Resident Physics Lectures Christensen, Chapter 8 Grids George David Associate Professor Department of Radiology Medical College of Georgia
Purpose Directional filter for photons Ideal grid passes all primary photons photons coming from focal spot blocks all secondary photons photons not coming from focal spot Focal Spot “Good” photon Patient “Bad” photon X Grid Film
Grid Construction Lead Interspace ~ .05“ thick upright strips (foil) material between lead strips maintains lead orientation materials fiber aluminum wood Lead Interspace
Grid Ratio h Grid ratio = h / w Ratio of interspace height to width w Lead Interspace h w Grid ratio = h / w
Grid Ratio Expressed as X:1 h Grid ratio = h / w Typical values 8:1 to 12:1 for general work 3:1 to 5:1 for mammography Grid function generally improves with higher ratios h w Grid ratio = h / w
Lines per Inch # lead strips per inch grid width Typical: 103 25.4 W + w w W w = thickness of interspace (mm) W = thickness of lead strips (mm)
Grid Structure
Grid Patterns Orientation of lead strips as seen from above Types Linear Cross hatched 2 stacked linear grids ratio is sum of ratios of two linear grids very sensitive to positioning & tilting Rare; only found in specials
Grid Styles Parallel Focused
Parallel Grid lead strips parallel useful only for small field sizes large source to image distances
Focused Grid Slightly angled lead strips Strip lines converge to a point in space called convergence line Focal distance distance from convergence line to grid plane Focal range working distance range width depends on grid ratio smaller ratio has greater range Focal range Focal distance
Ideal Grid passes all primary radiation Reality: lead strips block some primary Lead Interspace
Ideal Grid block all scattered radiation Reality: lead strips permit some scatter to get through to film Lead Interspace
Primary Transmission Fraction of a scatter-free beam passed by grid Ideally 100% (never achieved) Lead Interspace
Primary Transmission Typical values: 55 - 75% Theoretic calculation: (fraction of grid that is interspace) Tp (%)= 100 X W / (W+w) where W = Interspace thickness w = lead strip thickness actual transmission < theoretical primary attenuated by interspace material focusing imperfections W w W+w
Bucky Factor Radiation incident on grid ----------------------------------- transmitted radiation indicates actual increase in exposure because of grid’s presence due to attenuation of both primary & secondary radiation
Bucky Factor Measures fraction of radiation absorbed by grid high ratio grids have higher bucky factors
Bucky Factor higher x-ray technique higher patient dose typically 3-6 Higher bucky factor means higher x-ray technique higher patient dose typically 3-6
More Lines / inch at Same Ratio Means Less Lead Content & Less Contrast Improvement thinner lead & same ratio less lead (less thickness, same height) Same interspace dimensions same contrast improvement for 133 line 10:1 and 80 line 8:1 grids h d Grid ratio = h / d
Grid Disadvantages Increased patient dose Positioning critical poor positioning results in grid cutoff loss of primary radiation because images of lead strips projecte wider
Grid Cutoff focused grids used upside down lateral decentering (or angulation) focus- grid distance decentering combined lateral & focus-grid distance decentering
Upside Down Focused Grid Dark exposed band in center Severe peripheral cutoff
Lateral Decentering uniform loss of radiation over entire film uniformly light radiograph no recognizable characteristic (dangerous)
Lateral Decentering also occurs when grid tilted both result in uniform loss of intensity no other clinical clues may be mistaken for technique problems Can be compensated for by increasing patient exposure
Lateral Decentering Significant problem in portable radiography Compensate by over-exposing patient exact centering not possible minimizing lateral decentering low ratio grids long focal distances
Distance Decentering Grid too close or too far from focal spot Darker center All parallel grids have some degree of distance decentering Focused to infinity
Near focus-grid decentering Far focus-grid decentering cutoff at periphery dark center cutoff proportional to grid ratio decentering distance
Minimizing Distance Decentering Cutoff low grid ratio small fields
Combined lateral and focus-grid distance decentering Easy to recognize uneven exposure film light on one side, dark on the other
Moving Grids Motion starts with second trigger Grids move ~1- 3 inches must be fast enough not to see grid lines for short exposures Motion blurs out lead strip shadows
Moving Grid Disadvantages $$$ Vibration Potential May limit minimum exposure time Increases patient dose lateral decentering from motion up to 20% loss of primary evenly distributes radiation on film stationary grid makes interspace gaps darker for same amount of radiation
Grid Tradeoff Advantage Disadvantage cleanup / scatter rejection increased patient dose increased exposure time increase tube loading positioning & centering more critical $$$
Grid Selection use low ratios for low kVp, high ratios for high kVp book recommends 8:1 below 90 kVp 12:1 above 90 kVp
Air Gap Techniques Principle Negligible attenuation in air gap Angles of escape Principle radiation scatters uniformly decrease in scatter (most scatter misses film) air gap decreases angle of capture; increases angle of escape Negligible attenuation in air gap
Air Gap air gap very effective in removing scatter originating closest to film much of scatter nearest tube doesn’t reach film Much attenuation of scatter in the body Air gap decreases capture angle
Air Gap Applications Magnification Radiography including mammography geometry causes air gap Grid not used with air gap
Mammo Cellular Grid