Computed Tomography http://www.stabroeknews.com/images/2009/08/20090830ctscan.jpg http://upload.wikimedia.org/wikipedia/commons/archive/d/da/20060904231838!Head_CT_scan.jpg.

Slides:



Advertisements
Similar presentations
PRINCIPLES OF CT.
Advertisements

IMAGE RECONSTRUCTION.
ARTIFACTS IN COMPUTED TOMOGRAPHY
IMAGE GENERATION IN CT.
Computers and Computed Tomography
Image Reconstruction.
Wednesday night and Thursday Lab Mega Group.  1967: The first Computer Tomography (CT) theory was developed  1972: The CT scan was invented by Godfrey.
BIOE 220/RAD 220 REVIEW SESSION January 30, 2011.
A Technical Seminar On X-RAY AND CT SCAN.
Computed Tomography Basic principles V.G.Wimalasena Principal
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Image Artifacts Chapter 8 Bushong.
Computed Tomography II
Seeram Chapter 13: Single Slice Spiral - Helical CT
CT Multi-Slice CT.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
CT Physics V.G.Wimalasena Principal School of radiography.
Computed Tomography III
Computed Tomography RAD309
Computed Tomography. Introduced in 70’s Principle: Internal structures of an object can be reconstructed from multiple projections of the object.
tomos = slice, graphein = to write
Chapter 2 Stewart C. Bushong
CT scanning (f) Candidates should be able to show an understanding of the principles of CT scanning. (g) Candidates should be able to show an understanding.
Computed Tomography
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
Conventional and Computed Tomography
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
LEC ( 2 ) RAD 323. Reconstruction techniques dates back to (1917), when scientist (Radon) developed mathematical solutions to the problem of reconstructing.
Seeram Chapter 11: Image Quality
COMPUTED TOMOGRAPHY (I) RAD 365 CT - Scan
Basic principles Geometry and historical development
5.4.1 X-Rays. (a) describe the nature of X-rays Stowmarket Physics X-rays - nature Forms of electromagnetic radiation Short wavelength High frequency.
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
Digital Image Characteristic
Without reference, identify principles relating to Computed Tomography Clinical Applications with at least 70 percent accuracy.
Alhanouf Alshedi CT Physics and Instrumentation RAD
1 RT 255 C Cross Sectional Anatomy Week 1 FINAL
X-Ray Medical Imaging Physics – IB Objectives
Computed Tomography Physics, Instrumentation, and Imaging
Seeram Chapter 7: Image Reconstruction
Computed Tomography Q & A
COMPUTED TOMOGRAPHY HISTORICAL PERSPECTIVE
PRINCIPLE AND BASIC PHYSICS OF COMPUTED TOMOGRAPHY
Seeram Chapter 9: Image Manipulation in CT
X - RAYS IN DIAGNOSTICS D. Krilov
Factors affecting CT image RAD
LEC (1) Design/layout by E Bashir, CAMS, King Saud University, Ref: Seeram, Karthikeyan, internet.
Computed Tomography Diego Dreossi Silvia Pani University of Trieste and INFN, Trieste section.
CT IMAGE RECONSTRUCTION  Hounsfield envisioned dividing a slice into a matrix of 3-dimensional rectangular boxes (voxels) of material (tissue). Conventionally,
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
Part No...., Module No....Lesson No
Part No...., Module No....Lesson No
CT Chapter 4: Principles of Computed Tomography. Radiography vs. CT Both based on differential attenuation of x-rays passing through body Radiography.
Quality Assurance.
Interaction of x-ray photons (and gamma ray photons) with matter.
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. Chapter 26 Three-Dimensional Digital Imaging.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Single Slice Spiral - Helical CT
CT Multi-Slice CT.
Computed Tomography Basics
Computed Tomography Data Acquisition
Learning Objectives By the end of this lesson you should…
CT PRINCIPLES AYMAN OSAMA.
Basic principles Geometry and historical development
Basic Principles of CT Chapter 1.
Computed Tomography (C.T)
Computed Tomography (C.T)
Presentation transcript:

Computed Tomography http://www.stabroeknews.com/images/2009/08/20090830ctscan.jpg http://upload.wikimedia.org/wikipedia/commons/archive/d/da/20060904231838!Head_CT_scan.jpg http://www.capitalhealth.org/subpage.cfm?ref=36

Computed Tomography - Introduction Computed Tomography, CT for short (also referred to as CAT, for Computed Axial Tomography), utilizes X-ray technology and sophisticated computers to create images of cross-sectional “slices” through the body. CT exams and CAT scanning provide a quick overview of pathologies and enable rapid analysis and treatment plans. Tomography is a term that refers to the ability to view an anatomic section or slice through the body. Anatomic cross sections are most commonly referred to as transverse axial tomography. The CT scanner was developed by Godfrey Hounsfield in the late 1960s. This x-ray based system created projection information of x-ray beams passed through the object from many points across the object and from many angles (projections). CT produces cross-sectional images and also has the ability to differentiate tissue densities, which creates an improvement in contrast resolution.

Computed Tomography - Introduction The x-ray tube in a CT scanner is designed to produce a fan shaped beam of x-rays that is approximately as wide as your body. The x-ray tube on a CT scanner is more heavy duty than tubes used for standard film imaging since the unit rotates and they operate at slightly higher energies. Opposite the patient is an array of detectors that measure the intensity of the x-ray beam at points laterally across the patients body. Modern CT scanners use solid state detectors that have very high efficiencies. Solid state detectors are made of a variety of materials that create a semiconductor junction similar to a transistor. Ultrafast ceramic detectors use rare earth elements such as silicon, germanium, cadmium, yttrium or gadolinium, which create a semiconducting p-n junction. Ceramic solid-detectors are very fast, can be extremely stable, and are produced to form an array of very small, efficient detectors that can cover a large area.

