Understanding and Comparing Confocal Scanning Lasers to Optical Coherence Tomography for Optic Nerve Head Analysis (ONH) A presentation courtesy of Zeiss.

Slides:



Advertisements
Similar presentations
Detection of weak optical signals D.R. Selviah, R.C. Coutinho, H.A. French and H.D. Griffiths Department of Electronic and Electrical Engineering, University.
Advertisements

Image Reconstruction.
OCT= Optical Coherence Tomography
Bio-Rad Labs. Chromatic Aberration Sources and Measurement.
Monday, December 15, 2008 Record homework. Unpack. Get response pad. Permission slips on the stool. Agenda Agenda:  Quiz on 16.3 & 16.4  Completion of.
OPTICAL COHERENCE TOMOGRAPHY IN GLAUCOMA
LIGHT AND THE RETINAL IMAGE: KEY POINTS Light travels in (more or less) straight lines: the pinhole camera’s inverted image Enlarging the pinhole leads.
Motivation Spectroscopy is most important analysis tool in all natural sciences Astrophysics, chemical/material sciences, biomedicine, geophysics,… Industry.
SPECTRALIS® Glaucoma Module Premium Edition
Oct interpretation Ghanbari MD This is what we wanted.
POSTER TEMPLATE BY: Image Processing in Spectral Domain Optical Coherence Tomography (SD-OCT) Vasilios Morikis 1,2, Dan DeLahunta.
Scanning Confocal Basics n What’s so special ? n Confocals give higher resolution than normal microscopes. n Confocals can OPTICALLY SECTION a sample.
Presented by A. Rismanchian MD Farabi Hospital Clinical Aspects Of OPTICAL COHERENCE TOMOGRAPHY IN GLAUCOMA.
ABRF meeting 09 Light Microscopy Research Group. Why are there no standards? Imaging was largely an ultrastructure tool Digital imaging only common in.
Short pulses in optical microscopy Ivan Scheblykin, Chemical Physics, LU Outline: Introduction to traditional optical microscopy based on single photon.
3D Reconstruction of Anatomical Structures from Serial EM images.
Imaging Science FundamentalsChester F. Carlson Center for Imaging Science The Human Eye and Vision 1 (Producing The Image)
POSTER TEMPLATE BY: Optical Coherence Tomography Assembly of the System and Analysis of Sciatic Rat Nerve Daniel deLahunta.
METO 621 Lesson 27. Albedo 200 – 400 nm Solar Backscatter Ultraviolet (SBUV) The previous slide shows the albedo of the earth viewed from the nadir.
What is OCT? Ultrahigh resolution optical coherence tomography (UHR OCT) is an upcoming new technology that allows non-invasive, optical medical diagnostic.
Peripherals and Storage Looking at: Scanners Printers Why do we need storage devices anyway? What are magnetic disks? How do magnetic disks physically.
Microscopy.
D EDICATED S PECTROPHOTOMETER F OR L OCALIZED T RANSMITTANCE A ND R EFLECTANCE M EASUREMENTS Laetitia ABEL-TIBERINI, Frédéric LEMARQUIS, Michel LEQUIME.

