Chapt. 25 Computed Radiography Digital terms are NOT uniform, but vary from one brand to the next!

Similar presentations


Presentation on theme: "Chapt. 25 Computed Radiography Digital terms are NOT uniform, but vary from one brand to the next!"— Presentation transcript:

1 Chapt. 25 Computed Radiography Digital terms are NOT uniform, but vary from one brand to the next!

2 DR introduced in 1981 Fuji introduced “digital imaging” in 1981 –Many forms of digital CR DR

3 CR Terms PSL= photostimulable luminescence PSP=photostimulable phosphore SPS=storage phosphor screen IP=imaging plate SP=storage phosphore PMT=photomultiplier tube PD=photodiode

4 PSL & PSP Photostimulable luminescence (PSL) –Emit light promptly AND at a later time! Photostimulable phosphor (PSP) – barium fluorohalide looks physically like an intensifying screen BUT stores the latent image via metastable electrons (STORAGE PHOSPHOR SCREENS) SPSs

5 The “imaging plate” The PSP screen is inside a “rigid” (not really as they are brittle and do break easily – much more so than film/screen cassettes) cassette The actual imaging receptor is called the IMAGING PLATE

6 Light Stimulation Emission Process similar to Thermosluminescent dosimetry (TLD – film type badges for dosimetry) They are either “powder” (like an intensifying screen crystal) OR arranged like “needles” i.e. Powder PSP or Needle PSP

7 Thoughts on PSP’s Powder vs “needle” Highly sensitive to RADIATION = easily fogged!!!

8 CR Reader The physical receptor - cassette is “loaded” into the reader –IP is removed and moves along the LONG AXIS of the IP!!! (slow scan) –As the IP is moved, a deflection device is used to deflect the LASER beam across the IP (fast scan) IP doesn’t really separate from the cassette and doesn’t go around “rollers” etc.

9 CR Optical feature To properly and precisely translate EACH metastable electron (latent image forming charge) in a PRECISE fashion to provide for great SPATIAL RESOLUTION

10 Compare and contrast film-screen to CR techniques Film- screen has much LOWER latitude than CR!!! –As such – a WIDE variety of techniques can be used = Burn em up BURNIE!!!! If in doubt – “BURN IT OUT” Compare Characteristic curves for each:

11 Image noise Box 25-1 SAME as screen-film noise PLUS Box 25-2 –Mechanical defects – due to the scan drivers –Optical defects – laser, stimulating beam, light quanta emission and collected –Computer defects – electronic noise; too few sampling; too little quantization