Display of Motion & Doppler Ultrasound

Slides:



Advertisements
Similar presentations
Chapter3 Pulse-Echo Ultrasound Instrumentation
Advertisements

Doppler Echocardiography
Doppler Echocardiography
Doppler echo & colour doppler -fazil bishara.  Blood is not a uniform liquid blood flow is pulsatile and is a very complex phenomenon!!!  Density provides.
Doppler echocardiography & tissue doppler Dolly mathew.
Doppler Ultrasound: Principles
There are three major types of Blood flow Doppler patterns: 1) Continuous Wave 2) Pulse Wave 3) Color Flow DOPPLER ULTRASOUND.
01: Introduction to Ultrasound George David, M.S. Associate Professor of Radiology.
Resident Physics Lectures
Resident Physics Lectures 02: Sound Properties and Parameters.
James T. Shipman Jerry D. Wilson Charles A. Higgins, Jr. Waves and Sound Chapter 6.
Sound Waves Physics Chapter 13 Section 1. I. Production of sound waves Produced by an object vibrating Produced by an object vibrating -ex. Tuning fork.
Hospital Physics Group
Laser Anemometry P M V Subbarao Professor Mechanical Engineering Department Creation of A Picture of Complex Turbulent Flows…..
ECE 501 Introduction to BME
Ultrasound Medical Imaging Imaging Science Fundamentals.
1 Sinusoidal Waves The waves produced in SHM are sinusoidal, i.e., they can be described by a sine or cosine function with appropriate amplitude, frequency,
- Doppler Ultrasound & Obstetrical Ultrasound
Ultrasound Imaging Atam Dhawan.
Flow Sensors.
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
Development of an Ultrasound Lab Laura Wade April 4 th Z.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Doppler Radar From Josh Wurman Radar Meteorology M. D. Eastin.
Mechanical Waves and Sound Ch. 17 Physical Science.
Ultrasonography.
Your Name Your Title Your Organization (Line #1) Your Organization (Line #2) Week 4 Update Joe Hoatam Josh Merritt Aaron Nielsen.
James T. Shipman Jerry D. Wilson Charles A. Higgins, Jr. Waves and Sound Chapter 6.
Noadswood Science,  To understand how amplitude, wavelength and frequency affect sound waves Monday, September 07, 2015.
Lecture 9 (1) Physics in Life Sciences Fluid flow in human body2.
Display of Motion & Doppler Ultrasound
George David Associate Professor Ultrasound Physics 04: Scanner ‘97.
Resident Categorical Course
Ch 16. Waves and Sound Transverse Wave Longitudinal Wave
Pegasus Lectures, Inc. COPYRIGHT 2006 Volume II Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th.
Go to section Interest Grabber Vibrations A wave is a vibration that carries energy from one place to another. But not all vibrations are waves. Hold a.
Lesson 21 Laminar and Turbulent Flow
Mechanical Waves and Sound
Review of Ultrasonic Imaging
Sound Vibration and Motion.
Sarah Gillies Ultrasound Sarah Gillies
1 Chapter 6 WAVES Dr. Babar Ali. 2 CHAPTER OUTLINE  Wave Concept Wave Concept  Wave Properties Wave Properties  Wave Speed Wave Speed  Wave Types.
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
Resident Physics Lectures
HYDRAULICS of steady and pulsatile flow in systemic circulation Motivation Experimental techniques for measurement pressure and flowrate How to calculate.
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Ultrasound Physics Reflections & Attenuation ‘97.
Oscillations About Equilibrium. 7.1 Periodic Motion.
Doppler Physics Waves from a static source Wave peaks evenly spaced around the source at 1 wavelength intervals.
Artifacts Ultrasound Physics George David, M.S.
Fluids in Motion How does blood flow dislodge plaque from an artery? Why can a strong wind cause the roof to blow off a house? Why do people snore? © 2014.
Blood Flow Detection > by the Doppler principle > by Electromagnetic principles © D. J. McMahon 2014 rev
Fundamental Ultrasound Principles Karen Potts Clinical Scientist Review date Jan 2010 Department of Medical Physics Kent & Canterbury.
 Ultrasound waves are longitudinal with high frequencies ( ≈ > 20,000 Hz, though medical Ultrasound is between 1 to 15 MHz.)  When an ultrasound reaches.
Ultrasound Physics Image Formation ‘97. Real-time Scanning Each pulse generates one line Except for multiple focal zones frame one frame consists of many.
BME 312 BIOMEDICAL INSTRUMENTATION II LECTURER: ALİ IŞIN
Ultrasound Physics Image Formation ‘97.
Section 15.7 The Doppler Effect and Shock Waves (cont.)
Doppler Spectral Analysis
Lecture on Ultrasonography
Sound in medicine Lect.10.
Reflections & Attenuation
Ultrasound.
Ultrasound Physics Image Formation ‘97.
Ultrasound.
Ultrasound.
Resident Physics Lectures
Sound COS: #18 Use Models to determine how light and sound waves differ in how they are absorbed, reflected, and transmitted through different media.
Mechanical Waves and Sound
Ultrasonic Imaging Using Contrast Agents
Presentation transcript:

Display of Motion & Doppler Ultrasound Resident Class

Blood Flow Characterization Hemodynamics Blood Flow Characterization Plug Laminar Disturbed Turbulent

Plug Flow Type of normal flow Constant fluid speed across tube Occurs near entrance of flow into tube

Laminar Flow also called parabolic flow fluid layers slide over one another occurs further from entrance to tube central portion of fluid moves at maximum speed flow near vessel wall hardly moves at all friction with wall

Flow Disturbed Flow Turbulent Flow Normal parallel stream lines disturbed primarily forward particles still flow Turbulent Flow random & chaotic individual particles flow in all directions net flow is forward Often occurs beyond obstruction such as plaque on vessel wall