Computed Tomography - The basics The x-rays are produced in a part of the ring and the ring is able to rotate around the patient. The target ring contains an array of detectors and is internally cooled so the to reduce electronic noise and to cool the anode. The patient is put into the system using a precise high speed couch. http://www.themesotheliomalibrary.com/ct-scan.html http://www.endocrinesurgery.ucla.edu/images/adm_tst_ct_scan.jpg

- The basics of image formation Computed Tomography - The basics of image formation http://www.themesotheliomalibrary.com/ct-scan.html The x-ray tube and detectors rotate around the patient and the couch moves into the machine. This produces a helical sweep pattern around the patient. The patient opening is about 70cm in diameter. The data acquired by the detectors with each slice is electronically stored and are mathematically manipulated to compute a cross sectional slice of the body. Three dimensional information can be obtained by comparing slices taken at different points along the body. Or the computer can create a 3D image by stacking together slices. As the detector rotates around many cross sectional images are taken and after one complete orbit the couch moves forward incrementally. Willi Kalender, Computed Tomography, Publicis Corporate Publishing 2005

- The basics of image formation Computed Tomography - The basics of image formation Here the x-ray tube and detector array makes many sweeps past the patient. The x-ray tube and detector array is capable of rotating around the axis of the patient. Each scan tries to determine the composition of each transverse cross section.

- The basics of image formation Computed Tomography - The basics of image formation As the x-ray tube and detectors swing around an intensity profile mapping is created. This could also be written as an attenuation profile which is the incident intensity minus the transmitted intensity. This generates a set of N equations that will be solved simultaneously for m(x,y) in the image reconstruction system.

- The basics of image formation Computed Tomography - The basics of image formation Homogeneous object, monochromatic radiation Inhomogeneous object, monochromatic radiation Inhomogeneous object, polychromatic radiation m(x,y) = ? In a CT scan we measure the intensity of radiation. The attenuation value, m, is easily determined if you have a homogeneous object. The incident intensity needs to be known and for inhomogeneous objects we need many scans to determine m(x,y).

- The basics of image formation Computed Tomography - The basics of image formation Pixel – picture element – a 2D square shade of gray. Voxel – volume element – a 3D volume of gray. This is a result of a computer averaging of the attenuation coefficients across a small volume of material. This gives depth information. Each voxel is about 1mm on a side and is as thick as 2 – 10mm depending on the depth of the scanning x-ray beam.

- The basics of image formation Computed Tomography - The basics of image formation The detectors see the forward projected x-rays and measure the intensity, given that the x-ray intensity without the body present is known. The intensity Ni written as sum of attenuation coefficients along a given x-ray path. This generates a shade of gray and a number associated with this shade. Then the detector changes angles and the process repeats. The images are reconstructed by a method called back projection, or tracing backwards along the x-rays forward path to reconstruct the image and calculating the absorption due to a localized region. This a mathematically tedious process, but is handled easily with computers.

- The basics of image formation Computed Tomography - The basics of image formation The top scan we see that there are lighter and darker regions somewhere in it, but we don't know whether the light/dark regions is high, low, or in the middle. In other words, we know where the light region is horizontally but not vertically. So by stretching it out we're kind of saying, "We don't know where the light spot is vertically, so for now give it all vertical values!” Now do a vertical scan and now we've taken the light/ dark spots whose location we know vertically and "smeared" it out across all horizontal positions. You can see where the light areas cross and it gets even more light there and we can start to form an image. By "adding" more shadows medium light lines would eventually disappear and we’d have a more complete and higher resolution image.

- Hounsfield Units or CT numbers Computed Tomography - Hounsfield Units or CT numbers CT numbers (or Hounsfield units) represent the percent difference between the x-ray attenuation coefficient for a voxel and that of water multiplied by a constant. Water has a CT number of zero and the numbers can be positive or negative depending on the absorption coefficient. This is how we assign a shade of gray, and 1000 is just a scaling factor set by the CT manufacturer.

Computed Tomography - Image Quality Number of Pixels In images a and b we have an 80 x 80 images matrix and you can easily see the discrete pixels. In images c and d we have a 1024 x 1024 image matrix. Here the individual pixels are not seen and the image quality increases.

Computed Tomography - Image Quality Contrast Resolution – The ability to differentiate between different tissue densities in the image High Contrast - Ability to see small objects and details that have high density difference compared with background. - These have very high density differences from one another. - Ability to see a small, dense lesion in lung tissue and to see objects where bone and soft tissue are adjacent Low Contrast - Ability to visualize objects that have very little difference in density from one another. - Better when there is very low noise and for visualizing soft-tissue lesions within the liver. - Low contrast scans can differentiate gray matter from white matter in the brain.

Computed Tomography - Imaging artifacts Artifacts can degrade image quality and affect the perceptibility of detail. Includes Streaks – due to patient motion, metal, noise, mechanical failure. Rings and bands – due to bad detector channels. Shading - can occur due to incomplete projections. Streaks Rings and bands Shading

- Advantages & Disadvantages Computed Tomography - Advantages & Disadvantages Advantages: Desired image detail is obtained Fast image rendering Filters may sharpen or smooth reconstructed images Raw data may be reconstructed post-acquisition with a variety of filters Disadvantages Multiple reconstructions may be required if significant detail is required from areas of the study that contain bone and soft tissue Need for quality detectors and computer software X-ray exposure