What to compare against the validation templates (see templates in course webpage: Resources/Validation%20Templates%20from%20Red.

Slides:



Advertisements
Similar presentations
EPA Methods 3A, 6C, 7E, 10 & 20 Corrections to May 15, 2006 Final Rule That Updated the Methods That Updated the Methods Foston Curtis US EPA.
Advertisements

April 7: QA Systems, EPA definitions, PQAOs and common sense – Mike Papp April 14: Routine Quality Control and Data Management (1-pt QC, flow rate, and.
Institute for Tribal Environmental Professionals, Tribal Air Monitoring Support Center Guest Speaker: Mike Papp, OAQPS.
Module 61 Module 6: Uncertainty Don’t just calculate—first think about sources of error, and don’t double-count errors.
QA Strategy Workgroup Ambient Air Quality System Training Ambient Air Monitoring Quality Systems Requirements Mike Papp- OAQPS.
Update on Mercury Calibration Gas Standards and Traceability Scott Hedges US EPA, Clean Air Markets Division 2009 EPRI CEM User Group Conference St. Louis,
PM-2.5 FEM Requirements for Met One BAM Glenn Gehring, Technology Specialist Tribal Air Monitoring Support (TAMS) Center
Why do we lose analyzer data? Monitor malfunction DAS malfunction Power outages Environmental problems Wildlife damage Vandalism Operator error.
Data QA/QC Techniques. Copyright VIM Technologies, Inc. All Rights Reserved. VIM’s 10-Step Program To Compliance Success 2.
SRPMIC Ozone Monitoring Sites Senior Center SiteHigh School Site Red Mountain Site Lehi Site.
Statistical Process Control
Chapter 11 Quality Control.
Short Course on Introduction to Meteorological Instrumentation and Observations Techniques QA and QC Procedures Short Course on Introduction to Meteorological.
Technical requirements  Sampling  Analysis  Reporting of the results.
QA/QC FOR ENVIRONMENTAL MEASUREMENT
1 AQS Ambient Monitoring Topics AQS Conference August 20, 2008 David Lutz.
1 Guest Speaker: Bill Frietsche US EPA.  April 7: QA Systems, EPA definitions, PQAOs and common sense – Mike Papp  April 14: Routine Quality Control.
Presented by: Mike Hamdan South Coast Air Quality Management District Diamond Bar, CA Presented at: The Tribal Air Monitoring Training, Pechanga Reservation,
Guest Speaker: Jeremy Howe Little River Band Of Ottawa Indians The Measuring Sticks of Air Monitoring.
Field Analysis Quality Control
Approach of a UK auditor Paul Mudway Mudway Health, Safety & Environment.
Debra Waller NJDEP-Office of Quality Assurance
EPA Precursor Gas Training Workshop Precursor Gas Quality Assurance Implementation Dennis K. Mikel EPA Office of Air Quality Planning and Standards.
Introduction to BGI Instruments
Institute for Tribal Environmental Professionals Tribal Air Monitoring Support Center Melinda Ronca-Battista Brenda Sakizzie Jarrell Southern Ute Indian.
Quality Assurance/ Quality Control
Continuous Particulate Training Workshop
1 Dec. 8, 1997 LEADS Quality Assurance Summary Robert Brewer (512) Monitoring Operations Division Network QA Manager.
AMP255 – Precision and Bias Data Report 2008 AQS Conference The AMP255 Precision and Bias Data Report August 22, 2008 Presented by Jonathan Miller EPA.
Introduction to Standard Reports. Standard Reports 2 How to get information out of AQS Standard Reports Site / Monitor Metadata Detail Data Reports “
Melinda Ronca-Battista, ITEP, TAMS Center Tribal Air Monitoring and QA Requirements for NAAQS- Comparable Data.
1 Guest Speaker: Brandy Toft Leech Lake Ojibwe.  Overview of FRM/FEM/ARM status, requirements, and reporting  QC (routine checks, audits, and method-
