QC/PACS Artifact Identification. Artifact Causes Image receptor/Hardware –Dirty –Foreign material –Failures Software –Algorithm –Enhancements/manipulations.

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

IMAGE ARTIFACTS.
Image Reconstruction.
IMAGE QUALITY.
The Field of Digital Radiography
Primary Exposure Factors IV
Radiographic Quality Chapter 5.
Digital Radiography.
Radiology Artifacts.
Analyzing the Image Density. Density Overall blackening of the image.
Digital Image Manipulation I By Professor Stelmark.
Digital Imaging CHAPTERS 1, 4-7 CARTER.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
DARKROOM ,PROCESSING & Artifacts REVIEW
Technical Artifacts and Errors
FILM ARTIFACTS.
TIMERS.
Chapter 5 Cassette-Based Image Acquisition
Digital Radiography Chapter 22. History of Digital Radiography Slower process of conversion because no pressing need to convert to digital radiography.
Direct Digital Radiography or Direct Capture Radiography Bushong Ch. 27.
Dawn Guzman Charman, M.Ed., R.T. RAD TECH A
CR Image Acquisition By Professor Stelmark.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Computed Radiography and Digital Radiography
Unit III Creating the Image. Unit III Creating the Image.
Chapter 6: Digital Radiographic Imaging
RT Wk 14 Digital Artifacts & Imaging ERRORS
Quality Control.
CHAPTER 3 EQUIPMENT OPERATION AND QUALITY CONTROL
Image Quality Radiographic Resolution.
1 Digital Radiography Fall filmless’ radiology departments Diagnostic radiographers have traded their film and chemistry for a computer mouse.
Optimisation Strategies Ali B alhailiy
RAD 254 Digital Imaging Basic Elements of Digital Imaging CR/DR.
Factors affecting CT image RAD
1 Possible Radiographic Errors Radiology. 2 During Exposure Film is bent during set up and exposure Causing distortion Film is bent during set up and.
Computed Radiography (CR), & Digital Radiography (DR) Images
Computed Radiography By Prof. Stelmark. Presently, an acceleration in the conversion from screen-film radiography (analog) to digital radiography (DR)
Elsevier items and derived items © 2009 by Mosby, Inc., an affiliate of Elsevier Inc. Digital Image Artifacts Chapter 31.
Things they never tell you when the equipment is installed REV 11/2008
Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15
Registry Review Digital Imaging Considerations (Image Production and Evaluation)
1 ARTIFACTS: AN UNWANTED DENSITY ON THE FILM
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
Digital Imaging.
Image Production and Evaluation NOTE: QA AND QC ARE USED INTERCHANGABLY IN APPLETON AND LANGE.
Chapt. 25 Computed Radiography Digital terms are NOT uniform, but vary from one brand to the next!
IMAGE RECEPTORS. Follow the Sequence- Film System Tube Tube Patient Patient (bucky) or non-bucky (bucky) or non-bucky Cassette Cassette INTENSIFYING SCREEN.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
CR Image Acquisition By Professor Stelmark.
Digital Artifacts Edge Enhancement High contrast and edge definition –Lose fine detail on the edges of high contrast areas.
DIGITAL IMAGING.
Answer (slide 1/3): The correct answer is b) Black Surround Mask applied incorrectly. Black Surround Masking is an image processing feature used to reduce.
Direct Digital Radiography or Direct Capture Radiography
History of digital radiology 1980 Kinos L.T. ET all developed a portable radiographic X-ray camera made of tantalum and aluminum. Digital radiography.
B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS &
{ Digital Image Processing By Professor Stelmark.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Chapter 5 Cassette-Based Image Acquisition
Sunday Case of the Day Physics
Digital Artifacts.
Advantages/Disadvantages of CR
כיצד נרכשת התמונה בסרט הרנטגני?
מה צריך רנטגנאי לדעת על דימות דיגיטלית?
Challenges with Digital Imaging
Digital Imaging CHAPTERS 1, 4-7 CARTER.
Image Production and Evaluation
Chapter 6 Cassetteless Equipment and Image Acquisition
Digital radiography.
Presentation transcript:

QC/PACS Artifact Identification

Artifact Causes Image receptor/Hardware –Dirty –Foreign material –Failures Software –Algorithm –Enhancements/manipulations Operator/Patient –Collimation –Misused equipment –Incorrect patient data entry

Image Receptor/Hardware Dirt –In the CR cassette or reader Foreign material –Matter other than dirt within the CR cassette or reader Failures –Dead pixels –Mechanical failures with in the CR reader –Within the PACS network

CR Screen Artifacts These are caused by Dirt or other foreign matter within the cassette/plate apparatus

Dead Pixels Calibration algorithms detect individual pixels, clusters, and lines of contiguous pixels that fail to produce a usable output value. Algorithms are applied to the image to determine pixel values for the ‘dead’ or missing pixels.

CR Ghost Images Ghost or memory artifacts occur when the radiation remains trapped in the phosphor plate. Remember, radiation can remain trapped for several minutes. CR plates must be correctly erased to prevent ghost images.

Ghost Image This artifact can also occur with DR systems; particularly systems using photodiodes. The artifact occurs when exposures are made too rapidly and the unit was unable to clear its data completely.

CR Reader Artifacts Typically display as lines across the entire image –In contrast to dirt within the cassette or on the PSP which are shorter and/or curved lines.

Damaged laser beam head

Transmission errors in the network

Printer Distortion Occurs when the image size and the printed size are not equal.

Software Causes Algorithms –Incorrect selection –Age related –Pathology induced Post-processing –Edge enhancement, etc

Edge Enhancement High contrast and edge definition –Lose fine detail on the edges of high contrast areas

For the pediatric chest an adult chest algorithm was selected. Technologist error.

DR - RF Noise

Operator Errors Collimation Backscatter Moire artifact Quantum mottle –Insufficient mAs

Collimation Too much or too little collimation may effect image quality. When being reconstructed the image histogram has been programmed to search for a particular collimation pattern. If the collimation is faulty it may result in artifacts on the final image.

Light bulb Artifact Image A demonstrates backscatter on the image which produces a ‘light bulb’ effect within the image. Image B shows the same image with less kVp.

Moire Pattern Grid lines run in the same direction as the laser in the CR reader.

Quantum Mottle Image A demonstrates quantum mottle resulting from too few photons striking the imaging plate. Images B and C show increasing mAs values and decreasing quantum mottle.

Conclusion Digital artifacts can be caused by numerous means. It is the responsibility of the II/PACS professional to be the final monitor of image quality within the modern imaging department.

Future Possibilities Digital image monitoring