Introduction to NMR Spectroscopy and Imaging Lecture 10 Basics of Magnetic Resonance Imaging (Spring Term, 2011) Department of Chemistry National Sun Yat-sen.

Slides:



Advertisements
Similar presentations
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Advertisements

Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Ari Borthakur, PhD Associate Director, Center for Magnetic Resonance & Optical Imaging Department of Radiology Perelman school of Medicine, University.
Magnetic Resonance Imaging
Imaging Sequences part I
MRI Phillip W Patton, Ph.D..
BE 581 Lecture 3- Intro to MRI.
Magnetic Resonance Imaging - MRI
RF Pulse – generates a B 1 field that realigns the precessing spins in the low energy state In the case of a 90 o pulse the alignment is perpendicular.
MR Sequences and Techniques
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Statistical Parametric Mapping
Parameters and Trade-offs
Magnetic Resonance Imaging
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Principles of MRI: Image Formation
Chapter 9 Basic MRI I Mark D. Herbst, MD, PhD. Notice This lecture contained many drawings on the whiteboard, so get these from one of the other students.
Magnetic Resonance Imaging Maurice Goldman Member Académie des sciences.
Basic Principles MRI related to Neuroimaging Xiaoping Hu Department of Biomedical Engineering Emory University/Georgia Tech
The Basics of MRI The Basics of MRI. Current MRI technology displays images as multiple sets of gray tone images. Visualization and interpretation of.
Terry M. Button, Ph.D. Principals of Magnetic Resonance Image Formation.
Nuclear Magnetic Resonance (NMR) Magnetic Resonance Imaging (MRI)
Basics of Magnetic Resonance Imaging
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Psy 8960, Fall ‘06 EPI, Part 2: variants1 Segmentation Partial Fourier Spin echo vs. gradient echo Inversion recovery Long vs. short TE.
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
MRI: an Introduction By Mohammad Ali Ahmadi Pajouh Amirkabir University of Technology Biomedical Eng. Dep.
Medical Imaging Systems: MRI Image Formation
Magnetic Resonance Imaging 4
Principles of MRI Physics and Engineering
Imaging Sequences part II
ELEG 479 Lecture #12 Magnetic Resonance (MR) Imaging
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Medical Imaging Systems: MRI Image Formation
Introduction to NMR Spectroscopy and Imaging Review (Spring Term, 2011) Department of Chemistry National Sun Yat-sen University 核磁共振光譜與影像導論.
Introduction to Magnetic Resonance Imaging Bruno Quesson, CR1 CNRS.
Pulse Sequences Types of Pulse Sequences: Functional Techniques
Basic Concept of MRI Chun Yuan. Magnetic Moment Magnetic dipole and magnetic moment Nuclei with an odd number of protons or neutrons have a net magnetic.
Basic of Magnetic Resonance Imaging Seong-Gi Kim Paul C. Lauterbur Chair in Imaging Research Professor of Radiology, Neurobiology and Bioengineering University.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
September, 2003BME 1450 Introduction to NMR 1 Nuclear Magnetic Resonance (NMR) is a phenomenon discovered about 60 years ago.
fMRI Methods Lecture2 – MRI Physics
Quiz In a 2D spin warp or FT MR scan, aliasing should only occur
Contrast Mechanisms in MRI Introduction to Cardiovascular Engineering Michael Jay Schillaci, PhD Managing Director, Physicist Tuesday, September 16 th,
Allen W. Song, PhD Brain Imaging and Analysis Center Duke University MRI: Image Formation.
Contrast Mechanism and Pulse Sequences
Magnetic Resonance Imaging – Basic Principles –
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
BME1450 Intro to MRI February 2002 The Basics The Details – Physics The Details – Imaging.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
V.G.Wimalasena Principal School of Radiography
MR Image Formation FMRI Graduate Course (NBIO 381, PSY 362)
MRI: Contrast Mechanisms and Pulse Sequences
Magnetic Resonance Learning Objectives
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Charged particle. Moving charge = current Associated magnetic field - B.
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Lecture 1: Magnetic Resonance
MAGNETIC RESONANCE IMAGING by PRADEEP V.EPAKAYAL. Mem.no L.
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
FMRI data acquisition.
Imaging Science Department, IC Jo Hajnal
MRI Physics in a Nutshell Christian Schwarzbauer
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Magnetic Resonance Imaging: Physical Principles
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Basic MRI I Mark D. Herbst, MD, PhD
核磁共振光譜與影像導論 Introduction to NMR Spectroscopy and Imaging Lecture 05 Basic Two-Dimensional Experiments (Spring Term, 2011) (Fall Term, 2011) (Spring Term,
(4)ELECTRONIC SUPPORT SYSTEM
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
Presentation transcript:

