Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist.

Slides:



Advertisements
Similar presentations
Instruments for Radiation Detection and Measurement Lab # 4.
Advertisements

What are PET basics?.
Image Reconstruction.
6: Positron Emission Tomography
IMAGE QUALITY.
Chapter 8 Planar Scintigaraphy
Nuclear Medicine. Nuclear Medicine Physiological Imaging Radioactive isotopes which emit gamma rays or other ionizing forms (half life for most is hours.
Buxton & District Science Discussion Medical Scanners Marge Rose 16 th November 2012.
Fysisk institutt - Rikshospitalet 1. 2 Overview Gamma camera Positron emission technology (PET) Computer tomography (CT) Proton therapy Electrical impedance.
Nuclear Medicine Spring 2009 FINAL. 2 NM Team Nuclear medicine MD Nuclear medicine MD Physicist Physicist Pharmacist Pharmacist Technologist Technologist.
Medical Imaging Mohammad Dawood Department of Computer Science University of Münster Germany.
Medical Imaging Mohammad Dawood Department of Computer Science University of Münster Germany.
Gamma Camera Technology
Topics in Medical Physics Xiaoming Zheng, PhD. School of Dentistry and Health Sciences December 2009, Chengdu.
Nuclear imaging Instrumentation part Prepared by : Dr. Ali Saad, College of applied medical Science Dept of Biomed. Technology.
Introduction to Medical Imaging
tomos = slice, graphein = to write
PHYSICS IN NUCLEAR MEDICINE: QUANTITAITVE SPECT AND CLINICAL APPLICATIONS Kathy Willowson Department of Nuclear Medicine, Royal North Shore Hospital University.
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
Maurizio Conti, Siemens Molecular Imaging, Knoxville, Tennessee, USA
lecture 2, linear imaging systems Linear Imaging Systems Example: The Pinhole camera Outline  General goals, definitions  Linear Imaging Systems.
The content of these slides by John Galeotti, © Carnegie Mellon University (CMU), was made possible in part by NIH NLM contract# HHSN P,
Medical Image Analysis Introduction Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Basic principles Geometry and historical development
Positron Emission Tomography
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
Positron Emission Tomography Outline PET Examples Imaging Goal Reconstruction/Data Requirements Method of Data Acquisition in PET –Positron Decay/Annihilation.
Supplementary Material Emission Computed Tomography
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Coincidence to Image: PET Imaging Jennifer White Marketing Manager SNS Workshop October 13, 2003.
Lisa Rhoden (RT)(R) B.S. CNMT PET PET permits investigation and comparison of events occurring at the molecular and cellular level. This data is not.
Design and simulation of micro-SPECT: A small animal imaging system Freek Beekman and Brendan Vastenhouw Section tomographic reconstruction and instrumentation.
Innovation is in our genes. 1 Siemens Medical Solutions Molecular Imaging What are SPECT basics?
Fundamental Limits of Positron Emission Tomography
Medical Image Analysis Dr. Mohammad Dawood Department of Computer Science University of Münster Germany.
Computed Tomography Q & A
Medical Image Analysis Image Reconstruction Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Factors affecting CT image RAD
Single Photon Emission Computed Tomography
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.
BMI I FS05 – Class 4 “Nuclear Imaging: Math” Slide 1 Biomedical Imaging I Class 5 – Radionuclide Imaging (PET, SPECT), Part 3: Attenuation and Scatter.
1 INTRODUCTION TO THE PHYSICS OF DIAGNOSTIC IMAGING Outline of Course Brief History Common Terminology Imaging Modalities.
Professor Brian F Hutton Institute of Nuclear Medicine University College London Emission Tomography Principles and Reconstruction.
A Continuity Equation Based Optical Flow Method for Cardiac Motion Correction in 3D PET Data M Dawood, C Brune, X Jiang, F Büther, M Burger, O Schober,
Ultrasound, Positron Emission Tomography, and Single Photon Emission Tomography Allen T. Newton, Ph.D. PAVE 2014.
Nuclear Medicine Physics
Seoul National University Functional & Molecular Imaging System Lab Progress in Nuclear Imaging Mikiko Ito, PhD Dept. of Nuclear Medicine, Seoul National.
Nuclear Medicine Principles & Technology_I
Molecular Imaging & Positron Emission Tomography Nicholas Mulhern BME 281.
Impact of Axial Compression for the mMR Simultaneous PET-MR Scanner Martin A Belzunce, Jim O’Doherty and Andrew J Reader King's College London, Division.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Nuclear Medicine Physics Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Image Artifacts in Nuclear Medicine Part B.
Chapter-4 Single-Photon emission computed tomography (SPECT)
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Introduction to Medical Imaging SPECT, Introduction to Medical Imaging SPECT, PET and Lesion Detection Guy Gilboa Course
PET Imaging Positron Emission Tomography
Simulations in Medical Physics Y. TOUFIQUE*, R.CHERKAOUI EL MOURSLI*, M.KACI**, G.AMOROS**, *Université Mohammed V –Agdal, Faculté des Sciences de Rabat,
Chapter-5 Positron emission tomography (PET)
Nuclear Medicine Physics
P.E.T. Positron Emission Tomography
Learning Objectives By the end of this lesson you should…
Image quality and Performance Characteristics
Function and Structure in
Application of Nuclear Physics
Physics of Nuclear Medicine, SPECT and PET
Single Photon Emission Tomography
Sophie Lancelot, Luc Zimmer  Trends in Pharmacological Sciences 
Function and Structure in
Assist. Prof. Dr. Ilker Ozsahin Oct
Presentation transcript:

Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist

Image Acquisition Techniques  Static-(Bones, Lungs)  Dynamic-(Renography)  Gated-(Cardiac)  Tomography  SPECT  PET  List Mode-(Cardiac)

