Pulse sequences.

Slides:



Advertisements
Similar presentations
Durgesh Kumar Dwivedi Department of NMR & MRI AIIMS, New Delhi, India
Advertisements

Pulse Timing parameters & Weighting
Contrast T1 weighted – (MPRAGE-anatomical) T2 weighted – (fmri)
Imaging Sequences part I
Chapter 7: Gradient Echo Imaging Methods
PULSE SEQUENCES Emphasizing the differences among spin density, T1, and T2 relaxation time constants of the tissues is the key to the exquisite contrast.
Proton Spin In absence of a magnetic field, protons spin at random
M R I Pulse Sequences Jerry Allison Ph.D..
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
Institute for Biomedical Engineering EXCITE Afternoon Hands-On MRI Sessions: fMRI & DTI.
Magnetic Resonance Imaging - MRI
RF Pulse – generates a B 1 field that realigns the precessing spins in the low energy state In the case of a 90 o pulse the alignment is perpendicular.
MR Sequences and Techniques
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Fund BioImag : MRI contrast mechanisms 1.What is the mechanism of T 2 * weighted MRI ? BOLD fMRI 2.How are spin echoes generated ? 3.What are.
Parameters and Trade-offs
Magnetic Resonance Imaging
Fast spin echo 단국대학교 영상의학과 유 동 수.
The Basics of MRI The Basics of MRI. Current MRI technology displays images as multiple sets of gray tone images. Visualization and interpretation of.
Relaxation Exponential time constants T1 T2 T2*
Basics of Magnetic Resonance Imaging
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
Magnetic Resonance Imaging 4
Physics of Magnetic Resonance Chapter 12
Gradient echo pulse sequences
Principles of Magnetic Resonance
Imaging Sequences part II
MRI Physics: Just the Basics
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Chapter 4 Mark D. Herbst, M.D., Ph.D.. Magnetization MRI depends on magnetization of tissues Temporary magnetization occurs when the hydrogen nuclei are.
Pulse Sequences Types of Pulse Sequences: Functional Techniques
Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo.
Basics of MRI.
RT 4912 Review (A) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
Quiz In a 2D spin warp or FT MR scan, aliasing should only occur
Rad T 265 MRI Lecture. No Magnetic Field = No Net Magnetization Protons align with a magnetic field…
G Practical MRI 1 Basic pulse sequences.
Contrast Mechanism and Pulse Sequences
Magnetic Resonance Imaging – Basic Principles –
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
BIOE 220/RAD 220 REVIEW SESSION 6 March 5, What We’ll Cover Today General questions? Spinal cord anatomy review Fat in images T2* vs T2 decay Review.
Magnetic Resonance Imaging
Magnetic Resonance Imaging
MRI Physics: Pulse Sequences
Principles of MRI magnetic resonance imaging
V.G.Wimalasena Principal School of Radiography
RT 4912 Review (C) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
MRI: Contrast Mechanisms and Pulse Sequences
Spinning Nucleus Produces Magnetic Moment
Magnetic Resonance Learning Objectives
MRI Magnetic Resonance Imaging. Definition A non-ionizing technique with full three dimensional capabilities, excellent soft-tissue contrast, and high.
Comparison of T1 FLAIR and T1 FSE Images Andrew Allmendinger, DO Sylvie Destian, MD.
MAGNETIC RESONANCE IMAGING by PRADEEP V.EPAKAYAL. Mem.no L.
بسم الله الرحمن الرحيم Dr. Maged Ali Hegazy Assistant Professor Alazhar University.
MRI B 陳名揚.
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
Key sequences in MSK imaging
MRI Physics in a Nutshell Christian Schwarzbauer
بسم الله الرحمن الرحيم.
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Magnetic Resonance Imaging: Physical Principles
Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2005 MRI Lecture 5 Thomas Liu, BE280A, UCSD, Fall 2005.
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
MRI ARTEFACTS PG Student : Dr Aditi Shah PG Guide: Dr Rajendra Chavan
Examples of typical clinical MRI. (A).
(4)ELECTRONIC SUPPORT SYSTEM
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
T2 Relaxation Time T2 relaxation time is defined as the time needed to dephase up to 37% of the original value. T2 relaxation refers to the progressive.
Presentation transcript:

Pulse sequences

Categorization Spin echo Inversion recovery Gradient echo Conventional spin echo Fast spin echo Inversion recovery Gradient echo Coherent Incoherent Steady state free precession Ultra-fast imaging

Conventional spin echo Illustration The situation before the patient is placed inside the magnet