Introduction to NMR Spectroscopy and Imaging Lecture 10 Basics of Magnetic Resonance Imaging (Spring Term, 2011) Department of Chemistry National Sun Yat-sen University 核磁共振光譜與影像導論

Chapter 10 Basics of Magnetic Resonance Imaging Basic principle of MRI Back projection Fourier imaging Image contrast Fast 3D Imaging Functional MRI (fMRI)

Sources of Materials for this Chapter

Micro-imaging Animal imaging Whole-body (human) imaging ( MRI) Scanner = (MRI) Imager = MRI machine = MRI =NMR spectrometer with microimaging function

Golay Maxwell (Anti-Helmholtz)

Back Projection

Reconstruction

Broad band and narrow band pulses (Selective) (Nonselective)

Gaussian Half Gaussian EBURP2 Q5 (Gaussian cascade) Q3 (Gaussian cascade) There are many other selective pulses available. Customer designed pulse shapes can be synthesized on spectrometers at will. Other selective pulses (pulse shaping)

With a selective pulse of a definite excitation width of Δν, only a slice of thickness Z is excited (transverse magnetization generated)

Brute force back projection

Problems A sample contains water at two locations, x = 0 cm and x = 2.0 cm. A one- dimensional magnetic field gradient of 1 G/cm is applied along the x-axis during the acquisition of an FID. What frequencies (relative to the isocenter frequency) are contained in the Fourier transformed spectrum? An NMR spectrum is recorded from a sample containing two water locations. The frequency encoding gradient is 1 G/cm along the y-axis. The spectrum contains frequencies of Hz and -500 Hz relative to the isocenter frequency. What are the locations of the water? You want to excite spins in an xy-plane located at z = -5.0 cm. The resonance frequency at the isocenter is MHz and your slice selection gradient is 1 G/cm. Describe in detail the RF pulse which should be used. A sample contains water at two locations, y = 1.0 cm and y = -2.0 cm. A one- dimensional magnetic field gradient is applied along the y-axis during the acquisition of an FID. The frequency encoding gradient is 1 G/cm. What frequencies (relative to the isocenter frequency) are contained in the Fourier transformed spectrum? An NMR spectrum is recorded from a sample containing two water locations. The frequency encoding gradient is 0.5 G/cm along the z-axis. The spectrum contains frequencies of Hz and +500 Hz relative to the isocenter frequency. What are the locations of the water? You want to excite spins in an xy-plane located at z = -2.0 cm. The resonance frequency at the isocenter is MHz and your slice selection gradient is 2 G/cm. Describe in detail the RF pulse which should be used.

Single voxel image FT2 = FT1

Two-voxel image FT2 Zero-filling FT1

Field Of View (FOV) and Image Resolution

Resolution

Problems Two samples are located in a magnetic field at x=0 cm and x=5 cm. A 1 G/cm phase encoding gradient is applied in the +X-direction for 10 ms. How much phase will be acquired by the sample located at x=5 cm relative to that located at x=0 cm? You wish to produce an image of hydrogen nuclei in the zx-plane. What directions should the slice, phase, and frequency encoding gradients be applied in? A particular magnetic resonance imager uses a 1 G/cm frequency encoding gradient to produce an image with an 8 cm FOV. What quadrature sampling rate should be used to produce this FOV when imaging hydrogen? You wish to produce an image with a 8 cm FOV and 256 phase encoding gradient steps. The maximum phase encoding gradient you can produce is 1 G/cm. What should the width of the phase encoding gradient be? Show that: FOV = sw / ( γ Gf). Two samples are located in a magnetic field at x=0 cm and x=-4 cm. A 2 G/cm phase encoding gradient is applied in the +X-direction for 5 ms. How much phase will be acquired by the sample located at x=-4 cm relative to that located at x=0 cm? You are using an imaging sequence that applies a slice selection gradient in the x direction, a phase encoding gradient in the z direction, and a frequency encoding gradient in the y direction. What type of imaging plane will be produced? A particular magnetic resonance imager uses a 2 G/cm frequency encoding gradient and a quadrature sampling rate of 32 kHz. What field of view will be obtained when imaging hydrogen? You wish to produce an image with a 4 cm FOV and 512 phase encoding gradient steps. The maximum phase encoding gradient you can produce is 2 G/cm. What should the width of the phase encoding gradient be?