 April 7: QA Systems, EPA definitions, PQAOs and common sense – Mike Papp  April 14: Routine Quality Control and Data Management (1-pt QC, flow rate,
1 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop European Standards Requirements of.
Quality Control/ Quality Assurance Annabelle Allison ITEP/TAMS Center.
Quality Control – Part II Tim Hanley EPA Office of Air Quality Planning and Standards.
Emission Inventory Quality Assurance/Quality Control (QA/QC) Melinda Ronca-Battista ITEP/TAMS Center.
EPA Region 9 Meredith Kurpius August 19, Status of Tribal Air Monitoring Value of tribal monitoring Used to protect public health on tribal land.
Laboratory QA/QC An Overview.
Appendix A: Calculations for Data Quality Assessment QC check statistics Precision calcs Bias calcs PM stats Reporting: quarterly and annual.
CARB Continuous PM 2.5 Monitoring Activity (BAM’s) Reggie Smith California Air Resources Board.
WLTP Test Procedures for Chassis Dynamometer Testing Chris Laroo - U.S. EPA April 14, 2010 WLTP-DTP
1 Dec. 11, 1997 LEADS Quality Control Systems Robert Brewer (512) Monitoring Operations Division Network QA Manager.
1 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheBukarest National Reference Laboratory for Air.
Laboratory Certification Update Part 2 Common Findings KWWOA Louisville April 15, 2015 Presented by Frank Hall, Laboratory Certification Coordinator.
Data Management Concepts James Payne Environmental Protection Department Morongo Band of Mission Indians.
An Overview of EPA’s Quality Assurance Guidance for Ambient Air Quality Monitoring Data Data Analysis and Interpretation February 12 – 14, 2008, Tempe,
Dirk Felton RTP, NC February 12-13, 2008 Air Quality Data Summit: Session: Inventory of Data Systems Data provider perspectives.
 April 7: QA Systems, EPA definitions, PQAOs and common sense – Mike Papp  April 14: Routine Quality Control and Data Management (1-pt QC, flow rate,
LEADS/EMS DATA VALIDATION IPS MeteoStar December 11, 2006 WHAT IS VALIDATION? From The Dictionary: 1a. To Make Legally Valid 1b. To Grant Official.
Measurement Quality Objectives = Data Validation Requirements Excerpted from EPA QA Handbook Volume II (aka “Redbook”)
WHAT IS THE CHEROKEE NATION? Cherokee Nation Air Quality Data Management Concepts for Quality Data Collection Ryan Callison.
Module 11 Module I: Terminology— Data Quality Indicators (DQIs) Melinda Ronca-Battista ITEP Catherine Brown U.S. EPA.
Quality Management & Operations Support Branch Quality Assurance Section Presented By: Webster Tasat.
Diagnostic clinical chemistry
BACK TO BASICS Quality Assurance and Validation Gaseous NAAQS Pollutants (O3, NO2, SO2, CO) By Kendall Perkins May 17, 2006.
Annual Air Monitoring Data Certification and Concurrence Process 1.
Robert S. Wright US EPA, Research Triangle Park, NC EPA Protocol Gases Fall 2012 Update PurityPlus Specialty Gas Producers.
Statewide Compliance Monitoring and Local Monitoring Projects
Quality is a Lousy Idea-
Calibration Calibration of the analyzer establishes the relationship between actual pollutant concentration input and the analyzer response. Should be.
Quality is a Lousy Idea-
Back To Basics: PM2.5 NESCAUM MAC Meeting Newport, RI May 16-17, 2006
Proposed Ozone Monitoring Revisions Ozone Season and Methods
Calibration in the Ambient Air Quality Monitoring Network
Chapter 11 Quality Control.
Module 6: Uncertainty Don’t just calculate—first think about sources of error, and don’t double-count errors Presentation of Uncertainty: 1. Components.
Quality Control Lecture 3
Precision, Bias, and Total Error (Accuracy)
Presentation transcript:

