M. D. A NDERSON Cancer Center Nuclear Imaging (M.Cam) Richard E. Wendt III, Ph.D. and William D. Erwin, M.S. Small Animal Cancer Imaging Research Facility.

Slides:



Advertisements
Similar presentations
Topic 8. Gamma Camera (II)
Advertisements

Instruments for Radiation Detection and Measurement Lab # 4.
Gamma Camera Quality Control
Study of plastic scintillators for fast neutron measurements
Chapter 8 Planar Scintigaraphy
Nuclear Medicine. Nuclear Medicine Physiological Imaging Radioactive isotopes which emit gamma rays or other ionizing forms (half life for most is hours.
Comparison of Scintimammography and Dedicated Emission Mammotomography
Nuclear Medicine Spring 2009 FINAL. 2 NM Team Nuclear medicine MD Nuclear medicine MD Physicist Physicist Pharmacist Pharmacist Technologist Technologist.
QC And NEMA In The Nuclear Arena
Gamma Camera Technology
Introduction to Medical Imaging
Single Photon Computerized Tomography SPECT neuroimaging Seyed Kazem Malakouti, MD Faculty of Iran University of Medical Sciences Seyed Kazem Malakouti,
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
The content of these slides by John Galeotti, © Carnegie Mellon University (CMU), was made possible in part by NIH NLM contract# HHSN P,
Introduction to Nuclear Medicine
M. D. A NDERSON Cancer Center Magnetic Resonance Imaging: 4.7 T, 40 cm Bruker Biospec James A Bankson, Ph.D. and John D Hazle, Ph.D. Small Animal Cancer.
Coincidence imaging today
Design and simulation of micro-SPECT: A small animal imaging system Freek Beekman and Brendan Vastenhouw Section tomographic reconstruction and instrumentation.
November 8, 2005JLAB Awake Animal Project New Instrumentation for JLAB Awake Animal Project Smaller Pixel Size requires higher granularity in PMT (2.4.
Innovation is in our genes. 1 Siemens Medical Solutions Molecular Imaging What are SPECT basics?
Characterization of LaBr3:Ce scintillator optimized for spatial resolution in low-energy gamma detection F. Cusanno, E. Cisbani, S. Colilli, R. Fratoni,
Fundamental Limits of Positron Emission Tomography
Biomedical applications of molecular imaging Tony Lahoutte UMons Nov-Dec 2011.
Imaging of Ra-223 with a small- pixel CdTe detector: potential for improved image quantification for radionuclide dosimetry James Scuffham 1,2 Silvia Pani.
Single Photon Emission Computed Tomography
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.
Professor Brian F Hutton Institute of Nuclear Medicine University College London Emission Tomography Principles and Reconstruction.
Li HAN and Neal H. Clinthorne University of Michigan, Ann Arbor, MI, USA Performance comparison and system modeling of a Compton medical imaging system.
Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist.
Nuclear Medicine Principles & Technology_I
A Single Photon Emission Computer Tomograph for breast cancer imaging S. Vecchio a, N. Belcari a, P. Bennati b, M. Camarda a, R. Campanini c, M. N. Cinti.
Prostate probe with SPECT technique NSS – MIC November 5 - KnoxvilleF. Garibaldi- INFN – Roma1 – gr. Coll. ISS  the medical problem  the proposal.
1 Tri-Modality Imaging of Small Animals Mark B. Williams, PhD University of Virginia DOE Awake Animal Meeting, May 20, 2004.
Improved Hybrid PET Imaging Using Variable Focal Length Fan-Slat Collimators Thomas C. Rust and Dan J. Kadrmas, Ph.D. Medical Imaging Research Laboratory.
Active Pixel Sensors in Medical and Biologi The application of Large Area Active Pixel Sensor (LAS) to high resolution Nuclear Medicine imaging Bob Ott.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
Nuclear Medicine Technique I 355 RAD L.Aya Ahmed Saeed.
1 MPGD2009 Advancements of labelled radio-pharmaceutics imaging with the PIM-MPGD J. Donnard a, N.Arlicot b,
Introduction to nuclear medicine NMT 231 L.Aya Ahmed Saeed.
Introduction to NM Omima Adam NMT 231. What is nuclear medicine NM Is the branch of medicine that uses radiation and nuclear properties of radionuclide's.
Acquisition time6 min1 min 12 s Collimator height25 mm (Anger)12 mm (HiSens) Detector1 layer, 1 pixel / hole3 layers, 1 pixel / hole3 layers, 4 pixels.
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Biomarkers from Dynamic Images – Approaches and Challenges
Chapter-2 The Planar Imaging Gamma Camera
Chapter-4 Single-Photon emission computed tomography (SPECT)
Introduction to Medical Imaging SPECT, Introduction to Medical Imaging SPECT, PET and Lesion Detection Guy Gilboa Course
Nuclear Medicine Introduction
Prostate probe with SPECT technique NSS – MIC November 5 - KnoxvilleF. Garibaldi- INFN – Roma1 – gr. Coll. ISS  the medical problem  the proposal.
Simulations in Medical Physics Y. TOUFIQUE*, R.CHERKAOUI EL MOURSLI*, M.KACI**, G.AMOROS**, *Université Mohammed V –Agdal, Faculté des Sciences de Rabat,
A Study on the Determination of Scan Speed in Whole Body Bone Scan Applying Oncoflash Gwang Gil Yang Department of Nuclear Medicine, Asan Medical Center,
Date of download: 7/6/2016 Copyright © 2016 SPIE. All rights reserved. Optical setup and schematic description of the acquisition. (a) Optical setup inside.
Chapter-5 Positron emission tomography (PET)
Case of the Day The likely cause of the poor image quality is: A. Incorrect collimators used B. Energy window incorrectly set over scatter below the 140.
Nuclear Medicine Physics
SCINTILLATOR STURUCTURE
Imaging molecolare ad alta risoluzione spaziale ed alta efficienza
Quantitative Nuclear Medicine Imaging
Image quality and Performance Characteristics
Image quality and Performance Characteristics
The EDR Gamma Scanner José M. Pérez* , O. Vela, E. de Burgos CIEMAT
CLRS 321: Nuclear Medicine Physics and Instrumentation I
Unit V: Introduction to Gamma Camera
Application of Nuclear Physics
Physics of Nuclear Medicine, SPECT and PET
QC And NEMA In The Nuclear Arena
Function and Structure in
First demonstration of portable Compton camera to visualize 223-Ra concentration for radionuclide therapy Kazuya Fujieda (Waseda University) J. Kataoka,
ME instrument and in-orbit performance
Assist. Prof. Dr. Ilker Ozsahin Oct
The evaluation of clinical usefulness on application of Half-time acquisition factor in Gated Cardiac Blood Pool scan Dong-Hun Lee.
Presentation transcript:

M. D. A NDERSON Cancer Center Nuclear Imaging (M.Cam) Richard E. Wendt III, Ph.D. and William D. Erwin, M.S. Small Animal Cancer Imaging Research Facility A Overview of Modality The Siemens Medical Solutions, USA m.CAM ™ gamma camera is designed for imaging pharmaceuticals labeled with gamma emitting radionuclides, in a range of laboratory animals, from mice to small primates or dogs. It is an ideal instrument for biomedical research laboratories to quantitatively measure the biodistribution of radiolabeled diagnostic and therapeutic compounds in vivo, in either 2-D or 3-D (SPECT) mode. Instrument Description The Siemens m.CAM consists of a single High Definition Digital Detector (HD 3 ) mounted onto a portable stand. The HD 3 head may be rotated over 90 degrees, for imaging at any angle between horizontal (face up) to vertical (face out). The HD 3 technology provides improved spatial and energy resolution, uniformity and count rate performance, in an energy independent fashion, compared to conventional detector head designs. This is especially important when imaging small objects such as mice. Performance and Capabilities The maximum imaging field-of-view (FOV) is 53.3 cm (21”) x 38.7 cm (15.25”) with an intrinsic resolution of 3.2 mm. The m.CAM is equipped with a Low Energy High Resolution (LEHR) collimator for imaging radionuclides such as Tc-99m (140 keV), and a Medium Energy Low Penetration (MELP) collimator for radionuclides such as In-111 (172, 247 keV). The system resolution is 4-20 mm (6.3 mm at 10 cm, LEHR with Tc- 99m) depending upon gamma ray energy, subject size and collimator. Achievable image pixel sizes range from mm (32x32 image matrix, zoom factor = 1) down to mm (1024x1024 image matrix, zoom factor = 3.2). Inherent system deadtime is 5.74  sec, resulting in a count rate data loss of 20% at 38,400 counts per second (cps). C B A Gamma Imaging with DigiRad: 111 In-DTPA-PEG-Annexin V Mice with MDA-MB-468 tumors without treatment (A), treated with PG-TXL 4 d earlier (B), or C225 4 d earlier (C). Mice were then injected with 111 In-DTPA- PEG-Annexin V. Images were acquired 48 h after radiotracer injection. Arrow: Tumor. Gamma Imaging with m.CAM: 111 In-DTPA-Annexin V vs. 111 In-DTPA-PEG-Annexin V Mice with MDA-MB-468 tumors without treatment (A), treated with PG-TXL 4 d earlier (B), or PG-TXL + C225 1 d earlier (C). Mice were then injected with 111 In-DTPA-Annexin V (A&B) or with 111 In-DTPA-PEG-Annexin V (C). B A C System sensitivity (Tc-99m, 15% energy window) is 172 cpm/  Ci for the LEHR collimator, and 237 cpm/  Ci for the MELP collimator. Up to six independent energy windows (50 keV – 600 keV) can be selected. The system is capable of acquiring conventional 2-D (planar) static images and dynamic images (for cardiac imaging or perfusion imaging). A Tomographic Accessory for 3-D SPECT imaging of mice and rats is currently under evaluation. Data output formats are Siemens ICON, Interfile 3.3, and DICOM 3.0. Images can be converted to several standard formats as well (TIFF, JPEG, etc.) Data can be provided to investigators via ftp or on Zip (100 Mb), Jaz (1Gbyte), CD-R, and CD-RW physical media. Potential Applications Single animal imaging: mouse through small primate or dog Simultaneous imaging of several animals: Up to 12 mice have been imaged at once, yielding higher throughput, consistent deadtime characteristics across a cohort, and synchronized dynamic data across a cohort Gamma Imaging with m.CAM: Example Image First 2-D image acquired with the m.CAM. Twelve nude mice containing an 111 In radiolabeled compound were imaged simultaneously at 48 hr post-injection. Image acquisition parameters were: 5 min static, 1024x1024 matrix, zoom = 1 (pixel size = 0.6 mm), and three simultaneous energy window settings (top to bottom: 172 keV/15%, 247 keV/15% and combined).