Problems with Planar Imaging  Planar imaging  2D representation of 3D Distribution of activity  No depth information  Structures at different depths are superimposed  Loss of contrast

Problems with Planar Imaging  Planar imaging  2D representation of 3D Distribution of activity  No depth information  Structures at different depths are superimposed  Loss of contrast

Problems with Planar Imaging  Planar imaging  2D representation of 3D Distribution of activity  No depth information  Structures at different depths are superimposed  Loss of contrast

Problems with Planar Imaging  Planar imaging  2D representation of 3D Distribution of activity  No depth information  Structures at different depths are superimposed  Loss of contrast

Problems with Planar Imaging  Planar imaging  2D representation of 3D Distribution of activity  No depth information  Structures at different depths are superimposed  Loss of contrast

Problems with Planar Imaging  Planar imaging  2D representation of 3D Distribution of activity  No depth information  Structures at different depths are superimposed  Loss of contrast Image contrast 2:1 Object Contrast 4:1

Single Photon Emission Computed Tomography  Collect multiple planar images at several angles around the patient  Typically views over 360°  Can be views over 180°

Single Photon Emission Computed Tomography  Image Reconstruction  2D images of selected planes within the 3D object  Better Contrast  Lower Spatial Resolution  Normal reconstruction techniques are Filtered Back Projection or Iterative Reconstruction

Back Projection  Back Project each planar image onto three dimensional image matrix

Back Projection  Back Project each planar image onto three dimensional image matrix

Back Projection  Back Project each planar image onto three dimensional image matrix

Back Projection  Back Project each planar image onto three dimensional image matrix

Back Projection  Back Project each planar image onto three dimensional image matrix

Back Projection  More views – better reconstruction  1/r blurring, even with infinite number of views

Filtered Back Projection  Filter planar views prior to back projection  Correction of 1/r blurring requires ‘Ramp’ Filter  Gives increasing weight to higher spatial frequencies  Amplifies Noise

Filtered Back Projection  In Practice  Use modifications of Ramp Filter  Compromise between Noise and Spatial Resolution

Modified Ramp Filter  Multiplication of the ramp filter by another function  Often a gaussian shape  Width of the gaussian affects the “roll off” of the ramp

Filtered Back Projection

Problems with Filtered Back Projection  Back projection is mathematically correct, but real life images require Filtered Back Projection  Back Projection can introduce noise and streaking artefacts  Not good with attenuation correction  Filtered Back Projection can reduce noise and artefacts, but may degrade resolution

Iterative Reconstruction  NOT a new technique  Pre-dates Filtered Back Projection  Computationally Intensive  Long Reconstruction Times  Requires fast computers for reconstruction  Takes around 1 min for a 16-frame gated 128 x 128 matrix cardiac scan

What is Iterative Reconstruction?  Iteration is process of successively better “guesses”  The image processing computer creates an image by refining the expected projections in comparison to those recorded  This form of IR is known as “Maximum Likelyhood Expectation Maximisation” (MLEM)

MLEM Algorithm

Benefits of IR  More accurate modelling of emission/detection  Can include attenuation correction and other information from MR, CT etc  Lower noise

Image Fusion  “Unclear Medicine” images can be registered to CT  Reduces attenuation artefacts  Allows localisation of “fuzzy blob” images  Can improve diagnostic accuracy

Attenuation Correction  X-Ray imaging essentially provides an attenuation “map”  Images formed by different attenuation patterns  NM imaging does not need attenuation  In fact do not want it!  Hybrid imaging (e.g.SPECT-CT) takes attenuation map of CT images and uses to correct for attenuation in 3D NM images

“Jordan”  6 x 500ml saline bags strapped to torso phantom (3 each side) to simulate breast attenuation  Positioned to cover anterior LV

Normal Perfusion: “Jordan” FBPFBPSCIR IRSC IRACIRACSC

Resolution Recovery  Resolution worsens with increasing distance from the collimator  If we can model how this happens, we can build this into our Iterative projections

Resolution Recovery  Better modelling means better images  Fewer counts needed to get acceptable images  Shorter acquisitions  Lower doses

NM Imaging: The PET Camera  PET camera invented in the 1970s  Positron Camera 1959

Early positron study (1953)

Why use positron emitters?  Many of the positron emitters occur in biological molecules (C, N, O, etc.)  Many have small molecular weights relative to the biological molecules they may be used to label (e.g., F) even if they aren’t found there naturally.  PET isotopes can be attached to biologically interesting molecules with no or minimal impact on the behaviour of those molecules in the body.

Positron Emission Tomography  PET isotopes emit positrons rather than gamma rays  Coincidence Imaging  Better Spatial Resolution (Typically 4mm)  Requires Dedicated Equipment  Limited Availability

Annihilation conservation of momentum: before: system at rest; momentum ~ 0 after: two photons created; must have same energy and travel in opposite direction. conservation of energy before: 2 electrons, each with a rest mass of 511keV after: 2 photons, each with 511keV.     decay via positron emission electron/positron annihilation photon annihilation photon

detector Coincidence Imaging line of response (LOR)

Detector Array

Coincidence Imaging - Detector Ring

Types of Coincidence Event

2D to 3D Imaging  Stack multiple rings behind each other  Allows for true 3D imaging  Shorter imaging time so better throughput and fewer motion artefacts

Time of Flight (TOF) PET  Because we are “timing” the arrivals of the photons, we can tell how far apart they are  All photons travel at the speed of light  Simultaneous equation to work out point of origin  Makes “line of response” more like a point

PET Camera Crystals  NaI has too poor stopping power for 511keV  BGO is main material used  Siemens patented LSO *this table was provided by Siemens…