Lecture 2b. Electromagnetic Spectrum Visible range: =380-750 nm Ultraviolet: =190-380 nm Low energyHigh energy.

Slides:



Advertisements
Similar presentations
Introduction to Spectrophotometry & Beer’s Law
Advertisements

Spectroscopy and Beer’s Law
Infrared Spectroscopy A.Why are we doing this experiment? 1)Infrared spectra tell us only a limited amount about a transition metal complex 2)In very simple.
Electromagnetic spectrum Visible range: = nm Ultra-violet: = nm Low energyHigh energy.
Colorimetry & Spectrophotometry.
Chapter Eleven Ultraviolet-Visible Spectrophotometry
Spectrophotometry Chapter 17, Harris
Electronic Spectra: Ultraviolet and Visible Spectroscopy Ultraviolet and visible light give rise to electronic excitations. Spectroscopy of organic.
Spectrophotometric Analysis
Introduction to Spectrophotometry
Introduction to Spectrophotometry
Spectrophotometry: An Analytical Tool. PGCC CHM 103 Sinex IoIo I Cell with Pathlength, b, containing solution light source detector blank where I o =
Electromagnetic spectrum Visible range: = nm Ultra-violet: = nm Low energyHigh energy.
Lecture 3b. Electronic Transitions Most molecules absorb electromagnetic radiation in the visible and/or the ultraviolet range The absorption of electromagnetic.
Chapter 2 Quantitative.
SPECTROPHOTOMETRY cont.. Application qualitative A Absorbance spectrum.
Introduction to Instrumental Analysis - Spectrophotometry
Determination of Concentration Using Spectrophotometry
Spectrophotometry and Plotting of Calibration Curve
Let’s Talk About Beer Beer’s Law and Concentrations.
Colorimetry & Spectrophotometry.
SPECTROPHOTOMETRY. Determines concentration of a substance in solution by Measures light absorbed by solution at a specific wavelength by using spectrophotometer.
Spectrophotometry: An Analytical Tool
Colorimetry & Spectrophotometry
Spectrophotometry and Plotting of Calibration Curve BIO-2.
Determining the Concentration of a Solution: Beer’s Law
Spectrophotometry.
Introduction to Spectrophotometry
Lecture 6c. Introduction Electromagnetic spectrum Visible range: = nm Ultraviolet: = nm Low energyHigh energy.
Asa Arjoon U6 Chemistry Presentation on Visible and Ultraviolet Spectroscopy Name : Form : Subject :
EQUILIBRIUM CONSTANT DETERMINATION
Visible Spectroscopy Electromagnetic Radiation: Light & Color.
Spectrophotometry Electromagnetic Radiation = Light What is Light?
  Examining how much light is absorbed by a compound’s sample at various wavelengths  Spectrum peaks—  Indicates the wavelengths associated with electrons’
1 Spectroscopy  Atomic emission spectra  UV/Vis spectra  Infrared (IR)
SPECTROPHOTOMETRY. Principle : there is interaction between the light and sample particle, spectrophotometer is employed to measure the amount of light.
CLS 332 CLINICAL INSTRUMENTAL ANALYSIS. A VISIBLE ABSORPTION SPECTROMETER.
ATOMIC ABSORPTION SPECTROSCOPY (AAS) Atomization: It is the conversion of molecules to their component atoms in gaseous state using a source of heat (flame).
Determination of Concentration Using Spectrophotometry
Introduction to Spectrophotometry & Beer’s Law
UV SPECTROSCOPY Absorption spectra.
Photometry.
Atomic Absorption Spectroscopy.  Qualitative test for metals in solution  Can also be used for coloured lights  This emission of light is called an.
A darker color means a higher concentration of the colored component
COLORIMETRY AND SPECTROPHOTOMETRY
Investigating the relationship between Molar Concentration and Absorbance of Light Beer’s Law Link to spec 20 video.
Theory of Spectrophotometry
Introduction to Spectrophotometry
Absorption Spectrums 9/18/15.
SPECTROPHOTOMETRY.
Lab1 A VISIBLE ABSORPTION SPECTROMETER
Introduction to Spectrophotometry
Estimation of analyte concentration on the basis of light absorption
Introduction to Spectrophotometry
Spectrophotometer Dr . S. Jayakumar.
Spectrophotometry: An Analytical Tool
Spectroscopy Chem honors.
UV SPECTROSCOPY Absorption spectra.
COLORED SOLUTIONS A solution will appear a certain color if it absorbs its complementary color from the color wheel EX2-1 (of 24)
Lecture 2b Beer’s Lambert Law.
Cu2+ + 4NH3 → Cu(NH3)42+ (deep blue)
Can observe color using absorption spectroscopy (which looks at the intensity of light absorbed versus intensity of light transmitted)
SPECTROPHOTOMETRY Applied Chemistry.
STEP 1 – DETERMINE THE WAVELENGTH OF LIGHT TO USE
Beer's- Lambert Law and Standard Curves of concentrations
Cu2+ + 4NH3 → Cu(NH3)42+ (deep blue)
Sample AP Model Drawing Question
Spectrophotometry A method to determine concentration of a species exploiting the absorption of EMR.
Clinical instrumental analysis
Spectrophotometric Analysis
Presentation transcript:

Lecture 2b

Electromagnetic Spectrum Visible range: = nm Ultraviolet: = nm Low energyHigh energy

Emission vs. Absorption When determining a color, one has to know if the process that causes the color is due to emission or due to absorption of electromagnetic radiation Example 1: Sodium atoms emit light at =589 nm resulting in a yellow-orange flame Example 2: Indigo absorbs light at =605 nm which is in the orange range  the compound assumes the complementary color (blue-purple)

Beer’s Law Fundamental law regarding absorbance of electromagnetic radiation The cell dimension (l) is usually 1 cm (for standard cuvettes) The  -value is wavelength dependent. Thus, a spectrum is a plot of the  -values as the function of the wavelength (unit for  : M -1 *cm -1 ) The larger the  -value is, the larger the peak is going to be The data given in the literature only list the wavelengths and  -values (or its log value) of the peak maxima i.e., 331 (6460 or 3.81) The desirable concentration of the sample is determined by the largest and smallest  -values of the peaks in the spectral window to be measured

Practical Aspects The absorbance readings for the sample have to be in the range from A min =0.1 and A max =1 in order to be reliable Concentration limitations are due Association at higher concentrations (c>10 -4 M) Linear response of the detector in the UV-spectrophotometer Linear range for absorbance Concentration c min c max Linear concentration range

Iron Determination I The reaction of Fe 2+ -ions with bypyridyl leads to the red-violet complex. The complex is chiral and consists of equal amounts of the  - and  -isomer Note that only Fe 2+ -ions form the complex but not Fe 3+ -ions ( =620 nm,  =220). Thus, any Fe 3+ -ions have to be reduced first (with ascorbic acid) prior to the measurement The absorbance of the sample (via the transmission) at the wavelength of =520 nm (  = ~8660) can be used to determine the concentration of the Fe 2+ -ions in solution

Iron Determination II However, the proper response has to determined first by using standards to establish a calibration curve The student prepares several Fe 2+ - solution with known concentration and obtains the absorbance readings for the Fe 2+ -complex It is important to blank the spectrophotometer before each measurement (Why?) The slope of the best-fit line (Absorbance vs. concentration) should be close to the molar extinction coefficient