DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.

Similar presentations


Presentation on theme: "DOPPLER SONOGRAPHY By Dr/ Dina Metwaly."— Presentation transcript:

1 DOPPLER SONOGRAPHY By Dr/ Dina Metwaly

2 Principles of Doppler ultrasound
General principles Spectral-specific parameters Color-specific parameters Power Doppler imaging Normal flow in arteries Normal flow in veins

3 General principles of Doppler ultrasound

4 The Doppler effect Proposed by Christian Doppler
It defined :sound that reflects off a moving object undergoes a change in frequency. Commonly heard when a vehicle sounding a siren approaches, passes, & recedes from an observer Received frequency Higher during approach Identical at instant of passing by Lower during recession

5 The difference in the transmitted frequency (Ft) and the received frequency (Fr) is referred to as the Doppler frequency shift. The angle between the direction of sound and the direction of motion of the target is referred to as the Doppler angle.

6 Doppler effect applied to Echocardiography
Transducer emits ultrasound reflected from RBC. If RBC (flow of blood) moves toward transducer, frequency of the reflected sound’s increases If RBC (flow of blood) moves away from the transducer, frequency of the reflected sound’s decreases

7 Goals of Doppler Detection flow in a vessel
Detection direction of flow Detection type of flow: Arterial or venous Normal or abnormal Measurement the velocity of flow

8 Types of Doppler Continuous wave Doppler
Spectral Doppler (duplex): This combination of gray-scale sonography with pulsed Doppler sonography Spectral & color Doppler (triplex) Power Doppler

9 Pulse wave doppler Short intermittent busts of ultrasound are transmitted. By varying the delay time between the transmission and reception of the sound wave, it is possible to determine the location (Le., depth) from which the Doppler signal arises the sample volume or Doppler gate.

10 The waveform that displays the Doppler information has the Doppler frequency shift on the vertical axis and time on the horizontal axis. The Doppler shift from objects moving toward the transducer is positive and is typically displayed above the baseline, and the shift from objects moving away from the transducer is negative and typically displayed below the line.

11

12

13

14 As indicated in the Doppler equation, the frequency shift is proportional to the velocity.
So when analyzing blood flow, the size of the waveform varies with the flow velocity. The frequency shift is also proportional to the cosӨ. At a Doppler angle of 90 degrees (blood flow perpendicular to direction of sound), the cos Ө = 0 and no Doppler frequency shift will be detected. On the other hand, at a Doppler angle of 0 degrees (blood flow and sound direction are parallel), the cos Ө = 1 . Because this is the maximum possible value for the cos Ө, the Doppler frequency shift is maximum at an angle of 0 degrees. Therefore, orienting the transmitted Doppler pulse with respect to the blood vessel so that the Doppler angle is as close to 0 degrees as possible will obtain the largest Doppler signal

15 Optimizing Doppler angle :Larger the angle, greater the error
Ideally should be zero Usually not possible Do not use angle > Great error in velocity Angle Complete loss of flow Transducer position Obtain smaller angle Different US systems May be different results

16

17 Color Doppler It provides a real-time image, displaying tissue morphology in gray scale and blood flow in color. Color Doppler sonography analyzes the phase information, frequency, and amplitude of the returning echoes. Signals from moving red blood cells are assigned a color (red versus blue) based on the direction of the phase shift (i. e . , the direction o f blood flow toward o r away from the transducer). The color shade for each pixel is based on the mean frequency shift arising from that pixel. High-frequency shifts are assigned a lighter color, and lower-frequency shifts are assigned a darker color Stationary objects produce no phase shift and are assigned a gray-scale value, as in conventional gray-scale imaging.

18 Color Doppler Displayed as color information-
Direction- red vs blue (toward or away from transducer) Velocity- brightness (bright blue higher velocity) Variance (turbulence)- coded green to give a mosiac apperance. Overlays this information on 2D images Time consuming (temporal resolution is especially poor with a large sector window)

19 Example of Color Doppler

20 Power Doppler Power Doppler imaging estimates the power or strength of the Doppler signal rather than the mean frequency shift. The Doppler detection sequence used in power Doppler imaging is identical to that employed in frequency-based color Doppler imaging. However, once the Doppler shift has been detected, the frequency components are ignored in lieu of the total energy of the Doppler signal . The color relate to the moving blood volume rather than the direction or the velocity of flow

21 Disadvantages of power Doppler imaging
Do not provide velocity of flow Do not provide direction of flow New machines provide direction of flow in power mode Very motion sensitive (poor temporal resolution)Less suitable for rapid scan along vessels

22

23 Advantages of Color Doppler and Power Doppler
Determines flow direction Slightly more sensitive to detect low-velocity flow Determines relative flow Angle independen Useful in imaging tortuous vesselst Less affected by tissue motion No aliasing Less affected by probe motion Distinguish pre-occlusive from occlusive lesions & Increases accuracy of grading stenosis

24

25 DOPPLER OPTIMIZATION Transducer Frequency
Because the Doppler frequency shift is proportional to the transmitted frequency, higher-frequency transducers cause a higher Doppler frequency shift that is easier to detect. Therefore, higher-frequency probes result in a stronger reflection from red blood cells. These effects improve the sensitivity of higher-frequency probes.

26 Gain Doppler gain is a receiver end amplification of the Doppler signal that can be applied to either the Doppler waveform or to the color Doppler image. In most situations, the gain should be increased to a maximum value just before the point where random noise begins to obscure the pulsed Doppler waveform, or with color Doppler to the point where color starts to appear in nonvascular spaces of the color image. The Doppler gain affects only the Doppler portion of the image and does not affect the gray-scale background

27

28 Power: Stronger or more powerful sound pulses will produce stronger reflections that are more easily detected. Power output affects both the gray-scale and Doppler images. In general, increasing the power output improves Doppler sensitivity . This can be very important in deep abdominal applications where tissue attenuation significantly weakens the Doppler signal. However, increasing the power output also lead to a number of artifacts. Therefore, power levels should be kept as low as is reasonably achievable to obtain the desired information.