The patient is now inside the bore of the magnet The patient is now inside the bore of the magnet. The magnetization of the patient’s protons M0 is aligned along the Z axis

The 900 pulse is now applied The 900 pulse is now applied. M0 is now completely removed from the Z axis and lies along the Y axis

Relaxation is now taking place Relaxation is now taking place. Spin-lattice (T1)relaxation caused the magnetization to re-grow along the Z axis. Spin-spin relaxation causes the magnetization vectors to dephase (move apart) while still in the X-Y plane

1800 pulse is now given. All the vectors now point in the opposite direction. The magnetization vectors rephase (come together) in the X-Y plane. As they come together the echo is being formed

The magnetization is now completely rephased in the X-Y plane and points along the Y axis. This causes the full height of the echo. The actual MRI signal is taken here.

Relaxation continues to take place Relaxation continues to take place. The magnetic vectors again dephase in the X-Y plane while the regrowth along the Z axis continues

Complete relaxation has taken place Complete relaxation has taken place. There is no net vector in the X-Y plane, and the magnetization is full grown along the Z axis. This is identical to the situation before the 900 pulse was applied

The coordinate system in relation to the magnet

Short TE and short TR gives T1 weighted image Use two RF rephasing pulses generating two spin echoes to produce T2 and proton density weighting First echo has short TE and long TR – produce proton density weighting Second echo has a long TE and a long TR – produce T2 weighting

uses Gold standard for most imaging May be used for every examintion T1 images useful for demonstrating anatomy – because high SNR With contrast enhancement T1 images show pathology T2 images also demonstrate pathology Diseased tissues are generally more oedematous and/or vascular. They have increased water content and, have a high signal on T2 images

Parameters T1 weighting Proton density/T2 Short TE 10-20 ms Short TR 300 – 600 ms Typical scan time 4-6 min Proton density/T2 Short TE 20 ms/long TE 80 ms+ Long TR 2000 ms+ Typical scan time 7-15 min

Advantages Disadvantages Good image quality Very versatile True T2 weighting sensitive to pathology Disadvantages Scan times relatively long

Fast spin echo In contrast to conventional spin echo, fast spin echo applies a train of 1800 pulses per TR and different phase encoding steps are used. Each 1800 pulse produce an echo (proton density & T2) This drastically reduce the scan time The number of 1800 pulses in the train called the turbo factor or train length

E.g. if in conventional SE 256 phase matrix and 1 NEX is used, the scan time is 256TR In FSE if the turbo factor is 16, the scan time is 256TR/16 =16TR

Uses Useful in most clinical applications Central nervous system, pelvis, musculoskeletal regions Note Fat remains bright on T2 Unless fat saturation techniques are used Muscles appear darker in FSE images Artefacts from metal implants is significantly reduced

Parameters T1 weighting T2 weighting Short effective TE less than 20ms Short TR 300 – 600 ms Turbo factor 2-6 Typical scan time 30s to 1 min T2 weighting Long effective TE 100 ms Long TR 4000 ms+ Turbo factor 8-20 Typical scan time 2 min

Advantages & Disadvantages Reduced scan time High resolution matrices and multiple NEX can be used Image quality improved Increase T2 information Some flow and motion effects increased Incompatible with some imaging options Fat bright on T2 Reduces magnetic susceptibility effect, so should not be used when haemorrhage is suspected

Inversion Recovery Starts with a 1800 inversion pulse. This inverts NMV through 1800 into full saturation. When inverting pulse removed NMV begins to relax back to B0 A 900 excitation pulse is then applied at a time TI (Time from Inversion) from the 1800 inversion pulse

TI

TI determines the weighting & contrast Short TI gives T1 contrast Long TI gives proto density contrast After the 900 excitation pulse 1800 rephasing pulse is applied at a time TE This produces the spin echo TR is the time between each 1800 inverting pulse

Uses conventionally used to produce heavily T1 weighted images to demonstrate anatomy & in contrast enhanced imaging Now more widely used in conjunction with fast spin echo to produce T2 weighted images

Parameters

STIR (short TI inversion recovery) Uses TI that corresponds to the time it takes fat to recover from full inversion to the transverse plane so that there is no longitudinal magnetization corresponding to fat. As a result the signal from fat is nulled. Used to achieve suppression of fat in T1 weighted images. TI 150 – 175 ms

FLAIR (Fluid attenuated inversion Recovery) The signal from CSF is nulled by selecting a TI corresponding to the time of recovery of CSF from 180 to the transverse plane and there is no longitudinal magnetization present. Used to suppress signal from CSF in T2 weighted images TI - 1700 -2200 ms