Flow, Pressure & Resistance pressure difference between ends of tube drives fluid flow Resistance more resistance = lower flow rate resistance affected by fluid’s viscosity vessel length vessel diameter flow for a given pressure determined by resistance

Flow Variations pressure & flow in arteries fluctuate with pulse pressure & flow in veins much more constant pulse variations dampened by arterial system

Flow Rate Measurements Volume flow rate Volume of liquid passing a point per unit time Example 100 ml / second

Flow Rate Measurements Linear flow rate Distance liquid moves past a point per unit time Example 10 cm / second

Flow Rate Measurements Volume Flow Rate = Linear flow rate X Cross Sectional Area

Flow Rate Measurements Volume Flow Rate = Linear flow rate X Cross-sectional Area High Velocity Small Cross-section Low Velocity Large Cross-section Same Volume Flow Rate

Volume Flow Rates constant volume flow rate in all parts of closed system Sure! Any change in flow rate would mean you’re gaining or losing fluid.

Stenosis narrowing in a vessel fluid must speed up in stenosis to maintain constant flow volume no net gain or loss of flow turbulent flow common downstream of stenosis

Stenosis If narrowing is short in length If narrowing is long Little increase in overall resistance to flow Little effect on volume flow rate If narrowing is long Resistance to flow increased Volume flow rate decreased

Doppler Shift difference between received & transmitted frequency caused by relative motion between sound source & receiver Frequency shift indicative of reflector speed IN OUT

Doppler Examples change in pitch of as object approaches & leaves observer train Ambulance siren moving blood cells motion can be presented as sound or as an image

Doppler Angle angle between sound travel & flow 0 degrees 90 degrees q angle between sound travel & flow 0 degrees flow in direction of sound travel 90 degrees flow perpendicular to sound travel

Flow perpendicular to sound Flow Components Flow vector can be separated into two vectors Flow parallel to sound Flow perpendicular to sound

Flow perpendicular to sound Doppler Sensing Only flow parallel to sound sensed by scanner!!! Flow parallel to sound Flow perpendicular to sound

Doppler Sensing Sensed flow always < actual flow Actual flow

Doppler Sensing cos(q) = SF / AF Actual flow (AF) q Sensed flow (SF) q

Doppler Equation 2 X fo X v X cosq f D = fe - fo = ------------------------- c q where fD =Doppler Shift in MHz fe = echo of reflected frequency (MHz) fo = operating frequency (MHz) v = reflector speed (m/s) q = angle between flow & sound propagation c = speed of sound in soft tissue (m/s)

Relationships 2 X fo X v X cosq f D = fe - fo = ------------------------- c positive shift when reflector moving toward transducer echoed frequency > operating frequency negative shift when reflector moving away from transducer echoed frequency < operating frequency q q

Relationships Doppler angle affects measured Doppler shift cosq 2 X fo X v X cosq f D = fe - fo = ------------------------- c q Doppler angle affects measured Doppler shift q

Simplified (?) Equation 2 X fo X v X cosq f D = fe - fo = ------------------------- c 77 X fD (kHz) v (cm/s) = -------------------------- fo (MHz) X cosq Simplified: Solve for reflector velocity Insert speed of sound for soft tissue Stick in some units

Doppler Relationships 77 X fD (kHz) v (cm/s) = -------------------------- fo (MHz) X cos  higher reflector speed results in greater Doppler shift higher operating frequency results in greater Doppler shift larger Doppler angle results in lower Doppler shift

Continuous Wave Doppler Audio presentation only No image Useful as fetal dose monitor

Continuous Wave Doppler 2 transducers used one continuously transmits voltage frequency = transducer’s operating frequency typically 2-10 MHz one continuously receives Reception Area flow detected within overlap of transmit & receive sound beams

Continuous Wave Doppler: Receiver Function receives reflected sound waves Subtract signals detects frequency shift typical shift ~ 1/1000 th of source frequency usually in audible sound range Amplify subtracted signal Play directly on speaker - =

Pulse Wave vs. Continuous Wave Doppler No Image Image Sound on continuously Both imaging & Doppler sound pulses generated

Doppler Pulses short pulses required for imaging minimizes spatial pulse length optimizes axial resolution longer pulses required for Doppler analysis reduces bandwidth provide purer transmitted frequency important for accurate measurement of frequency differences needed to calculate speed

Color-Flow Display Features Imaged electronically scanned twice imaging scan processes echo intensity Doppler scan calculates Doppler shifts Reduced frame rates only 1 pulse required for imaging additional pulses required when multiple focuses used several pulses may be required along a scan line to determine Doppler shift

Duplex Doppler Gates operator indicates active Doppler region on display regions are called gates only sound in gate analyzed for frequency shift can be isolated based on delay time after pulse Gate

Spectral Display shows range of frequencies received amplitude of each frequency indicated by gray shade can be displayed real time fast Fourier Transform (FFT) technique frequency range Frequency Elapsed Time

Spectral Broadening display indicates range of frequencies corresponds to range of speeds of blood cells range indicative of type of flow laminar, disturbed, turbulent frequency range Frequency Time

Pulse Wave Doppler Allows range selectivity monitor Doppler shift (frequency difference) at only selected depth(s) ability to separate flow from >1 vessel or localize flow within vessel

Spectral vs. Color-Flow spectral Display shows frequency range directly Color Doppler’s color represents complete spectrum at each pixel frequency range Frequency Elapsed Time

"Color Power Angio" of the Circle of Willis Power Doppler AKA Energy Doppler Amplitude Doppler Doppler angiography Magnitude of color flow output displayed rather than Doppler frequency signal flow direction or different velocities not displayed "Color Power Angio" of the Circle of Willis