what to compare against the validation templates (see templates in course webpage: Resources/Validation%20Templates%20from%20Red book%20QA%20Handbook%20vol%20II.pdf Resources/Validation%20Templates%20from%20Red book%20QA%20Handbook%20vol%20II.pdf ) 1

 April 7: QA Systems, EPA definitions, PQAOs and common sense – Mike Papp  April 14: Routine Quality Control and Data Management (1-pt QC, flow rate, and instrument stability checks) – Travis Maki  April 21: Audits Overview (NPAP, PEP, Annual PE, Flow Rate Audits) – Jeremy Howe  April 28: Calculating Bias and Precision and AQS reports – Angie Shatas  May 5: 40 CFR 58 App. A- Gaseous Pollutants – Glenn Gehring  May 12: 40 CFR 58 App. A- Ozone – Brenda Jarrell  May 19: 40 CFR 58 App. A- PM filter and continuous methods –Brandy Toft 2

 Measuring  Comparing with MQO  Graphing it/ watching for its drift  Fixing it when needed Everything must be documented and, when significant, reported 3

 Instruments operating stably  Measure what you can, verify it is: ◦ Within limits ◦ Not slowly drifting up or down  Verify you are meeting your QAPP specs  QAPP specs assess bias and imprecision  Ideal is to estimate bias and imprecision, so that you can know and limit your overall error  Write it down! 4

 Preparing for the field  Checking your equipment ◦ Analyzer working, QC within limits ◦ Transfer standards also need to be working  Supporting stuff: ◦ Site, power, weather, frequency of your checks ok? Look to the future  Data recording system working ◦ Routine data ◦ QC data  Downloading the data 5

…the difference between your answer and the “truth” Two components of total error (accuracy):  Bias (jump)  Precision (wiggle) 6

 Minimized by calibrating your equipment against a standard ◦ Make sure the standard has TRACEABILITY to a NIST standard ◦ Compare result of the standard(s) to what your instruments read (pollutant, flow rate, temp, pressure) Measures any bias in your equipment at that time 7

 Some imprecision is unavoidable  Sometimes up, sometimes down– “random”  Estimated by measuring the same thing several times  Or by side-by-side collocated  Minimized by carefully following procedures 8

 The term “transfer” means that you are “transferring” the reliability of a measurement from a primary calibration standard to your field instrument  The primary calibration standard can be the place where you send your transfer standard to be recalibrated annually  Pressure, temperature, and flow rate of your instruments must be checked with your transfer standards 9

 For example, from CFR 50 Appendix L: “An authoritative flow rate standard shall be used for calibrating or verifying the sampler's flow rate measurement device with an accuracy of ±2 percent. The flow rate standard shall be a separate, stand-alone device... This flow rate standard must have its own certification and be traceable to a National Institute of Standards and Technology (NIST) primary standard for volume or flow rate. 10

11

 Single-point QC check is when you conduct a check using one flow rate or concentration or value  1-point QC (used to be called span) is an example of a one-point check  Verifications DO NOT involve any changes to your equipment or settings  See list of QC checks in validation templates for each pollutant (see pg 184 of QA Handbook Vol II in course website esources/ ) esources/ 12

 Perform a check of your equipment: ◦ flow rate for PM ◦ concentration for gas ◦ internal mass for TEOM  This may be single-point (one flow rate) or multi-point (several flow rates)  If this is within specifications, record this and continue 13

 If results are NOT within specs, then the instrument must be adjusted  This adjustment means that the response of the instrument is changed, which is defined as a CALIBRATION  Calibrations must be multi-point  After calibrating, do QC check again 14

 Corrective Action if the flow rate difference between the transfer standard and the sampler is > 4%  Check sampler for internal and external leaks  Ensure that temp. and pressure are within normal ranges  Run check a 2 nd time  If still > 4%, perform a multipoint calibration followed by a single-point verification 15

 1-point QC check must show within 7% for O3, 10% for others (if starting out, every week)  Zero must not drift more than 2% of full scale  Span concentration must not drift more than 7% since last check 16

Gaseous Verifications  Gaseous instrument verifications can be conducted: ◦ Through-the-probe (is what you will be audited on—do this if possible) ◦ Through-the-manifold ◦ Through the back of the analyzer  To evaluate the sampling system it is recommended that at least some verifications be conducted through- the-probe 17