Sensation & Perception
Chapter 23 Properties of Light. Section 1: Objectives Use ray diagrams to show how light is reflected or refracted. Compare plane mirrors, concave mirrors,
Visual structure & Blind spot. Question 1 What do these devices have in common?
Austin Roorda, Ph.D. University of Houston College of Optometry
Lithological and Structural Mapping Using Satellite Data Mohamed Abdelsalam Missouri S&T.
Dr. Ayesha Abdullah Learning outcomes By the end of this lecture the students would be able to; Correlate the structural organization of the.
1)Adaptive optics: optimization and wavefront sensing 2)Novel microscope enhancements.
Detectors and Cross Talk Presented below are cross talk measurements carried out on 2 Burle and 1 Hamamatsu MCP PMTs and 1 Hamamatsu MultiAnode PMT (MAPMT).
Grading Rubric Total Points = 30 TABLE OF CONTENTS (1 pt) NEATNESS (3 pts): Can I identify each section?? No scratching out of mistakes. Am I having a.
How do I see color? Photochemical receptors receive the light (____ and _____) Rods-brightness cones-the color They release a ________ signal to the brain.
ECE 299 Holography and Coherent Imaging
1 Computational Vision CSCI 363, Fall 2012 Lecture 5 The Retina.
Optical Coherence Tomography (OCT) Gella Laxmi 2009PHXF013P.
Molecular Cell Biology Light Microscopy in Cell Biology Cooper Modified from a 2010 lecture by Richard McIntosh, University of Colorado.
Image Compression in Optical Coherence Tomography Biomedical Engineering, Tambov State Technical University, Russia K.E.S.Ghaleb, A.Yu.Potlov, S.N.Abdulkareem,
Designing a Microscopy Experiment Kurt Thorn, PhD Director, Image from Susanne Rafelski, Marshall lab.
Date of download: 6/1/2016 The Association for Research in Vision and Ophthalmology Copyright © All rights reserved. From: Optical Coherence Tomography.
Date of download: 6/2/2016 Copyright © 2016 American Medical Association. All rights reserved. From: Spectral Domain Optical Coherence Tomography: Ultra-high.
Imaging.
Date of download: 6/21/2016 Copyright © 2016 American Medical Association. All rights reserved. From: Effect of Peripapillary Vitreous Opacity on Retinal.
Optical Instruments II Instruments for Imaging the Retina.
Date of download: 7/5/2016 Copyright © 2016 SPIE. All rights reserved. Spatial light modulator-two-photon microscope (SLM-2PM) scheme: (1) Ti:Sa laser.
Metallurgha.ir1. Lecture 5 Advanced Topics II Signal, Noise and Bandwidth. Fundamental Limitations of force measurement metallurgha.ir2.
Optical Non-Invasive Approaches to Diagnosis of Skin Diseases
Assessment of trans-laminar cribrosa pressure gradient and lamina cribrosa thickness in Normal Tension Glaucoma V. V. Volkov, I.L. Simakova, A.N. Kulikov,
Structure-Function Relationship Between the Bruch Membrane Opening-based Minimum Rim Width and Visual Field Defects in Advanced Glaucoma Serhat Imamoglu,
Copyright © 2013 American Medical Association. All rights reserved.
Volume 122, Issue 7, Pages (July 2015)
Glaucoma Diagnosis Ophthalmology Glaucoma
P.J. Caspers, G.W. Lucassen, G.J. Puppels  Biophysical Journal 
A new way to see abnormal blood vessels in AMD
Optical Non-Invasive Approaches to Diagnosis of Skin Diseases
Mohamed S. Sayed, MD, Michael Margolis, MD, Jessica L
Mont-Carlo simulation of OCT structural images of subcutaneous
How to interpret the HRT
Figure 2 Optical coherence tomography of the human retina A) Detailed retinal segmentation sample on spectral-domain optical coherence tomography (OCT)
Figure 1 Example OCT scan
Multiphoton Excitation Provides Optical Sections from Deeper within Scattering Specimens than Confocal Imaging  Victoria E. Centonze, John G. White  Biophysical.
Fault Identification: Time Slice View
Atsuya Miki, MD, PhD, Yasushi Ikuno, MD, Robert N
(case 6)  (A) Fundus photography showing subtle discrete areas of RPE atrophy (green areas). (case 6)  (A) Fundus photography showing subtle discrete areas.
In vitro glycation of articular cartilage alters the biomechanical response of chondrocytes in a depth-dependent manner  J.M. Fick, M.R.J. Huttu, M.J.
P.J. Caspers, G.W. Lucassen, G.J. Puppels  Biophysical Journal 
Fig. 4 Pupil shape and image quality in the model sheep eye.
Fig. 3. Low power hemi retinal image of the RPE surface showing the albino central and equatorial retina and pigment distribution in the periphery. Low.
Fig. 6. Confocal image series at 10-μm intervals through the full retinal thickness at P48 in WT and vldlr-null mice. Confocal image series at 10-μm intervals.
Presentation transcript:

Understanding and Comparing Confocal Scanning Lasers to Optical Coherence Tomography for Optic Nerve Head Analysis (ONH) A presentation courtesy of Zeiss

Third Generation Products LDT abandoned TopSS in 1999 to continue develop- ment of GDx Access HDT released their HRT II instrument, which is smaller, faster, easier and cheaper

How they Work A nm LASER light source plus (2) Pinhole (20-25 micron) apertures detect light reflected off the retina LASER DETECTOR PINHOLE APERTURE FOCAL POINT

Signal Confocal Scanning Lasers sense the shape of the surface of the Hyloid Face or the Vitreous to Internal Limiting Membrane interface. At this point the peak reflection is recorded Difficulties with some common conditions as PVD can cause artifacts. Have excellent “X-Y” resolution, but coarse “Z” resolution.

Scanning Lasers make Slices Typically 32 “Slices” are acquired over a 4mm depth. Each slice has a depth of 25 microns. A non sampled gap of 100 microns is between slices. Y X Z

Scanning Typically a 6mm by 6mm area of retinal surface is scanned. Each slice is about 20 –25 microns thick. Each gap between slices is about 100 microns thick. Only a few slices fall on the ONH.

Examples of Data Two of the 32 slices are displayed at the right. Lets plot the paths of light of the 3 data- points. At different areas on the disk in slice 14 of 32 notice: –Yellow is on a blood Vessel –Green is on the rim –Purple is in the cup Slice 14 of 32 Slice 18 of 32

How the slices determine topography A graph illustrates REFLECTANCE for the same location at each of the 32 slices. Since the Yellow curve peaked before the Green or Purple curve, a RELATIVE difference in Topography can be calculated.

Establishing a Reference Plane A Publication by Weinreb et al shows the Papilomacular bundle as an area adjacent to the rim showing the least amount of change. So these instruments use this area and other areas as a peripheral band concentric around the disk to manufacture “ZERO” or a reference plane.

Volumetric Analysis The operator must draw the outline of the disk. This is not easy as the image is not taken in white light. Notice how two different operators outline the same disk with different shapes. Volumetric parameters are operator dependent.

Optical Coherence Tomography To scan the ONH, Optical Coherence Tomography acquires (6) 4mm long by 2mm deep “Slices” in the X-Z axis. Datapoints are sampled every 2 microns providing excellent resolution in the Z axis. Confocal Scanning Lasers typically sample every 100 microns. Z X Y

Image Analysis Using a proprietary algorithm, OCT identifies the Retina Pigment Epithelium and places a marker.

The markers show the boundaries of the disk in each individual slice. The next slide shows all six slices.

ONH Tomography Examples Slice 1 Slice 2 Slice 3 Slice 4 Slice 5 Slice 6

ONH Analysis Using the algorithm OCT objectively finds the margin of the disk. OCT objectively finds the cup by using an offset value of 150 microns up from the RPE. No reference plane is required.

OCT Printout Details volumetric analysis of cup and Disk. Provides Direct Cross Section of anatomy.

“Z” Dimension Datapoints Confocal Scanning Lasers acquire 32 datapoints over 4mm Optical Coherence Tomography acquire 1024 datapoints over 2mm

Datapoint Comparrison Confocal Scanning Lasers have 100 micron gaps between datapoints in “Z” dimension. Optical Coherence Tomography has only 2 micron gap. 150 Microns 100 Micron Gap

Conclusion Confocal Scanning Laser Technology is not capable of seeing small changes in the ONH. Glaucomatous change occurs very slowly. Progression may not be observed with sampling points with 100 micron gaps. OCT with higher “Z” resolution is more likely to see glaucomatous progression. Objective placement of Cup and Disk insures accuracy between operators. Viewing the anatomy confirms ONH Analysis from artifacts seen in confocal scanning lasers.