Brute force back projection T

FT in n-dimension space Image = Spatial FT of original FIDs

Brute force back projection T k-space: (t, θ)

Te t k-space: (t, G x )

Signal in k-space:M(G x,t) = t GxGx Stack representation Contour representation

t GxGx

t θ k-space: (t, θ)

What does re-grid mean? Original signal: M(Δθ, Δt), M(Δθ,2 Δt), M(Δθ,3 Δt),…M(Δθ, (np/2)*Δt), M(2Δθ, Δt), M(2Δθ,2 Δt), M(2Δθ,3 Δt),…M(2Δθ, np*Δt), …… M(nv*Δθ, Δt), M(nv*Δθ,2 Δt), M(nv*Δθ,3 Δt),…M(nv*Δθ, np*Δt), Regrided signal: M(ΔtcosΔθ, ΔtsinΔθ), M(ΔtcosΔθ, 2ΔtsinΔθ,…, M(ΔtcosΔθ,( np/2)*ΔtsinΔθ M(2ΔtcosΔθ, ΔtsinΔθ), M(2ΔtcosΔθ, 2ΔtsinΔθ,…, M(2ΔtcosΔθ,( np/2)*ΔtsinΔθ …… M(nv*ΔtcosΔθ, ΔtsinΔθ), M(nv*ΔtcosΔθ, 2ΔtsinΔθ,…,M(nv*ΔtcosΔθ,( np/2)*ΔtsinΔθ Just coordinate transformation!

Re-griding y x

3D image of neck

Breathing lung

Beating heart

μMRI applications Single Cell Gene Expression Materials

Diffusion coefficient images of isolated Aplysia neurons acquired at 8.5 T. The cells are visible as a dark region with low diffusion coefficient within the brighter circle of the capillary tube containing artificial sea water. There is a strong contrast in these images between the nuclear region and the perikaryon, indicating that water molecules in these regions have strongly differing diffusion coefficients. A variety of irregular nuclear morphologies are apparent from the shapes of the bright regions present within the cell bodies J.S. Schoeniger, N. Aiken, E. Hsu, S.J. Blackband, J. Magn.Reson. Ser. B 103 (1994) 261.

“Smooth” CD surface in the nano-eye Not MRI mimage!

Some microimages made in this lab Unknown flower Sesame seed Fuel cell God-fish bone Onion epidermis

Weighted images Density image is the most basic Relaxation weighting Diffusion weighting Chemical shift weighting Temperature/pressure weighting Multi-quantum relaxation weighting Biochemical/physical weighting Psychological weighting

Question: what is the purpose of each of the white gradient fields?

Answer: restores the initial alignment of transverse magnetization because the slice- selection gradient field dephase the transeverse magnetization. Answer: shifts the echo (during acquisition) so that the first point is not the echo peak when phase-encoding gradient field is zero.

Spin Echo Imaging

Spin Echo Multiple Slice Imaging (SEMS) d1 can be short because each next scan is on a different slice.

Spin Echo Multiple Slice Imaging (SEMS) Tss Tpe Tef=atTec Maximum slices per TR:

TE (ms) TE (ms) TR (ms) Spin echo images

Inversion-recovery imaging

Inversion-recovery imaging: slice selection

Inversion-recovery imaging

TR (s) 1 2 TI(ms) Inversion-recovery images

Gradient recalled echo imaging

One additional gradient is applied between the 90o and 180o pulses. This gradient is along the same direction as the frequency encoding gradient. It dephases the spins so that they will rephase by the center of the echo. This gradient in effect prepares the signal to be at the edge of k-space by the start of the acquisition of the echo.

