image voxel: 2.2 x 2.2 x 2.2 mm 3 MammoSPECT Reconstruction details MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion

S. Vecchio et al., XI th VCI, Feb MammoSPECT Figures of merit for reconstructed images BKG : background average  BKG : background standard deviation  ROI : average over a ROI centered on the tumour MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion Region of Interest (ROI) : 3 x 3 pixels ( 6.6 x 6.6 mm 2 ) Signal to Noise Ratio (SNR) and Image Contrast (IC) formulas for tomographic images:

N. Lanconelli et al., Conference Records of 2006 IEEE NSS/MIC/RTSD, San Diego, USA, M MammoSPECT Results for 5 mm Ø tumour T/B 10:1 8 cm10 cm13 cm T/B 15:1 MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion SNR IC 36% 39% 37% SNR IC 34% 38% 30% breast

N. Lanconelli et al., Conference Records of 2006 IEEE NSS/MIC/RTSD, San Diego, USA, M MammoSPECT Results for 8 mm Ø tumour T/B 15:1 MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion T/B 10:1 8 cm10 cm13 cm SNR IC 43% 44% 42% SNR IC 60% 59% 57% breast

S. Vecchio et al., XI th VCI, Feb MammoSPECT Results for 8 mm Ø tumour Centered1 cm2 cm %56%52% SNR IC 10 cm breast - T/B 10:1 MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion

S. Vecchio et al., XI th VCI, Feb II a head I a head PMT H8500 PMT H8500 PMT H8500 PMT H8500 PMT H8500 MammoSPECT Readout for the entire tomograph MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion SCD OP-AMP Chain+ Pre-amp PMT H8500 SCD OP-AMP Chain+ Pre-amp each acquisition board manage two side-by-side PMTs: if one PMT is above threshold both PMT are acquired Time correlation available only for the side-by-side PMTs

S. Vecchio et al., XI th VCI, Feb Hamamatsu 64-anode PMT H8500 (5 cm x 5 cm) : the readout of each head manages 3 x 64 = 192 channels. MammoSPECT Head Design (2) MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion PMTs scintillator crystal collimator electronics boards

S. Vecchio et al., XI th VCI, Feb PMT H8500 USB 2X + 2Y + timing SCD OP-AMP 64 anodes + last dynode 8X + 8Y Chain+ Pre-amp Electronics partially inherited from YAP-(S)PET Small Animal Scanner MammoSPECT PMT multiplexed read-out The Symmetric Charge Division circuit divides the charge into an X and Y decoder network. (8 x 8)  (8+8)  The SCD is followed by simple resistive chains to further reduce the number of signals to two for each direction. MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion Acquisition board

S. Vecchio et al., XI th VCI, Feb MammoSPECT PMT multiplexed read-out PMT H8500 (8X + 8Y)  2X + 2Y With 4 (ch) + 1 (OR-Dy) OP-AMP USB 2X + 2Y + timing SCD OP-AMP 64 anodes + last dynode 8X + 8Y Chain+ Pre-amp MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion  Constant fraction discrimination  Energy threshold: about 50 keV  Max acquisition rate per board: 100 kHz  up to 300 kHz can be managed by the whole acquisition system

N. Belcari et al., Nucl. Instr. and Meth. A 572 (2007), NaI:Tl 23  23 matrix 1.8 mm  1.8 mm side Separated by 200mm epoxy 45.8 mm  45.8 mm area n° AA0325, 1.5 mm window, 122 keV ( 57 Co) NaI: 122 keV MammoSPECT SCD-based multiplexed readout for a single PMT MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion Resolution (mm)0.8 Peak/Valley6.8

S. Vecchio et al., XI th VCI, Feb The event reconstruction never involves more than two PMT of one head: PMT # 3 PMT # 2 MammoSPECT Readout for a 3-PMTs assembly Time correlation between each side-by-side PMT pair. The PMT in the center can be separately correlated with both the other ones. PMT #1 PMT # 1&2 PMT # 2&3 MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion

S. Vecchio et al., XI th VCI, Feb MammoSPECT Recover of the dead area MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion anticoincidence events coincidence events total events NaI:Tl 23  23 matrix 1.8 mm side, 2 mm pitch 1.5 mm window, 122 keV ( 57 Co)

S. Vecchio et al., XI th VCI, Feb MammoSPECT Imaging performance MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion 57 Co flood field irradiation (no collimator) Energy 122 keV: 16 %<18 %

S. Vecchio et al., XI th VCI, Feb MammoSPECT Imaging performance MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion Horizontal profile 57 Co flood field irradiation Vertical profiles Horizontal profiles detector areadead area total area Resolution (mm) 1,11,21,3 Peak/Valley 4,63,22,5 Vertical profile Vertical profile in the dead area

S. Vecchio et al., XI th VCI, Feb MammoSPECT Summary and future work The dead area in the H8500 PMT triplet assembly is successfully recovered with reasonable values of P/V and energy resolution Measurements of detection efficiency are coming soon MC simulation estimate a sensitivity limit of 8 mm Ø tumour (T/B 5:1) in 13 cm breast The realization of the rotating ring has being completed The SPECT ring will be transferred at the Nuclear Medicine Division of University of Pisa for the validation on breast phantoms. This work is developed within the framework of the PRIN2004 Italian project: “New Techniques for Breast Cancer Imaging”. MammoSPECT Design Head design Reconstruction Monte Carlo Readout Results Conclusion