MethodPollutantsFrequencyMQO One-Point QCSO 2, NO 2, O 3, COEvery 2 Weeks O 3 : Precision = 7% Bias = 7% SO 2, NO 2, CO : Precision = 10% Bias = 10% Annual Performance Evaluation (Audit) SO 2, NO 2, O 3, COOnce per Year<= 15% for each audit concentration Flow Rate Verification (QC check) PM 10 (lo-Vol), PM 2.5 Once every 4 weeks <= 4% of Standard & 5% of Design Value Flow Rate Verification (QC check) PM 10 (high-Vol), TSP Once per quarter <= 7% of Standard & 10% of Design Value Semi-Annual Flow Rate Audit PM 10, PM 2.5 Every 6 Months <= 4% of Standard & 5% of Design Value Semi-Annual Flow Rate Audit PM 10 (high-Vol), TSP Every 6 Months <= 10% of Standard & 10% of Design Value Collocated SamplingPM 10, TSP, PM % of Network Every 12 Days PM 2.5, TSP, PM 10 : Precision = 10% PM 2.5 PEP Program NPAP PM 2.5 SO 2, NO 2, O 3, CO Quarter Year (see QA Requirements. xls) Bias = 10% 18

MethodPollutantsFrequencyWho? With What? One-Point QCSO 2, NO 2, O 3, CO Every 2 WeeksSite operatorAnnually calibration standard, owned by tribe Annual Performance Evaluation SO 2, NO 2, O 3, CO Once per YearNot site operator, no national certification required, not the same standard that is used for routine QC checks Flow Rate Verification (not required to be reported to AQS) PM 10, PM 2.5 Once every 4 weeks Can be site operator, no national certification required, not the same standards that are used to calibrate your analyzer Semi-Annual Flow Rate Audit PM 10, PM 2.5 Every 6 MonthsPreferably not site operator, no national certification required, not the same standards that are used for monthly QC checks PM PEP, NPAPPM 2.5, All gaseous See QA Requirements.xls OAQPS/TAMS certified auditor OAQPS/TAMS equipment 19

Gaseous Standards  All calibrations, verifications, and audits should be conducted with certified gas standards or a certified transfer standard (ozone).  Gas standards are certified by the vendor and can be recertified (EPA Protocol 600/R- 97/121)  Ozone transfer standards should be annually certified and traceable to a Standard Reference Photometer (SRP) 20

 DASC (Data Assessment Statistical Calculator) ◦ In Excel, easy to use ◦ Calculates what EPA calculates in AMP255 ◦ Can be used with any parameters ◦ Cites equation #s from CFR  AQS: Data Quality Indicators Report (AMP255) ◦ Covers all of the Requirements ◦ ml ml 21

 Consider the standard to be the “truth” for your equipment  Difference between the “truth” and your equipment is the instrument error (this may include bias [error either usually high or usually low], and imprecision [wiggle sometimes high and sometimes low])  How is instrument error quantified? 22

Basis for All Calculations (divide by best estimate of the truth, to get a %-off-from- truth):  For standard gas  For Collocated Samples 23

If you are outside the limits:  Review procedures & logs to identify problem  Go back and review data—may have to throw out (invalidate—never delete) data to the last passing check (see validation templates for critical criteria that require invalidation— may not have to invalidate data if it’s a systematic or operational criterion)  Fix the problem, document it and report (when significant) 24

 “Accuracy” for EPA means “total error”  Comparison conducted with completely different system can be used to estimate total error  For PM2.5, this is a performance evaluation conducted with a side-by-side FRM device, with the filter weighed by a different lab  Total error includes both precision and bias errors  Accuracy estimates are verified thru audits 25

Resources  AMTIC  Transfer Standards for Calibration of Air Monitoring Analyzers for Ozone: Technical Assistance Document df  U. S. EPA Ambient Air Monitoring Protocol Gas Verification Program: Implementation Plan 2nd DRAFT 7/ EPA NPAP TTP PE SOP COMPENDIUM (DRAFT FINAL, ) pttpsop.pdf 26

Resources  Field Standard Operating Procedures for the Federal PM2.5 Performance Evaluation Program OP.pdf  Pb Performance Evaluation Program PEPImplementationPlan_7_09.pdf  Model Quality Assurance Project Plan For the National Air Toxics Trends Stations QAPP.pdf Resources—National QA Workgroup  QA National Meeting Presentations 27

 QC checks are necessary to both do, and  Track results of  Subsequent webinars will cover pollutant specifics QC checks  Course website: A101_Resources/ A101_Resources/  Our s: ◦ Travis Maki: ◦ ◦ 28