2

Gradient recalled echo imaging

Gradient recalled echo images (TE=5 ms) Θ(o) TR (ms)

Spin-Echo S = k r (1-exp(-TR/T1)) exp(-TE/T2) Inversion Recovery (180-90) S = k r (1-2exp(-TI/T1)+exp(-TR/T1)) Inversion Recovery ( ) S = k r (1-2exp(-TI/T1)+exp(-TR/T1)) exp(-TE/T2) Gradient Recalled Echo S = k r (1-exp(-TR/T1)) Sinq exp(-TE/T2*) / (1 -Cosq exp(-TR/T1))

Multiple Slice Imaging with Other Sequences Obviously, inversion-recovery, gradient recalled echo sequences can also be performed in multi-slice acquisition mode. Maximum slices per TR for gradient echo sequence: Maximum slices per TR for inversion-recovery sequence:

Chemical shift imaging (CSI)

WeightingTR ValueTE Value T1T1 < = T 1 < < T 2 T2T2 > > T 1 > = T 2  > > T 1 < < T 2

Tissue T 1 (s) T 2 (ms) ** CSF White Gray Meninges Muscle Adipose *Based on  =111 for 12mM aqueous NiCl 2

Activations (red–white color scale) and deactivations (blue–green color scale) in a group of 11 healthy elderly subjects during visual word encoding and retrieval (p 0.01, uncorrected). The Talairach coordinates of axial slices are indicated on the left-hand bottom of each image.

Concepts in Magnetic Resonance Part A, Vol. 16A(1) 35–49 (2003) Concepts in Magnetic Resonance Part A, Vol. 16A(1) 5–15 (2003)

Problems How many slices could you image with a 90-FID sequence which uses a 20 ms slice selection gradient, 10 ms phase encoding gradient, 100 ms frequency encoding gradient, and TR of 1 second? Draw a timing diagram for an inversion recovery imaging sequence which uses a 90-FID sequence rather than a spin-echo sequence to detect the signal present TI after the inversion (180 degree) pulse. There are two adjacent tissues in an image with a T2 of 30 and 50 ms. You wish to produce a spin-echo image of the tissues wuch that the contrast between the tissues is maximized. What TE should be used? You wish to image an oblique slice located at an angle of 45o to the +Z-axis and 45o to the +Y-axis. Since gradients can only be produced with the three gradient coil systems located around the x, y, and z-axes, what combination of the three gradients should be applied to produce a slice selection, phase encoding, and frequency encoding gradient? How many slices could you image with a gradient echo sequence that uses a 20 ms slice selection gradient, 10 ms phase encoding gradient, 10 ms dephasing frequency gradient, 100 ms frequency encoding gradient, and TR = 1 s? Draw a timing diagram for an inversion recovery imaging sequence which uses a gradient echo sequence rather than a spin-echo sequence to detect the signal present TI after the inversion (180 degree) pulse. There are two adjacent tissues in an image with T1a = 300 ms, T2a* = 20 ms, ra = 50 ASDU, and T1b = 150 ms, T2b* = 20 ms, rb = 50 ASDU. (ASDU = arbitrary spin density units.) You wish to produce a 90 o gradient echo image of the tissues such that the contrast between the tissues is maximized. What TR should be used? Assuming the standard magnetic coordinate system, what components of the gradients Gx, Gy, and Gz should be applied to create the gradients Gslice, Gphase, and Gfrequency for the following slice selection? A slice located 30o? away from the Z-axis and forming equal angles with the +X and +Y axes.

Magnetic Resonance Microimaging (μMRI) The principle is basically the same as that of ordinary MRI. Sample size is in mm or smaller but resolution might be 10 times higher or more. Higher gradients are needed. Being at fast developing stage, with promise of single-spin imaging. A few challenges need to be addressed.

Microcoil is used (OD= 470 μm) Human hair (75 μm)

For μNMR microcoil array and surface coils Surface coils Coil arrays 100 μm Human hair

Concepts in Magnetic Resonance, Vol. 13, 128, Concepts in Magnetic Resonance, Vol. 13, 190, 2001.

0.2 mm I.D. NMR microcoil (Full of de- ionized water) Copper wire 5.5 mm Top mm 4.4 mm 1.4 mm 5.5 mm Side 2.45 mm 0.35 mm mm The Gradient set holder Micro-Coil Gradient System

Other magnetic resonance imaging techniques Magnetic force imaging Quantum imaging (single-spin imaging) Ferromagnetic resonance force imaging

Would MRI do these? Yes! We will make it!